Duranext Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Device Story
Duranext Abutments are titanium alloy (Ti-6Al-4V ELI) machinable blanks used to fabricate patient-specific dental implant abutments via CAD/CAM processes. The apical end is pre-manufactured to interface with specific endosseous dental implant platforms (NobelActive, NobelReplace, NobelParallel, Zimmer Screw-Vent/Tapered Screw-Vent). Clinicians order custom abutments; designs are sent to an Elegant Direct Corp. validated milling center for manufacture. The final abutment supports cement-retained single crowns or multi-unit restorations. The device is used in dental clinics; the milling process occurs at a registered manufacturing facility. The output is a custom abutment that restores chewing function for the patient.
Clinical Evidence
Bench testing only. No clinical data included. Evidence includes biocompatibility testing (ISO 10993-5, ISO 10993-12), sterilization validation (ISO 17665-1), reverse engineering of OEM implant bodies/abutments to confirm interface compatibility, and static compression and compression fatigue testing per ISO 14801.
Technological Characteristics
Material: Ti-6Al-4V ELI. Form factor: Machinable blanks (9.5 mm diameter, 20 mm length). Interface: Internal connection. Manufacturing: CAD/CAM milling at validated centers. Design parameters: 0-30° angulation, 0.5-4.0 mm gingival height, 4.0 mm minimum post height, 20.0 mm maximum abutment height. Sterilization: ISO 17665-1.
Indications for Use
Indicated for partially or fully edentulous patients requiring single or multiple tooth prosthetic restorations supported by dental implants in the maxilla or mandible.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
July 22, 2020
Elegant Dental Corp. % Floyd Larson President PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K200355
Trade/Device Name: Duranext Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: June 18, 2020 Received: June 22, 2020
Dear Floyd Larson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200355
#### Device Name
Duranext Abutments
#### Indications for Use (Describe)
Duranext Abutments are intended for use with dental implants as a support for single or multiple toth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
| Implant System Compatibility | Implant Body<br>Diameter, mm | Implant Platform |
|--------------------------------------------|------------------------------|------------------|
| NobelActive® (conical connection) | 3.0 | 3.0 (3.0 mm) |
| | 3.5 | NP (3.5 mm) |
| | 4.3, 5.0 | RP (3.9 mm) |
| | 5.5 | WP (5.1 mm) |
| NobelReplace Conical Connection | 3.5 | NP (3.5 mm) |
| | 4.3, 5.0 | RP (3.9 mm) |
| NobelParallel Conical Connection | 3.75 | NP (3.5 mm) |
| | 4.3, 5.0 | RP (3.9 mm) |
| | 5.5 | WP (5.1 mm) |
| NobelReplace® (Internal tri-channel) | 3.5 | NP (3.5 mm) |
| | 4.3 | RP (4.3 mm) |
| | 5.0 | WP (5.0 mm) |
| | 6.0 | 6.0 (6.0 mm) |
| Zimmer Screw Vent®/<br>Tapered Screw-Vent® | 3.7, 4.1 | 3.5 mm |
| | 4.7 | 4.5 mm |
| | 6.0 | 5.7 mm |
All digitally designed custom abutments for use with Duranext Abutments are to be sent to an Elegant Direct Corp. validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Duranext Abutments Elegant Direct Corp.
July 21, 2020
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | Elegant Direct Corp.<br>2308 McDonald Avenue<br>Brooklyn, NY 11223 | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| | Telephone: | +1-718-375-5999 |
| | Fax: | +1-718-375-5983 |
| Official Contact | Gary Fingerman, President | |
| Representative/Consultant | Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone:<br>Fax:<br>Email: | +1 858-792-1235<br>+1 858-792-1236<br>flarson@paxmed.com<br>kthomas@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Duranext Abutments |
|----------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental Devices |
# PREDICATE DEVICE INFORMATION
| Primary predicate device: | K150203 Medentika CAD/CAM Abutments, Medentika GmbH |
|---------------------------|-------------------------------------------------------------------|
| Reference devices: | K160784 CAM Titanium Blanks, Altatec GmbH |
| | K171799 Elos Accurate Customized Abutment, Elos Medtech Pinol A/S |
| | K134045 Zimmer Zfx Ti Abutment for NobelActive, NobelReplace CC |
| | Zimmer Dental Inc. |
{4}------------------------------------------------
K183518, Preat Abutments, Preat Corporation K071370, NobelActive Internal Connection Implant, Nobel Biocare AB K142260, NobelActive®, Nobel Biocare AB K102436, NobelActive 3.0, Nobel Biocare AB K062566, NobelReplace Tapered Conical Connection, Nobel Biocare USA LLC K173418, NobelParallel™ Concial Connection, Nobel Biocare AB K050258, Groovy Implants, Nobel Biocare AB K011028, Screw-Vent Dental Implant System, Sulzer Dental, Inc. K112160, Tapered Screw-Vent® X Implant, Zimmer Dental, Inc.
