BPM16B automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging 125 ~ 210 mm (approx. 4.9 ~ 8.3 inch) and for home use
Device Story
Wrist-worn blood pressure monitor; uses oscillometric method during cuff inflation to measure systolic/diastolic blood pressure and heart rate. Intended for home use by adults. Device features LCD panel for output display. Detects irregular heartbeats and provides visual warning symbol. Operates via battery power. Healthcare providers use readings to monitor patient blood pressure trends; aids in hypertension management and cardiovascular health assessment.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2013. Results demonstrate device meets required limits for mean difference and standard deviation for automated non-invasive sphygmomanometers.
Indicated for adults aged 18+ with wrist circumference 125-210 mm for home-based automated measurement of systolic/diastolic blood pressure and heart rate via oscillometric method.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
AVITA BPM16 Wrist Type Blood Pressure Monitor (K102624)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 17, 2020
Avita Corporation % Anita Chen Advisor ZhengCheng Consulting Corporation 238, No.19, 335 Lane, Fu-Xi Road, Shulin District New Taipei City, Taiwan, 238
Re: K200346
Trade/Device Name: Wrist Type Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 7, 2020 Received: November 19, 2020
Dear Anita Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200346
Device Name
Wrist Type Blood Pressure Monitor
#### Indications for Use (Describe)
BPM16B automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's wrist. The intended use of this over-the-counter device is for adults aged 18 years and older with wrist circumference ranging 125 ~ 210 mm (approx. 4.9 ~ 8.3 inch) and for home use
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| | <div style="display:flex; flex-direction:row;"> <div style="margin-right: 10px;"> </div> <div style="margin-right: 10px;"> </div> </div> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Section 5 510(k) Summary
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92
## The assigned 510(k) Number: TBD
| 1. | Submitter | | | |
|----|----------------------------------------|------------------------------------------------------|--|--|
| | Mailing Address | AVITA Corporation | | |
| | | 9F, NO.78, SEC.1, Kwang-Fu Road, San-Chung District, | | |
| | | New Taipei City, 24158, Taiwan, R.O.C. | | |
| | | Phone: +886-2-85121568 | | |
| | | Establishment Registration No.: 9617543 | | |
| | Contact Person | Anita Chen /Amber Dong | | |
| | Phone: | +886-939855759/ +8862-8512-1568 #5306 | | |
| | Fax: | +886-2-85121347 | | |
| | E-mail: | M9104303@gmail.com /Amber_dong@avita.com.tw | | |
| | Date Prepared | 2018.01.05 | | |
| 2 | Device Name | | | |
| | Proprietary Name: | Wrist Type Blood Pressure Monitor | | |
| | Common or usual name | Blood Pressure Monitor | | |
| | Product Code | DXN | | |
| | Device | Blood Pressure Monitor | | |
| | CFR Classification | CFR Part 870.1130 | | |
| | Device Class | II | | |
| | Classification Panel | Cardiovascular | | |
| 3 | Predicate Device Name | | | |
| | 510(k) number: | K102624 | | |
| | Trade or proprietary or<br>model name: | AVITA BPM16 Wrist Type Blood PressureMonitor | | |
| | Manufacturer: | AVITA Corporation | | |
| | 510(k) number: | K182166 | | |
| | Trade or proprietary or<br>model name: | BP4350 Automatic Wrist Blood Pressure Monitor | | |
| | Manufacturer: | Omron Healthcare, Inc.. | | |
{4}------------------------------------------------
| 4. Device Description: | Blood Pressure Monitor is a device intended for use in<br>automatically measures human's Systolic, Diastolic blood<br>pressure and heart rate. |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5. Intended Use: | BPM16B automatically measures human's Systolic, Diastolic<br>blood pressure and heart rate by using the oscillometric method<br>during inflation. All values can be read out in one LCD panel.<br>Measurement position is at human being's wrist. The intended<br>use of th years and older with wrist circumference ranging 125<br>~ 210 mm (approx. 4.9 ~ 8.3 inch) and for home use. |
| Special Conditions for<br>Use Statement(s): | For patient only |
- Technological 6. Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the Product name is substantially equivalent to the predicate device as summarized in Table 1. The differences raise no new question of safety and effectiveness.
