K200169 · FHC, Inc. · GEI · May 14, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200169
Device Name
VFS1 Bipolar Electrosurgical Generator
Applicant
FHC, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
May 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VFS1 Bipolar Generator is intended for use in surgical procedures for cutting tissue and coagulating blood vessels. Procedures must be performed by a surgeon with experience in the use of electrosurgical devices and generators.
Device Story
VFS1 Bipolar Electrosurgical Generator provides 1 MHz RF energy for cutting and coagulating living human tissue; includes integrated variable-rate irrigation pump for saline delivery. Used in hospitals and clinical settings by surgeons. System inputs include user-selected power settings via rotary/pushbutton switches and foot pedal activation. Device provides audible feedback via tone generator and visual feedback via LCD display (power settings, irrigation, system errors/warnings). Earth-isolated bipolar output minimizes leakage current. Safety features include continuous monitoring of critical parameters with alerts. Benefits include precise tissue management and vessel coagulation during surgery.
Clinical Evidence
No clinical data. Evidence consists of bench testing including hardware verification (input/output, RF power, overcurrent protection, irrigation, mechanical, environmental, and shipping tests per ISTA-3A) and ex vivo comparability studies using tissue samples (steak, liver, kidney) to verify system operation.
Indicated for micro, macro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 14, 2020
FHC, Inc. % Craig Pagan, Consultant C2C Development, LLC 1135 W NASA Blvd, Suite 500 Melbourne, Florida 32901
Re: K200169
Trade/Device Name: VFS1 Bipolar Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 10, 2020 Received: April 14, 2020
Dear Craig Pagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200169
#### Device Name
VFS1 Bipolar Electrosurgical Generator
#### Indications for Use (Describe)
The VFS1 Bipolar Electrosurgical Generator is indicated for use in micro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY SECTION 5
1. Submitter's name, address, telephone number, a contact person, and the date the summary was prepared:
FHC, Inc. 1201 Main Street Bowdoin, ME-04287 Tel: 207-666-5651 Fax: 207-666-8292
Contact: Kelly Moeykens
Preparation Date: April 6th, 2020
### 2. Name(s) of the Device:
| Proprietary/Trade Name: | VFS1 Bipolar Electrosurgical Generator |
|-------------------------|---------------------------------------------------------------|
| Common Name: | Bipolar Electrosurgical Generator or Electrosurgical Generato |
| Regulation Number | 21 CFR 878.4400 |
| Regulation Name: | Electrosurgical device |
| Classification Panel | General and Plastic Surgery |
| Product Code: | GEI |
| Regulatory Class: | II |
### 3. Legally Marked Predicate Device to which the submitter claims substantial equivalence:
| Trade Name: | Malis Bipolar 2000 Electrosurgical Generator |
|----------------|----------------------------------------------|
| 510(k) #: | K050364 |
| Product Code: | GEI |
| Decision Date: | March 9th, 2005 |
| Trade Name: | MALISTM Irrigation Module 1000 |
| 510(k) #: | K033499 |
| Product Code: | GEI |
| Decision Date: | November 28th, 2003 |
{4}------------------------------------------------
## 4. Description of device:
The VFS1 Bipolar Electrosurgical Generator (referred to as generator or VFS1 Generator) is designed to cut and coagulate living human tissue during surgical procedures. The system provides 1 MHz Radio Frequency (RF) energy that is user/operator selectable based on the surgical procedure undertaken. The generator is configured for bipolar output and is earth isolated, to minimize leakage current, thus providing increased patient comfort and safety. The system will be provided with safety features that are continuously monitored and will provide warnings and/ or alerts when critical parameters exceed specified limits.
The VFS1 Generator is designed to cut and coagulate living tissue during surgical procedures. The system includes a bipedal footswitch. The generator is equipped with a tone generator to provide audible feedback each time cutting, or coagulation is performed. The generator incorporates a variable-rate irrigation pump to provide saline flow during procedures.
The VFS1 Generator is compatible with a variety of reusable bipolar products; i.e., insulated mirror finish bipolar forceps, disposable nonstick bipolar instruments as well as disposable bipolar cords and tubing products.
ltems that are included in the system are:
- VFS1 generator -CO 840
- . Footpedal- CO 841
| C0840 | Valley Forge Scientific<br>VFS1 Bipolar<br>Electrosurgical<br>Generator |
|-------|-------------------------------------------------------------------------|
| C0841 | Electrosurgical Foot<br>Pedal |
{5}------------------------------------------------
# 5. Statement of Indications for Use:
The VFS1 Bipolar Electrosurgical Generator is indicated for use in micro, macro, and endoscopic bipolar irrigation, cutting and coagulating of tissue and the coagulation of blood vessels in all types of surgery.
