K200216 · Scivita Medical Technology Co., Ltd. · HET · Aug 21, 2020 · Obstetrics/Gynecology
Device Facts
Record ID
K200216
Device Name
4K UHD Camera System
Applicant
Scivita Medical Technology Co., Ltd.
Product Code
HET · Obstetrics/Gynecology
Decision Date
Aug 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1720
Device Class
Class 2
Indications for Use
4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery. 4K Camera Control Unit The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only. 4K Camera Head The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
Device Story
4K UHD Camera System captures endoscopic images using a CMOS sensor-equipped camera head; converts optical input to electronic signals; processes signals via a 4K camera control unit (CCU). System outputs 4K/2K video via HDMI/SDI to external monitors for real-time visualization. Used in hospitals/clinics by physicians/trained professionals during endoscopic surgery. Provides high-resolution visualization of operative area; aids clinical diagnosis, treatment, and observation. Data can be recorded to external USB devices. Reusable system; requires cleaning/disinfection between uses.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via depth of field, resolution, direction/field of view, color reproduction, signal-to-noise ratio, geometric distortion, frame frequency, system delay, and image intensity uniformity tests. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and reprocessing standards (AAMI TIR 30, AAMI TIR 12) confirmed.
Technological Characteristics
Reusable endoscopic imaging system. Components: 4K camera head (1/3" 3-chip CMOS), CCU (models ES-CS4K100/200). Connectivity: HDMI/SDI video output. Power: 100-240V AC. Reprocessing: Disinfection via glutaraldehyde. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM D4169, AAMI TIR 30/12.
Indications for Use
Indicated for use in endoscopic surgery to provide imaging of the operative area for diagnosis, treatment, and observation. Intended for use in hospitals, clinics, and doctor's offices by physicians and trained healthcare professionals.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 21, 2020
Scivita Medical Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co, Ltd P.O. Box 120-119 Shanghai. 200120 CHINA
Re: K200216
> Trade/Device Name: 4K UHD Camera System Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: II Product Code: HET, EOB, FGB, GCJ, NWB, FET, HRX Dated: July 9, 2020 Received: July 22, 2020
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200216
Device Name 4K UHD Camera System
#### Indications for Use (Describe)
4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery.
## 4K Camera Control Unit
The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only.
4K Camera Head
The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements per 21 CFR 807.92.
1. Sponsor Identification
Sponsor Identification: Scivita Medical Technology Co., Ltd. No.8, Zhong Tian Xiang, Suzhou Industrial Park, Suzhou, Jiangsu, 215000, China.
Establishment Registration Number: Not yet registered
Contact Person: Ruqin Wu Position: Quality Manager Tel: +86-512-81877788 Fax: +86-512-85187285 Email: wuruqin@scivitamedical.com
#### Designated Submission Correspondent:
Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
Date of Preparation: August 20, 2020
{4}------------------------------------------------
## 2. Device Information
Trade Name: 4K UHD Camera System Common Name: Endoscopic Video Imaging System Model: ES-CS4K100/100C ES-CS4K200/100C Regulation Number: 21 CFR 884.1720 Regulation Name: Gynecologic Laparoscope And Accessories Regulatory Class: II Product Code: HET, EOB, FGB, GCJ, NWB, FET, HRX
#### 3. Identification of Predicate Device
510(k) Number: K151011 Product Name: Olympus Medical Systems Corp. The predicate device has not been subject to a design-related recall.
#### 4. Device Description
The proposed system, 4K UHD Camera System, comprises the 4K camera head, objective lens and power cable and video cables. The proposed system is reusable device, and provided non-sterile. The proposed camera head and should be cleaned and disinfected after each use.
The video cables include an SDI video cable and an HDMI video cable. The 4K camera control unit has two models (ES-CS4K100 and ES-CS4K200) with the only difference on signal output terminals, the detail difference refers to Table 3 of this document; the 4K camera head only has one model (ES-CS4K100C). Therefore, the 4K UHD Camera System is available in two models in combination of the 4K camera control unit and 4K camera head. The system model and primary components of the proposed system are provided in Table 1.
| 4K UHD Camera System | System model | Primary component |
|----------------------|-----------------|---------------------------------------------------------------------|
| | ES-CS4K100/100C | 4K camera control unit (ES-CS4K100)<br>4K camera head (ES-CS4K100C) |
| | ES-CS4K200/100C | 4K camera control unit (ES-CS4K200)<br>4K camera head (ES-CS4K100C) |
Table 1 System model and primary component
Working Principle and/or Mechanism of Action:
The 4K UHD Camera System can provide 4K images via different signal output terminals. The 4K camera head incorporates a complementary metal oxide semiconductor (CMOS) image sensor to convert optical images into electronical signals. The objective lens on end of the 4K camera head is used to connect the camera head with the endoscope.
