K200323 · Radformation, Inc. · QKB · Oct 30, 2020 · Radiology
Device Facts
Record ID
K200323
Device Name
AutoContour
Applicant
Radformation, Inc.
Product Code
QKB · Radiology
Decision Date
Oct 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
AutoContour is intended to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning.
Device Story
AutoContour assists radiation treatment planners in contouring medical images for radiation therapy. System comprises three components: Windows-based agent service monitoring network storage for new CT datasets; cloud-based automatic contouring service; and web-based application for manual registration, review, editing, and export of structure sets. Input consists of CT images; output is DICOM RTSTRUCT file. Device used in clinical settings by radiation oncology personnel. Machine learning algorithms generate initial contours; user reviews and edits output. Output facilitates radiation treatment planning by automating structure delineation, potentially improving workflow efficiency and consistency.
Clinical Evidence
No clinical trials performed. Bench testing only. Verification and validation testing conducted per internal protocol to ensure software performance against requirements. Machine learning model validated on independent datasets separate from training/validation sets to demonstrate generalizability.
Technological Characteristics
Software-only application; no energy delivered to patients. Components: Windows-based agent, cloud-based contouring service (Linux), and web-based application (Linux/modern browsers). Connectivity via TCP/IP. Machine learning-based contouring algorithm. Supports CT input for contouring; CT, MR, PET for registration. Outputs DICOM RTSTRUCT.
Indications for Use
Indicated for radiation treatment planners assisting in contouring and reviewing structures within medical images for radiation therapy planning; applicable to any patient type for whom relevant modality scan data is available.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 30, 2020
Radformation, Inc. % Mr. Kurt Sysock Co-founder/CEO 335 Madison Avenue, 16th Floor NEW YORK NY 10017
Re: K200323
Trade/Device Name: AutoContour Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: September 19, 2020 Received: September 22, 2020
Dear Mr. Sysock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K200323
Device Name AutoContour
Indications for Use (Describe)
AutoContour is intended to assist radiation treatment planners in contouring structures within medical images in preparation for radiation therapy treatment planning.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5. 510(k) Summary
{4}------------------------------------------------
This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.
## 5.1. Submitter's Information
| Table 1 : Submitter's Information | |
|-----------------------------------|------------------------------------------------------|
| Submitter's Name: | Kurt Sysock |
| Company: | Radformation, Inc. |
| Address: | 335 Madison Avenue, 16th Floor<br>New York, NY 10017 |
| Contact Person: | Alan Nelson<br>Chief Science Officer, Radformation |
| Phone: | 518-888-5727 |
| Fax: | ---------- |
| Email: | anelson@radformation.com |
| Date of Summary Preparation | 9/16/2020 |
## 5.2. Device Information
| Table 2 : Device Information | |
|------------------------------|--------------------------------------------------------------------------------|
| Trade Name: | AutoContour |
| Common Name: | AutoContour, AutoContouring, AutoContour Agent,<br>AutoContour Web Application |
| Classification Name: | Class II |
| Classification: | Picture archiving and communications system |
| Regulation Number: | 892.2050 |
| Product Code: | QKB |
| Classification Panel: | Radiology |
# 5.3. Predicate Devices Information
Mirada RTx (K130393) & Workflow Box (K181572)
{5}------------------------------------------------
## 5.4. Device Description
AutoContour consists of 3 main components:
- 1. An "agent" service designed to run on the Windows Operating System that is configured by the user to monitor a network storage location for new CT datasets that are to be automatically uploaded to:
- 2. A cloud-based AutoContour automatic contouring service that produces initial contours and
- 3. A web application accessed via web browser which allows the user to perform registration with other image sets as well as review, edit, and export the structure set containing the contours.
#### 5.5. Indications for Use
AutoContour is intended to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning.
#### 5.6. Technological Characteristics
AutoContour (Subject Device) makes use of a Predicate Devices, Mirada RTx (K130393) & Workflow Box (K181572) for substantial equivalence comparison. Note that Workflow Box utilizes Mirada RTx as its predicate device.
## 5.6.1. AutoContour vs. Mirada RTx (K130393) & Workflow Box (K181572)
RTx is "intended to display and visualize 2D & 3D multi-modal medical image data" and "supports the loading and saving of DICOM RT objects and allows the user to define, import, display, transform, store and export such objects including regions of interest structure and dose volumes to radiation therapy planning systems" (https://www.accessdata.fda.gov/cdrh docs/pdf13/K130393.pdf, accessed 1/20/2020).
Workflow Box "is a system designed to allow users to route DICOM-compliant data to and from automated processing components... [including] processing components for automatically contouring imaging data using … machine learning based algorithms."