# INDICATIONS FOR USE STATEMENT
Duranext Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
| Implant System Compatibility | Implant Body<br>Diameter, mm | Implant Platform |
|--------------------------------------------|------------------------------|------------------|
| NobelActive® (conical connection) | 3.0 | 3.0 (3.0 mm) |
| NobelActive® (conical connection) | 3.5 | NP (3.5 mm) |
| NobelActive® (conical connection) | 4.3, 5.0 | RP (3.9 mm) |
| NobelActive® (conical connection) | 5.5 | WP (5.1 mm) |
| NobelReplace Conical Connection | 3.5 | NP (3.5 mm) |
| NobelReplace Conical Connection | 4.3, 5.0 | RP (3.9 mm) |
| NobelParallel Conical Connection | 3.75 | NP (3.5 mm) |
| NobelParallel Conical Connection | 4.3, 5.0 | RP (3.9 mm) |
| NobelParallel Conical Connection | 5.5 | WP (5.1 mm) |
| NobelReplace® (Internal tri-channel) | 3.5 | NP (3.5 mm) |
| NobelReplace® (Internal tri-channel) | 4.3 | RP (4.3 mm) |
| NobelReplace® (Internal tri-channel) | 5.0 | WP (5.0 mm) |
| NobelReplace® (Internal tri-channel) | 6.0 | 6.0 (6.0 mm) |
| Zimmer Screw Vent®/<br>Tapered Screw-Vent® | 3.7, 4.1 | 3.5 mm |
| Zimmer Screw Vent®/<br>Tapered Screw-Vent® | 4.7 | 4.5 mm |
| Zimmer Screw Vent®/<br>Tapered Screw-Vent® | 6.0 | 5.7 mm |
All digitally designed custom abutments for use with Duranext Abutments are to be sent to an Elegant Direct Corp. validated milling center for manufacture.
### SUBJECT DEVICE DESCRIPTION
Duranext Abutments from Elegant Direct Corp. are a line of machinable blanks incorporating interface features compatible with eleven (11) endosseous dental implant system platforms (three (3) designs from two (2) manufacturers) and intended to be milled at an Elegant Direct Corp. validated milling center to produce patientspecific dental implant abutments. The subject device platform diameters range from 3.0 mm to 6.0 mm, and the corresponding compatible implant body diameters range from 3.0 mm to 6.0 mm.
{5}------------------------------------------------
Duranext Abutments are designed for fabrication of custom titanium alloy dental implant abutments by a CAD/CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. The portion of each abutment available for milling is 9.5 mm in diameter and 20 mm long. The apical end is premanufactured to fit the compatible implant platform, as shown above, and is available in an engaging (anti-rotation) design. A feature at the coronal end of the abutment is provided to interface with the milling equipment. Each abutment is provided with a screw designed to fit the compatible implant. The patient-specific abutment is intended to support a cement-retained single crown or multi-unit restoration.
The design parameters for the CAD/CAM fabrication of custom abutments from Duranext Abutment are:
Minimum wall thickness - 0.5 mm Minimum post height - 4.0 mm Maximum abutment height from the prosthetic platform - 20.0 mm Maximum gingival height - 4.0 mm Minimum gingival height - 0.5 mm Angulation - 0° to 30°
Manufacture of CAD/CAM custom abutments from Duranext Abutments is to be performed at an Elegant Direct Corp. validated milling center that is registered with FDA as a medical device manufacturing establishment.
# PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included: sterilization according to ISO 17665-1; biocompatibility according to ISO 10993-12; reverse engineering of OEM implant bodies, OEM abutments, and OEM abutment screws to confirm compatibility; and static compression and compression fatigue testing according to ISO 14801. No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.
