CT F/S/D Compatible Applicators (Fletcher), HDR Compatible Tandem and Ovoid Applicators (Henschke), HDR CT Compatible Split Ring Applicator, Segmented Vaginal Applicator, HDR Miami Applicator HDR Brachytherapy Applicator, 2/3 Endometrial Applicator Sets, CT HDR Ring & Tandem Applicator with Rectal Retractor
K200221 · Mick Radio-Nuclear Instruments, Inc. · JAQ · Feb 28, 2020 · Radiology
Device Facts
Record ID
K200221
Device Name
CT F/S/D Compatible Applicators (Fletcher), HDR Compatible Tandem and Ovoid Applicators (Henschke), HDR CT Compatible Split Ring Applicator, Segmented Vaginal Applicator, HDR Miami Applicator HDR Brachytherapy Applicator, 2/3 Endometrial Applicator Sets, CT HDR Ring & Tandem Applicator with Rectal Retractor
Applicant
Mick Radio-Nuclear Instruments, Inc.
Product Code
JAQ · Radiology
Decision Date
Feb 28, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
2/3 Endometrial Applicator Sets: A manual radionuclide applicator system, a manually operated device intended to apply a radionuclide source into the body or to the surface of the body for radiation therapy. Segmented Vaginal Applicator: The use of sealed Radioisotopes to treat tumors within the body has been documented and published since the turn of the century. Modem era Radiation Therapy has developed delivery systems using isotopes of Cesium, Iridium, Iodine, and Gold to name a few examples. Many tumors are now treated by internal exposure to radiation emitted from sealed radioactive sources. Two common modalities of this are Low Dose Rate remote afterloading (Brachytherapy). One common use of Brachytherapy is in the treatment of cancer of the vaginal process. The system described in this 510(k) has been developed to function as an applicator for the positioning of sealed sources in the intracavitary treatment of the vagina. HDR Miami Applicator: This applicator is designed as an accessory to the Varisource System (Varian Associates K952913) and the Gammamed System (K89113) which uses a single radioactive source of Iridium-192 to treat cancer in a wide range of body sites. The Miami Applicator is placed in the vicinity of the cervix via the vagina just as described for the predicate device (Nucletron Miami Vaginal Applicator, K953946) and different diameter sleeves and intrauterine tubes, can be optimized to best meet the clinical needs of the patient along with minimization of dose to the mucosa. HDR Compatible Tandern and Ovoid Applicators: The applicators presented in this 510(k) notification have been developed to function as Applicators for the positioning of HDR Remote After-Loader sealed sources in the intra-cavitary treatment of cancer of the vagina and cervix. CT Compatible F/S/D Applicators: CT Compatible F/S/D applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice. HDR CT Compatible Split Ring Applicator: The HDR CT Compatible Split Ring type applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice. CCT HDR Tandem Ring Applicator with Rectal Retractor: The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/ Ring Applicator with Rectal Retractor is indicated for High Dose Rate irradiation of the uterus and cervix.
Device Story
Manual radionuclide applicator systems for intracavitary brachytherapy; used to position sealed radioactive sources (e.g., Iridium-192) for treating cervical, uterine, or vaginal cancer. Applicators include various geometries (tandem, ovoid, ring, segmented vaginal, endometrial tubes) to optimize dose distribution and minimize mucosal exposure. Operated by radiation oncologists/physicists in clinical settings. Modification adds MR Conditional labeling based on computational modeling of radiofrequency-induced heating. Devices remain physically identical to previously cleared predicates; no changes to design, materials, or manufacturing. Output is physical positioning of radiation source; clinical benefit is targeted tumor irradiation.
Clinical Evidence
No clinical data. Bench testing only: computational modeling performed to evaluate radiofrequency-induced heating for MR Conditional labeling.
Technological Characteristics
Manual radionuclide applicator systems; various geometries (tandem, ovoid, ring, segmented). Materials and construction identical to predicates. MR Conditional status confirmed via computational modeling. No energy source or software components.
Indications for Use
Indicated for patients requiring high dose rate (HDR) or low dose rate (LDR) radiation therapy for cancer of the cervix, uterus, or vagina. Used for intracavitary positioning of sealed radioactive sources.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA acronym is in a blue square, and the full name is in blue text. The seal features a stylized caduceus, a symbol associated with medicine and healing.
