K200305 · Guard Medical, Inc. · OKO · May 26, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200305
Device Name
NPseal
Applicant
Guard Medical, Inc.
Product Code
OKO · General, Plastic Surgery
Decision Date
May 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4683
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
Device Story
NPseal is a single-use, non-powered negative pressure wound therapy (NPWT) system for closed surgical incisions up to 5 cm x 0.5 cm. Device maintains negative pressure in -75 mmHg to -125 mmHg range. System includes integrated mechanical pump and high moisture vapor transmission rate (MVTR) breathable polyurethane film dressing. Intended for 3-day wear, with maximum 6-day total duration if replaced once. Used in clinical or home settings; operated by healthcare providers or patients. Device removes exudate from incision site to promote healing. Output is mechanical suction; provider monitors wound status and exudate levels to guide therapy. Benefits include simplified wound management and potential for improved healing outcomes.
Clinical Evidence
Bench testing only. Performance data included biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation, systemic toxicity, implantation), pressure/exudate handling, foam characterization, adhesion, peel testing, MVTR, shelf life, and benchtop usability. Human factors user validation study performed.
Technological Characteristics
Non-powered mechanical suction pump; breathable polyurethane film dressing with high MVTR. Operating pressure: -75 to -125 mmHg. Sterile (gamma irradiation). Single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring wound management via negative pressure to promote healing by removing small amounts of exudates from closed surgical incisions.
Regulatory Classification
Identification
A non-powered suction apparatus device intended for negative pressure wound therapy is a device that is indicated for wound management via application of negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. It is further indicated for management of wounds, burns, flaps, and grafts.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Non-powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy (NPWT).” See § 878.1(e) for the availability of this guidance document.
PICO Single Use Negative Pressure Wound Therapy System (K180618)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 26, 2020
Guard Medical Inc. % Eric Bannon Consultant AlvaMed. Inc. 935 Great Plain Avenue, #166 Needham, Massachusetts 02492
Re: K200305
Trade/Device Name: NPseal Regulation Number: 21 CFR 878.4683 Regulation Name: Non-Powered Suction Apparatus Device Intended for Negative Pressure Wound Therapy Regulatory Class: Class II Product Code: OKO Dated: May 1, 2020 Received: May 4, 2020
Dear Eric Bannon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200305
Device Name NPseal
Indications for Use (Describe)
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
#### I. SUBMITTER
Guard Medical, Inc. 10 East 40th Street Suite 3310 New York, NY 10016
Phone: (888) 417-3644 Fax: (617) 249-0955
Contact Person: Eric Bannon
Date Prepared: May 18, 2020
#### II. DEVICE
Name of Device: NPseal
Common or Usual Name: Negative Pressure Wound Therapy Non-powered Suction Apparatus
Classification Name: Non-Powered suction apparatus device intended for negative pressure wound therapy. (21 CFR 878.4683)
Regulatory Class: II
Product Code: OKO
#### III. PREDICATE & REFERENCE DEVICES
Predicate Device: SNaP Wound Care System, K151710
Reference Device: PICO Single Use Negative Pressure Wound Therapy System, K180618
#### IV. DEVICE DESCRIPTION
The NPseal Negative Pressure Advanced System is a single-use device that includes an integrated, mechanical pump system. The NPseal maintains Negative Pressure Wound Therapy (NPWT) in the -75 mmHq to -125 mmHq nominal range.
The NPseal is intended for 3 days of use. Therapy duration of the system may be less than indicated if clinical practice or other factors such as wound size, rate or volume of exudate, or orientation of the dressing results in earlier removal or need for system change. The NPseal can be replaced only one time for a total maximum wear time of 6 days.
The NPseal is intended for surgically closed incisions up to 5 cm x 0.5 cm.
#### V. INDICATIONS FOR USE
The NPseal is indicated for patients who would benefit from wound management via application of negative pressure, particularly as the device may promote wound healing through the removal of small amounts of exudates from closed surgical incisions.