| Table I predicate device | | Predicate Devices | Predicate Devices | | |
|-----------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| | Comparison table<br>between test | | | | |
| SE<br>Comparisons | Subject(AViTA<br>BPM16B Wrist Type<br>Blood Pressure<br>Monitor) | AViTA BPM16 Wrist<br>Type Blood Pressure<br>Monitor<br>K102624 | Predicate (BP4350<br>Automatic Wrist<br>Blood Pressure<br>Monitor)<br>K182166 | Comment | |
| Classification | 21CFR 870.1130 | 21CFR 870.1130 | 21CFR 870.1130 | Same | |
| Product Code | DXN | DXN | DXN | Same | |
| FDA Class | II | II | II | Same | |
| | Comparison table<br>between test | Predicate Devices | Predicate Devices | Predicate Devices | |
| | Intended Use | BPM16B device<br>automatically<br>measures systolic<br>and diastolic blood<br>pressure and pulse<br>rate by the<br>oscillometric<br>method. The<br>measurement<br>position is at the<br>wrist. The device<br>is intended to be<br>used by adults<br>with a wrist<br>circumference<br>ranging from 125<br>mm to 210 mm<br>(approx. 4.9 ~ 8.3<br>inches). The<br>device is intended<br>for home use.<br>When the device<br>detects irregular<br>heartbeats during<br>measurement, an<br>irregular heartbeat<br>symbol will<br>appear along with<br>the measured<br>readings. | The AViTA BPM<br>lx Series device<br>automatically<br>measures systolic<br>and diastolic blood<br>pressure and pulse<br>rate by the<br>oscillometric<br>method. The<br>measurement<br>position is at the<br>wrist. The device<br>is intended to be<br>used by adults<br>with a wrist<br>circumference<br>ranging from 125<br>mm to 210 mm<br>(approx. 4.9 ~ 8.3<br>inches). The<br>device is intended<br>for home use.<br>When the device<br>detects irregular<br>heartbeats during<br>measurement, an<br>irregular heartbeat<br>symbol will appear<br>along with the<br>measured<br>readings. | The device is a<br>digital monitor<br>intended for use in<br>measuring blood<br>pressure and pulse<br>rate in adult<br>patient population<br>with wrist<br>circumference<br>ranging from 5.3<br>inches to 8.5<br>inches (13.5cm to<br>21.5cm). The<br>device detects the<br>appearance of<br>irregular<br>heartbeats<br>during<br>measurement and<br>gives a warning<br>signal with<br>readings. | Same as The AViTA<br>BPM 1x Series |
| | Method of<br>measurement | Oscillimetric | Oscillimetric | Oscillimetric | Same |
| | Measurement<br>Type | During inflation | During deflation | During inflation | Same as BP4350 |
| | Range of<br>measurement | Rated Range of Cuff<br>Pressure: 0~<br>300mmHg,<br>Rated Range of<br>Determination:<br>40~255mmHg,<br>Pulse 40~199<br>Beats/minute | Rated Range of Cuff<br>Pressure: 0~<br>300mmHg,<br>Rated Range of<br>Determination:<br>30~280mmHg,<br>Pulse 40~199<br>Beats/minute | Rated Range of Cuff<br>Pressure: 0~<br>299mmHg,<br>Rated Range of<br>Determination:<br>40~260mmHg,<br>Pulse 40~180<br>Beats/minute | Similar |
| | Accuracy | Pressure ± 3mmHg<br>Pulse ± 4% | Pressure ± 3mmHg<br>Pulse ± 4% | Pressure ± 3mmHg<br>Pulse ± 5% | Same as The AViTA<br>BPM 1x Series |
| | Inflation | Automatic | Automatic | Automatic | Same |
| | Comparison table<br>between test | Predicate Devices | Predicate Devices | Predicate Devices | |
| Deflation | Automatic | Automatic | Automatic | Same | |
| Pressure<br>Changed<br>Rate | 2-5 mmHg/sec. | 2-5 mmHg/sec. | 2-5 mmHg/sec. | Same | |
| Display | Liquid Crystal<br>Digital | Liquid Crystal<br>Digital | Liquid Crystal<br>Digital | Same | |