{5}------------------------------------------------
The 4K UHD Camera System is designed to be used with endoscopes, a light source (ES-LS110D), monitors, light guide cables and other ancillary equipment for endoscopic diagnosis, treatment and observation.
# Conditions of Use:
4K UHD Camera System is intended to be used in hospitals, clinics and doctor's office by doctors and trained healthcare professionals. Compatible equipment of the proposed system is provided in the Table 2.
| Equipment | Manufacturer | Model | K Number |
|------------------------------------------------------------------------------------------------------------------------|--------------|----------------------------------------------------------|------------------|
| 4K UHD Camera System | SCIVITA | ES-CS4K 100/100C | This application |
| HSV LAPAROSCOPY SET<br>(optional select one from all endoscope<br>when connecting with the proposed<br>device to test) | WOLF | Laparoscope<br>(0°, 30°,45°) | K941541 |
| HENKE SASS WOLF OF AMERICA<br>ARTHROSCOPE | WOLF | Arthroscope<br>(0°,30°,70°) | K080560 |
| SINUSCOPE AND ACCESSORIES | WOLF | Sinuscope<br>(0°,30°,70°) | K981751 |
| LED LIGHT SOURCE | SCIVITA | ES-LS110D | |
| Monitor | SONY | LMD-X2705MD/<br>LMD-X550MD/<br>LMD-X310MD/<br>LMD-X310MT | K150377 |
Table 2 Recommended compatible equipment of the proposed system
The key performance specifications of the proposed system are provided in the Table 3.
## Table 3 General specifications
| Weight | 5Kg | |
|--------------------------------------|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimension(W×H×D) | 370 (W) ×95 (H)×425.5(D) mm | |
| Camera Head | 1/3" Three Chip CMOS | |
| Signal output (resolution) | 4096×2160, 1920×1080 | |
| Signal<br>output<br>(interface type) | ES-CS4K100/100C | There are HDMI×2, SDI-1 BNC terminal×4, SDI-2 BNC<br>terminal×1 seven interfaces in total. This model can output<br>the following three types: 4K SDI × 1, 2K SDI × 1, 4K<br>HDMI × 2 |
| | ES-CS4K200/100C | There are HDMI×2, SDI-1 BNC terminal×4, SDI-2 BNC<br>terminal×1 seven interfaces in total. This model can |
{6}------------------------------------------------
| | output the following four types: 4K SDI × 1, 2K SDI × 1,<br>4K HDMI × 1, 2K HDMI × 1 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Video output format | HDMI output: 2160/59.94p, 2160/50p, 1080/59.94p,<br>1080/59.94i, 1080/50p, 1080/50i<br>SDI output: 2160/59.94p, 2160/50p, 1080/59.94p,<br>1080/59.94i, 1080/50p, 1080/50i |
| Image recording | The data is stored in real time to the external access USB<br>interface device by photographing or recording functions. |
| Operation mode | Continuous operation |
| White balance | Automatic white balance |
The level of concern of the software contained in the subject device is determined to be Moderate Level of Concern.
## 5. Indications for use:
4K UHD Camera System is used to provide imaging of the operative area in endoscopic surgery.
## 4K Camera Control Unit
The 4K camera control unit has been designed to be used with endoscopes, 4K camera head, light source, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and observation. The 4K camera head is compatible with the 4K camera control unit only.
#### 4K Camera Head
The 4K camera head has been designed to be used with endoscopes, 4K camera control unit, and other ancillary equipment for endoscopic diagnosis, treatment, and observation
The indications for use statement of the subject device is similar to that of the video processor component of the predicate device. The main difference in the indication for use is the subject device doesn't include indications for the light source because a light source is not included in this subject device submissions for use of the camera control unit and camera head of the subject device and predicate device are the differences in the indications do not represent a new intended use.
