K200168 · Puregraft, LLC · QPB · Apr 22, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200168
Device Name
Dermapose Access
Applicant
Puregraft, LLC
Product Code
QPB · General, Plastic Surgery
Decision Date
Apr 22, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dermapose Access fat harvest system is intended for use in aesthetic body contouring.
Device Story
Dermapose Access is a sterile, single-use system for harvesting autologous fat for aesthetic body contouring. It consists of a polycarbonate housing with a linear needle guidance system, a 12G lancet point needle, a 14G harvesting cannula, and a silicone tube set with a regulator valve. The device is operated by a physician in a clinical setting. The system connects to a user-provided vacuum pump (min 18 inHg) to lift the patient's skin into a guide cavity, allowing for a controlled 10 mm depth incision and subsequent cannula introduction. The physician performs the harvest manually using a user-provided syringe. The device includes a protective filter to prevent fluid ingress into the vacuum pump. By providing a guided, vacuum-assisted approach to skin stabilization and incision, the device facilitates consistent adipose tissue collection, potentially improving procedural efficiency and patient outcomes in body contouring.
Clinical Evidence
Bench testing only. Performance testing included skin puncture tests, reliability testing, skin hold and needle depth analysis, and cannula analysis. Biocompatibility testing was conducted per ISO 10993-1:2018, including cytotoxicity (ISO 10993-5), sensitization (ISO 10993-10), intracutaneous reactivity (ISO 10993-10), acute systemic toxicity (ISO 10993-11), pyrogen testing (ISO 10993-11), and extractable elements (ISO 10993-18). Human factors and usability engineering were evaluated to confirm safety and effectiveness for intended users. Shelf-life was validated via accelerated aging and ISTA 2A ship testing.
Technological Characteristics
Polycarbonate housing, 12G lancet needle, 14G cannula, silicone tubing. Vacuum-assisted skin hold (requires external pump). Sterile (ethylene oxide, SAL 10^-6). Single-use. Biocompatible per ISO 10993. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for aesthetic body contouring in patients requiring small volume autologous adipose tissue harvest.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Puregraft LLC Olivia Kim RA/QA Director 420 Stevens Avenue, Suite 220 Solana Beach, California 92075 June 9, 2021
Re: K200168
Trade/Device Name: Dermapose Access Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: QPB
Dear Olivia Kim:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated April 22, 2020. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Cindy Chowdhury, OHT4: Office of Surgical and Infection Control Devices, 240-402-6647, Cindy.Chowdhury@fda.hhs.gov.
Sincerely,
# Cindy Chowdhury -S
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue rectangle with the letters "FDA" in white. To the right of the rectangle, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
April 22, 2020
Puregraft LLC Olivia Kim RA/QA Director 420 Stevens Avenue, Suite 220 Solana Beach, California 92075
## Re: K200168
Trade/Device Name: Dermapose Access Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: January 21, 2020 Received: January 23, 2020
Dear Olivia Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/2/Picture/5 description: The image shows the name "Joseph A. Nielsen" in a large, bold font. To the right of the name is a digital signature block. The signature block includes information such as the DN, OU, email, country, and date of the signature.
Digitally signed by Joseph A. Nielser A/CDRH/ODF/DSD/PRSB1.
For Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200168
Device Name Dermapose Access
Indications for Use (Describe)
The Dermapose Access fat harvest system is intended for use in aesthetic body contouring.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
# Submitter Information
| Applicant Name: | Puregraft LLC<br>420 Stevens Avenue, Suite 220<br>Solana Beach, CA 92075 USA |
|------------------|-----------------------------------------------------------------------------------------------------------|
| Company Contact: | Olivia Kim<br>RA/QA Director<br>Phone: (858) 386-4140<br>Fax: (858) 217-5134<br>Email: okim@puregraft.com |
Date Prepared: April 20, 2020
#### Device Information
| Submission Number: | K200168 |
|-----------------------------|-----------------------------|
| Trade or Proprietary Name: | Dermapose™ Access |
| Common Name: | Suction lipoplasty system |
| Device Classification Name: | System, Suction, Lipoplasty |
| Product Code: | MUU |
| Regulatory Class: | Class II |
| Classification Regulation: | 21 CFR 878.5040 |
| Panel: | General & Plastic Surgery |
#### Predicate Device
st'rim, K142073, Thiebaud SAS
#### Indications for Use
The Dermapose™ Access fat harvest system is intended for use in aesthetic body contouring.
#### Device Description
The Dermapose™ Access is a pre-assembled, sterile, single use system intended to assist the harvesting of autologous fat from a patient for aesthetic body contouring. It creates a guided incision for cannula introduction at a controlled depth (10 mm) under the skin for the purpose of small volume adipose tissue harvest.
The following are provided sterile with the Dermapose™ Access:
{5}------------------------------------------------
- A pre-assembled polycarbonate housing with a linear needle guidance system and a 12G x ● 51 mm lancet point needle to allow passage of a 14G cannula through the bore;
- A 14G x 150 mm Harvesting/Tumescent Cannula; and
- A silicone tube set with regulator valve, which connects to a user-provided vacuum source o
The Dermapose™ Access is also intended to be used with a user-provided vacuum pump capable of at least 18 inHg (60 kPa) of vacuum in order to lift the patient's skin into the guide cavity. The pump should have a standard barb fitting of approximately 1/4 inch. The use of a collection canister is recommended but is not necessary as fluid should not enter the tubing. To protect the pump from possible fluid ingress, the Dermapose™ Access includes a protective filter.