March 20, 2020 Mick Radio-Nuclear Instruments, Inc. James Hurlman Director QA/RA 521 Homestead Avenue Mount Vernon, New York 10550 Re: K200221 Trade/Device Name: CT/MR 2/3 Channel Endometrial, CT/MR Segmented Vaginal, CT/MR Miami, CT/MR Fletcher, CT/MR Henschke FSD, CT/MR Split Ring Applicator Sets, CT/MR Ring & Tandem, including Vienna Accessory Kit Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAQ, IWJ
Dear James Hurlman:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 28, 2020. Specifically, FDA is updating this SE Letter as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Thalia Mills, OHT7: Office of In Vitro Diagnostics and Radiological Health, 301-796-6641, Thalia.Mills(@fda.hhs.gov.
Sincerely,
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional in appearance.
Mick Radio-Nuclear Instruments, Inc. February 28, 2020 % Mr. James Hurlman Director QA/RA 521 Homestead Avenue MOUNT VERNON NY 10550 Re: K200221 Trade/Device Name: 2/3 Endometrial Applicator Sets, Segmented Vaginal Applicator, HDR Miami Applicator, HDR Compatible Tandem and Ovoid Applicators, CT Compatible F/S/D Applicators, HDR CT Compatible Split Ring Applicator, and CT HDR Tandem Ring Applicator with Rectal Retractor Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: Class II Product Code: JAO, IWJ Dated: January 22, 2020 Received: January 29, 2020
Dear Mr. Hurlman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200221
### Device Name
2/3 Endometrial Applicator Sets, Segmented Vaginal Applicator, HDR Compatible Tandem and Ovoid Applicators, CT Compatible F/S/D Applicators, HDR CT Compatible Split Ring Applicator, and CT HDR Tandem Ring Applicator with Rectal Retractor
### Indications for Use (Describe)
2/3 Endometrial Applicator Sets: A manual radionuclide applicator system, a manually operated device intended to apply a radionuclide source into the body or to the surface of the body for radiation therapy.
Segmented Vaginal Applicator: The use of sealed Radioisotopes to treat tumors within the body has been documented and published since the turn of the century. Modem era Radiation Therapy has developed delivery systems using isotopes of Cesium, Iridium, Iodine, and Gold to name a few examples. Many tumors are now treated by internal exposure to radiation emitted from sealed radioactive sources.
Two common modalities of this are Low Dose Rate remote afterloading (Brachytherapy). One common use of Brachytherapy is in the treatment of cancer of the vaginal process. The system described in this 510(k) has been developed to function as an applicator for the positioning of sealed sources in the intracavitary treatment of the vagina.
HDR Miami Applicator: This applicator is designed as an accessory to the Varisource System (Varian Associates K952913) and the Gammamed System (K89113) which uses a single radioactive source of Iridium-192 to treat cancer in a wide range of body sites. The Miami Applicator is placed in the vicinity of the cervix via the vagina just as described for the predicate device (Nucletron Miami Vaginal Applicator, K953946) and different diameter sleeves and intrauterine tubes, can be optimized to best meet the clinical needs of the patient along with minimization of dose to the mucosa.
HDR Compatible Tandern and Ovoid Applicators: The applicators presented in this 510(k) notification have been developed to function as Applicators for the positioning of HDR Remote After-Loader sealed sources in the intra-cavitary treatment of cancer of the vagina and cervix.
CT Compatible F/S/D Applicators: CT Compatible F/S/D applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice.
HDR CT Compatible Split Ring Applicator: The HDR CT Compatible Split Ring type applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of cancer in the cervix and uterus is an accepted clinical practice.