With regard to intended use, the subject Duranext Abutment is substantially equivalent to the primary predicate device, K 150203 Medentika CAD/CAM Abutments and to the abutment reference devices K160784 CAM Titanium Blanks, K171799 Elos Accurate Customized Abutment and K134045 Zimmer Zfx Ti Abutment for NobelReplace CC. All are intended for use with dental implants as a support for single tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. While the primary predicate device is intended for use with the Straumann CARES System, digitally designed abutments for use with the abutment reference devices are intended to be sent to a validated milling center for manufacture, as is the case for the subject device. With regard to OEM implant system compatibility, all subject device compatibilities are included in those for the primary predicate device, with the exception of the WP (5.5) platform of the NobelParallel Conical Connection implants. Slight differences in the wording of the Indications for Use do not affect the intended use of the device to support a single-tooth or multiple-unit prosthesis for restoration of chewing function.
With regard to design, the subject device, the primary predicate device and abutment reference devices all are titanium alloy blanks intended to be used for the manufacture of patient-specific abutments using CAD/CAM techniques, with manufacture of the final finished device performed at a validated milling center - an Establishment that is registered with FDA as a manufacturer. The design parameters intended for the subject device are substantially equivalent to those of the primary predicate device; maximum angulation for each is 30°, maximum abutment height is 20 mm and minimum post height is 4 mm, maximum gingival height for the subject
{6}------------------------------------------------
510(k) Summary Page 4 of 7
device is 4 mm, which is not a worse case than the 6 mm maximum gingival height of the primary predicate device.
Testing to support substantial equivalence included biocompatibility testing according to ISO 10993-5 and ISO 10993-12. In addition, performance testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants was performed to demonstrate that the subject device has sufficient strength for its intended use when tested in dynamic loading.
## CONCLUSION
The subject device, the primary predicate device, and the abutment reference devices have intended use, have similar technological characteristics, and are materials. The subject device, the primary predicate, and reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{7}------------------------------------------------
| Subject Device | | | | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|---------------------------|--------------------------------|
| Duranext Abutments<br>Elegant Direct Corp. | Duranext Abutments are intended for use with dental implants as a support for single or multiple tooth prostheses in the maxilla or<br>mandible of a partially or fully edentulous patient | | | |
| | Implant System Compatibility | | Implant Body Diameter, mm | Implant Platform |
| | NobelActive® (conical connection) | | 3.0 | 3.0 (3.0 mm) |
| | | | 3.5 | NP (3.5 mm) |
| | | | 4.3, 5.0 | RP (3.9 mm) |
| | | | 5.5 | WP (5.1 mm) |
| | NobelReplace Conical Connection | | 3.5 | NP (3.5 mm) |
| | | | 4.3, 5.0 | RP (3.9 mm) |
| | NobelParallel Conical Connection | | 3.75 | NP (3.5 mm) |
| | | | 4.3, 5.0 | RP (3.9 mm) |
| | NobelReplace® (Internal tri-channel) | | 3.5 | NP (3.5 mm) |
| | | | 4.3 | RP (4.3 mm) |
| | | | 5.0 | WP (5.0 mm) |
| | | | 6.0 | 6.0 (6.0 mm) |
| | Zimmer Screw Vent®/<br>Tapered Screw-Vent® | | 3.7, 4.1 | 3.5 mm |
| | | | 4.7 | 4.5 mm |
| | | | 6.0 | 5.7 mm |
| Primary Predicate Device | All digitally designed custom abutments for use with Duranext Abutments are to be sent to an Elegant Direct validated milling<br>center for manufacture. | | | |
| K150203<br>Medentika CAD/CAM Abutments | Medentika Preface CAD/CAM Abutments are intended for use with dental implants as a support for single or multiple<br>tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient. | | | |
| (Preface) | Implant System Compatibility | Series | Implant Diameter (mm) | Platform Diameter (mm) |
| Medentika GmbH | Nobel Biocare Replace™ Select | E | 3.5, 4.3, 5.0, 6.0 | 3.5, 4.3, 5.0, 6.0 |
| | Nobel Biocare NobelActive™ | F | 3.0, 3.5, 4.3, 5.0 | 3.0, 3.5, 3.9 (4.3), 3.9 (5.0) |
| | Biomet 3i Osseotite® Certain® | H | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | Biomet 3i Osseotite® | I | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 |
