← Product Code [ODG](/productcode/ODG) · K200090

# PENTAX Medical EG-J10U Endoscopic Ultrasound System - EG34-J10U Ultrasound Upper GI Video Scope (Convex Array Type), EG36-J10UR Ultrasound Upper GI Video Scope (Radial Array Type), EG38-J10UT Ultrasound Upper GI Video Scope (Convex Array Type) (K200090)

_Pentax of America, Inc. · ODG · Jul 23, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K200090

## Device Facts

- **Applicant:** Pentax of America, Inc.
- **Product Code:** [ODG](/productcode/ODG.md)
- **Decision Date:** Jul 23, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The PENTAX Medical EG-J10U Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

## Device Story

Endoscopic ultrasound system provides optical/ultrasonic visualization and therapeutic access to upper GI tract; includes video scopes (EG34-J10U, EG36-J10UR, EG38-J10UT) with flexible insertion tubes, control bodies, and ultrasound transducers (convex or radial array). Used with cleared Pentax video processors and Hitachi ultrasound scanners. Input: endoscopic image data (via CCD) and ultrasonic pulses (via transducer). Output: visual display on monitor. Procedure: scope introduced per-orally; natural rubber latex balloon fitted over transducer, inflated with water to create acoustic coupling field. Physician operates controls for angulation, air/water delivery, and accessory insertion (e.g., biopsy devices). System facilitates diagnosis and therapeutic intervention. Benefits: enhanced visualization of GI organs/tissues for clinical decision-making.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including human factors study for reprocessing instructions, simulated use testing, soil accumulation analysis, cleaning/disinfection validation, sterilization validation (Gamma Ray and liquid chemical), and optical/electrical safety verification (IEC 60601 series).

## Technological Characteristics

Flexible endoscope with CCD for optical imaging and linear/radial array ultrasound transducer. Materials include natural rubber latex (balloons). Connectivity: proprietary umbilical connectors to video processors and ultrasound scanners (via junction box PUN-JBP1 for Preirus). Sterilization: System 1E liquid chemical sterilization for scopes; Gamma Ray for balloons. Software: Class A (IEC 62304), Moderate level of concern.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- PENTAX Medical Endoscopic Ultrasound System ([K182004](/device/K182004.md))

## Reference Devices

- PENTAX Video Processor EPK-i5010 ([K122470](/device/K122470.md))
- PENTAX Video Processor EPK-i7010 ([K150618](/device/K150618.md))
- Hitachi HI VISION Preirus ([K093466](/device/K093466.md))
- Hitachi ARIETTA 70 ([K134016](/device/K134016.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 23, 2020

PENTAX of America, Inc. William Goeller Vice President, Quality/Regulatory Affairs 3 Paragon Drive Montvale, NJ 07645-1782

Re: K200090

Trade/Device Name: PENTAX Medical EG-J10U Endoscopic Ultrasound System EG34-J10U Ultrasound Upper GI Video Scope (Convex Array Type), EG36-J10UR Ultrasound Upper GI Video Scope (Radial Array Type), EG38-J10UT Ultrasound Upper GI Video Scope (Convex Array Type) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODG, ITX Dated: June 29, 2020 Received: July 2, 2020

Dear William Goeller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200090

#### Device Name

PENTAX Medical EG-J10U Endoscopic Ultrasound System EG34-J10U Ultrasound Upper GI Video Scope (Convex Array Type), EG36-J10UR Ultrasound

Upper GI Video Scope (Radial Array Type) , EG38-J10UT Ultrasound Upper GI Video Scope (Convex Array Type)

#### Indications for Use (Describe)

The PENTAX Medical EG-J10U Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D)    | Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold letters. Below the word "PENTAX" is a gray line, and below the line is the word "MEDICAL" in gray, bold letters. 

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.

Applicant: PENTAX Medical HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782

Contact: William Goeller Vice President, Quality and Regulatory Affairs PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645-1782 Telephone: (201)571-2300 ext. 2318 FAX: (201)391-4189 Email: william.goeller@pentaxmedical.com

Date Prepared: 1/14/2020

Common Name: Endoscopic Ultrasound / Ultrasound Gastroscope Device/Trade Name: PENTAX Medical EG-J10U Endoscopic Ultrasound System-EG34-J 10U Ultrasound Upper GI Video Scope (Convex Array Type), EG36-J10UR Ultrasound Upper Gl Video Scope(Radial Array Type), EG38-J10UT Ultrasound Upper Gl Video Scope (Convex Array Type)