# 6. Intended Use
The VFS1 Bipolar Generator is intended for use in surgical procedures for cutting tissue and coagulating blood vessels. Procedures must be performed by a surgeon with experience in the use of electrosurgical devices and generators.
| Technological<br>Characteristics | Valley Forge Scientific<br>(division of FHC Inc.)<br>VFS1 Electrosurgical<br>Generator | Valley Forge Scientific<br>Corp.<br>Malis 2000 Bipolar<br>Electrosurgical System | PREDICATE<br>Malis Irrigation Module<br>1000 |
|----------------------------------|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Modalities | Electrosurgical Cutting<br>and Coagulation<br>Irrigation | Electrosurgical Cutting<br>and Coagulation | Irrigation |
| Environment of<br>Use | Hospitals and Clinical<br>Settings where<br>electrosurgery is used. | Hospitals and Clinical<br>Settings where<br>electrosurgery is used. | Hospitals and Clinical<br>Settings where<br>electrosurgery is used. |
| Power Source | AC Voltage from wall<br>receptacle | AC Voltage from wall<br>receptacle | AC Voltage from wall<br>receptacle |
| Device (System)<br>Components | - Generator<br>- Foot pedal<br>- Power Cord<br>- Cord/Tubing Set<br>- Irrigation<br>Module(check) | - Generator<br>- Foot pedal<br>- Power Cord<br>- Cord/Tubing Set | - Peristaltic Pump<br>- Foot Pedal |
| | LCD Display, displays: | LCD Display, displays: | |
| | Cut and Coag Power<br>Settings | Cut and Coag Power<br>Settings | |
| | Power Units | Power Units | |
| Display Screen<br>Details | Tone Volume and<br>Brightness Settings | Tone and Voice Volume<br>and Brightness Settings | Built in Numeric display |
| | System Errors and<br>Warnings | System Errors and<br>Warnings | |
| | Irrigation Setting | Absent | |
# 7. Comparison of Technological Characteristics to Predicate Devices
{6}------------------------------------------------
| Technological<br>Characteristics | Valley Forge Scientific<br>(division of FHC Inc.)<br>VFS1 Electrosurgical<br>Generator | PREDICATE<br>Valley Forge Scientific<br>Malis 2000 Bipolar<br>Electrosurgical System | PREDICATE<br>Malis Irrigation Module<br>1000 |
|----------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------|
| Pump Head Type | Peristaltic | N/A | Peristaltic |
| Motor Type | DC electric motor | N/A | DC electric motor |
| Irrigation<br>Delivery Method | Tubing set with industry<br>standard connectors | N/A | Tubing set with industry<br>standard connectors |
| Patient interface | Bipolar forceps | N/A | Bipolar forceps |
| Liquid Source | Saline bag | N/A | Saline bag |
| Delivery Rate | Slow drip, medium drip,<br>fast drip | N/A | Slow drip, medium drip,<br>fast drip |
| RF Output Mode | Bipolar | Bipolar | N/A |
| RF Output Power<br>Range | Cut: 0 to 100 Watts<br>Coag: 0 to 50 Watts | Cut: 0 to 300 Watts<br>Coag: 0 to 50 Watts | N/A |
| RF Output<br>Waveforms | Cut: 1 MHz sinewave<br>Coag: 1 MHz damped<br>aperiodic | Cut: 1 MHz sinewave<br>Coag: 1 MHz damped<br>aperiodic | N/A |
| Output Power<br>Controls | Rotary Switches<br>Pushbutton Switches | Rotary Switches | Rotary Switches |
| Power Activation | Foot pedal | Foot pedal and Finger<br>switches | Foot pedal |
## 8. Performance Data
Non-clinical: Basic mechanical and functional capabilities were tested for the subject electrosurgical generator to determine substantial equivalence to the predicate device. This also includes a Comparibility Study of the subject device and the predicate device. Report Refer to section 18 for details.
The following is a summary of components and testing performed:
#### . VCP-09-103 – VFS1 Hardware Verification Testing
Testing and inspection performed on the VFS1 Electrosurgical Generator, to verify compliance with the Product Development Specification, DSI-09-100. Testing included:
- о Input Voltage
- Input Frequency o
- Input Protection o
{7}------------------------------------------------
- Output Waveform O
- RF Power Output O
- RF Output vs Load Resistance O
- O RF Overcurrent Protection
- Over-dosage Protection O
- Irrigation O
- Real Time Clock O
- Mechanical O
- Operating Conditions O
- Storage Conditions O
- O Cooling
- Transportation O
- Shock O
- Cleaning O
- Front Panel Connections O
- Rear Panel Connections O
- Power Cord O
- O Front Panel Controls
- O Rear Panel Controls
- Display O
All testing met specifications.
#### VCP-09-109 – VFS1 Comparibility Study Report ●
Testing was performed on the VFS1 Electrosurgical Generator to verify proper operation of the generator with all the components and accessories working together as a system. Testing included Ex Vivo testing for Cut and Coag using steak, liver and kidney tissue samples. All testing was found to be acceptable.
#### ● ROH-09-100 - VFS1 ROHS Verification Report
RoHS Compliance was verified for all components and subassemblies of the VFS1 Surgical Generator, Model CO840.
#### VDP-09-100 – Shipping Test Report ●
Testing was performed to confirm that the external packaging configuration for the VFS1 Electrosurgical Generator provides a high probability of safe and intact arrival at the destination by confirming that the integrity of the packaging and its components withstand simulated shipping and handling stress and by demonstrating that the shipping and handling stresses have no adverse effect on the VFS1 Generator performance. Testing was performed per ISTA-3A Shipping Test and met all requirements.
Clinical: No clinical data was used to determine substantial equivalence.
{8}------------------------------------------------
# 9. Conclusion
Based on the non-clinical performance data performed comparing the VFS1 Electrosurgical Generator to the predicate devices (Malis 2000 Bipolar Electrosurgical System and the Malis Irrigation Module 1000), it is concluded that the device is as safe, as effective, and performs as well as or better than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.