Microlife Upper Arm Automatic Digital Blood Pressure Monitor
Applicant
Microlife Intellectual Property GmbH
Product Code
DXN · Cardiovascular
Decision Date
Jun 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor,Model WatchBP Office (BP3SK1-3B) is a noninvasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm. The device can accurately measure in pregnant patients including those with known or suspected preeclampsia. The device provides aortic blood pressure parameters. includes central systolic blood pressure (cSBP), central pulse pressure (cPP) and central diastolic blood pressure (cDBP), non-invasively through the use of a brachial cuff. The device detects the appearance of atrial fibrillation during measurement and gives a warning signal together with the measured blood pressure value if atrial fibrillation is detected. The memory data can be transferred to the PC (personal computer) running the WatchBP Analyzer software by connecting the monitor via USB cable or Bluetooth. The device is for hospital use only.
Device Story
Non-invasive digital blood pressure monitor; uses oscillometric technique with upper-arm cuff. Inputs: cuff pressure oscillations via resistive pressure sensor. Processing: converts pressure changes to electrical signals; calculates systolic/diastolic BP, pulse rate, MAP, and central BP parameters (cSBP, cPP, cDBP); detects atrial fibrillation. Outputs: BP values, pulse rate, central BP parameters, and AFIB warning signal on LCD. Used in hospital settings by clinicians. Data transfer via USB or Bluetooth to PC running WatchBP Analyzer software. Features AUTO mode (1-6 consecutive measurements) and MANUAL mode (auscultatory method using stethoscope). Benefits: provides automated BP monitoring, central BP assessment, and AFIB screening to assist clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing and leveraging clinical validation data from predicate devices (K152770 and K171937). The subject device shares identical core algorithms for BP, AFIB, and central BP measurement with the predicates.
Indicated for adults and pediatrics (excluding neonates) with arm circumferences 14-52 cm. Used for non-invasive measurement of systolic/diastolic BP, pulse rate, MAP, and aortic BP parameters (cSBP, cPP, cDBP). Indicated for use in pregnant patients, including those with known or suspected preeclampsia. Includes atrial fibrillation detection.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model ProBP 2400 (K152770)
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Central (TWIN200 CBP) (K171937)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
June 18, 2020
Microlife Intellectual Property GmbH % Vaibhav Rajal Official Correspondent for Microlife Intellectual Property, GmbH mdi Consultants, Inc. 55 Northern Blvd. Suite 200 Great Neck. New York 11021
Re: K200297
Trade/Device Name: Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (BP3SK1-3B) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 14, 2020 Received: May 20, 2020
Dear Vaibhav Rajal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200297
#### Device Name
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (BP3SK1-3B)
#### Indications for Use (Describe)
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor,Model WatchBP Office (BP3SK1-3B) is a noninvasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm.
The device can accurately measure in pregnant patients including those with known or suspected preeclampsia.
The device provides aortic blood pressure parameters. includes central systolic blood pressure (cSBP), central pulse pressure (cPP) and central diastolic blood pressure (cDBP), non-invasively through the use of a brachial cuff.
The device detects the appearance of atrial fibrillation during measurement and gives a warning signal together with the measured blood pressure value if atrial fibrillation is detected.
The memory data can be transferred to the PC (personal computer) running the WatchBP Analyzer software by connecting the monitor via USB cable or Bluetooth.
The device is for hospital use only.
Type of Use (Select one or both, as applicable)
| <span></span> | <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|---------------|-----------------------------------------------------------------------------------------------------|
| <span></span> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## VOL 4, 001
## 510(k) SUMMARY
#### The assigned 510(k) number is: K200297
#### 1. Submitter's Identification:
Microlife Intellectual Property GmbH, Switzerland Espenstrasse 139 9443 Widnau / Switzerland
### Date Summary Prepared: June 17, 2020
Contact: Mr. Gerhard Frick Vice President of Technical and Service Microlife Intellectual Property GmbH, Switzerland Tel: +41 79 216 0070 E-Mail: gerhard.frick@microlife.ch
#### 2. Name of the Device:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (BP3SK1-3B)
Regulation Number: 21 CFR Part 870.1130 Regulation Name: Non-Invasive Blood Pressure Measurement System Regulatory Class: II Product Code: DXN
#### 3. Information for the 510(k) Cleared Device (Predicate Device):
#### Primary Predicate:
- a. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model ProBP 2400, K152770, Microlife Intellectual Property GmbH.