| Power<br>Supply | 2 "AAA(LR03)(1.5<br>V)"<br>Alkaline Batteries | 2 "AAA(LR03)(1.5<br>V)"<br>Alkaline Batteries | 2 "AAA(LR03)(1.5<br>V)"<br>Alkaline Batteries | Same | |
| Storage/<br>Transportatio<br>n<br>Environment | - 20°C ~ + 50°C<br>≤ 85% R.H. | - 20°C ~ + 50°C<br>≤ 85% R.H. | - 20°C ~ + 60°C,<br>15% ~ 90% R.H. | Same as The AVITA<br>BPM lx Series | |
| Operating<br>Environment | 10°C ~ 40°C<br>≤ 85% R.H.<br>700~1060 hPa | 10°C ~ 40°C<br>≤ 85% R.H.<br>700~1060 hPa | 10°C ~ 40°C<br>15% ~ 90% R.H.<br>800~1060 hPa | Same as The AVITA<br>BPM lx Series | |
| Material | ABS housing | ABS housing | ABS housing | Same | |
| Sets<br>memory | of 1*90 | 2*90, total 180 | 2*100, total 200 | Similar | |
| Number of<br>Push Button | 2 keys | 4 keys | 5 keys | Similar | |
| Storage<br>pouch | Yes | Yes | Yes | Same | |
| Cuff size | Wrist circumference<br>approx.<br>125 ~ 210 mm<br>(Approx. 4.9~8.3<br>inches) | Wrist circumference<br>approx.<br>125 ~ 210 mm<br>(Approx. 4.9~8.3<br>inches) | Wrist circumference<br>approx.<br>135 ~ 215 mm<br>(approx. 5.3 ~ 8.4<br>inches) | Same as The AVITA<br>BPM lx Series | |
| Unit Weight | approx.<br>106g<br>(exclude batteries) | approx.<br>120g<br>(exclude batteries) | Approx. 91 g | Similar | |
Table 1 predicate device
AViTA Corporation
{5}------------------------------------------------
Section 5. 510(k) Summary
{6}------------------------------------------------
# 7. Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the device. Compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device.
# 8. Non-Clinical Tests Performed:
- a. EMC Test: IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests
{7}------------------------------------------------
- b. Radio Frequency Wireless Test: The EUT was performed according to FCC Part 15 Subpart C Section 15.247 procedure and setup followed by ANSI C63.10.2013 requirements.
- c. Safety Test:
-IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
-IEC 60601-1-11:, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance - Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment
- d. Biocompatibility testing
The biocompatibility evaluation and testing of the Product name was conducted in accordance with the following standards and guidance, as recognized by the FDA:
- FDA Draft Guidance Use of International Standard ISO- 10993, "Biological Evaluation . of Medical Devices, Part 1: Evaluation and Testing".
- . ISO 10993-5, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- . 10993-10, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11, Biological evaluation of medical devices- Part 11: Tests for systemic toxicity.
- Performance Test: e.
IEC 80601-2-30 Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
- f. Software Verification and Validation: IEC 62304 standard and FDA Guidance for the Content of Pre-Market Submission for Software Contained in Medical Devices standard.
- g. Usability & risk management:
IEC 60601-1-6: Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance -Collateral Standard: Usability.
IEC 62366: Medical Devices-Application of usability engineering to medical device.
EN ISO 14971 - Medical devices - Application of risk management to medical devices
## 9. Conclusion
Based on the intended use and/or indications for use, technological characteristics, performance testing and comparison to the predicate device, the Product name is substantially equivalent to the predicate device and raises no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.