Regulation Number: 21 CFR Part 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Codes: ODG, ITX Predicate Device: PENTAX Medical Endoscopic Ultrasound System (K182004) PENTAX Medical is seeking clearance of a new line of ultrasound upper GI video scopes EG-J10U with the Pentax Video Processors EPK-i5010 and EPK-i7010, for use with Hitachi's HI VISION Preirus and ARIETTA 70. PENTAX Video Processors EPK-i5010 (K122470) and EPK-i7010 (K150618) as well Hitachi HI VISION Preirus (K093466) and ARIETTA 70 (K134016) have been previously cleared.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for PENTAX Medical. The word "PENTAX" is in red, bold letters, and it is positioned above a gray line. Below the line, the word "MEDICAL" is written in gray, bold letters. The logo is simple and professional.

This 510(k) also captures some minor design changes that have occurred during the evolution of the product line. Although the changes are believed to be minor, the 510(k) is being submitted to account for technological advances in associated compatible devices and to ensure that FDA has the most current information concerning the PENTAX Medical Ultrasound Upper Gl Video Scopes

The subject devices have virtually the same indications for use, viewing directions, and image size as the predicates. The subject devices use the same processors and peripherals as the predicate device.

The main differences between the subject devices and predicate devices are as follows:

- . Introduction of a new line of ultrasound upper GI video scopes EG-J10U, which includes EG34-J10U, EG36-J10UR and EG38-J10UT as compatible Ultrasound Upper GI Video Scopes.
- . A Reprocessing Instructions for Use (rIFU) that demonstrates a streamlined, easy-to-follow format that is the output of Human Factors Testing.

## Device Description:

The EG34-J10U, EG36-J10UR and EG38-J10UT Ultrasound Upper GI Video Scopes, are endoscopes used to provide visualization of, and therapeutic access to, the upper gastrointestinal tract. They are used with cleared Pentax Video Processors (a softwarecontrolled device) and cleared Hitachi Ultrasound Scanners (a software-controlled device). The endoscopes have a flexible insertion tube, a control body, PVE umbilical connector, and ultrasound scanner umbilical connector. The PVE umbilical connector will be attached to the Video Processor and has connections for illumination, video signals, air/ water/ and suction.

The ultrasound scanner umbilical connector will be attached to the ultrasound scanner unit. A sterile, single use disposable natural rubber latex balloon is fitted over the convex array ultrasound transducer prior to the procedure. During an ultrasound endoscopy procedure, the latex balloon is inflated with water. The water that is contained within the balloon creates a water field that covers the transducer. The water field enables more effectively transport of ultrasonic pulses from the ultrasound transducer to the target anatomical site and back to the ultrasound transducer.

The control body includes controls for up/ down/ left/ right angulation, air/ water delivery, and an accessory inlet port. The endoscope contains light carrying bundles to illuminate the body cavity, a charge couple device (CCD) to collect endoscopic image data, and a linear or radial array ultrasound transducer to collect ultrasonic image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif letters. Below "PENTAX" is the word "MEDICAL" in gray, bold, sans-serif letters. There is a gray line separating the two words.

The video processor contains a lamp that provides white light and is focused at the PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display.

The ultrasound transducer delivers ultrasonic pulses, reflections of the pulses are received, and the signals are passed to the ultrasound scanner for processing and display. EG34-J10U, EG36-J10UR, and EG38-J10UT Ultrasound Upper GI Video Scopes are connected to the ultrasound scanner Arietta 70 via the scanning unit connector of the endoscope directly to the probe connector of the scanning unit. In order to connect to the Preirus scanning unit, junction box PUN-JBP1 is required to connect the scanning unit connector to the probe connector.

The instrument is immersible (with the use of supplied cleaning accessories) except for the ultrasound scanner connector (as described in the endoscope operator manual cleaning instructions)

### Intended Use

The PENTAX Medical Ultrasound Upper Gl Video Scopes (EG34-J10U, EG36-J10UR, and EG38-J10UT) are intended to provide optical and ultrasound visualization of (via a video monitor), and therapeutic access to the upper gastrointestinal tract including the organs; tissues; and subsystems: esophagus, stomach and duodenum.

