K200386 · Z-Systems AG · DZE · Oct 14, 2020 · Dental
Device Facts
Record ID
K200386
Device Name
Z5-BL
Applicant
Z-Systems AG
Product Code
DZE · Dental
Decision Date
Oct 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.
Device Story
Z5-BL is a root-form, endosseous threaded dental implant system for surgical placement in the maxilla or mandible. System includes 3.6 mm diameter implants (8 mm, 10 mm lengths), gingiva formers, and straight/15° angled abutments for single-unit cemented restorations. Compatible with previously cleared components (K190243) including healing caps, multi-unit abutments, and Locator-type abutments. Implants manufactured from Y-TZP zirconia (ISO 13356); abutments from Y-TZP or PEEK. Implants provided sterile (plasma gas); other components sterilized by end-user (moist heat). Used by dental clinicians for functional and esthetic rehabilitation of edentulous patients. Output is a stable base for prosthetic appliances; clinical decision-making involves surgical site preparation and delayed loading protocols. Benefits include metal-free alternative for patients with metal allergies.
Clinical Evidence
Bench testing only. No clinical data included. Testing included: sterilization validation (ISO 14937, ISO 17665-1), shelf life (ISO 11607, ASTM F1886, ASTM D3078), bacterial endotoxin (USP <85>), biocompatibility (ISO 10993, USP Class VI), static/dynamic compression fatigue (ISO 14801), abutment screw loosening/removal torque (ISO 18130), and wear particle analysis (SEM/EDX).
Technological Characteristics
Materials: Y-TZP zirconia (ISO 13356), PEEK. Design: Root-form, endosseous threaded, 3.6 mm diameter. Surface: Grit-blasted and laser modified. Sterilization: Plasma gas (implants), moist heat (abutments). Mechanical: Static/dynamic fatigue testing per ISO 14801. Connectivity: None.
Indications for Use
Indicated for patients requiring dental implants in the upper or lower jaw for prosthodontic appliance attachment to replace missing teeth, including patients with metal allergies and associated chronic diseases. Intended for delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a bold, sans-serif font and the full name in a smaller, sans-serif font below it. The overall design is clean and professional, reflecting the agency's role in regulating food and drugs in the United States.
October 14, 2020
Z-Systems AG % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K200386
Trade/Device Name: Z5-BL Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: Class II Product Code: DZE, NHA Dated: September 11, 2020 Received: September 14, 2020
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K200386
Device Name
Z5-BL
Indications for Use (Describe)
Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div></td><td><div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div></td></tr></table> | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### 510(k) Summary
### K200386
#### Z5-BL
#### Z-Systems AG
#### September 11, 2020
kthomas@paxmed.com flarson@paxmed.com
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Z-Systems AG<br>Werkhofstrasse 5<br>CH-4702 Oensingen<br>Switzerland<br>Telephone: +41 62 388 69 69 |
|-------------------|-----------------------------------------------------------------------------------------------------|
| Official Contact | Rubino DiGirolamo, CEO |
Representative/Consultant Kevin A. Thomas, PhD Floyd G. Larson, MS, MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1 858-792-1235 Fax: +1 858-792-1236
Email:
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Z5-BL |
|------------------------|----------------|
| Common Name | Dental implant |
| Regulation Number | 21 CFR 872.3640 |
|------------------------|---------------------------|
| Regulation Name | Endosseous dental implant |
| Regulatory Class | Class II |
| Product Code | DZE |
| Secondary Product Code | NHA |
| Classification Panel | Dental |
|----------------------|------------------------------------------------------------------------------------------------|
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory,<br>ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device: K190243, Z5-BL, Z-Systems AG Reference Devices: K132881, Z5c, Z-Systems AG K131701, Z5mlb and Z5mlc, Z-Systems AG K120793, Z-Look3 Evo SLM, Z-Systems AG
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
Z5-BL implants are designed for surgical implantation into the upper and lower jaw for the attachment of prosthodontic appliances to replace missing teeth. Z5-BL implants are suitable for patients with metal allergies and the chronic diseases resulting from them. Z5-BL implants are intended for delayed loading.
#### SUBJECT DEVICE DESCRIPTION
Z5-BL is a dental implant system that includes root-form, endosseous threaded dental implants indicated for bone level placement. The subject device includes dental implants provided in one body diameter of 3.6 mm in lengths of 8 mm, 10 mm. The subject device also includes a gingiva former, straight abutments for single-unit cemented restorations, and a 15° angled abutment also for single-unit cemented restorations.
The subject device is compatible with components cleared previously in K190243, including a healing cap, straight abutments for multi-unit cemented restorations, Locator-type abutments for retention of overdentures, and occlusal (abutment) screws.
