ClearEndoclip is intended to be used with FDA-cleared endoscope for endoscopic clip placement within the gastrointestinal tract. It is indicated to be used for (1) Endoscopic marking (2) Hemostasis for (a) Mucosal/sub-mucosal defects < 3cm (b) Bleeding ulcers (c) Arteries < 2 mm (d) Polyps < 1.5 cm in diameter (e) Diverticula in the colon (3) Anchoring to affix jejunal feeding tubes to the wall of the small bowel (4) As a supplementary method, closure of GI tract luminal perforations <20 mm that can be treated conservatively
Device Story
ClearEndoclip is a sterile, single-use endoscopic clipping device used for hemostasis, marking, and tissue anchoring within the gastrointestinal tract. The device consists of a stainless steel clip pre-loaded into a delivery system comprising a handle, catheter, and slider mechanism. Operated by a physician via an endoscope, the user pushes the slider to open the clip and pulls it to close or reposition; the clip can be opened and closed up to five times before final deployment. Once positioned, the clip provides physical force to hold tissue together, mimicking surgical sutures. The device is MR Conditional, allowing for use in 3.0T MR environments. It benefits patients by enabling non-surgical, minimally invasive closure of mucosal defects and bleeding sites under direct visualization.
Clinical Evidence
Bench testing only. Performance data included magnetically induced displacement force (ASTM F2052-15), torque (ASTM F2213-17), MR image artifact (ASTM F2119), RF-induced heating (ASTM F2182), and MR pull-off force testing. All tests met established criteria.
Indicated for patients requiring endoscopic marking, hemostasis for GI defects/ulcers/arteries/polyps/diverticula, anchoring of jejunal feeding tubes, or closure of GI luminal perforations <20mm.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
October 5, 2020
Finemedix Co., Ltd. % Kyungyoon Kang CEO K-Bio Solutions 589 Oakwood Drive Santa Clara, CA 95054
Re: K200217
Trade/Device Name: ClearEndoclip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal ligator Regulatory Class: II Product Code: PKL, FHN, MND Dated: August 26, 2020 Received: September 1, 2020
Dear Kyungyoon Kang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200217
Device Name
ClearEndoclip
Indications for Use (Describe)
ClearEndoclip is intended to be used with FDA-cleared endoscopic clip placement within the gastrointestinal tract. It is indicated to be used for
- (1) Endoscopic marking
- (2) Hemostasis for
- (a) Mucosal/sub-mucosal defects < 3cm
- (b) Bleeding ulcers
- (c) Arteries < 2 mm
- (d) Polyps < 1.5 cm in diameter
- (e) Diverticula in the colon
- (3) Anchoring to affix jejunal feeding tubes to the wall of the small bowel
(4) As a supplementary method, closure of GI tract luminal perforations <20 mm that can be treated conservatively
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
#### I. SUBMITTER
Company: Finemedix Co., Ltd. 60, Maeyeo-ro, Dong-gu, Daegu Postal code: 41065, Republic of Korea Tel: 82-053-741-8388 Fax: 82-053-741-8168
510(k) Correspondent: Kyungyoon Kang, CEO: K-Bio Solutions (Tel: 812-345-7485, Kyungyoon.kang@kbiotechsolutions.com)
Manufacturer Contact: Heon-Sik Lee, RA Manager, Finemedix (hslee@finemedix.com) Date Prepared: September 30th , 2020
#### II. DEVICE
Trade Name: ClearEndoclip Common Name: Endoscopic Clipping Device Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Product Code: PKL (hemostatic metal clip for the GI tract), FHN (ligator, hemorrhoidal), MND (ligator, esophageal) Regulatory Class: II
#### III. PREDICATE DEVICE
Predicate Device: ClearEndoclip (K183021), Manufacturer: Finemedix Co., Ltd. This predicate has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
ClearEndoclip is a sterile device consisting of a pre-loaded, single-use, endoscopic clipping device with two main components: the (clip-fixing) delivery system and the clip. Clip is preloaded in the clip fixing delivery system, connected with an operation wire. Clip will open when the slider of the clip fixing device is pushed, and closed when it is pulled towards the operator. The clip could be closed and re-opened up to five times. When the slider is pulled further, the clip closes completely. The clip will be released when the slider is pushed. The delivery system consists of a handle and delivery catheter. The delivery system is constructed using stainless steel. HDPE outer sheath, and polypropylene stopper materials. The delivery system will allow
{4}------------------------------------------------
for the device to rotate at the distal end. The ClearEndoclip is offered in 165cm and 230cm working lengths.