CCT HDR Tandem Ring Applicator with Rectal Retractor: The Mick Radio-Nuclear Instruments, Inc. CT HDR Tandem/ Ring Applicator with Rectal Retractor is indicated for High Dose Rate irradiation of the uterus and cervix.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:10pt"><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Special 510(k) Summary for compliance with CFR 807.92
March 3, 2020
### 1. General Provisions
| Common Name: | Radionuclide brachytherapy source (JAQ) and Manual<br>radionuclide applicator system (IWJ). |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | The GYN 1 Applicator Family - Intracavitary<br>- CT/MR Fletcher,<br>- CT/MR Henschke FSD,<br>- CT/MR Split Ring Applicator Sets<br><br>The GYN II Applicator Family<br>- CT/MR Segmented Vaginal,<br>- CT/MR Miami,<br>- CT/MR 2/3 Channel Endometrial,<br>- CT/MR Ring & Tandem, including Vienna Accessory Kit |
| Owner Name | Mick Radio-Nuclear Instruments, Inc. |
| Address | 521 Homestead Avenue<br>Mount Vernon, New York 10550 |
| Telephone | (914) 667-3999 |
| Fax | (914) 665-8834 |
| Contact Name | Isaac Strickland |
| Contact email | isaac.strickland@micknuclear.com |
{6}------------------------------------------------
### 2. Name of Predicate Device:
| Original<br>510(k) # | Trade/ Device Name | Regulation<br>Number | Regulatory<br>Classification | Product<br>Code | Review Panel |
|----------------------|--------------------------------------------------------|----------------------|------------------------------|-----------------|--------------|
| K993472 | Segmented Vaginal Applicator | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K020176 | HDR Miami Applicator HDR Brachytherapy Applicator | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K040704 | HDR Compatible Tandem and Ovoid Applicators (Henschke) | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K063381 | CT F/S/D Compatible Applicators | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K063382 | HDR CT Compatible Split Ring Applicator | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K122840 | CT HDR Tandem/Ring Applicator with Rectal Retractor | 21 CFR 892.5700 | Class II | JAQ | Radiology |
| K871216 | 2/3 Endometrial Applicator Sets | 21 CFR 892.5650 | Class I | IWJ | Radiology |
### 3. Reason for this Special 510(k) Premarket Notification
The purpose for this bundled submission is the modification of the labeling and Instructions for Use for each of the listed devices to add MR Conditional labeling to each of them.
### 4. Classification
This devices are classified as listed in the table above.
### 5. Intended Use and Device Description
2/3 Channel Endometrial Applicator Sets: The 2/3 Channel Endometrial Applicator is designed for the treatment of the endometrium and cervix. The applicator set includes 3 intrauterine tubes that are available in 3 cm and 5 cm configurations. These intrauterine tubes can be assembled using only 2 or all 3 intrauterine tubes, depending on the patient requirements. The only change to the device will be the labeling that the device is now MR Conditional.
Henschke Applicator Set: The Henschke Applicator is based on "Henschke style" geometry for brachytherapy of the endometrium and cervix. This applicator is similar in design to the Fletcher Applicator with the exception of semispherical ovoids. These ovoids are ideal for patients with narrow vaginas due to their small size, the ovoid tube is closer to the surface on the side of the cervix while maintaining distance within the fornices. The tandem is inserted into the endometrium at a pre-determined depth and secured in place with the cervical stop. The only change to the device will be the labeling that the device is now MR Conditional.
Fletcher (FSD) Applicator Set: The Fletcher Applicators are based on "Fletcher style" geometry for brachytherapy of the endometrium and cervix. The ovoids are cylindrical to create equal spacing in the vaginal fornices for a symmetrical dose distribution in this area. The tandem is inserted to into the endometrium at a pre-determined depth and secured in place with the cervical stop. The only change to the device will be the labeling that the device is now MR Conditional.
{7}------------------------------------------------
#### 5. Intended Use and Device Description con't.:
Miami Applicator Set: The Miami Applicator is designed for treatment of the vagina and cervix and includes an intrauterine tube for the treatment of the endometrium, as required. The design incorporates 7 treatment channels around the periphery of the cylinder body and includes a center channel that can accommodate an intrauterine tube. The intrauterine tubes are available with either a 30° angled tip or in a straight (0° angle) configuration. A stump plug is provided to seal the end of the applicator when an intrauterine tube is not required. Build-up caps are available for the cylinder body to provide added spacing between the mucosa and radioactive source. This applicator provides additional treatment options, with the peripheral channels, to increase the dose to the target area while limiting the dose to normal tissue. The only change to the device will be the labeling that the device is now MR Conditional.
Ring & Tandem Applicator Sets: The Ring & Tandem Applicator is based on the Stockholm technique for brachytherapy of the endometrium and cervix. The applicator consists of a ring tube and intrauterine tube that, when combined, create a fixed geometry and form a 90° angle. Build up caps of different diameters are provided for the ring tube that provide additional spacing between the radioactive source and mucosa. Including the Vienna Ring and Tandem Accessory Kit. The only change to the device will be the labeling that the device is now MR Conditional.