#### Reference Predicate:
- b. Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office Central (TWIN200 CBP), K171937, Microlife Intellectual Property GmbH.
#### 4. Device Description:
Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (BP3SK1-3B) is designed to measure systolic and diastolic blood pressure, pulse rate, and mean arterial pressure (MAP) of the adults and pediatrics (but not neonates) populations with arm circumference sizes ranging from 14 -52 cm by
{4}------------------------------------------------
using a non-invasive technique in which one inflatable cuff is wrapped around the single upper arm. Our method to define systolic and diastolic pressure is similar to the auscultatory method but use a resistive pressure sensor rather than a stethoscope and mercury manometer. The sensor convert tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating pulse rate, mean arterial pressure (MAP), central systolic blood pressure (cSBP), central pulse pressure (cPP) and central diastolic pressure (cDBP) which is a well - known technique in the market called the "oscillometric method".
The device has two measurement modes that can be used :<< AUTO>> mode and << MANUAL>> mode. It has number of measurements setting function, resting time setting function, measurement intervals setting function, atrial fibrillation detection function, Central Blood Pressure (CBP) measurement function etc. In addition, the device can be used in connection with your personal computer (PC) running the WatchBP Analyzer software. The memory data can be transferred to the PC by connecting the monitor with the PC via USB cable or Bluetooth.
The measurement program in << AUTO>> mode of the device can be set, includes Number of Measurements, Resting Time (Countdown time), Interval Time, AFIB detector, CBP measurement, HIDE and Average calculation (Discard
1st measurement). Select << AUTO>> mode, press the Ubutton to perform automatic measurements based on the settings of << AUTO>> mode. The device shows all the settings and then starts counting down the Resting Time before the first measurement. The average measurement reading is displayed and saved after the measurements are complete.
The measurement program in << MANUAL>> mode can be set to preferences. The program includes setting the Highest Cuff Pressure and Hide Cuff Pressure during deflation. Select the << MANUAL>>mode if auscultatory blood pressure measurement is preferred above oscillometric blood pressure measurement. In << MANUAL>> mode, the device serves as a pressure gauge. No oscillometric measurements will be taken. Systolic and diastolic Korotkoff sounds are determined by the physician using a stethoscope placed over the brachial artery.
The device can accurately measure blood pressure in pregnant patients including those with known or suspected pre-eclampsia.
The device detects the appearance of atrial fibrillation during measurement and the atrial fibrillation symbol "is displayed on the LCD screen if any atrial fibrillation signal has been detected.
The device is for hospital use only.
#### 5. Indications for Use:
The Microlife Upper Arm Automatic Digital Blood Pressure Monitor,Model WatchBP Office (BP3SK1-3B) is a non-invasive digital blood pressure device using oscillometric technique and an upper-arm blood pressure cuff to measure
{5}------------------------------------------------
systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm.
The device can accurately measure blood pressure in pregnant patients including those with known or suspected pre-eclampsia.
The device provides aortic blood pressure parameters, includes central systolic blood pressure (cSBP), central pulse pressure (cPP) and central diastolic blood pressure (cDBP), non-invasively through the use of a brachial cuff.
The device detects the appearance of atrial fibrillation during measurement and gives a warning signal together with the measured blood pressure value if atrial fibrillation is detected.
The memory data can be transferred to the PC (personal computer) running the WatchBP Analyzer software by connecting the monitor via USB cable or Bluetooth.
The device is for hospital use only.
#### 6. Comparison to the 510(k) Cleared Device (Predicate Device):
### Subject (Modified) Device Compared to Primary Predicate ProBP 2400 (K152770):
Based on information from the Comparison Chart:
The modified device model WatchBP Office (BP3SK1-3B) uses the same oscillometric method as the predicate device ProBP 2400 with the same software algorithm to determine the systolic and diastolic blood pressure, pulse rate and mean arterial pressure (MAP). Upper arm cuff is inflated automatically by pump, the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units. They are both intended to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in pregnant patients including those with known or suspected pre-eclampsia, adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm.