## Technological Characteristics
The Dermapose™ Access is substantially equivalent to the predicate device, st'rim. A comparison of technological characteristics is provided in the table below. Both the subject device and the predicate device provide a method of cannula introduction for harvesting autologous adipose tissue. In addition, both the subject and predicate devices are provided sterile (via ethylene oxide) and intended for single use.
| Device Name | Subject Device<br>Dermapose™ Access | Predicate Device<br>St'rim (K142073) | Similarity |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use | The Dermapose™ Access fat harvest<br>system is intended for use in aesthetic<br>body contouring. | The st'rim™ fat tissue harvest and<br>injection cannula set is intended for<br>use in aesthetic body contouring. | Identical |
| User Environment | Physician | Physician | Identical |
| Design Concept | Guided needle and cannula kit | Assorted needle and cannula kit | Different |
| Components | 14G Harvest/Tumescent cannula 12G incision needle housed in a linear guidance and vacuum-assisted skin hold system. A silicone tube set for connection of vacuum | 14G Harvest/Tumescent cannula 14G incision needle 21G injection cannula (short) 21G injection cannula (long) 21G incision needle Luer-to-luer connector | Different |
| Method of<br>Harvesting | Cannula (provided) and Syringe (user-provided) | Cannula (provided) and Syringe (user-provided) | Identical |
| Incision | Guided incision | Non-guided incision (physician pinch) | Different |
| Source of Harvest<br>Energy | None - Manual by operator | None - Manual by operator | Identical |
| Source of Skin<br>Hold Energy | User provided vacuum source | None - Manual by operator (pinch) | Different |
| Device performs<br>Harvesting | No – requires syringe capable of<br>holding vacuum | No - requires syringe capable of<br>holding vacuum | Identical |
{6}------------------------------------------------
| Mechanical<br>operation (entry) | Needle incision followed by<br>concentric cannula introduction, then<br>needle removal | Needle incision, then needle removal,<br>followed by cannula introduction | Similar |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------|
| Mechanical<br>operation<br>(harvest) | Requires syringe capable of holding<br>vacuum, motion is back and forth and<br>sweeping about entry point | Requires syringe capable of holding<br>vacuum, motion is back and forth and<br>sweeping about entry point | Identical |
| Mechanical<br>operation<br>(process) | Not part of system | Not part of system | Identical |
| Mechanical<br>operation<br>(delivery) | Injection via small syringes and<br>injection cannula (not included) | Injection via small syringes (not<br>included) and injection cannula<br>(included) | Similar |
| Construction | Preassembled except for luer<br>connections | Preassembled except for luer<br>connections | Identical |
| Sterility | Supplied sterile for single use | Supplied sterile for single use | Identical |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Identical |
| Sterility<br>Assurance Level | SAL = 10-6 | SAL = 10-6 | Identical |
| Biocompatibility | Biocompatible | Biocompatible | Identical |
## Performance Testing
The Dermapose "1" Access was tested in accordance with established protocols and met the acceptance criteria for all tests performed.
Performance testing included:
- Skin Puncture Test: Evaluated the ability to penetrate skin using the Dermapose™ Access. ●
- Reliability Test: Evaluated the reliability of the Dermapose™ Access. ●
- Skin Hold and Needle Depth Analysis: Evaluated the performance of the Dermapose™ ● Access to hold and puncture skin.
- Cannula Analysis: Evaluated the laboratory testing of the cannula provided with the ● Dermapose™ Access.
Biocompatibility testing for the Dermapose™ Access was performed in accordance with the requirements of ISO 10993-1:2018 for external communicating devices having contact < 24 hours. In all instances, the test articles were found to be biocompatible. The battery of testing included the following tests:
- Cytotoxicity per ISO 10993-5:2009
- Sensitization per ISO 10993-10:2010 ●
- Intracutaneous per ISO 10993-10:2010 ●
- Acute Systemic Toxicity per ISO 10993-11:2017 ●
- Pyrogen Testing per ISO 10993-11:2017 ●
- Determination of Extractable Elements per ISO 10993-18:2005
{7}------------------------------------------------
Human Factors and Usability Engineering was evaluated. The Dermapose™ Access was found to be safe and effective for the intended users, uses, and use environments.
Sterilization of the Dermapose™ Access is conducted via the Single Lot Release Method by Ethylene Oxide according to Annex E: Single Lot Release of ISO 11135:2014/Amd.1:2018.
The Dermapose "1 Access has an expiry date of 24 months based on integrity testing conducted post completion of accelerated aging and ISTA 2A ship testing.
#### Statement of Substantial Equivalence
Based on the information contained in this submission, it is concluded that the Dermapose™ Access is substantially equivalent to the predicate device, st'rim. The Dermapose™ Access has the same indications for use, as well as similar technological characteristics and principles of operation as its predicate device. Thus, the Dermapose™ Access is substantially equivalent to the st rim.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.