K200175 · DiaDent Group International · KIF · Jun 25, 2020 · Dental
Device Facts
Record ID
K200175
Device Name
DIA-ROOT BIO Sealer
Applicant
DiaDent Group International
Product Code
KIF · Dental
Decision Date
Jun 25, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DIA-ROOT BIO Sealer is a MTA (mineral trioxide aggregate) based root canal sealer that provides complete and permanent sealing of root canals. It can be used with or without root canal obturation materials.
Device Story
DIA-ROOT BIO Sealer is a premixed, hydraulic MTA-based root canal sealer supplied in a water-blocked syringe. It requires no mixing; it hardens upon reaction with moisture present in the oral cavity/dentin tubules. Used by dentists in clinical settings to provide permanent root canal sealing. The device is applied via disposable tips. It functions as an insoluble, radiopaque filling material. Benefits include simplified application and permanent sealing of the root canal system.
Clinical Evidence
No clinical data was collected or provided. Substantial equivalence is supported by bench testing and biocompatibility evaluation per ISO 6876:2012, ISO 7405:2018, and various ISO 10993 standards.
Technological Characteristics
Premixed hydraulic MTA-based cement. Materials: Calcium Silicate, Calcium Aluminate, Ytterbium Trifluoride, Zirconium Oxide, Silanamine, Hydroxypropyl Methylcellulose, Polyethylene glycol, Sorbitan, White Mineral Oil. Form factor: Syringe with disposable tips. Conforms to ISO 6876:2012 (flow, film thickness, solubility, radio-opacity). Non-sterile, 2-year shelf life.
Indications for Use
Indicated for permanent sealing of root canals in patients requiring root canal therapy. Can be used with or without obturation materials.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 25, 2020
Diadent Group International Kab Sun Lee Quality Assurance Manager 16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu Cheongju-si, Chungcheongbuk-do, 28161 Republic of Korea
Re: K200175
Trade/Device Name: DIA-ROOT BIO Sealer Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: March 25, 2020 Received: March 27, 2020
Dear Kab Sun Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200175
Device Name DIA-ROOT BIO Sealer
Indications for Use (Describe)
DIA-ROOT BIO Sealer is a MTA (mineral trioxide aggregate) based root canal sealer that provides complete and permanent sealing of root canals. It can be used with or without root canal obturation materials.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary (K200175)
This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Application Information
| Date Prepared | Jun 24, 2020 |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name and Address | DiaDent Group International<br>16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu,<br>Cheongju-si, Chungcheongbuk-do, 28161, Republic of Korea |
| Contact Person | Kab Sun Lee<br>Quality Assurance Manager<br>Phone: +82-43-266-2315<br>FAX: +82-43-235-2315<br>Email: diadent32@diadent.co.kr |
## 2. Device Information
| 510(k) Number | K200175 |
|------------------------|---------------------------|
| Device Type | Resin, Root Canal Filling |
| Regulation Description | Root canal filling resin. |
| Review Panel | Dental |
| Regulation Number | 21 CFR 872.3820 |
| Product Code | KIF |
| Device Class | II |
| Device Name | DIA-ROOT BIO Sealer |
## 3. Primary Predicate Device
| 510(k) Number | K170950 |
|-------------------|-------------------|
| Applicant | Vericom Co., Ltd. |
| Device Name | Well-Root ST |
| Regulation Number | 21 CFR 872.3820 |
| Product Code | KIF |
| Device Class | II |
## 4. Device Description
DIA-ROOT BIO Sealer is a hydraulic material, and a premixed form that does not require mixing. It blocks the root canal by hardening by reacting with water in the oral cavity. It is contained in a waterblocked syringe and corresponds to ISO 6876:2012, Dentistry-Root canal sealing materials. DIA-ROOT BIO Sealer has two models and they are packaged with components; Disposable tip, Silicone сар.
| Model Name | Composition |
|--------------------------|-----------------------------------------------------------|
| DIA-ROOT BIO Sealer 2.0g | 2.0g syringe 1ea + Disposable tip 20ea + Silicone cap 1ea |
| DIA-ROOT BIO Sealer 0.5g | 0.5g syringe 1ea + Disposable tip 4ea + Silicone cap 1ea |
{4}------------------------------------------------
#### 5. Indications for Use
DIA-ROOT BIO Sealer is a MTA (mineral trioxide aggregate) based root canal sealer that provides complete and permanent sealing of root canals. It can be used with or without root canal obturation materials.
#### 6. Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and primary predicate device.
#### 7. Non-Clinical Performance Data
The performance and biological tests were conducted on the subject device; DIA-ROOT BIO Sealer according to the following standards.
ISO 6876:2012, Dentistry - Root canal sealing materials ISO 7405:2018, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry ISO 10993-1:2018, Evaluation and testing within a risk management process ISO 10993-2:2006. Animal welfare requirements ISO 10993-3:2014, Tests for genotoxicity, carcinogenicity and reproductive toxicity ISO 10993-5:2009, Tests for in vitro cytotoxicity ISO 10993-6:2016, Tests for local effects after implantation ISO 10993-10:2010, Tests for irritation and skin sensitization ISO 10993-11:2017, Tests for systemic toxicity ISO 10993-12:2012, Sample preparation and reference materials
The test results corresponded the requirements of standards. Therefore, the subject device is substantially equivalent to the primary predicate device.
#### 8. Technological Characteristics
The subject device, DIA-ROOT BIO Sealer has similar characteristics to the primary predicate device, Well-Root ST.
First, the indications for use of the subject device and primary predicate device is to seal root canal permanently. Our indications for use is more detailed than the predicate device, but it does not contain new indications.
