← Product Code [NHA](/productcode/NHA) · K200189

# Luna Dental Implant System- Healing Abutment (K200189)

_Shinhung Mst Co., Ltd. · NHA · Feb 26, 2020 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200189

## Device Facts

- **Applicant:** Shinhung Mst Co., Ltd.
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Feb 26, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.

## Device Story

Healing abutment for Luna Dental Implant System; connects to previously cleared dental implant fixtures (K123155, K160106) to facilitate gingival healing prior to final abutment placement. Device is surgically placed in oral cavity by dental professionals. Composed of commercially pure titanium alloy (Grade 4); available in narrow and regular platform types with varying diameters and lengths. Narrow type features purple anodized surface for identification. Supplied sterile via gamma radiation. Acts as temporary interface to maintain gingival contour during healing phase; supports subsequent prosthetic restoration.

## Clinical Evidence

Bench testing only. Risk analysis conducted per ISO 14971 confirmed that design modifications did not negatively impact biocompatibility, sterilization, or shelf life, rendering prior predicate testing applicable.

## Technological Characteristics

Material: Commercially pure titanium alloy (Grade 4). Sterilization: Gamma radiation. Form factor: Healing abutment, narrow and regular platforms. Surface treatment: Anodizing (purple for narrow type). Connectivity: Mechanical interface with dental implant fixture.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- Luna Dental Implant System ([K123155](/device/K123155.md))
- TS IMPLANT SYSTEM ([K121585](/device/K121585.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 26, 2020

SHINHUNG MST Co., Ltd. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA. Inc. 1150 Roosevelt STE 200 Irvine, California 92620

Re: K200189

Trade/Device Name: Luna Dental Implant System - Healing Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 13, 2020 Received: January 27, 2020

Dear Priscilla Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Srinivas Nandkumar, Ph.D. Director Implantable Dental Devices Team DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

Device Name

Luna Dental Implant System - Healing Abutment

Indications for Use (Describe)

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.

| Type of Use (Select one or both, as applicable)                                                  |                                                                                                 |
|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary (K200189)

This summary of 510(K) information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: 02/26/2020

## 1. Submitter

SHINHUNG MST Co., Ltd. 110-2, Donghwagongdan-Ro Munmak-Eup Wonju Si, Gangweondo, Republic of Korea 220-801 Tel. +82-33-730-1901

## 2. Official Correspondent

Priscilla Chung LK Consulting Group USA, Inc. 1150 Roosevelt STE 200, Irvine CA 92620 Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com

## 3. Device

- . Trade Name: Luna Dental Implant System - Healing Abutment
- Common Name: Implant Abutment
- Classification Name: Endosseous Dental Implant Abutment
- Product Code: NHA
- . Classification regulation: 21CFR 872.3630

## 4. Predicate Device:

- Luna Dental Implant System (K123155) by SHINHUNG MST Co., Ltd.
- TS IMPLANT SYSTEM (K121585) by Osstem Co., Ltd.

## 5. Description:

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function. This submission is to get clearance on the modification of the healing abutments that are used with the fixture cleared under K123155 and K160106.

The healing abutment is to be connected to the implant and is to heal gingiva before

{4}------------------------------------------------

setting abutment on the implant in the oral cavity. It is made of commercially pure titanium alloy Gr4. It offers narrow and regular platform types, and the narrow type has anodizing surface treatment in purple to be distinguished from the regular type.

N (Narrow) Type

- . 4.3mm(Dia) x 9.2/10.2/11.2/12.2/13.2mm (Length)
- . 4.8mm(Dia) x 9.2/10.2/11.2/12.2/13.2mm (Length)

R (Regular) Type

- 4.3mm(Dia) x 10/11/12/13/14mm (Length) ●
- . 4.8mm(Dia) x 10/11/12/13/14mm (Length)
- 5.8mm(Dia) x 10/11/12/13/14mm (Length)
- 6.3mm(Dia) x 10/11/12/13/14mm (Length)

#### 6. Indication for use:

The Luna Dental Implant System is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, and to restore the patient's chewing function.