### Indications for use

The PENTAX Medical EG-J10U Endoscopic Ultrasound System is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Tract including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

### Summary of Technology Characteristics:

The PENTAX Medical EG-J10U Endoscopic Ultrasound System is functionally equivalent to its predicate device, the PENTAX Medical Endoscopic Ultrasound System cleared by FDA in 2018. The only difference between the two devices is that the predicate is used with the Legacy Ultrasound Video Gastroscopes, and the subject is used with the additional EG-J10U series of Ultrasound GI Video Scopes that includes EG34-J10U, EG36-J10UR and EG38-J10UT

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below "PENTAX" is a horizontal line, and below that is the word "MEDICAL" in gray, bold, sans-serif font. 

The additional devices have been evaluated through performance testing and raise no issue of safety and effectiveness of the device as these differences have no effect on the performance, function or general intended use of the device.

## Non-Clinical Performance Data

The PENTAX Medical EG-J10U Endoscopic Ultrasound System has been successfully tested for its functions, performance and safety as per FDA recognized consensus standards. The following performance data are provided in support of the substantial equivalence determination.

Operational and Reprocessing Instructions for Use are provided for the scopes.

### Human Factors

A human factors study to assess the ability of reprocessing staff to carry out the reprocessing instructions for use (rIFU) was conducted. The device was been found to be safe and effective for the intended patients, intended users, and in their use environments.

## Reprocessing Validation

Simulated use testing, soil accumulation analysis, cleaning, and high-level disinfection validation studies of the EG34-J10U, EG36-J10UR and EG38-J10UT Upper GI Video Scopes and accessories were conducted and confirmed the effectiveness of reprocessing procedures.

## Sterilization and Shelf Life

PENTAX Medical coordinated with STERIS Corporation to validate the use of System 1E liquid chemical sterilization for the sterilization of the EG-J10U series Ultrasound Upper GI Video Scopes. The devices are not provided sterile.

PENTAX Medical conducted Gamma Ray Sterilization for packaging and for PENTAX Medical Ultrasound Balloons made of natural rubber latex. A shelf-life of 2 years after sterilization was verified.

The labeling for the PENTAX Medical Ultrasound Balloons and their packaging prominently bears the 21 CFR 801.437 warning in bold: Caution: This product contains natural rubber latex which may cause allergic reactions. Also, the Operating Instructions for Use of the EG-J10U series of Ultrasound Upper Gl Video Scopes prominently bears the warning in bold: Caution: The balloon contains natural rubber latex which may cause allergic reactions.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold letters on the top line. Below that, there is a gray line, and then the word "MEDICAL" is in gray, bold letters. The logo is simple and clean, with a focus on the company name.

## Software

Software verification and validation tests were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software is classified as CLASS A under the Software Safety Classification per IEC 62304:2006, Medical device software- Software life cycle processes. The software level of concern is "Moderate" based on FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."

Cybersecurity risks have been assessed and mitigated according to the FDA Guidances for Industry and Staff "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" issued October 2, 2014 and "Postmarket Management of Cybersecurity in Medical Devices." issued December 28,2016.

## EMC and Electrical Safety

The acceptable level of electromagnetic compatibility (EMC) and electrical safety (ES) for the PENTAX Medical Endoscopic EG-J10U Ultrasound System were confirmed by the following standards:

IEC 60601-1-2:2007; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+A1:2012; IEC 60601-2-18:2009; and IEC 60601-2-37:2007.

# Optical Testing

As a part of Design Verification and Validation, optical properties including signal to noise ratio, spatial resolution (MTF), distortion, light distribution, color (IEEE), spectral distribution and photobiological safety were measured for the EG34-J10U, EG36-J10UR and EG38-J10UT in conjunction with the EPK - i5010 and EPK-i7010 video processors. All results show that there are no differences between the subject device, and the predicate device.

## Substantial Equivalence Discussion:

After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labeling, and sterilization method, we conclude that the subject device PENTAX Medical EG-J10U Endoscopic Ultrasound System is as safe and effective as the predicate device. There are no differences in indications for use and intended use between the subject and predicate device and are therefore, substantially equivalent. The technological differences in terms of design features, performance characteristics and constituent materials are not substantive.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold letters on the top line. Below that, there is a gray line, and then the word "MEDICAL" is in gray, bold letters on the bottom line.

## Conclusion:

Accordingly, PENTAX Medical believes the PENTAX Medical EG-J10U Endoscopic Ultrasound System is substantially equivalent to the identified predicate, the PENTAX Medical Endoscopic Ultrasound System, cleared by FDA in 2018.

---

**Source:** [https://fda-staging.innolitics.com/device/K200090](https://fda-staging.innolitics.com/device/K200090)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
