Altus Spine Pedicle Screw System

K200322 · Altus Partners, LLC · NKB · Jun 2, 2020 · Orthopedic

Device Facts

Record IDK200322
Device NameAltus Spine Pedicle Screw System
ApplicantAltus Partners, LLC
Product CodeNKB · Orthopedic
Decision DateJun 2, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Altus Spine Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The Altus Spine Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The Altus Spine Pedicle Screw System may be used for noncervical pedicle fixation via in posterior percutaneous approach with MIS instrumentation for the indications listed above. When used as an anterolateral thoracic/umbar system the Altus Spine Pedicle Screw System may also be used for the same indications listed above as an adjunct to fusion.

Device Story

System of implantable screws, rods, and connectors; used to stabilize thoracic, lumbar, and sacral spinal segments during fusion. Surgeon implants components into vertebral body via posterior percutaneous or anterolateral approach. Provides rigid construct to immobilize spinal segments; facilitates bone graft fusion. Used in clinical settings by surgeons. Non-sterile; requires hospital sterilization.

Clinical Evidence

Bench testing only. Performed static compression bending, static torsion, and dynamic compression bending per ASTM F1717-15. Results demonstrate substantial equivalence to the predicate device.

Technological Characteristics

Materials: Titanium alloy (ASTM F136) and Cobalt Chrome (ASTM F1537). Components: Pedicle screws, rods, rod-to-rod connectors, lateral connectors, dominos, cross connectors. Non-sterile, hospital-sterilized. Mechanical fixation system.

Indications for Use

Indicated for skeletally mature patients requiring noncervical pedicle fixation for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion. Used as an adjunct to spinal fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ June 2, 2020 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that. Altus Partners, LLC Mr. Mark Melton QA/RA Manager 1340 Enterprise Drive West Chester, Pennsylvania 19380 Re: K200322 Trade/Device Name: Altus Spine Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWO Dated: March 23, 2020 Received: March 26, 2020 Dear Mr. Melton: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ## K200322 Device Name Altus Spine Pedicle Screw System ### Indications for Use (Describe) The Altus Spine Pedicle Screw System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine. The Altus Spine Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion in skeletally mature patients. The Altus Spine Pedicle Screw System may be used for noncervical pedicle fixation via in posterior percutaneous approach with MIS instrumentation for the indications listed above. When used as an anterolateral thoracic/umbar system the Altus Spine Pedicle Screw System may also be used for the same indications listed above as an adjunct to fusion. ### Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | SUBMITTER: | Altus Partners, LLC<br>1340 Enterprise Drive<br>West Chester, PA 19380<br>Phone: 610-356-6148<br>Fax: 610-300-3049 | |----------------------|-----------------------------------------------------------------------------------------------------------------------------| | CONTACT PERSON: | Mark Melton<br>Regulatory Affairs Specialist<br>mmelton@-spine.com | | DATE PREPARED: | May 21, 2020 | | COMMON NAME: | Pedicle Screw Spinal System | | PROPRIETARY NAME: | Altus Spine Pedicle Screw System | | PREDICATE DEVICES: | The Primary predicate device is: Monaco<br>Pedicle Screw System (K181339) | | CLASSIFICATION NAME: | 21 CFR §888.3070 Thoracolumbosacral Pedicle Screw System | | PRODUCT CODES: | NKB, KWQ | | DEVICE CLASS: | Class II | | MATERIAL: | The materials used are Titanium Alloy material that conforms to ASTM<br>F136 and Cobalt Chrome that conforms to ASTM F1537. | ### DEVICE DESCRIPTION: The Altus Spine Pedicle Screw System consists of a system of implantable screws to be used with implantable rods for the purpose of aiding in spinal fusion. The Altus Spine Pedicle Screw System attaches to the vertebral body by means of screws to the non-cervical spine. Which allows a surgeon to build a spinal implant construct with the intent to stabilize the spinal operative site during the fusion process of bone graft in the disc space. Implantable components are Rod-to-Rod Connectors, Lateral Connectors, Dominos and Cross Connectors. All composed of titanium alloy meeting the requirements of ASTM F136 and cobalt chrome meeting the requirements of ASTM F1537. The device is supplied non-sterile and is intended for sterilization by hospital personnel. #### INDICATIONS FOR USE: The Altus Spine Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine. The Altus Spine Pedicle Screw System is intended for noncervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origins with deneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients. {4}------------------------------------------------ The Altus Spine Pedicle Screw System may be used for noncervical pedicle fixation via in posterior percutaneous approach with MIS instrumentation for the indications listed above. When used as an anterolateral thoracic/lumbar system the Altus Spine Pedicle Screw System may also be used for the same indications listed above as an adjunct to fusion. ### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS: The Altus Spine Pedicle Screw System is the same as the primary predicate in regards to implant Indications for use/intended use, technological characteristics, materials and surgical technique. #### SUMMARY OF NON-CLINAL TESTS SUBMITTED: Tests performed were Static Compression Bending, Static Torsion, and Dynamic Compression Bending per ASTM F1717-15. The testing demonstrated that the Altus Spine Pedicle Screw System is substantially equivalent to the predicate device. ### SUBSTANTIAL EQUIVALENCE CONCLUSION: The Altus Spine Pedicle Screw System is the same as the primary predicate (K181339) in regards to implant materials and surgical technique. Altus Partners has determined that the modifications to the Altus Spine Pedicle Screw System do not alter the system function, strength and stability or materials. Therefore, the Altus Spine Pedicle Screw System is substantially equivalent to the predicate devices, and raises no new questions of safety or effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...