{4}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE AND REFERENCE DEVICES
| | Subject Device:<br>NPseal | Predicate Device:<br>SnAP Wound Care System | Reference Device:<br>PICO Single Use Negative<br>Pressure Wound Therapy<br>System |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Guard Medical | Spiracur | Smith & Nephew |
| 510(k) Number | K200305 | K151710 | K180618 |
| Product Code | OKO | OKO | OMP |
| Regulation<br>Number | 878.4683 | 878.4683 | 878.4780 |
| Regulation<br>Description | Non-Powered suction<br>apparatus device intended<br>for negative pressure<br>wound therapy | Non-Powered suction<br>apparatus device intended for<br>negative pressure wound<br>therapy | Powered suction pump |
| Common<br>Name | Negative Pressure Wound<br>Therapy non-powered suction<br>apparatus | Negative Pressure Wound<br>Therapy non-powered suction<br>apparatus | Powered suction pump and<br>wound dressing kit |
| | Subject Device:<br>NPseal | Predicate Device:<br>SNaP® Wound Care System | Reference Device:<br>PICO Single Use Negative<br>Pressure Wound Therapy<br>System |
| Intended Use | The NPseal is indicated for<br>patients who would benefit<br>from wound management<br>via application of negative<br>pressure, particularly as the<br>device may promote wound<br>healing through the removal<br>of small amounts of<br>exudates from closed<br>surgical incisions. | The SNaP® Wound Care<br>System is indicated for<br>patients who would benefit<br>from wound management<br>via the application of<br>negative pressure,<br>particularly as the device<br>may promote wound healing<br>through the removal of<br>excess exudate, infectious<br>material and tissue debris.<br>The SNaP® Wound Care<br>System is indicated for<br>removal of small amounts of<br>exudate from chronic, acute,<br>traumatic, subacute and<br>dehisced wounds, partial-<br>thickness burns, ulcers<br>(such as diabetic, venous or<br>pressure), surgically closed<br>incisions, flaps and grafts. | PICO is indicated for<br>patients who would benefit<br>from a suction device<br>(negative pressure wound<br>therapy) as it may promote<br>wound healing via removal of<br>low to moderate levels of<br>exudate and infectious<br>materials.<br>Appropriate wound types<br>include:<br>- Chronic<br>- Acute<br>- Traumatic<br>- Subacute and dehisced<br>wounds<br>- Partial-thickness burns<br>- Ulcers (such as diabetic or<br>pressure)<br>- Flaps and grafts<br>- Closed surgical incisions<br>PICO Single Use Negative<br>Pressure Wound Therapy<br>System is suitable for use<br>both in a hospital and<br>homecare setting. |
| Sterility | Sterile - Gamma irradiation | Sterile - Gamma irradiation | Sterile - EO |
| Single Use | Yes | Yes | Yes |
| Wear Time | 3 days | 7 days | 7 days |
| Dressing<br>film(s) | High Moisture Vapor<br>Transmission Rate (MVTR),<br>breathable polyurethane film | Single layer hydrocolloid<br>dressing, non-breathable | High Moisture Vapor<br>Transmission Rate (MVTR)<br>upper film, perforated silicone<br>adhesive skin layer |
| | Subject Device:<br>NPseal | Predicate Device:<br>SnaP Wound Care System | Reference Device:<br>PICO Single Use Negative<br>Pressure Wound<br>Therapy System |
| Negative<br>Pressure<br>Operating<br>Ranges | -75 to -125 mmHg | -75, -100, or -125 mmHg | -80 mmHg |
{5}------------------------------------------------
{6}------------------------------------------------
The subject, predicate and reference devices are all NPWT dressing systems.
The subject and predicate devices are substantially equivalent, as evidenced by their shared characteristics:
- . Single use
- . Nonpowered
- Identical negative pressure operating ranges (-75 to -125 mmHg)
- Indicated for removal of small amounts of exudates. ●
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The biocompatibility evaluation for the NPseal was conducted in accordance with the international standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA along with the agency's quidance document. "Use of International Standard ISO 10993-1. 'Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process:' Guidance for Industry and Food and Drug Administration Staff" (issued June 16, 2016). The tests performed included the following:
- . Cytotoxicity
- . Sensitization
- Irritation or Intracutaneous Reactivity
- Acute Systemic Toxicity
- . Material-Mediated Pyrogenicity
- . Subacute Systemic Toxicity
- . Subchronic Systemic Toxicity
- . Implantation
#### Bench Testing
- . Pressure and exudate handling over time test
- · Curved surface pressure over time
- Foam characterization test .
- . Pump body adhesion test
- Dressing peel test .
- Exudate handling without negative pressure test .
- Moisture vapor transmission rate (MVTR) test .
{7}------------------------------------------------
- Shelf Life test .
- . Benchtop Usability Verification
- Set Negative Pressure test
#### Human Factors Testing
- User validation study .
# VIII. CONCLUSIONS
The test data support the safety of the device and the hardware verification and validation demonstrate that the NPseal shall perform as intended in the specified use conditions. The data demonstrate that the NPSeal is substantially equivalent to the predicate device with respect to the intended use and technological comparison. There were no new questions regarding safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.