{5}------------------------------------------------

#### 7. Basis for Substantial Equivalence

The Luna Dental Implant System - Healing Abutment is substantially equivalent to its predicate device (K123155) in terms of indications for use, raw material, and design. The differences are in body design and size range. The design change does not raise a question in substantial equivalence since the change is not significant. We have identified a reference device that encompass the size range of the subject device. Based on the information submitted here in we conclude that the subject device is substantially equivalent to the predicate devices.

| Device Name            | Subject Device                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                               | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer           | SHINHUNG MST Co.,<br>Ltd                                                                                                                                                                                                                                       | SHINHUNG MST Co.,<br>Ltd                                                                                                                                                                                                                                       | Osstem Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 510(k) Number          | K200189                                                                                                                                                                                                                                                        | K123155                                                                                                                                                                                                                                                        | K121585                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Device Name            | Luna Dental Implant<br>System - Healing<br>Abutment                                                                                                                                                                                                            | Luna Dental Implant<br>System                                                                                                                                                                                                                                  | TS IMPLANT SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Form                   | Abutment                                                                                                                                                                                                                                                       | Implant and Abutment                                                                                                                                                                                                                                           | Implant and Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for<br>use | The Luna Dental Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such as<br>artificial teeth, and to<br>restore the patient's<br>chewing function. | The Luna Dental Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw arches to<br>provide support for<br>prosthetic devices, such as<br>artificial teeth, and to<br>restore the patient's<br>chewing function. | The TS Implant System<br>is indicated for use in<br>partially or roily<br>edentulous mandibles<br>and maxillae, in support<br>of single or multiple-unit<br>restorations including:<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is<br>intended for delayed<br>loading. The abutment is<br>intended for use with a<br>dental implant fixture to<br>provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or overdenture. TS<br>Implant System is<br>compatible with<br>abutment in the ET/SS<br>Implant System. |
| Sterilization          | It is supplied sterile<br>(Gamma radiation)                                                                                                                                                                                                                    | It is supplied sterile<br>(Gamma radiation)                                                                                                                                                                                                                    | It is supplied sterile<br>(Gamma radiation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |

{6}------------------------------------------------

|          | NP(Anodizing)                                                              | NP (Anodizing)                                                        | (Non-Anodizing)                                                         |
|----------|----------------------------------------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------|
| Shape    | Image: NP(Anodizing)                                                       | Image: NP (Anodizing)                                                 | Image: (Non-Anodizing)                                                  |
|          | RP(Non-Anodizing)                                                          | RP(Anodizing)                                                         |                                                                         |
| Material | Ti-Gr4                                                                     | Ti-Gr4                                                                | Ti-Gr4                                                                  |
| Size     | 1) N type<br>-Diameter: 4.3~4.8mm<br>-Height:<br>4.0, 5.0, 6.0, 7.0, 8.0mm | 1) N type<br>-Diameter: 4.5~6.0mm<br>-Height:<br>2.0, 3.5, 5.0, 7.0mm | 1) N type<br>-Diameter: 4.0~4.5mm<br>-Height: 3.0, 4.0, 5.0, 7.0, 9.0mm |
|          | 2) R type<br>-Diameter: 4.3~6.3mm<br>-Height:<br>4.0, 5.0, 6.0, 7.0, 8.0mm | 2) R type<br>-Diameter: 4.5~7.0mm<br>-Height:<br>2.0, 3.5, 5.0, 7.0mm | 2) R type<br>-Diameter: 4.0~8.0mm<br>-Height: 3.0, 4.0, 5.0, 7.0, 9.0mm |

#### 8. Non-Clinical Testing

Risk analysis was conducted according to ISO14971 to evaluate the effect of the modification. The risk assessment did not show that the device changes affected biocompatibility, sterilization, and shelf life such that the testing in the prior (predicate) file is still applicable.

#### 9. Conclusion

The subject devices and the predicate device have the same intended use and have the same technological characteristics.

Overall, the Luna Dental Implant System - Healing Abutment has the following similarities to the predicate device:

- * have the same intended use,
- * use the same operating principle,
- * incorporate the similar design,
- * incorporate the same material and the sterilization method.

Based on the similarities, we conclude that the Luna Dental Implant System - Healing Abutment is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K200189](https://fda-staging.innolitics.com/device/K200189)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).