The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is a sensorenabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart. The catheter is used with the EnSite Precision™ System to combine and display magnetic processed patient positioning and navigation mapping information.
Device Story
Steerable, flexible electrophysiology catheter; used for recording intracardiac signals and cardiac stimulation; maps atrial regions of heart. Input: intracardiac electrical signals and magnetic positioning data. Operation: physician manipulates handle-based control mechanism to deflect distal loop; catheter integrates with EnSite Precision System (impedance and magnetic field technology) to display 3D cardiac mapping and patient positioning. Output: electrical signals for diagnostic analysis and navigation data for 3D mapping. Used in clinical electrophysiology labs by physicians. Benefits: enables precise cardiac mapping and signal acquisition for diagnostic studies. Reprocessed device limited to one cycle; includes cleaning, decontamination, and functional testing.
Thermoplastic elastomer construction; noble metal electrodes. Bi-directional steerable shaft with handle-based deflection. Sensor-enabled for magnetic navigation. Compatible with EnSite Precision System. Reprocessed via cleaning and decontamination; sterilized. Single-use limit (max 1 cycle).
Indications for Use
Indicated for use in diagnostic electrophysiology studies for recording intracardiac signals, cardiac stimulation, and mapping of atrial regions of the heart in patients undergoing such procedures.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
May 13, 2020
Stryker Sustainability Solutions Mia McCorkel Senior Regulatory Affairs Specialist 18101 W Drake Drive Tempe, Arizona 85283
Re: K200205
Trade/Device Name: Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: April 14, 2020 Received: April 15, 2020
Dear Mia McCorkel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Model numbers intended for reprocessing:
| Model Number | Description | Curve | Number of<br>Loop<br>Electrodes | Electrode<br>Spacing | Loop<br>Diameter |
|-----------------|------------------------------------------------|-------|---------------------------------|----------------------|------------------|
| D-AVSE-DF10-F15 | Bi-Directional<br>Circular Mapping<br>Catheter | DF | 10 | 3-3-3 | 15 |
| D-AVSE-DF10-F20 | Bi-Directional<br>Circular Mapping<br>Catheter | DF | 10 | 5-5-5 | 20 |
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K200205
Device Name
Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled
Indications for Use (Describe)
The Reprocessed Advisor FL Circular Mapping Catheter, Sensor Enabled is a sensor-enabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart.
The catheter is used with the EnSite Precision System to combine and display magnetic processed patient positioning and navigation mapping information.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "stryker" is in bold, black, sans-serif font on the top line. Below that, the words "Sustainability Solutions" are also in bold, black, sans-serif font, but in a smaller size.
# SECTION 5: 510(k) SUMMARY
## Submitter:
Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283
## Contact:
Mia McCorkel Senior Regulatory Affairs Specialist 480-343-1855 (c) 480-763-2965 (f) mia.mccorkel@stryker.com
#### Date of Preparation: January 27, 2020
### Name of Device:
| | Trade/Proprietary Name: | Reprocessed Advisor FL™ Circular Mapping Catheter,<br>Sensor Enabled™ |
|--|-------------------------|-----------------------------------------------------------------------------------------------------|
| | Common Name: | Diagnostic Electrophysiology Catheter, Electrode Recording<br>Catheter or Electrode Recording Probe |
| | | Classification Information: Cardiovascular (21 CFR § 870.1220, NLH, Class II) |
| | Model Numbers: | D-AVSE-DF10-F15, D-AVSE-DF10-F20 |
### Predicate Device:
| Model Numbers | 510(k) Number | 510(k) Title | Original Manufacturer |
|------------------------------------|---------------|---------------------------------------------------------|-----------------------|
| D-AVSE-DF10-F15<br>D-AVSE-DF10-F20 | K160335 | Advisor FL Circular Mapping<br>Catheter, Sensor Enabled | St. Jude Medical |
### Device Description:
The Reprocessed Advisor™ FL Circular Mapping Catheters, Sensor Enabled™ are steerable, flexible, insulated electrophysiology catheters constructed of thermoplastic elastomer material and noble metal electrodes. The shaft curvature is manipulated by the control mechanism located on the handle at the catheter's proximal end. To adjust the curve on the bi-directional catheter, use the actuator to deflect the catheter in either direction. The distal loop is oriented counter-clockwise as viewed from the handle.
K200205
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "stryker" is written in a bold, sans-serif font in black. Below the word "stryker" is the phrase "Sustainability Solutions," also written in a bold, sans-serif font in black. The logo is simple and modern, and it conveys a sense of strength and innovation.
This device is compatible with the EnSite Precision™ System which is a three dimensional (3-D) cardiac mapping system that combines impedance and magnetic field technology.
| Model Numbers | Description | Curve | Loop Diameter |
|-----------------|---------------------------------------------------------------------|-------|---------------|
| D-AVSE-DF10-F15 | Reprocessed Advisor FL Circular<br>Mapping Catheter, Sensor Enabled | DF | 15mm |
| D-AVSE-DF10-F20 | Reprocessed Advisor FL Circular<br>Mapping Catheter, Sensor Enabled | DF | 20mm |
The model numbers included in the scope of this submission are as follows:
# Intended Use:
The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is a sensorenabled steerable electrophysiology catheter used for recording intracardiac signals and cardiac stimulation during diagnostic electrophysiology studies. The catheter can be used to map the atrial regions of the heart.
The catheter is used with the EnSite Precision™ System to combine and display magnetic processed patient positioning and navigation mapping information.
# Summary of Technological Characteristics:
The design, materials, and intended use of Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ are equivalent to the predicate device. The mechanism of action of the reprocessed device is identical to the predicate device in that the same standard mechanical design, materials, and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. In addition, Stryker Sustainability Solutions' reprocessing of the device includes removal of adherent visible soil and decontamination. Each individual device is tested for appropriate function of its components prior to packaging and labeling operations.
# Performance Data:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™. This included the following tests:
- Biocompatibility ●
- Validation of Reprocessing ●
- Sterilization Validation
- Functional Performance Testing ●
- о EEPROM verification
- Location sensor testing o
- Visual inspection O
- Dimensional testing O
- Electrode continuity and isolation testing O
- Curve and in-plane deflection testing O
- Rigidity testing o
- Freedom from leakage O
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Stryker Sustainability Solutions. The word "Stryker" is in a bold, sans-serif font, with the letters slightly overlapping. Below "Stryker" is the phrase "Sustainability Solutions" in a smaller, bold, sans-serif font. The logo is black on a white background.
- o Tensile testina
- Locking mechanism testing
- . Electrical Safety Testing
The performance testing demonstrates that reprocessed devices are as safe and effective as the predicate and operate as originally intended.
The Reprocessed Advisor™ FL Circular Mapping Catheter, Sensor Enabled™ is reprocessed no more than one (1) time. Each reprocessed device is tracked with a serial number label which is affixed to the catheter shaft. Once the device reaches the maximum number or reprocessing cycles, it is rejected and taken out of service. Reprocessing is conducted only by Stryker Sustainability Solutions. Stryker Sustainability Solutions restricts its reprocessing to exclude devises previously reprocessed by other reprocessors.
# Conclusion:
The results of bench and laboratory testing demonstrate that the Reprocessed Advisor™ FL Circular Mapping Catheters, Sensor Enabled™ are at least as safe and effective and perform as well as the identified legally marketed predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.