{7}------------------------------------------------
6. Comparison of Technological Characteristics
| ITEM | Subject Device K200216 | Predicate Device K151011 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | 4K UHD Camera System | VISERA 4K UHD SYSTEM |
| Product Code | HET, EOB, FGB, GCJ, NWB, FET,<br>HRX | HET, GCJ, EOB, EOQ, FGB, NWB |
| Regulation Number | 21 CFR 884.1720 | 21 CFR 884.1720, |
| Class | II | II |
| Indications for Use | 4K UHD Camera System is used to<br>provide imaging of the operative area in<br>endoscopic surgery.<br><br>4K Camera Control Unit<br>The 4K camera control unit has been<br>designed to be used with endoscopes, 4K<br>camera head, light source, monitors, and<br>other ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br>The 4K camera head is compatible with<br>the 4K camera control unit only.<br><br>4K Camera Head<br>The 4K camera head has been designed<br>to be used with endoscopes, 4K camera<br>control unit, and other ancillary<br>equipment for endoscopic diagnosis,<br>treatment, and observation | VISERA 4K UHD CAMERA CONTROL<br>UNIT OLYMPUS OTV-S400<br>The camera control unit has been designed<br>to be used with Olympus endoscopes,<br>camera heads, light source, monitors, and<br>other ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br><br>VISERA 4K UHD XENON LIGHT<br>SOURCE OLYMPUS CLV-S400<br>The light source has been designed to be<br>used with Olympus endoscopes, camera<br>control unit, light guide cables, and other<br>ancillary equipment for endoscopic<br>diagnosis, treatment, and observation.<br><br>4K CAMERA HEAD OLYMPUS<br>CH-S400-XZ-EB<br>The camera head has been designed to be<br>used with Olympus endoscopes, camera<br>control unit, and other ancillary equipment<br>for endoscopic diagnosis, treatment, and<br>observation. |
| Single use / Reusable | Reusable | Reusable |
| Sterile | No | No |
| Configuration<br>(primary components) | 4K Camera Control Unit<br>4K Camera Head | light source<br>4K Camera Control Unit<br>4K Camera Head |
| ITEM | Subject Device K200216 | Predicate device |
| Model | ES-CS4K100<br>ES-CS4K200 | K151011<br>VISERA 4K UHD CAMERA CONTROL<br>UNIT OLYMPUS OTV-S400 |
| Power supply | 100-240V AC, 50/60Hz | 100-240V~ ±10%,<br>50/60Hz±1Hz |
| Dimension | 370(W) ×95(H)×425.5(D) mm | 390(W) × 160(H) × 506(D) mm |
| Weight | 5Kg | 13.5Kg |
| Input power | 150VA | 350VA |
| Resolution | 4096×2160pxl,<br>1920×1080pxl | 4096×2160,<br>3840×2160,<br>1920×1080 |
| Image<br>signal<br>output | ES-CS4K100 | Total seven signal output terminals<br>HDMI (4K HMDI) ×2,<br>SDI (4K SDI) terminal×4,<br>SDI (3G/HD-SDI) terminal×1;<br>The seven signal output terminals can<br>output three types of signal output: 4K<br>SDI , 2K SDI , 4K HDMI |
| | ES-CS4K200 | Total seven signal output terminals:<br>HDMI (4K HMDI) ×1,<br>HDMI (HMDI)×1,<br>SDI (4K) terminal×4,<br>SDI (3G/HD-SDI) terminal×1;<br>The seven signal output terminals can<br>output four types of signal output: 4K SDI, 2K SDI , 4K HDMI, 2K HDMI |
| Standard Color Chart<br>Output | Color bar image | Color bar image |
| Observation light imaging | White light imaging(WLI)<br>Six imaging function | White light imaging (WLI)<br>Narrow band imaging (NBI) |
Table 4 General Comparison
Table 5 Comparison of ES-CS4K100/200 Camera Control Unit
{8}------------------------------------------------
| Table 6 Comparison of Camera Head | |
|-----------------------------------|--|
| | |
| ITEM | Subject Device K200216 | Predicate Device K151011 |
|-----------------------|---------------------------|-----------------------------------------|
| Model | ES-CS4K100C | 4K CAMERA HEAD OLYMPUS<br>CH-S400-XZ-EB |
| Camera Head dimension | W48 mm × H64 mm × L168 mm | W43.6 mm × H49.5 mm × L122.5 mm |
| Cable dimension | φ5mm× 2.9m | 5.1 mm x 3 m |
{9}------------------------------------------------
| Wight(excluding cable) | 231.4g (excluding cable) | 280 g (excluding cable) |
|------------------------|--------------------------------------------------|-------------------------|
| Reprocessing | End user reprocess | End user sterilized |
| | Disinfection by using glutaraldehyde<br>solution | EOG/ STERRAD |
The differences in technological characteristics do not raise different questions of safety and effectiveness.
- 7. Summary of Non-Clinical Performance Testing
Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > IEC 60601-1:2005+AMD1:2012 Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance.
- > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests.
- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular requirements for the basic safety and > essential performance of endoscopic equipment
- > ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Systems.
- > AAMI TIR 30:2016 A compendium of processes, materials, and acceptance criteria for cleaning reusable medical devices
- > AAMI TIR 12:2010 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers
The software of the proposed device was validated as Moderate level of concern (LoC) in accordance with the following guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices -Guidance for Industry and FDA Staff
The following performance testing were also conducted on the subject device and the subject device met all predefined acceptance criteria:
- A Depth of Field Test
- A Resolution Test
- A Direction of View and Field of View Test
- Color Reproduction and Signal Noise Ratio Test >
- > Geometric Distortion Test
- > Image Frame Frequency and System Delay
- > Image Intensity Uniformity
{10}------------------------------------------------
# 8. Conclusion
The performance testing summarized above support a substantial equivalence determination. The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.