The subject device dental implants are manufactured from Y-TZP zirconia conforming to ISO 13356. The subject device abutments are manufactured from Y-TZP zirconia conforming to ISO 13356, or from polyetheretherketone (PEEK).
The subject device implants are provided sterilized by plasma gas. All other components are to be sterilized by the end-user by moist heat (steam).
#### PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included:
sterilization validation according to ISO 14937 (referenced from K132881) and ISO 17665-1 (referenced from K120793);
sterile product shelf life testing according to ISO 11607-1, ISO 11607-2, ASTM F1886/F1886M, and ASTM D3078 (referenced from K132881);
bacterial endotoxin testing for subject devices provided sterile (Limulus Amebocyte Lysate (LAL) test according to USP 40-NF 35 <85>, referenced from K190243);
biocompatibility of the yttria-stabilized zirconia (Y-TZP) conforming to ISO 13356 used to manufacture the subject device dental implants and abutments (referenced from identical material used for dental implants and abutments cleared in K190243);
biocompatibility endosseous threaded surface of the subject device implants (referenced from identical Al2O3 grit blasted and laser modified surface applied to the dental implants cleared in K190243);
biocompatibility of the PEEK material according to ISO 10993-5, ISO 10993-12, ISO 10993-10, ISO 19003-11. ISO 10993-18, and USP Class VI testing according to USP 37-NF32 <88> (referenced from K190243);
static compression and compression fatigue testing according to ISO 14801 of the subject device zirconia implants and zirconia abutments, with both zirconia and titanium alloy abutment screws;
assessment of abutment screw loosening and abutment screw removal torque testing (according to ISO 18130) after insertion to the torque recommended in the subject device labeling, and after dynamic testing
{5}------------------------------------------------
according to ISO 14801, including microscopic examination of the abutment screws and internal threads of the implant bodies (referenced from K190243);
assessment of wear particles associated with the Ti-6A1-4V alloy abutment screw used with zirconia implant bodies and zirconia abutments after dynamic testing according to ISO 14801, including examination by light microscopy and scanning electron microscopy with energy dispersive X-ray analysis (SEM / EDX) (referenced from K190243).
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.
The subject device is substantially equivalent in intended use to the primary predicate K190243, and to the reference devices K132881 and K131701. All are intended for use in the maxilla and mandible to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. The subject device and the primary predicate device K190243 are made of identical materials (implants and abutments). The subject device dental implants and the primary predicate device K190243 dental implants have an identical endosseous threaded surface (grit-blasted and laser modified), have the same internal implant connection features, and the same implant lengths. The subject device and the primary predicate device K190243 also include straight and angled abutments for cement-retained, single-unit restorations. Differences between subject device and the primary predicate device K190243 include the implant diameter (subject device 3.6 mm) and implant platform diameter (subject device also 3.6 mm). None of these minor design differences impact safety or effectiveness or change the intended use of the device.
The reference device K132881 is for support of substantial equivalence of the subject device in terms of sterilization and shelf life (also referenced in K190243). The subject device dental implants are provided to the end user sterilized by plasma gas. The subject device components that are provided sterile use the same packaging and have the same sterile barrier shelf life as the sterile components cleared in K132881 (referenced in K190243). The processes to monitor, control, and test for bacterial endotoxins for the subject device are the same as K190243.
The reference device K131701 is for support of substantial equivalence of the subject device in terms of the smaller dental implant body diameter and corresponding implant platform diameter (both 3.6 mm). The subject device dental implants and the reference device K131701 dental implants also are made of identical materials, and have the identical endosseous threaded surface (grit-blaster and laser modified).
Mechanical performance testing of the subject device was performed according to ISO 14801 Dentistry -Implants – Dynamic fatigue test for endosseous dental implants. The fatigue limit data demonstrated that constructs of the device have sufficient strength for their intended use.
Minor differences in the designs, dimensions, sizes, or materials among the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence. These minor differences do not impact safety or effectiveness because these differences are related to the specific designs features and system components, and are mitigated by the mechanical performance testing.
