K200237 · Shenzhen Kentro Medical Electronics Co., Ltd. · NUH · Oct 19, 2020 · Neurology
Device Facts
Record ID
K200237
Device Name
Transcutaneous Electronic Nerve Stimulator
Applicant
Shenzhen Kentro Medical Electronics Co., Ltd.
Product Code
NUH · Neurology
Decision Date
Oct 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Transcutaneous Electronic Nerve Stimulator is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
Device Story
Portable, battery-powered transcutaneous electrical nerve stimulator (TENS) for OTC pain relief. Device delivers electrical pulses via skin-contact electrode pads to stimulate muscles. Models KTR-249X series include remote control with LCD for mode/time display; other models feature integrated buttons. Users operate device to select modes and adjust intensity. Output affects nerve/muscle stimulation to provide temporary pain relief. Device is intended for home use.
Clinical Evidence
Bench testing only. Evaluated for electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1, 5, 10), usability (IEC 62366-1), and software verification/validation. Waveform testing verified output specifications against FDA guidance for TENS and powered muscle stimulators.
Technological Characteristics
Portable, battery-powered (CR2032 or Li-polymer) TENS device. ABS plastic housing. Pulsed, symmetric, biphasic waveform. 1 channel output. 3 to 15 operation modes. 16 intensity steps. Voltage-controlled. Biocompatible electrode pads (ISO 10993). Connectivity: optional remote control (KTR-249X series).
Indications for Use
Indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 19, 2020
Shenzhen Kentro Medical Electronics Co., Ltd Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Room 2231, Building 1, Ruifeng center, Kaichuang road, Huangpu district Guangzhou, Guangdong 51006 China
Re: K200237
Trade/Device Name: Transcutaneous Electronic Nerve Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: July 15, 2020 Received: July 21, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Pamela Scott, MS. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200237
#### Device Name
(Model: KTR-2401, KTR-2412, KTR-2412, KTR-2412, KTR-2302, KTR-2341, KTR-2342, KTR-2491, KTR-2492, KTR-2493, KTR-2494)
#### Indications for Use (Describe)
Transcutaneous Electronic Nerve Stimulator is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Sponsor: | Shenzhen Kentro Medical Electronics Co., Ltd |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device: | Transcutaneous Electronic Nerve Stimulator, Model: KTR-2401, KTR-2402, KTR-2411, KTR-2412, KTR-2301, KTR-2302, KTR-2341, KTR-2342, KTR-2491, KTR-2492, KTR-2493, KTR-2494 |
### 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### Submitter's Information 1.
- 510(k) Owner's Name: Shenzhen Kentro Medical Electronics Co., Ltd �
- Establishment Registration Number: 3013671142 �
- Address: No.3, Xihu industry zone, Xikeng Village, Henggang Town, Longgang District, Shenzhen, � Guangdong, China