Second, both the subject device and primary predicate device are premixed hydraulic materials, and to seal the root canal by hardening with water in the oral cavity. Also, they are contained in a syringe and supplied with disposable tips.
Third, both the subject device and primary predicate device confirm to ISO 6876 and have similar physical and mechanical properties; Flow, Film thickness, Solubility, Radio-opacity. Also, they are biocompatible.
Finally, they are supplied non-sterile state and have 2 years shelf-life.
The raw materials composition of the subject device is slightly different from the primary predicate device. However, the main material, Calcium silicate of the subject device is similar to Calcium aluminosilicate compound of the primary predicate device. Also, Zirconium Oxide and Polyethylene glycol are contained in both subject device and primary predicate device. Through the results of bench and biocompatibility tests, this difference does not affect substantial equivalance.
#### [Comparison table]
| | Subject Device | Primary Predicate Device | Discuss |
|---------------|----------------|--------------------------|------------|
| 510(k) Number | K200175 | K170950 | - |
| Product Code | KIF | KIF | Equivalent |
| Device Class | II | II | Equivalent |
{5}------------------------------------------------
## DiaDent Group International Product Name: DIA-ROOT BIO Sealer
| Manufacturer | DiaDent Group<br>International | Vericom Co., Ltd. | - |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | DIA-ROOT BIO Sealer | Well-Root ST | - |
| Indications for<br>Use | DIA-ROOT BIO Sealer is<br>a MTA (mineral trioxide<br>aggregate) based root<br>canal sealer that provides<br>complete and permanent<br>sealing of root canals. It<br>can be used with or<br>without root canal<br>obturation materials. | Permanent sealing of root<br>canal | Equivalent |
| Description | DIA-ROOT BIO Sealer is<br>a hydraulic material, and a<br>premixed form that does<br>not require mixing. It<br>blocks the root canal by<br>hardening by reacting with<br>water in the oral cavity. It<br>is contained in a water-<br>blocked syringe and<br>corresponds to ISO<br>6876:2012, Dentistry-Root<br>canal sealing materials. | Well-Root ST is a<br>convenient premixed<br>ready-to use composition<br>which requires the<br>presence of water to set<br>and harden. The device is<br>contained in a plastic<br>syringe and the system<br>includes a plunger,<br>disposable tips, and a<br>holder. | Equivalent |
| Package Contents | -Syringe<br>-Disposable tip<br>-Silicone Cap | -Syringe<br>-Disposable tip<br>-Holder | Equivalent |
| Raw Materials | - Calcium Silicate<br>- Calcium Aluminate<br>- Ytterbium Trifluoride<br>- Zirconium Oxide<br>- Silanamine, 1,1,1-<br>trimethyl-N-<br>(trimethylsilyl)-,<br>hydrolysis products with<br>silica<br>- Hydroxypropyl<br>Methylcellulose<br>- Polyethylene glycol 400<br>- Polyethylene glycol 200<br>- Sorbitan<br>- White Mineral Oil- | -Calcium aluminosilicate<br>compound<br>- Calcium Sulfate<br>dihydrate<br>- Calcium sodium<br>phosphosilicate<br>- Zirconium Oxide<br>- Titanium Dioxide<br>- Polyethylene glycol<br>- Polypropylene glycol<br>- Propylene glycol | The main material,<br>calcium silicate of<br>the subject device<br>is similar to<br>calcium<br>aluminosilicate<br>compound of the<br>primary predicate<br>device. Also,<br>zirconium oxide<br>and polyethylene<br>glycol are<br>contained in both<br>subject device and<br>primary predicate<br>device. Through<br>the results of bench<br>and<br>biocompatibility<br>tests, this<br>difference does not<br>affect substantial<br>equivalence. |
| | | | |
| Principle of<br>operation | DIA-ROOT BIO Sealer is<br>a premixed form that does<br>not require mixing, and a<br>hydraulic material that<br>hardens by reacting water.<br>Also, it blocks the root<br>canal by hardening by<br>reacting with water in the<br>oral cavity. It is contained<br>in a water-blocked syringe<br>and is supplied with<br>disposable tips and<br>silicone caps. | Well-Root ST is a<br>convenient premixed<br>ready-to-use injectable<br>white hydraulic cement<br>paste developed for<br>permanent root canal<br>filling and sealing<br>applications. Well-Root<br>ST is an insoluble,<br>radiopaque material which<br>sets and hardens with<br>moisture providing from<br>dentin tubules during<br>hydration reaction. Well-<br>Root ST is packaged in a<br>pre-loaded syringe and is<br>supplied with disposable<br>tips. | Equivalent |
| Performance<br>Standard<br>Conformance | Conformed ISO 6876 | Conformed ISO 6876 | Equivalent |
| Physical and<br>Mechanical<br>properties | -Flow: Not less than 17<br>mm<br>-Film thickness: Not more<br>than 50 µm<br>-Solubility: Not more than<br>3 %<br>-Radio-opacity: Not less<br>than 3 mm | -Flow: Not less than 17<br>mm<br>-Film thickness: Not more<br>than 50 µm<br>-Solubility: Not more than<br>3 %<br>-Radio-opacity: Not less<br>than 3mm | Equivalent |
| Biocompatibility | Biocompatible | Biocompatible | Equivalent |
| Use | Prescription / Hospital | Prescription / Hospital | Equivalent |
| Period of use | Permanent | Permanent | Equivalent |
| Sterility | Non-sterile | Non-sterile | Equivalent |
| Shelf-life | 2 years | 2 years | Equivalent |
{6}------------------------------------------------
## 9. Conclusions
Based on the above information and all data provided in this submission, the subject device is substantially equivalent to the legally marketed device identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.