ClearEndoclip consists of two main components, first, the endoscopic clip, which gets physically deployed and placed as a hemostatic purpose clip in the patient's gastrointestinal tract and second, the delivery system, known as a clip fixing device used to deliver the endoscopic clip under the use of an endoscope.
| Component Name | Function |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clip | |
| Clip | It is detached from the Inner Sheath and holds the tissue by physical<br>force and fixes it to perform hemispheres, indication of lesion and<br>treatment of perforation substantially for the user's intended purpose. |
| Clip Ring | The clip ring moves according to the slider forward / backward. The<br>Clip is opened / closed by the Clip Ring, and when the Slider is pushed<br>all the way, the Clip is clamped. |
| Delivery System/Clip Fixing Device | |
| Inner Sheath (Coil) | The Clip is attached to the end of the Inner Sheath and is connected to<br>the handle to transmit the force and movement from the handle to the<br>Clip. |
| Outer Sheath<br>(Polyethylene<br>Catheter Tube) | A plastic wrapping around the Inner Sheath and Clip helps protect the<br>endoscope channel. |
| Tube Joint | When using the product, the outer sheath moves forward and backward<br>to expose the clip from the outer sheath, and to recover the inner sheath<br>into the outer sheath when the product is recovered. |
| Stopper | It prevents the outer sheath from moving in the direction of Handle<br>when the product is moved and stored, thereby preventing the clip from<br>opening. |
| Slider | It is used for open/close and detachment of clip. |
| Handle | It serves to fix this device with user's hand |
Functional Descriptions for Critical Components of ClearEndoclip
As far as the raw material compositions are concerned, Clip of ClearEndoclip is constructed with stainless steel material and deployed from the delivery system during the clinical use. ClearEndoclip is engineered such that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy.
{5}------------------------------------------------
#### V. INDICATIONS FOR USE
ClearEndoclip is intended to be used with FDA-cleared endoscope for endoscopic clip placement within the gastrointestinal tract. It is indicated to be used for
- (1) Endoscopic marking
- (2) Hemostasis for
- (a) Mucosal/sub-mucosal defects < 3cm
- (b) Bleeding ulcers
- (c) Arteries < 2 mm
- (d) Polyps < 1.5 cm in diameter
- (e) Diverticula in the colon
- (3) Anchoring to affix jejunal feeding tubes to the wall of the small bowel
- (4) As a supplementary method, closure of GI tract luminal perforations <20 mm that can be treated conservatively
- COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDI VI. CATE DEVICE
Fundamental technological characteristics of the ClearEndoclip are the same as those of the predicate device, ClearEndoclip. The operation principle, clinical applications, and instructions for use of the ClearEndoclip is the same as the predicate device, as both devices are used for endoscopic marking, hemostasis for mucosal/submucosal defects in digestive tract by placing an endoscopic clip for the purpose of achieving hemostasis for gastrointestinal bleeding. Both the proposed and predicate devices are provided sterile with ethylene oxide sterilization and used for single use only. Both the proposed and predicate devices consist of the endoscopic clip, which gets physically deployed and placed as hemostatic purpose clip in the patient's gastrointestinal tract and the delivery system, known as a clip fixing device used to deliver the endoscopic clip under the use of an endoscope.
The clinical application and operation principle of the proposed, ClearEndoclip are the same as the predicate device as the stainless mechanical clipping device is used in endoscopy in order to close two mucosal surfaces without the need for surgery or suturing. For both the proposed ClearEndoclip and predicate device, from the mechanical operation- point of view, the endoscopic clipping function is similar to a suture in gross surgical hemostatic applications as it is used to join together disjointed mucosal surfaces that lead to bleeding, however, ClearEndoclip can be applied through the channel of an endoscope under direct visualization, based upon the same clinical and operation principle as the predicate device.