(https://www.accessdata.fda.gov/cdrh docs/pdf18/K181572.pdf, accessed 1/20/2020)
| Table 3: Substantial Equivalence AutoContour vs. RTx & Workflow Box | | | |
|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Subject Device:<br>AutoContour<br>Radformation | Predicate Device:<br>Workflow Box (K181572) | Predicate Device: Mirada<br>RTx (K130393) |
| Target<br>Population | Any patient type for<br>whom relevant modality<br>scan data is available. | Any patient type for<br>whom relevant<br>modality scan data is | Any patient type for whom<br>relevant modality scan data<br>is available. |
| | (Substantially<br>Equivalent) | available. | |
| Energy Used<br>and/or<br>Delivered | None - software only<br>application. The software<br>application does not<br>deliver or depend on<br>energy delivered to or<br>from patients<br>(Substantially<br>Equivalent) | None - software only<br>application. The<br>software application<br>does not deliver or<br> | None - software only<br>application. The software<br>application does not<br>deliver or depend on<br>energy delivered to or from<br>patients |
| Intended users | Trained radiation<br>oncology personnel<br>(Substantially<br>Equivalent) | Designed to be used by<br>trained clinicians | Designed to be used by<br>trained clinicians |
| Design: Data<br>Visualisation/Gr<br>aphical User<br>Interface | Contains both an<br>automated processing<br>component and Data<br>Visualisation / Graphical<br>User Interface<br>(Substantially<br>Equivalent) | None - the proposed<br>device has no data<br>visualization<br>functionality. All data<br>processing is<br>automated and does<br>not require user<br>interaction. A control<br>interface is provided for<br>system administration<br>and configuration only | Yes. |
| Design: View<br>manipulation<br>and Volume<br>rendering | Window and level, pan,<br>zoom, cross-hairs, slice<br>navigation, fused views.<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices) | None - Not applicable | Window and level, pan,<br>zoom, cross-hairs, slice<br>navigation. Maximum or<br>minimum intensity projection<br>(MIP), volume rendering,<br>color rendering, surface<br>rendering, multi-planar<br>reconstruction (MPR), fused<br>views, gallery views. |
| Design: Image<br>registration | Manual Rigid<br>registration.<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices) | Registration for the<br>purposes of<br>replanning/re-contouring<br>and atlas based<br>contouring. The<br>algorithms used for<br>image registration are<br>the same for both<br>predicate and proposed<br>devices. | Manual and Landmark<br>Rigid. Automatic multi-modal<br>rigid. Mono-modal and<br>multi-modal deformable<br>registration. Motion correction<br>in hybrid scans and gated<br>scans. Registration for the<br>purposes of<br>re-planning/recontouring and<br>atlas based contouring. |
| Regions and<br>Volumes of<br>interest (ROI) | Machine learning based<br>contouring and manual<br>ROI manipulation.<br>(Subject Device | Atlas Based contouring,<br>registration based<br>re-contouring, machine | 2D and 3D ROIs,<br>semi-automatic ROI<br>definition, isocontour ROIs<br>using threshold and |
| | functionality is a subset<br>of Predicate Devices) | learning based<br>contouring | percentage of maximum,<br>one-click seed-pointing<br>contouring, <b>manual ROI</b><br><b>manipulation</b> , <b>ROI</b><br>transformation, Atlas-based<br>contouring. |
| Design:<br>Region/volume<br>of interest<br>measurements<br>and size<br>measurements | None - not applicable<br>(Substantially<br>Equivalent) | None - not applicable | Intensity, Hounsfield units,<br>activity and SUV<br>measurements including min,<br>max, mean, peak, standard<br>deviation, total glycolytic<br>activity, median, histogram,<br>max and mean ratio to<br>reference region. Gray for RT<br>Dose. Size measurements<br>include 2D and 3D<br>measurements including<br>rulers and volume, line<br>profile. |
| Design:<br>Region/Volume<br>Quantification | None - not applicable<br>(Substantially<br>Equivalent) | None - not applicable | Regions table with charting<br>supports analysis of<br>measurement over multiple<br>studies using standard<br>protocols such as RECIST,<br>PERCIST and WHO |
| Design:<br>Supported<br>modalities | CT input for contouring<br>or manual<br>registration/fusion.<br>MR, PET input for<br>manual<br>registration/fusion only.<br>DICOM RTSTRUCT for<br>output<br>(Minor differences) | <b>CT, MR, DICOM</b><br><b>RTSTRUCT</b> for image<br>processing. Any valid<br>DICOM data for data<br>routing | Static and gated <b>CT and</b><br><b>PET</b> , and static MR, SPECT,<br>NM, DICOM RT |
| Design:<br>Reporting and<br>data routing | No built-in reporting,<br>supports exporting<br>DICOM RTSTRUCT file.<br>(Minor differences) | Supports routing and<br>distribution of images to<br>other DICOM nodes<br>including to custom<br>executables determined<br>by the user. | Yes- Distribution of DICOM<br>compliant Images into other<br>DICOM compliant systems.<br>Built-in basic reporting |
| Compatibility<br>with the<br>environment<br>and other<br>devices | Compatible with data<br>from any DICOM<br>compliant scanners for<br>the applicable modalities.<br>Agent Uploader<br>component compatible<br>with Microsoft Windows. | Compatible with data<br>from any DICOM<br>compliant scanners for<br>the applicable<br>modalities. Compatible<br>with <b>Microsoft</b><br><b>Windows</b> . Integration<br>with Mirada DBx | Compatible with data from<br>any DICOM compliant<br>scanners for the applicable<br>modalities. Compatible<br>with <b>Microsoft Windows</b> .<br>Integration with Mirada DBx<br>application launcher and data<br>browser. |
| | Cloud-based automatic<br>contouring service<br>compatible with Linux.<br><br>Web application Server<br>based application<br>compatible with Linux<br>with frontend compatible<br>with all modern web<br>browsers..<br>(Minor differences) | application launcher and<br>data browser | |