{6}------------------------------------------------
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical materials. The subject device, the primary predicate, and reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{7}------------------------------------------------
# Table of Substantial Equivalence
| Comparison | Subject Device | Primary Predicate | Reference Device | Reference Device |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use Statement | K200386<br>Z5-BL<br>Z-Systems AG<br><br>Z5-BL implants are designed for surgical implantation<br>into the upper and lower jaw for the attachment of<br>prosthodontic appliances to replace missing teeth.<br>Z5-BL implants are suitable for patients with metal<br>allergies and the chronic diseases resulting from them.<br>Z5-BL implants are intended for delayed loading. | K190243<br>Z5-BL<br>Z-Systems AG<br><br>Z5-BL implants are designed for surgical implantation<br>into the upper and lower jaw for the attachment of<br>prosthodontic appliances to replace missing teeth.<br>Z5-BL implants are suitable for patients with metal<br>allergies and the chronic diseases resulting from them.<br>Z5-BL implants are intended for delayed loading. | K132881<br>Z5c<br>Z-Systems AG<br><br>Z5c implants are designed for surgical implantation<br>into the upper and lower jaw for the attachment of<br>prosthodontic appliances to replace missing teeth.<br>Z5c implant system is also suitable for patients with<br>metal allergies and the chronic diseases resulting from<br>them. Z5c implants are intended for delayed loading. | K131701<br>Z5mlb and Z5mlc<br>Z-Systems AG<br><br>Z5mlb:<br>Z5mlb implants arc designed for surgical implantation<br>into the edentulous upper and lower jaw for the<br>attachment of dentures to replace missing teeth. Z5mlb<br>implant system is also suitable for patients with metal<br>allergies and the chronic diseases resulting from them.<br>Z5mlc:<br>Z5mlc implants are designed for surgical implantation<br>into the edentulous upper and lower jaw for the<br>attachment of dentures to replace missing teeth. Z5mlc<br>implant system is also suitable for patients with metal<br>allergies and the chronic diseases resulting from them. |
| Intended Use | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible |
| Reason for Predicate/<br>Reference Device | Not applicable | IFU Statement; implant design, material, surface;<br>abutment design | Reference for sterilization and shelf life | Reference for implant diameter (3.6 mm) |
| Product Codes | DZE, NHA | DZE, NHA | DZE, NHA | DZE |
| Components | Implants, abutments | Implants, abutments,<br>abutment screws | Implants, abutments<br>(abutment cemented to implant) | One-piece implants |
| Implant Designs | | | | |
| Implant Diameter, mm | 3.6 | 4.0, 5.0 | 4.0, 5.0 | 3.6, 4.0 |
| Implant Endosseous Length, mm | 8, 10, 12 | 8, 10, 12 | 8, 10, 12 | 8, 10 |
| Platform diameter, mm | 3.6 | 4.0, 5.0 | 4.0, 5.0 | Not applicable |
| Implant Material | Y-TZP | Y-TZP | Y-TZP | Y-TZP |
| Implant Surface | Grit-blasted and laser modified | Grit-blasted and laser modified | Grit-blasted and laser modified | Grit-blasted and laser modified |
| Abutment Designs | | | | |
| Features | Subject device<br>Indexed, straight and angled<br>Previously cleared compatible<br>Non-indexed straight;<br>Indexed and Non-indexed Locator-type | Indexed, straight and angled;<br>Non-indexed straight;<br>Indexed and Non-indexed Locator-type | Non-indexed straight<br>Non-indexed angled 15° | Ball-type attachment;<br>Locator-type attachment |
| Prosthesis Attachment | Subject device<br>Cement-retained<br>Previously cleared compatible<br>Cement-retained;<br>Locator-type attachment | Cement-retained;<br>Locator-type attachment | Cement-retained | Ball-type attachment;<br>Locator-type attachment |
| Restoration | Single-unit; Multi-unit | Single-unit; Multi-unit | Single-unit; Multi-unit | Multi-unit |
| Comparison | Subject Device | Primary Predicate | Reference Device | Reference Device |
| | K200386<br>Z5-BL<br>Z-Systems AG | K190243<br>Z5-BL<br>Z-Systems AG | K132881<br>Z5c<br>Z-Systems AG | K131701<br>Z5mlb and Z5mlc<br>Z-Systems AG |
| Prosthetic platform diameter, mm | Subject device<br>3.8 mm – 4.3 mm<br>Previously cleared compatible<br>4.5 mm | 4.5 mm – 6.5 mm | 3.6 mm | Not applicable |
| Abutment angle | Straight (0°), 15° | Straight (0°), 15° | Straight (0°), 15° | Straight (0°) |
| Abutment material | Y-TZP | Y-TZP | Y-TZP | Y-TZP |
| Abutment Screw Material | Previously cleared compatible<br>Y-TZP<br>Ti-6Al-4V alloy | Y-TZP<br>Ti-6Al-4V alloy | Not applicable | Not applicable |
| Sterilization Status/Method | | | | |
| Implants | Sterile / plasma gas | Sterile / plasma gas | Sterile / plasma gas | Sterile / plasma gas |
| Abutments | Non-sterile / moist heat (steam) | Healing Cap: Sterile / plasma gas<br>All other:<br>Non-sterile / moist heat (steam) | Non-sterile / Dry heat | Not applicable |
| Other Components | Not applicable | Non-sterile / moist heat (steam) | Non-sterile / moist heat (steam) | Not applicable |
{8}------------------------------------------------
510(k) Summary
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.