The differences between the devices are:
- 1) Measurement mode
The subject device has two measurement modes (auto mode, manual mode) while the primary predicate device has three measurement modes(1x mode › 3x mode, manual mode). With a press of start/ Stop button, the <auto mode> of the subject device allows 1 to 6 consecutive measurements and give the average readings after all the measurements are finished. The number of measurement of <Auto mode> is selectable from the button operation. For the primary predicate device, <1 x mode> allows single measurement, <3x mode> allows 3 consecutive measurements. Number of measurements are just the U/I relative feature, not relative to the accuracy of BP measurement or AFIB detection. The manual mode
{6}------------------------------------------------
of both the subject device and primary predicate device is designed to have the deflation rate of 2-3 mmHg/s for the professionals can use the device and stethoscope to take manual blood pressure measurement. The firmware revision is due to the need of different U/I design but not relative to the algorithm in BP measurement, atrial fibrillation detection. The change does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. The changes of the U/I of the subject device with the flexibility to take 1-6 consecutive measurement is designed complying to AAMI/ANSI/IEC 80601-2-30. Therefore, the changes do not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
#### 2) Microprocessor
They are different in some features of microprocessor. However different features of microprocessor are all U/I related features (Microprocessor #2, #3, #4), which do not affect the clinical performance of the device. And all features related to brachial blood pressure algorithm are identical (Microprocessor #6, #7, #8).The added features of atrial fibrillation detection and the central blood pressure measurement are identical to reference predicate device (#9). Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
#### 3) F/W reversion of Algorithm (base on the items of "Microprocessor")
For F/W reversion of algorithm, they are only different in U/I, including measurement modes, memories, number measurement settings, resting time, interval time settings, hide reading functions, USB and Bluetooth connectivity which are not relative to clinical performance of the devices. The brachial blood pressure measurement algorithm related specifications (#1, #2, #3, #4) are identical to both predicate devices. The added features of atrial fibrillation detection and the central blood pressure measurement are identical to reference predicate device (#4, #5). Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
The F/W revision is due to the need of different U/I design but not relative to the algorithm in BP measurement, atrial fibrillation detection. U/I difference that needs F/W revision are addressed in other separated items, includes bluetooth connectivity, measurement modes, Number of Measurements Setting Function, Resting Time Setting Function, Measurement interval time Setting Function.
### 4) Atrial Fibrillation Detection Function
The Atrial Fibrillation Detection Function is added to the subject device, and the change does not affect the clinical performance for the brachial blood pressure measurement of the subject device. The function is identical to reference
{7}------------------------------------------------
predicate device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
### 5) Atrial Fibrillation Detected Indicator
It's only a visual indicator (ion: arral Fibrillation Detection Function.
#### 6) Review Memory
Due to the differences of the measurement modes between the subject device and predicate device, the subject device allows the user to review the last set of Auto mode measurements that consist average and individual 1- 6 measurements and the predicate device allows the user to review the last measurement of 1x mode and average of 3 measurements and individual measurements of 3X mode. The difference is U/I related features and not affect the safety or effectiveness of the subject device.
#### 7) PC-link Function
The PC-link Function is added to the subject device, it makes the device use series port to USB cable to connect with the PC, but this function is only a way to transfer the data and will not affect the accuracy and efficacy of the use. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
#### 8) Blood Pressure Analyzer
Blood Pressure Analyzer is added to the subject device. The memory data can be transferred to the PC (personal computer) running the Blood Pressure Analyzer software (WatchBP Analyzer) by connecting the monitor via USB cable or Bluetooth. This function is only a way to transfer the data and will not affect the accuracy and efficacy of the use. Therefore, the change does not affect the safety or effectiveness of the subject device according to the Verification and Validation Documentation of the Blood Pressure Analyzer software.
#### 9) Bluetooth Function
The Bluetooth Function is added to the subject device, it makes the device use Bluetooth (BT4.2) to connect with the PC running the Blood Pressure Analyzer software (WatchBP Analyzer), but this function is only a way to transfer the data and will not affect the accuracy and efficacy of the use. Therefore, the change does not affect the safety or effectiveness of the subject device according to the FCC Certification Bluetooth RF Test Report, and Bluetooth RF Exposure Evaluation Report.