| Communication<br>s/Networking | TCP/IP<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices) | TCP/IP and SCP | TCP/IP and SCP |
| Computer<br>platform &<br>operating<br>system | Agent Uploader<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows<br>Server 2016.<br><br>Cloud-based Server<br>based automatic<br>contouring application<br>compatible with Linux.<br><br>Web application Server<br>based application<br>compatible with Linux<br>with frontend compatible<br>with all modern web<br>browsers.<br>(Minor differences) | Server based application<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows<br>Server 2016. | Workstation and Server<br>based application supporting<br>Windows Server 2008 R2,<br>SP1 and Windows 7 (64-bit) |
| Support for<br>radiation<br>treatment<br>planning | AutoContour is intended<br>to assist radiation<br>treatment planners in<br>contouring and reviewing<br>structures within medical<br>images in preparation for<br>radiation therapy<br>treatment planning<br>(Subject Device intended<br>use is a subset of<br>predicate devices) | "Contours generated by<br>Workflow Box may be<br>used as an input to<br>clinical workflows<br>including, but not limited<br>to, radiation therapy<br>treatment planning" | "RTx supports the loading<br>and saving of DICOM RT<br>objects and allows the user to<br>define, import, display,<br>transform, store and export<br>such objects including<br>regions of interest structures<br>and dose volumes to<br>radiation therapy planning<br>systems." |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
Source for RTx & Workflow Box: (https://www.accessdata.fda.gov/cdrh_docs/pdf18/K181572.pdf, accessed 1/20/2020)
> As shown in Table 3, almost all technological characteristics are either substantially equivalent or a subset of the predicate devices' technological characteristics.
{9}------------------------------------------------
#### 5.7. Discussion of differences
#### Subset of the Predicate Device
The comparison table above shows that several features of AutoContour are a subset of the features of the predicate device. This is because AutoContour's indication for use is narrower in scope than the predicate device in that it is intended only "to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning". The design of AutoContour reflects that narrower scope and the features not shared between AutoContour are such that they can operate independently of the features that are not shared with the predicate devices, and therefore these differences do not create new questions regarding the safety and effectiveness of the device relative to the predicate devices.
#### Minor differences
The following minor differences exist, but do not represent any significant additional risks or decreased effectiveness for the device for its intended use:
- . Design: Supported modalities: AutoContour only supports CT image sets as inputs specifically for the contouring feature, while MR and PET image set inputs are only supported for registration purposes. In the predicate devices, all modalities are available for the workflows. This does not affect the safety and effectiveness of the device as it relates to its indications for use as the vast majority of radiation treatment planning is performed on CT images, and where MR or PET modalities are necessary for localization, AutoContour does support manual registration with those modalities which can be used to guide the contouring of the CT image set.
- . Design: Reporting and data routing: AutoContour does not include reporting features or direct data-routing with DICOM compliant systems instead AutoContour simply produces a DICOM compliant RT Structure Set file that the user may import into a DICOM compliant treatment planning system. This does not raise new questions regarding the safety and effectiveness of the device since the scope of AutoContour's indication for use is limited to making the contoured structures available for radiation treatment planning purposes and does not include managing or routing of the image data to various systems. The feature of managing or routing of the image data in this way is not a necessary component to achieve this goal, so not having this feature does not affect the safety or effectiveness of the device.
- Compatibility with the environment and other devices / Computer platform & operating system: AutoContour utilizes different computer platform & operating system for its 3 components, but this does not raise new questions regarding the safety and effectiveness of the device relative to the predicate device.
{10}------------------------------------------------
### 5.7. Performance Data
As with the Predicate Devices, no clinical trials were performed for AutoContour. Nonclinical tests were performed according to Radformation's AutoContour Complete Test Protocol and Report, which demonstrates that AutoContour performs as intended per its indications for use. Further tests were performed on independent datasets from those included in training and validation sets in order to validate the generalizability of the machine learning model.
Verification and validation tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.
#### 5.8. Conclusion
AutoContour is deemed substantially equivalent to the Predicate Devices, Mirada RTx (K130393) & Workflow Box (K181572). Verification and Validation testing and Hazard Analysis demonstrate that AutoContour is as safe and effective as the Predicate Devices. The minor technological differences between AutoContour and the Predicate Devices with regard to the intended use do not raise any questions on the safety and effectiveness of the Subject Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.