- Tel: +86-755-33825998 �
- Fax: +86-755-33825996 �
- Contact Person: Zewu Zhang �
- � Email: kentro@kentro.com.cn
#### Application Correspondent: 2.
- � Contact Person: Ms. Cassie Lee
- Guangzhou GLOMED Biological Technology Co., Ltd. �
- Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, � Guangdong, China
- � Tel: +86 20 8266 2446
- � Email: regulatory@glomed-info.com
#### 3. Subject Device Information
- Transcutaneous Electronic Nerve Stimulator � Trade Name:
- KTR-2401. KTR-2402. KTR-2411. KTR-2412. KTR-2301. KTR-2302, KTR-2341, KTR-2342, KTR-2491, KTR-2492, � Model: KTR-2493, KTR-2494 Electronic Stimulator � Common Name:
- Stimulator, Nerve, Transcutaneous, Muscle, Powered, For Classification name: � Muscle Conditioning, Over-The-Counter Neurology, Physical Medicine
- Review Panel: �
- NUH Product Code: �
- ll � Regulation Class:
- 890.5850 � Regulation Number:
#### 4. Predicate Device Information
| Sponsor | M.I.TECH Co., Ltd. | Shenzhen Kentro Medical<br>Electronics Co., Ltd | Guangzhou Xinbo<br>Electronic Co., Ltd. |
|--------------------------|--------------------|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| Device Name<br>and Model | HANAROCare ReJu | Transcutaneous Electrical<br>Nerve Stimulator<br>Model:KTR-206, KTR-208,<br>KTR-209 | Pain Therapy Device<br>Model: P.T.S-II, P.T.S-IIA,<br>P.T.S-IIB, CP-I |
{4}------------------------------------------------
#### Shenzhen Kentro Medical Electronics Co., Ltd Sponsor:
Transcutaneous Electronic Nerve Stimulator, Model: KTR-2401, KTR-2402, KTR-2411. KTR-2412. KTR-2301. KTR-2302. KTR-2341. KTR-2342. KTR-2491. Subject Device: KTR-2492, KTR-2493, KTR-2494
| 510(k) Number | K160893 (Primary) | K183288 | K163611 |
|-------------------|-------------------|---------------|---------------|
| Product Code | NUH | NUH, NGX, NYN | NUH, NGX, NYN |
| Regulation Number | 882.5890 | 882.5890 | 882.5890 |
| Regulation Class | II | II | II |
### 5. Device Description
Transcutaneous Electronic Nerve Stimulator (Models:
KTR-2401, KTR-2402, KTR-2411, KTR-2412, KTR-2301, KTR-2302, KTR-2341, KTR-2342, KTR-2491, KTR-2492, KTR-2493, KTR-2494) is a portable and battery powered multifunctional device, offering Transcutaneous Electronic Nerve Stimulator. They can give certain electrical pulse through electrode pads placed on the skin to help users to enjoy body stimulation.
For models: KTR-2401, KTR-2402, KTR-2411, KTR-2412, KTR-2301, KTR-2302, KTR-2342 have 3 operation modes:
For models: KTR-2491, KTR-2493, KTR-2493, KTR-2494 have 15 operation modes, 9 manual mode, and 6 automatic mode. They can be remotely controlled with an accessory remote control. And there is a LCD displays on remote control, which display mode and treatment time.
The electronic stimulator module has unified the operating elements of ON/OFF button. Intensity adjust button and Mode selection button.
The device is equipped with accessories of electrode pads, batteries and remote control (For KTR-249X series).
The electrode pads are complying with the biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization), are interchangeable.
### 6. Intended Use / Indications for Use
Transcutaneous Electronic Nerve Stimulator is indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm) and lower extremities (leg) due to strain from exercise or normal household work activities.
### 7. Test Summary
Transcutaneous Electronic Nerve Stimulator has been evaluated the safety and performance by lab bench testing as following:
- Electrical safety test according to IEC 60601-1. IEC 60601-1-11 and IEC 60601-2-10 standards �
- � Electromagnetic compatibility test according to IEC 60601-1-2 standard
- � Biocompatibility test according to ISO 10993-1, ISO 10993-5 and ISO 10993-10 standards
- Usability test according to IEC 62366-1 standard �
- � Software verification and validation test according to the requirements of the FDA "Guidance for Pre-Market Submissions and for Software Contained in Medical Devices"
{5}------------------------------------------------
| Sponsor: | Shenzhen Kentro Medical Electronics Co., Ltd |
|-----------------|---------------------------------------------------------------------------------------------------|
| | Transcutaneous Electronic Nerve Stimulator, Model: KTR-2401, KTR-2402, KTR- |
| Subject Device: | 2411, KTR-2412, KTR-2301, KTR-2302, KTR-2341, KTR-2342, KTR-2491,<br>KTR-2492, KTR-2493, KTR-2494 |
- ◆ The waveform test report has also been conducted to verify the output specifications of the device according to Guidance for Transcutaneous Electrical Nerve Stimulator for Pain Relief Intended for Over the Counter Use and Guidance for Powered Muscle Stimulator for Muscle Conditioning
### 8. Comparison to predicate device and conclusion
The technological characteristics, features, specifications, materials, mode of operation, and intended use of Transcutaneous Electronic Nerve Stimulator is substantially equivalent to the primary predicate device quoted above.