{8}------------------------------------------------
#### 10) Number of Measurements Setting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Number of Measurements Setting Function of 1 to maximum 6 measurements while the primary predicate has fixed number of measurement at each measurement modes. The difference is U/I related features. The change does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
#### 11) Resting Time Setting Function
The subject device has the Number of Resting Time Setting Function while the primary predicate has not the function. The resting time of subject device includes 15, 30, 60, 120, 160, 240 and 300 seconds, are selectable by the user. The difference is U/I related features of F/W reversion of algorithm. The change does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
### 12) Measurement Intervals Setting Function
The Measurement Intervals Setting Function of the predicate devices has the interval of 15, 30, 45 and 60 seconds and the subject device has 15, 30, 60, 120, 180, 240 and 300 seconds. The difference is U/I related features of F/W reversion of algorithm. The change does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
- 13) CBP&CPP (Central Blood Pressure& Central Pulse Pressure) Measurement Function
The CBP&CPP (Central Blood Pressure& Central Pulse Pressure) Measurement Function is added to the subject device, and the change does not affect the clinical performance for the brachial blood pressure measurement of the subject device. The function is identical to reference predicate device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
{9}------------------------------------------------
#### 14) Hide Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Hide Function while the primary predicate has not the function. But the change does not affect the use of the subject device, it does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
#### 15) Irreqular Heartbeat Detection Function
The Irregular Heartbeat Detection Function is removed from the subject device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
Based upon the aforementioned information, the two devices are substantially equivalent.
## Subject (Modified) Device Compared to Reference Predicate WatchBP Office Central (TWIN200 CBP) (K171937):
Based on information from the Comparison Chart:
The subject WatchBP Office (BP3SK1-3B) uses the same oscillometric method as the predicate WatchBP Office Central (TWIN200 CBP) with the same software alqorithm to determine the systolic and diastolic blood pressure, pulse rate, mean arterial pressure (MAP), central pulse pressure (cPP), central systolic blood pressure (cSBP) and central diastolic pressure (cDBP). Upper arm cuff is inflated automatically by pump, the deflation rate is controlled by factory set exhaust valve and the deflation pressures are transferred via tubing to a sensor in these two units. They both detect the appearance of atrial fibrillation during measurement and give a warning signal with the reading once the atrial fibrillation is detected.
The differences between the devices are:
- 1) Measuring Location
The subject device has two measuring locations (upper arm and ankle) while the primary predicate device only has one measuring location (upper arm).Their upper arm measuring function is identical. The ankle measuring function is removed from the subject device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
{10}------------------------------------------------
### 2) Measurement mode
The subject device has two measurement modes (auto mode, manual mode) while the primary predicate device has three measurement modes (SCREEN/ CENTRAL /ABI mode). With a press of start/ Stop button, the <auto mode> of the subject device allows 1 to 6 consecutive measurements and give the average readings after all the measurements are finished. The number of measurement of <Auto mode> is selectable from the button operation. For the reference predicate device, the <SCREEN mode> allow three consecutive measurements simultaneously on both arms. The CENTRAL mode allows 2 BP measurements, includes central BP indices, on single arm. The ABI mode allows the BP measurement on one arm and one ankle and calculate ankle brachial index after the finished of the measurement. The firmware revisions to suit for different modes, number of measurements, singe cuff or double cuffs (arm or ankle) are due to the need of different U/I design but not relative to the algorithm in BP measurement, atrial fibrillation detection or central BP indices, not affect the clinical performance of the subject device. The changes of the U/I of the subject device with the flexibility to take 1-6 consecutive measurement is designed complying to AAMI/ANSI/IEC 80601-2-30. Therefore, the changes do not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
### 3) Microprocessor
They are different in some features of microprocessor. However different features of microprocessor are all U/I related features (Microprocessor #2, #3, #4), which do not affect the clinical performance of the device. And all features related to brachial blood pressure algorithm are identical (Microprocessor #6, #7, #8).Their features of atrial fibrillation detection and the central blood pressure measurement are also identical (#9). Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement, atrial fibrillation detection and the central blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
4) F/W reversion of Algorithm (base on the items of "Microprocessor") For F/W reversion of algorithm, they are only different in U/I, including measurement modes, memories, number measurement settings, resting time, interval time settings, hide reading functions, USB and Bluetooth connectivity which are not relative to clinical performance of the devices. The brachial blood pressure measurement algorithm related specifications (#1, #2. #3, #4) are identical to both predicate devices. Their features of atrial fibrillation detection and the central blood pressure measurement are also identical (#4, #5). Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement, atrial fibrillation detection and the central blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
{11}------------------------------------------------
The F/W revisions to suit for different modes, number of measurements, singe cuff or double cuffs (arm or ankle) are due to the need of different U/I design but not relative to the algorithm in BP measurement, atrial fibrillation detection or central BP indices, not affect the clinical performance of the subject device. U/I difference that needs F/W revision are addressed in other separated items, includes bluetooth connectivity, measurement modes, Number of Measurements Setting Function, Resting Time Setting Function, Measurement interval time Setting Function, inflation pressure setting, inflation cuff selection, ABI calculation function.
