K200392 · Sm Eng Co., Ltd. · GAM · Jul 7, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200392
Device Name
REXSIN
Applicant
Sm Eng Co., Ltd.
Product Code
GAM · General, Plastic Surgery
Decision Date
Jul 7, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4493
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Device Story
REXSIN is a synthetic, absorbable, braided, coated surgical suture, provided with or without a stainless steel (SUS 304) needle. The suture material is polyglycolic acid (PGA), coated with polycaprolactone and calcium stearate. It is available in various sizes (8-0 to 2) and lengths. The device is used by surgeons in clinical settings for tissue approximation and ligation. It functions by providing temporary wound support during healing; it retains over 65% of tensile strength after 2 weeks and is completely absorbed in tissue within approximately 90 days. The device is supplied sterile for single use.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with USP <861> (diameter), USP <871> (needle attachment), USP <881> (tensile strength), and USP monograph for absorbable surgical sutures. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, implantation, blood interaction). Sterile barrier system validated per ISO 11607 and ASTM standards.
Technological Characteristics
Synthetic, braided, coated PGA suture. Coating: polycaprolactone and calcium stearate. Needle: stainless steel (SUS 304). Sizes: 8-0 to 2. Sterilization: terminally sterilized. Standards: USP <861>, <871>, <881>; ISO 10993 (biocompatibility); ISO 11607 (packaging); ASTM F88, F1929, F1980.
Indications for Use
Indicated for general soft tissue approximation and/or ligation, including ophthalmic procedures. Contraindicated for cardiovascular or neural tissue.
Regulatory Classification
Identification
An absorbable poly(glycolide/l-lactide) surgical suture (PGL suture) is an absorbable sterile, flexible strand as prepared and synthesized from homopolymers of glycolide and copolymers made from 90 percent glycolide and 10 percent l-lactide, and is indicated for use in soft tissue approximation. A PGL suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. “Monograph for Absorbable Surgical Sutures;” it may be monofilament or multifilament (braided) in form; it may be uncoated or coated; and it may be undyed or dyed with an FDA-approved color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
July 7, 2020
Sm Eng Co., Ltd % Sanglok Lee Manager Wise Company Inc. #303, 142, Gasan digital 1-ro Geumcheon-gu, 08507 Kr
Re: K200392
Trade/Device Name: Rexsin Regulation Number: 21 CFR 878.4493 Regulation Name: Absorbable Poly(Glycolide/L-Lactide) Surgical Suture Regulatory Class: Class II Product Code: GAM Dated: June 4, 2020 Received: June 11, 2020
Dear Sanglok Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhurv, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200392
Device Name REXSIN
Indications for Use (Describe)
REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for SM Eng, which is a company that specializes in engineering. The logo is purple and white, and it features the company's name and website address. The website address is www.sm-eng.net. The logo is simple and modern, and it is likely used on the company's website and marketing materials.
46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea TEL.82-51-305-8016 FAX.82-51-305-8021
## 510(k) Summary
### The assigned 510(k) Number: K200392
### Date of Submission: February 12, 2020
### Applicant
Company name: SM ENG CO., LTD Address: 46, Nakdong-daero 1302beon-gil, Sasang-gu, Busan, Korea TEL: +82 51 3058016 FAX: +82513058021 Email: leesg@sm-eng.net
### Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. #303, 142, Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 70 8812 3619 / +82 2 831 3615 FAX: +82 50 4031 3619 Email: info@wisecompany.org
### Subject Device Identification
Proprietary Name: REXSIN Common Name: Synthetic Absorbable PGA Suture With or Without Needle Device Class: Class II Regulation Number:21 C.F.R. 878.4493 Product Code: GAM Classification Name: Suture, Absorbable, Synthetic, Polyglycolic Acid
### Indication for use
REXSIN is indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue
### Predicate devices
510(k) Number: K073614 Device Name: WG-Surgical Sutures with Needle Manufacturer: FOOSIN MEDICAL SUPPLIES INC.LTD
#### De vice Description
Synthetic Absorbable PGA Suture With or Without Needle (REXSIN) are produced and provided by SM ENG Co., Ltd. REXSINis indicated for use in general soft tissue approximation and/or ligation, including use in ophthalmic procedures, but not for use in cardiovascular or neural tissue
{4}------------------------------------------------
SM Eng
www.sm-eng.net
46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea
TEL.82-51-305-8016 FAX.82-51-305-8021
Suture of REXSIN is Trisorb manufactured by SAMY ANG Biopharmaceuticals Corporation. SM Eng is receiving bulk Trisorb from Samyang, and then go through cutting, adhesion of need and thread, winding and sterilization process according to SM Eng procedure, which becomes REXSIN suture.
REXSIN is a sterilized medical device composed with absorbable, braided, coated suture with needle, stainless steel SUS 304
REXSIN Suture is a synthetic absorbable, braided, coated suture composed of a Polyglycolic acid (PGA). The suture is available undyed (natural) or dyed (D&C Violet No.2). Approximately over 65% of tensile strength remain after 2week. Complete absorption in tissues takes around 90 days.