The differences between the subject device and primary predicate device do not raise new issues of safety or effectiveness.
| Elements<br>of<br>Comparis<br>on | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remark |
|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device<br>Name and<br>Model | Transcutaneous<br>Electronic Nerve<br>Stimulator<br>Model:<br>KTR-2401,KTR-2402,KTR-2411,KTR-<br>2412,KTR-2301,KTR-2302,KTR-2341,KTR-<br>2342,KTR-2491,KTR-2492,KTR-2493,KTR-<br>2494 | HANAROCare ReJu | Transcutaneous<br>Electrical Nerve<br>Stimulator<br>Model: KTR-206,<br>KTR-208, KTR-209 | Pain Therapy<br>Device, Models:<br>P.T.S-II, P.T.S-IIA,<br>P.T.S-IIB, CP-I | -- |
| 510(k)<br>Number | Applying | K160893 | K183288 | K163611 | -- |
| Product<br>code | NUH | NUH | NUH, NGX, NYN | NUH, NGX, NYN | SE |
| Intended<br>Use | Transcutaneous<br>Electronic Nerve<br>Stimulator is<br>indicated for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist, back,<br>neck, upper<br>extremities (arm) and<br>lower extremities<br>(leg) due to strain<br>from exercise or<br>normal household<br>work activities. | HANAROCare ReJu<br>is indicated for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck, upper<br>extremities (arm)<br>and lower<br>extremities (leg) due<br>to strain from<br>exercise or normal<br>household work<br>activities. | To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>shoulder, back,<br>arm, leg, foot, due<br>to strain from<br>exercise or normal<br>household and work<br>activities. | To be used for<br>temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the<br>upper and lower<br>back, back of the<br>neck, upper<br>extremities<br>(shoulder and arm),<br>lower extremities<br>(leg and feet) due to<br>strain from exercise<br>or normal<br>household work<br>activities by<br>applying current to<br>stimulate nerve.<br>To be used for<br>symptomatic relief<br>and management of | SE |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remark |
| | | | | chronic, intractable pain and relief of pain associated with arthritis (Choose Mode B or C).<br>To stimulate healthy muscles in order to improve and facilitate muscle performance (Choose Mode A).<br>To temporarily increase local blood | |
| Power<br>Source(s) | For KTR-23XX series: CR2032;<br>3Vdc; 240mAh<br>For KTR-24XX series: PL301526;<br>3.7Vdc, 250mAh | Lithium-polymer, 3.7V | KTR-206: 2 AAA batteries (DC3V) | DC 3.0V, 2 x AAA | SE<br>Note 1 |
| Patient<br>Leakage<br>Current | NC DC: 0.5µA | DC: 0.5µA | | DC: 0.5µA | |
| | SFC DC: 0.6μΑ | DC: 0.6μΑ | N/A (Battery operated) | DC: 0.6μΑ | SE |
| Average DC current through electrodes when device is on but no pulses are being applied | < 0.01μΑ | < 0.01μΑ | -- | < 0.01μΑ | SE |
| Number of Output Channels: | 1 channel | 1 channel | 2 channels | 2 channels | SE |
| Number of Output Modes | For KTR-23XX series: 3 modes<br>For KTR-240X series | 4 | 5 | 3 | SE<br>Note 2 |