| | Nobel Biocare Brånemark | K | 3.3, 3.75, 4.0, 5.0 | 3.5, 4.1, 4.1, 5.1 |
| | Straumann Bone Level | L | 3.3, 4.1, 4.8 | 3.3, 4.1, 4.8 |
| | Straumann Standard | N | 3.3, 4.1, 4.8 | 3.5(NNC), 4.8, 6.5 |
| | Zimmer Tapered Screw-vent® | R | 3.3, 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 |
| | Astra Tech OsseoSpeed™ | S | 3.0, 3.5, 4.0, 4.5, 5.0 | 3.0, 3.5, 4.0, 4.5, 5.0 |
| | Dentsply Friadent® Frialit/XiVE® | T | 3.4, 3.8, 4.5, 5.5 | 3.4, 3.8, 4.5, 5.5 |
| | Medentika PreFace is intended for use with the Straumann® CARES® System. All digitally designed abutments for use with<br>Medentika CAD/CAM Abutments are intended to be manufactured at a Straumann® CARES® validated milling center | | | |
# Device Comparison Table of Substantial Equivalence – Indications for Use Statements
{8}------------------------------------------------
| Reference Devices | | | |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------|
| K160784<br>CAM Titanium Blanks<br>Altatec GmbH | CAM Titanium Blanks are intended for the fabrication of abutments and healing caps on CAMLOG® SCREWLINE<br>and CAMLOG® ROOT-LINE [, on CONELOG SCREWLINE or on iSy] implants in the maxilla and mandible for the purpose of<br>supporting single or multiple tooth prostheses of a partially or fully edentulous patient. Digitally designed abutments fabricated with<br>CAD/CAM techniques for use with CAMLOG® CAM Titanium Blanks are intended to be sent to a CAMLOG validated milling<br>center for manufacture. | | |
| K171799<br>Elos Accurate® Customized Abutment<br>Elos Medtech Pinol A/S | The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or<br>multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the<br>included Elos Prosthetic screw. The Elos Accurate® Customized Abutments are compatible with the following implant systems: | | |
| | Ref. No. | Platform compatibility | Implant diameter |
| | AB-BRA411213-US | Nobel Biocare® / Brånemark® RP | 3.75 mm & 4 mm |
| | AB-BRA351213-US | Nobel Biocare® / Brånemark® NP | 3.3 mm |
| | AB-BRA511213-US | Nobel Biocare® / Brånemark® WP | 5 mm |
| | All digitally designed Elos Accurate® Customized Abutments are intended to be manufactured at an Elos Medtech approved milling<br>facility. | | |
| K134045<br>Zimmer Zfx Titanium Abutment for<br>NobelActive Implant System<br>Zimmer Dental Inc. | The Zimmer Zfx Titanium Abutment for NobelActive Implant System is designed for use as a terminal or intermediate abutment for<br>cement retained prostheses. The abutment can be used with NobelActive and NobelReplace Conical Connection implants with a<br>Narrow Platform (NP) Ø 3.5mm or a Regular Platform (RP) Ø 3.9mm. | | |
{9}------------------------------------------------
| Comparison | Subject Device | Primary Predicate<br>Device | Reference Devices | | |
|----------------------------------------------|----------------------------|------------------------------------------|----------------------------|---------------------------------------|--------------------------------------------|
| | | K150203 | K160784 | K171799 | K134045<br>Zimmer Zfx Titanium |
| | Duranext Abutments | Medentika CAD/CAM<br>Abutments (Preface) | CAM Titanium Blanks | Elos Accurate® Customized<br>Abutment | Abutment for NobelActive<br>Implant System |
| | Elegant Direct Corp. | Medentika GmbH | Altatec GmbH | Elos Medtech Pinol A/S | Zimmer Dental Inc. |
| Design | | | | | |
| Abutment Blank | CAD/CAM Blank | CAD/CAM Blank | CAD/CAM Blank | CAD/CAM Blank | CAD/CAM Blank |
| CAD/CAM Abutment Design Parameters | | | | | |
| Minimum post height, mm | 4 | 4 | Not stated | Not stated | 3 |
| Maximum gingival height, mm | 4 | 6 | Not stated | Not stated | Not stated |
| Maximum abutment height, mm | 20 | 20 | Not stated | Not stated | 12 |
| Maximum abutment angulation | 30° | 30° | 30° | 20° | 25° |
| Abutment attachment to implant | Screw | Screw | Screw | Screw | Screw |
| Prosthesis attachment to abutment | Cement | Cement | Cement | Cement | Cement |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit |
| Abutment/implant<br>platform<br>diameter, mm | 3.0 - 5.7 | 3.0 - 6.0 | 3.3 - 6.0 | 3.5 - 5.1 | 3.5, 3.9 |
| Abutment/ Implant Interface | Internal | Internal | Internal | External | Internal |
| Material | | | | | |
| Abutment | Ti-6Al-4V ELI | Ti alloy | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V |
| Screw | Ti-6Al-4V ELI | Ti alloy | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti alloy |
# Device Comparison Table of Substantial Equivalence – Technological Characteristics
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.