Segmented Vaginal Applicator Set: The Segmented Vaginal Applicator is designed for treatment of the vagina and cervix and includes an intrauterine tube for the treatment of the endometrium, as required. The intrauterine tubes are available in different angles and lengths to meet anatomical requirements. The four individual segment design allows the applicator length to be adjustable based on the treatment volume. Additionally, the segments are available in different diameters to match the patient's anatomy. The only change to the device will be the labeling that the device is now MR Conditional.
Split Ring Applicator Sets: The Splicator's patented design combines the benefits of several other intracavitary applicators. The ring tubes can be configured as a closed/complete ring or separated/" split' into four different distances, symmetrically or asymmetrically with the spacing bracket. Build-up caps of different diameters fit over the ring tubes to provide additional tissue spacing distance between the radioactive source and the cervix. The only change to the device will be the labeling that the device is now MR Conditional.
Vienna System: The Mick Radio-Nuclear Instruments, Inc. Vienna System is intended to be used as an accessory to the Mick CT HDR Tandem I Ring Applicator and is indicated for High Dose Rate irradiation of the uterus and cervix. The Vienna System consists of perforated Build-Up Caps and complementary Needle Collectors which connect to the Mick CT HDR Tandem Ring Applicator. Pre-bent interstitial needles are intended to be used with the Vienna System but they
{8}------------------------------------------------
are not manufactured by Mick Radio-Nuclear Instruments. Inc. and are not part of this submission.
## Build-Up Caps
When used with the CT HDR Tandem/Ring Applicator, the Vienna Build-Up Caps enable the introduction of up to nine (9) interstitial needles around the circumference of the ring to enhance the standard HDR treatment. The Vienna System is not designed to be used with any Rectal Retractor due to the introduction of the interstitial needles.
## Needle Collectors
When used in conjunction with the Vienna Build-Up Caps, the Needle Collectors will maintain and control the positioning of an array of up to nine (9) interstitial needles. The Needle Collectors are part of the Vienna System and as such, are not designed to be used with any
### Rectal Retractor.
The Vienna System is designed to be used as an accessory with the Applicator and this does not alter the indications for use and is MR Conditional.
#### 6. Drawings
Drawings of the devices are in the Device Description section of this submission.
#### 7. Manufacturing Process
This device is manufactured according to Good Manufacturing Practices (GMPs) as defined in 21CFR part 820. The processes used to fabricate these devices are identical to those used for the predicate device described in this 510(k) notification.
#### 8. Biocompatibility
No new issues of biocompatibility are raised with regard to the modification of the devices for MR Conditional labeling.
#### 9. Summary of Substantial Equivalence
There have been no changes to the devices or their intended use. Performance Test Results
### In vitro Testing
Not applicable to the modification for the device. By introducing this device, no new issues of safety or effectiveness are raised.
### In vivo Testing
Not applicable for the modification for the device.
## Functional Testing
Computational Modeling of Mick Radio-Nuclear Instruments HDR Brachytherapy Intracavitary Applicators for Radiofrequency-Induced Heating Evaluation Final Report. Results = MR Conditional.
{9}------------------------------------------------
#### 10. Summary of Similarities and Differences
These devices are identical in design, manufacture, construction and materials and have the same intended use and performance characteristics to their original 510(k)'s. The modification to the devices will be to put MR Conditional labeling on both the individual product labeling and the IFU's. The fundamental scientific technology is unchanged from the original predicate device.
#### 11. Comparison Table
The Table below compares the modification of these devices to the predicate device (Original 510(k). Included in these tables are a comparison of the materials and intended uses of these devices.
| Original<br>510(k) # | Trade/ Device Name | Predicate Device<br>(Original 510(k)) | Modified Device |
|----------------------|--------------------------------------------------------|---------------------------------------|-------------------------------|
| K993472 | Segmented Vaginal Applicator | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing - Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K020176 | HDR Miami Applicator HDR Brachytherapy Applicator | Materials - MR Conditional | Same |
| | | MR Testing – Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K040704 | HDR Compatible Tandem and Ovoid Applicators (Henschke) | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K063381 | CT F/S/D Compatible Applicators | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K063382 | HDR CT Compatible Split Ring Applicator | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K122840 | CT HDR Tandem/Ring Applicator with Rectal Retractor | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
| K871216 | 2/3 Endometrial Applicator Sets | Materials - MR Conditional | Same |
| | | MR Testing - Not Complete | MR Testing – Complete |
| | | MR Conditional Labeling - No | MR Conditional Labeling - Yes |
| | | Intended Use | Same |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.