| Elements of<br>Comparison | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remark |
| | & KTR-241X series: 3<br>modes<br>For KTR-249X series:<br>15 modes | | | | |
| Output<br>Intensity<br>Level | 16 steps | 15steps | 16 steps | 5 steps | SE |
| Synchronous or<br>Alternating? | Synchronous | Unknown | Synchronous | Synchronous | SE |
| Regulated<br>Current or<br>Regulated<br>Voltage? | Voltage Control | Unknown | Voltage Control | Voltage Control | SE |
| Software/<br>Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Yes | Yes | SE |
| Automatic<br>Overload<br>Trip | No | Unknown | No | No | SE |
| Automatic<br>No-Load<br>Trip | No | Unknown | No | No | SE |
| Automatic<br>Shut Off | Yes | Unknown | Yes | Yes | SE |
| User<br>Override<br>Control | Yes | Unknown | Yes | Yes | SE |
| On/Off<br>Status | Yes | Unknown | Yes | Yes | SE |
| Indicator<br>Low<br>Battery | No | Unknown | No | No | SE |
| Display<br>Voltage/<br>Current<br>Level | Yes | Unknown | Yes | Yes | SE |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remarks |
| Timer<br>Range | 15min | 20 minutes fixed | 25min15min default<br>KTR-<br>206:5/10/15min<br>KTR-<br>208:5/10/15/20/25/30 min<br>KTR-<br>209:5/10/15/20/25/30 min | 10, 20, 40 min | SE |
| Weight | Main Unit:31g<br>Electrode:<br>EPAD-H01: 13g,<br>EPAD-H02: 13g,<br>EPAD-F01:15g,<br>EPAD-F02: 10g,<br>EPAD-F03: 12g,<br>EPAD-B01: 8g,<br>EPAD-T01: 8g,<br>EPAD-Z01: 55g | 11g | KTR-206: 1.68oz<br>KTR-208: 1.79oz<br>KTR-209: 2.70oz | Main Unit:<br>P.T.S-II: 75g<br>P.T.S-IIA: 100g<br>P.T.S-IIB: 100g<br>CP-I: 66g<br>Electrode:<br>Big Patch<br>Electrode: 40g<br>Small Patch<br>Electrode:10g<br>Insole Electrode:<br>200g<br>Sole Plant<br>Electrode A (only<br>for CP-I): 900g<br>Sole Plant<br>Electrode B: 920g | SE<br>Note 1 |
| Dimensions of main<br>unit (mm) | Model KTR-2401:<br>Φ46.3x12.07;<br>Model KTR-2402:<br>Φ46.28x11.69<br>Model KTR-2411:<br>46.29x46.29x12.08;<br>Model KTR-2412:<br>46.29x46.29x11.59;<br>Model KTR-2301:<br>Φ49.8x12.44;<br>Model KTR-2302:<br>Φ49.8x12.48<br>Model KTR-2341:<br>48.8x48.8x12.42;<br>Model KTR-2342:<br>48.8x48.8x12.45; | 36 x 35 x 13.7 mm | KTR-206:<br>112.5x59x33.3mm<br>KTR-208:<br>112.5x59x29.5mm | Main Unit:<br>P.T.S-II: 110 x 78 x 20 mm<br>P.T.S-IIA: 135 x 82 x 20 mm<br>P.T.S-IIB: 135 x 82 x 20 mm<br>CP-I: 92 x 78 x 20 mm<br>Electrode:<br>Large Patch<br>Electrode: 120 x 80 mm<br>Small Patch<br>Electrode: 46 x 46 mm<br>Insole Electrode:<br>260 x 110 mm<br>Sole Plant | SE<br>Note 1 |
| Elements<br>of<br>Comparison | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remarks |
| | Model KTR-2491:<br>Φ49.8x12.44;<br>Model KTR-2492:<br>Φ49.8x12.48<br>Model KTR-2493:<br>48.8x48.8x12.42;<br>Model KTR-2494:<br>48.8x48.8x12.45; | | KTR-209:<br>129.7x60x17.<br>8mm | Electrode<br>A (only for CP-I):<br>450 x 450 x 90 mm<br>Sole Plant<br>Electrode B: 450 x<br>450 x 90 mm | |
| Housing<br>Materials<br>and<br>Construction | Main unit: ABS plastic | Retardant<br>Polycarbonate | Main unit: ABS<br>plastic | Main unit: ABS<br>plastic | SE |