REXSIN Sutures meets requirements established by the United States Pharmacopoeia (USP) for Absorbable Surgical Sutures. REXSIN Sutures is available in 8-0, 7-0, 6-0, 5-0, 2-0, 0, 1 and 2, which are the sizes identified in the currently recognized United States.
### Non-Clinical Test Conclusion
Bench tests were conducted to verify that the subject device (REXSIN) met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
USP <861> SUTURES - DIAMETER USP <871> SUTURES - NEEDLE ATTACHMENT
USP <881> TENSILE STRENGTH USP MONOGRAPH OF ABSORBABLE SURGICAL SUTURE
Sterile barrier system testing
ISO 11607-1, Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging Systems ISO 11607-2, Packaging For Terminally Sterilized Medical Devices - Part 2: Validation Requirements For Forming, Sealing And Assembly Processes ASTM F 88, Standard Test Method for Seal Strength of Flexible Barrier Materials ASTM F 1929, Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
Accelerated aging stability testing
- ASTM F 1980, Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices. (Sterility)
Biocompatibility testing
- 1) Suture Needle
- ISO 10993-12: Sample Preparation
- ISO 10993-5: Test for Cytotoxicity
- ISO 10993-10: Test for Irritation and Sensitization
2) Suture
- ISO 10993-1: Selection of Tests
{5}------------------------------------------------
46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea
TEL.82-51-305-8016 FAX.82-51-305-8021
- ISO 10993-2: Animal Welfare
- ISO 10993-12: Sample Preparation
- ISO 10993-5: Test for Cytotoxicity
- ISO 10993-10: Test for Irritation and Sensitization
- ISO 10993-11: Testfor Systemic Toxicity
- ISO 10993-3: Tests for Genotoxicity
- ISO 10993-6: Test for Local Effects after Implantation
- ISO 10993-4: Selection of Tests for Interaction with Blood
Real-time and accelerated aging stability testing was performed to support shelf life of REXS IN. Additionally, the residual strength and absorption rate studies were evaluated in accordance with the requirements outlined in FDA's Class II Special Controls Guidance Document: Surgical
Sutures.
#### Substantially Equivalent Conclusion
The following table compares the subject device (REXSIN) to the predicate devices with respect to intended use, technological characteristics and principles of operation, etc.
| Property | Subject device:<br>REXSIN | Predicate device:<br>WG-Surgical Sutures with<br>Needle |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)Number | K200392 | K073614 |
| Product Code | GAM | Same |
| Indicationsfor use<br>statements | REXSIN is indicated for use in<br>general soft tissue approximation<br>and/or ligation, including use in<br>ophthalmic procedures, but not<br>for use in cardiovascular or<br>neural tissue. | Same |
| Class | ll | Same |
| Sterile | Yes | Same |
| Single Use | Yes | Same |
| Configuration | PGA Suture and Needle | Same |
| Suture | | |
| Material | Polyglycolic acid | Same |
| Coating material | Polycaprolactone and Calcium<br>stearate | Same |
| Color | Undyed (natural) and dyed (D&C<br>Violet No.2) | Same |
| Absorbable/Nonabsorbable | Absorbable | Same |
| Braided/Monofilament | Braided | Same |
| Barbed/Not Barbed | Not Barbed | Same |
| Suture Size | The subject device is available in<br>8-0, 7-0, 6-0, 5-0, 4-0,3-0,2-0, 0,<br>1 and 2., which are the sizes<br>identified in the currently<br>recognized United States<br>Pharmacopoeia. | Same. The subject device is<br>available in 10-0 through 3 or<br>4, which are the sizes<br>identified in the currently<br>recognized United States<br>Pharmacopoeia. |
| Length of Suture | 15, 30, 45, 50, 60, 75, 90, 100,<br>120, 125, 140, 250cm | Unknown |
Table. Comparison of Technology Characteristics
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for SM Eng. The logo is in a light gray color and has a stylized design. The website address, www.sm-eng.net, is written in a smaller font below the logo. The logo is simple and modern.
46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea TEL.82-51-305-8016 FAX.82-51-305-8021
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for SM Eng. The logo is in a serif font and is a dark gray color. The website address, www.sm-eng.net, is written in a smaller font below the logo. A purple line underlines the logo and website address. There is a gray swirl to the left of the logo.
46,Nakdong-daero,1302beon-gil, Sasang-Gu, Busan, Korea
TEL.82-51-305-8016 FAX.82-51-305-8021
| Diameter of Suture | The tensile strengths of subject<br>device comply with the tensile<br>requirement listed in USP <861> | Same | | |
|--------------------|-----------------------------------------------------------------------------------------------------------------|---------|--|--|
| Tensile strength | The tensile strengths of subject<br>device comply with the tensile<br>requirement listed in USP <881> | Same | | |
| Needle Attachment | The bond between suture and<br>needle of the applicant device<br>meet the requirements defined in<br>USP <871>. | Same | | |
| Needle | | | | |
| Material | Stainless Steel | Same | | |
| Needle type | Taper point, Reverse Cutting,<br>Conventional cutting, Taper<br>cutting, Spatula, Blunt point | Unknown | | |
| Biocompatibility | Comply with ISO 10993-5,<br>10993-10 | Same | | |
The subject devices, REXSIN are determined to be Substantially Equivalent (SE) to the predicate device, and as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.