Both ClearEndoclip and predicate device consist of two main components, first, the endoscopic clip, which gets physically deployed and placed as a hemostatic purpose clip in the patient's gastrointestinal tract and second, the delivery system, known as a clip fixing device used to deliver the endoscopic clip under the use of an endoscope.
{6}------------------------------------------------
Using the delivery system (a clip fixing device) that utilizes the same technological characteristics as the predicate device, in terms of the delivery mechanism to facilitate release of the endoscopic clip. the Delivery System is designed to control closing and re-opening of the endoscopic clip. As the same as the predicate device, a clip of ClearEndoclip is pre-loaded in its Delivery System (clip fixing device), connected with an operation wire. The manipulation mechanism of opening or closing the clip is also the same as the predicate device, as the ClearEndoclip will open when the handle slider placed in the delivery system (clip fixing device) is pushed and closed when the handle slide is pulled towards the operator. The clip could be closed and re-opened up to five times. When the slider of the delivery system is pulled further, the clip closes completely. The clip will be released when the slider is pushed. As demonstrated herein, the fundamental technological characteristics of ClearEndoclip including the device system construction and operation mechanism, which have been designed in order to meet the indications and user needs are the same as the predicate device.
As for technological characteristics, the raw material constructions of ClearEndoclip are the same as the predicate device. The clinical application and operation principle of the proposed. ClearEndoclip are the same as the predicate device as the stainless mechanical clipping device is used in endoscopy in order to close two mucosal surfaces without the need for surgery or suturing. For both the proposed ClearEndoclip and predicate device, from the mechanical operation- point of view, the endoscopic clipping function is similar to a suture in gross surgical hemostatic applications as it is used to join together disjointed mucosal surfaces that lead to bleeding, however, ClearEndoclip can be applied through the channel of an endoscope under direct visualization, based upon the same clinical and operation principle as the predicate device.
Additional MR Testing which demonstrates Finemedix ClearEndoclip will remain in place when exposed to worst case magnetically induced displacement force and torque in 3.0T MR system is being proposed in order to meet the FDA's requirements for "MR Conditional" This particular labeling update which is addressed by the additional MR testing verification does not raise different questions in terms of safety and effectiveness compared to the predicate device.
| | ClearEndoclip | Predicate Device:<br>ClearEndoclip<br>(K183021) |
|-----------------|----------------|-------------------------------------------------|
| MRI Information | MR Conditional | MR Unsafe |
The aforementioned raw material difference with the predicate device is not assessed to raise different questions of safety and effectiveness as ClearEndoclip also demonstrated its conformance to the FDA's same recognized consensus standards applied to both the predicate and reference devices.
{7}------------------------------------------------
#### VII. PERFORMANCE DATA
The following performance data for MR-related verification testing were provided in support of this substantial equivalence determination.
| Testing | Test Objective/Standard | Result |
|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Magnetically<br>Induced,<br>Maximum<br>Displacement<br>Force | Evaluation of the magnetically induced<br>displacement force was completed in<br>accordance with standard, ASTM F2052-<br>15. | Pass,<br>Met criteria |
| Magnetically<br>Induced<br>Maximum<br>Torque | Magnetically induced torque evaluation<br>was completed in accordance with<br>ASTM F2213-17. | Pass, Met<br>criteria |
| MR Image<br>Artifact Test | MR image artifact test was completed in<br>accordance with ASTM F2119. | Pass, Met<br>criteria |
| Radiofrequency<br>(RF) Induced<br>Heating | RF induced heating test was completed in<br>accordance with ASTM F2182. | Pass, Met<br>criteria |
| MR Pull-off<br>Test | Evaluation of the force needed to pull the<br>ClearEndoclip off the attached tissue in<br>comparison to the maximum<br>magnetically induced force was<br>completed. | Pass, Met<br>criteria |
MR Performance Testing of ClearEndoclip
#### VIII. CONCLUSIONS
The proposed, ClearEndoclip is assessed substantially equivalent to the predicate device given the fact that its indications for use/intended use are identical, and fundamental technological characteristics are the same as the predicate device. The favorable results of the aforementioned safety and performance testing demonstrate conformance to the appropriate recognized standards of FDA, and further demonstrate that no different questions in safety and effectiveness assessment are being raised compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.