| Waveform | Pulsed, symmetric,<br>biphasic | Monophasic | Pulsed, symmetric,<br>biphasic | Pulsed, symmetric,<br>biphasic | SE |
| Shape | Rectangular, with<br>interphase interval | Rectangular | Rectangular, with<br>interphase interval | Rectangular | SE |
| | 43.6V±10% @ 500Ω | (±10%) @ 500Ω<br>• Mode 1: 64V<br>• Mode 2: 67V<br>• Mode 3: 59V<br>• Combination : This<br>mode cycles the<br>above modes | KTR-206:<br>49.6V@ 500Ω<br>68.5V @ 2kΩ<br>73V @ 10kΩ | 40V±10% @ 500Ω | |
| Maximum<br>Output<br>Voltage | 59V±10% @ 2KΩ | (±10%) @ 2KΩ<br>• Mode 1: 113 V<br>• Mode 2: 119 V<br>• Mode 3: 108 V<br>• Combination: This<br>mode cycles the<br>above modes. | KTR-208:<br>58.5V @ 500Ω<br>70V @ 2kΩ<br>70.5V @ 10kΩ | 80V±10% @ 2KΩ | SE<br>Note 2 |
| | 66.5V±10% @ 10KΩ | (±10%) @ 10KΩ<br><15 | KTR-209:<br>62V @ 500Ω<br>80V @ 2KΩ<br>84V @ 10KΩ | 95V±10% @ 10KΩ | |
| Maximum<br>Output<br>Current | 87.2mA±10% @<br>500Ω | (±10%) @ 500Ω<br>• Mode 1 : 128 mA<br>• Mode 2 : 134 mA<br>• Mode 3 : 118 mA<br>• Combination: This<br>mode cycles the<br>above modes. | KTR-206:<br>99.2mA @ 500Ω<br>34.25mA @ 2KΩ<br>7.3mA @ 10KΩ | 80mA±10% @<br>500Ω | SE<br>Note 2 |
| | 29.5mA±10% @ 2KΩ | (±10%) @ 2KΩ<br>• Mode 1: 57 mA | KTR-208:<br>117mA @ 500Ω | 40mA±10% @ 2KΩ | |
| Elements<br>of<br>Comparis<br>on | Subject Device | Predicate Device 1<br>(primary) | Predicate Device 2 | Predicate Device 3 | Remark |
| | | • Mode 2: 60 mA<br>• Mode 3: 54 mA<br>• Combination: This<br>mode cycles the<br>above modes. | 35mA @ 2KΩ<br>7.05mA @ 10KΩ | | |
| | 6.65mA±10% @<br>10KΩ | (±10%) @ 10KΩ<br><15 | KTR-209:<br>124mA @ 500Ω<br>40mA @ 2KΩ<br>8.4mA @ 10KΩ | 9.5mA±10%<br>@10KΩ | |
| Pulse<br>Duration | 120µs | (±10%)<br>• Mode 1: 115 µs<br>• Mode 2: 75 µs<br>• Mode 3: 65 µs<br>• Combination: This<br>mode cycles the<br>above modes. | KTR-206: 84µs-<br>134µs<br>KTR-208: 82µs-<br>128µs<br>KTR-209: 80µs-<br>224µs | 200µs | SE<br>Note 2 |
| Pulse<br>frequency | 20-100Hz | (±10%)<br>• Mode 1: 2Hz<br>• Mode 2: 16.7Hz<br>• Mode 3: 33.3Hz<br>• Combination: This<br>mode cycles the<br>above modes. | KTR-206: 1Hz-<br>108Hz<br>KTR-208: 1Hz-<br>109Hz<br>KTR-209: 1Hz-<br>110Hz | 13.7~48.5Hz | SE<br>Note 2 |
| Net<br>Charge<br>(per<br>pulse) | 0µC @ 500Ω,<br>Method: Balanced<br>waveform | Unknown | 0µC @ 500Ω,<br>Method: Balanced<br>waveform | 0µC @ 500Ω,<br>Method: Balanced<br>waveform | SE |
| Maximum<br>Phase<br>Charge | 12.66µC @ 500Ω | (±10%) @ 500Ω<br>• Mode 1: 14.72 µC<br>• Mode 2: 10.05 µC<br>• Mode 3: 7.67 µC<br>• Combination: This<br>mode cycles the<br>above modes. | KTR-206: 12.32µC<br>@ 500Ω<br>KTR-208: 18.12µC<br>@ 500Ω<br>KTR-209: 33.07µC<br>@ 500Ω | 19.3µC @ 500Ω | SE<br>Note 2 |
| Maximum<br>Current<br>Density(m<br>A/cm2,<br>r.m.s.)<br>(@500Ω) | 0.058 (mA/cm²) | (±10%)<br><Standard<br>Electrode:<br>12.57cm²><br>• Mode 1: 0.154<br>• Mode…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.