#### 5) Review Memory
Due to the differences of the measurement modes between the subject device and reference predicate device, the subject device allows the user to review the last set of Auto mode measurements that consist average and individual 1- 6 measurements and the predicate device allows the user to review the average and individual reading of SCREEN and CENTRAL mode and the last measurement of ABI. The difference is U/I related features and not affect the safety or effectiveness of the subject device.
### 6) Blood Pressure Analyzer Software
The subject device and the reference predicate device both have Blood Pressure Analyzer Software, they are iust different in the version of the Blood Pressure Analyzer Software. The change will not affect the accuracy and efficacy of the use. Therefore, the change does not affect the safety or effectiveness of the subject device according to the Verification and Validation Documentation of the Blood Pressure Analyzer software.
### 7) Bluetooth Function
The Bluetooth Function is added to the subject device, it makes the device use Bluetooth (BT4.2) to connect with the PC running the Blood Pressure Analyzer software (WatchBP Analyzer), but this function is only a way to transfer the data and will not affect the accuracy and efficacy of the use. Therefore, the change does not affect the safety or effectiveness of the subject device according to the FCC Certification Bluetooth RF Test Report, and Bluetooth RF Exposure Evaluation Report.
#### 8) Number of Measurements Setting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Number of Measurements Setting Function while the primary predicate has not the function. But the change does not affect the use of the subject device, it does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
{12}------------------------------------------------
## 9) Resting Time Setting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Resting Time Setting Function while the primary predicate has not the function. But the change does not affect the use of the subject device, it does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
## 10) Measurement Intervals Setting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Measurement Intervals Setting Function which can be set to 15, 30, 60, 120, 180, 240 and 300 seconds for Auto mode while the reference predicate device can be set to 15, 30, 45 or 60 seconds only in ROUTINE Mode. The change does not cause any new or significantly modified risks according to the Risk Management File. It does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
## 11) Inflation Pressure Setting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has the Inflation Pressure Setting Function while the primary predicate has not the function. But the change does not affect the use of the subject device, it does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
## 12) Inflation Cuff Selecting Function
Due to different U/I related features of F/W reversion of algorithm, the subject device has not the Inflation Cuff Selecting Function while the primary predicate has the function. But the change does not affect the use of the subject device, it does not cause any new or significantly modified risks according to the Risk Management File and it does not affect the clinical performance of the subject device. Therefore, the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
### 13) PP Automatically Calculation Function
The PP Automatically Calculation Function is removed from the subject device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
{13}------------------------------------------------
- 14) ABI(Ankle Brachial Index) Calculating Function
The ABI (Ankle Brachial Index) Calculating Function is removed from the subject device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
### 15) Validated in preeclampsia
The subject device can accurately measure blood pressure in pregnant patients including those with known or suspected pre-eclampsia while the reference predicate device can't. But the function is the same as the primary predicate device. Therefore, the change does not affect the accuracy and efficacy for brachial blood pressure measurement. And the change does not affect the safety or effectiveness of the subject device according to the safety and essential performance testing.
Based upon the aforementioned information, the two devices are substantially equivalent.
#### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Testing information demonstrating safety and effectiveness of the Microlife Upper Arm Automatic Digital Blood Pressure Monitor. Model WatchBP Office (BP3SK1-3B) in the intended environment of use is supported by testing that was conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", DCRND, which outlines Electrical, Mechanical and Environmental Performance requirements.
The following testing was conducted to prove safety and effectiveness as well as substantial equivalence to the predicate devices:
The following National and International Standards were utilized for testing the subject device:
- 1) IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance AAMI / ANSI ES60601-1:2005/(R) 2012 And A1:2012, C1:2009/(R) 2012 And A2:2010/(R) 2012
- 2) IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements And Tests.
- 3) ISO 14971: 2007 Medical devices Application of risk management o medical devices.
- 4) AAMI/ANSI/ISO 10993-1:2009/(R)2013, Biological evaluation of medical
{14}------------------------------------------------
devices – Part 1: Evaluation And Testing Within A Risk Management Process.
