K200314 · Vascular Medcure, Inc. · QEW · Mar 4, 2020 · Cardiovascular
Device Facts
Record ID
K200314
Device Name
Capere Thrombectomy System
Applicant
Vascular Medcure, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Mar 4, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CAPERE® Thrombectomy System is indicated for: - · Non-surgical removal of soft emboli and thrombi from blood vessels. - · Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
Device Story
Percutaneous embolectomy system; used in peripheral vasculature; operated by physician via transfemoral or jugular venous access. System comprises 12Fr delivery catheter and 15Fr guide catheter. Mechanism: nitinol wire basket deployed at therapy site to capture emboli/thrombi; basket retracted and extended to surround clot; clot-loaded delivery catheter retracted into guide catheter funnel for removal. Side port allows fluid injection/aspiration. Device provides mechanical clot removal without primary aspiration. Benefits: non-surgical intervention for thrombus clearance.
Clinical Evidence
Bench testing only. Biocompatibility confirmed via material equivalence to predicate. Functional testing (ISO 10555-1, ISO 594-2) and simulated use testing demonstrated successful clot retrieval without basket rupture or catheter damage. Packaging integrity and shelf-life supported by predicate data.
Technological Characteristics
Components: 12Fr delivery catheter, 15Fr guide catheter, nitinol wire basket (18mm diameter, 11.4cm length). Materials identical to predicate. Principle: mechanical capture via nitinol basket. Connectivity: none. Sterilization: EO. Standards: ISO 10555-1, ISO 594-2.
Indications for Use
Indicated for non-surgical removal of soft emboli and thrombi from peripheral vasculature and injection/infusion/aspiration of fluids/contrast media. Contraindicated for use in pulmonary arteries.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 4, 2020
Vascular Medcure, Inc. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1735 Market Street, Suite 2300 Philadelphia, Pennsylvania 19103
Re: K200314
Trade/Device Name: Capere Thrombectomy System Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: February 6, 2020 Received: February 6, 2020
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200314
Device Name CAPERE® Thrombectomy System
| <b>Indications for Use (Describe)</b> |
|---------------------------------------|
|---------------------------------------|
The CAPERE® Thrombectomy System is indicated for:
- · Non-surgical removal of soft emboli and thrombi from blood vessels.
- · Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel.
The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K200314 - 510(k) SUMMARY
## Vascular Medcure, Inc.'s CAPERE® Thrombectomy System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Jeffrey P. DuMontelle
Vascular Medcure, Inc.
1500 S. Sunkist Street, Suite H
Anaheim, CA 92806
Office: 657-549-5175
714-915-0886 Cell:
Date Prepared: February 20, 2020
#### Name of Device
CAPERE® Thrombectomy System
#### Common or Usual Name
Embolectomy Catheter
#### Classification
21 CFR 870.5150, Class II, product code QEW, KRA
#### Predicate Devices
CAPERE® Thrombectomy System (K180722)
## Intended Use / Indications for Use
The CAPERE® Thrombectomy System is indicated for:
- Non-surgical removal of soft emboli and thrombi from blood vessels. o
- Injection, infusion and/or aspiration of contrast media and other fluids into blood vessel. O
The CAPERE® Thrombectomy System is intended only for use in the peripheral vasculature and is not intended for use in the pulmonary arteries.
#### Device Description
The CAPERE® Thrombectomy System primarily consists of an 12Fr Delivery Catheter and 15Fr Guide Catheter. The CAPERE® Thrombectomy System is delivered percutaneously via transfemoral or jugular venous access. Once delivered, the System's fine mesh nitinol wire basket is used to capture and mechanically remove emboli and thrombi. The CAPERE® System does not use aspiration to pull out the thrombus but does have a side port in the funnel catheter that allows aspiration or injection of saline or fluids if needed.
{4}------------------------------------------------
## Technological Characteristics
The CAPERE® Thrombectomy System has the same technological characteristics as its Both CAPERE® predicate device CAPERE Thrombectomy Catheter (K180722). Thrombectomy Systems have either the same components or substantially equivalent components (length change). Each device system includes a guide catheter, delivery catheter, and a mechanism for capturing/removing the soft emboli or thrombi. The predicate and CAPERE® Thrombectomy System both use a 12Fr delivery catheter. The CAPERE® Thrombectomy System utilizes a15Fr guide catheter while the predicate utilizes a 20Fr guide catheter. In addition, both systems use 0.035" quidewires during the procedure.
The CAPERE® System includes a Guide Catheter and a Delivery Catheter. In the same manner as the previously cleared device, the Guide Catheter and Delivery Catheter are advanced to the therapy site. The Guide Catheter is unsheathed, and the Delivery Catheter is then advanced passed the obstruction. A nitinol basket connected to the Delivery Catheter is deployed by retracting the outer sheath of the catheter.
To capture the soft emboli or thrombi ("clot"), the deploved basket is retracted while simultaneously being extended to surround and capture the length of the clot. The Delivery Catheter and clot-load is then retracted into the funnel of the Guide Catheter where the clot is captured and can be removed.
The basket design in the CAPERE® Thrombectomy Systems is exactly the same (nitinol, 18mm diameter, 11.4cm length) as the basket design in the predicate.
Therefore, the subject CAPERE® Thrombectorny System has the same technological characteristics as its predicate.
## Performance Data
The following nonclinical performance testing has been conducted, using the established test methods used in the cleared predicate device, to support the substantial equivalence of the CAPERE® Thrombectomy System to its predicate device. In all instances, the CAPERE® Thrombectomy System functioned as intended under the previously established test methods.
- . Biocompatibility of the patient-contacting components of the device was confirmed as there are no differences between the materials in the CAPERE® Thrombectomy System and the predicate.
- . Packaging of the CAPERE® Thrombectomy System is the same as the predicate, therefore, the predicate package integrity and accelerated aging studies demonstrate that the packaging was capable of maintaining a sterile barrier and protecting the device after 2X EO sterilization, simulated handling and shipping, environmental conditioning, and 6-months of accelerated aging in support of a 6-month shelf-life.
- . Functional bench testing was conducted (including demonstrated compliance with relevant standards such as ISO 10555-1 and ISO 594-2) on the bond joints affected by the reduced diameter of the CAPERE® Thrombectomy System with respect to the predicate and demonstrated that the CAPERE® Thrombectomy System maintained the predicate characteristics.
- . Simulated Use Testing was completed to demonstrate that simulated clot can be retrieved under simulated conditions without capture basket rupture, catheter
{5}------------------------------------------------
damage, or other adverse device effect and confirmed that the predicate characteristics were maintained.
## Substantial Equivalence
The CAPERE® Thrombectomy System has the same intended use and indications for use, technological characteristics, and principles of operation as its predicate devices. Nonclinical testing, including functional testing and simulated use testing demonstrated that the minor differences between the device and predicate (i.e. device diameter and lengths) do not raise new types of safety or effectiveness questions. Thus, the CAPERE® Thrombectomy System is substantially equivalent to the predicate devices.
#### Conclusion
Vascular Medcure's CAPERE® is an Embolectomy Catheter, Class II device that has been evaluated in nonclinical testing in accordance with FDA's recognized standards and preestablished acceptance criteria. Testing demonstrated that the device performs as intended. The CAPERE® is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.