- 5) AAMI/ANSI/ISO 10993-5:2009/(R)2014. Biological evaluation of medical devices – Part 5: Tests for In Vitro Cytotoxicity.
- 6) AAMI / ANSI / ISO 10993-10:2010/(R)2014,, Biological evaluation of medical devices - Part 10: Tests for Irritation and Skin Sensitization
- 7) AAMI/ANSI/IEC 80601-2-30 Medical electrical equipment Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. 2013
8) AAMI/ANSI/ISO 81060-2 Non-Invasive Sphygmomanometers - Part 2: Clinical Validation of Automated Measurement Type. 2013
None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Microlife Upper Arm Automatic Digital Blood Pressure Monitor, Model WatchBP Office (BP3SK1-3B) tested met all relevant requirements of the aforementioned tests.
#### 8. Discussion of Clinical Tests Performed:
Clinical Validation Concerning the Compliance of ANSI/AAMI/ ISO 81060-2: The subject device Model WatchBP Office (BP3SK1-3B) is from the technical point of view, identical to the predicate blood pressure monitor ProBP 2400. They both intended to measure systolic and diastolic blood pressures, pulse rate, mean arterial pressure (MAP) for use in pregnant patients including those with known or suspected pre-eclampsia, adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm. Although they are different in some features of microprocessor and F/W reversion of algorithm, the different features are all U/I related features which are not relative to clinical performance of the devices and all features related to brachial blood pressure algorithm are identical. The fundamental scientific technology of the modified WatchBP Office (BP3SK1-3B) device is the same as the predicate device ProBP 2400. Therefore the brachial blood pressure measurement algorithm detection in pregnant patients including those with known or suspected pre-eclampsia, adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm of WatchBP Office (BP3SK1-3B) is substantially equivalent with the predicate device ProBP 2400 and repeated clinical testing for aforementioned parameters is not required. There was no repeated clinical testing required for brachial blood pressure in pregnant patients including those with known or suspected pre-eclampsia, adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm to support WatchBP Office as the subject device WatchBP Office (BP3SK1-3B) can leverage the clinical validation of ProBP 2400 that was proven in K152770. Repeat clinical testing in accordance with the standard AAMI / ANSI /ISO81060-2 for the subject device WatchBP Office (BP3SK1-3B) regarding brachial blood pressure measurement in pregnant patients including those with known or suspected preeclampsia, adults and pediatrics (but not neonates) with arm cuff circumference sizes ranging from 14 -52 cm is therefore not necessary.
{15}------------------------------------------------
Moreover, the subject device Model WatchBP Office (BP3SK1-3B) is also intended to detect AFIB and measure CBP&CPP (Central Blood Pressure& Central Pulse. The subject device Model WatchBP Office (BP3SK1-3B) is from the technical point of view, identical to the predicate blood pressure monitor WatchBP Office Central (TWIN200 CBP).Although they are different in some features of microprocessor and F/W reversion of algorithm, the different features are all U/I related features which are not relative to clinical performance of the devices and all features related to brachial blood pressure algorithm , atrial fibrillation detection and the central blood pressure measurement are identical. The fundamental scientific technology of the modified WatchBP Office (BP3SK1-3B) device is the same as the predicate device WatchBP Office Central (TWIN200 CBP).Therefore the performance of the WatchBP Office Central (TWIN200 CBP) in terms of brachial blood pressure measurement, AFIB detection and CBP&CPP measurement would be essential equivalent with performance of the predicate device WatchBP Office Central (TWIN200 CBP). There was no repeated clinical testing required for brachial blood pressure, AFIB detection and CBP&CPP measurement to support WatchBP Office as the subject device WatchBP Office (BP3SK1-3B) can leverage the clinical validation of WatchBP Office Central (TWIN200 CBP) that was proven in K171937.Repeat clinical testinq in accordance with the standard AAMI / ANSI/IEC81060-2 for the subject device WatchBP Office (BP3SK1-3B) regarding brachial blood pressure measurement, AFIB detection and CBP&CPP measurement for use in adults is therefore not necessary..
#### 9. Software information:
Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". In addition, since our device requires the use of off-the-shelf software to operate the PC-link function, we adhered to the FDA September 1999 document "Guidance for Off-The-Shelf Software Use in Medical Devices".
#### 10. Conclusions:
Conclusions drawn from the non-clinical and clinical tests demonstrate that the subject device is as safe, effective, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.