Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air, it is not intended for patients where supplementary oxygen is being provided. Breezing Med is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional.
Device Story
Breezing Med is a wearable, mask-based indirect calorimeter for professional healthcare use. It measures resting metabolic rate by capturing exhaled air to determine oxygen uptake (VO2) and carbon dioxide production (VCO2). The device consists of a body, mask, headgear, and a single-use sensor cartridge. It connects to a mobile application for data processing and display. The system calculates Resting Energy Expenditure (REE) and Respiratory Quotient (RQ) from measured gas volumes. Healthcare professionals use these outputs to assess energy expenditure, guide nutritional planning, and assist in diagnosing metabolic disorders. The device provides objective data to support clinical decision-making regarding weight management and metabolic health.
Clinical Evidence
Bench testing only. Verification testing compared Breezing Med performance against the predicate Express Series. Results showed O2 and CO2 volume measurement accuracy within +/-0.8% and flow rate accuracy within +/-0.3 Lpm (<10 Lpm) or +/-3% (>=10 Lpm). Biocompatibility (ISO 10993), electrical safety (IEC 60601-1), and EMC (IEC 60601-1-2) testing were performed. A simulated clinical use study confirmed the device functions according to its intended use.
Technological Characteristics
Wearable indirect calorimeter; mask-based; fixed-orifice differential pressure pneumotach flowmeter; colorimetric O2 and CO2 sensors; single-use sensor cartridge; rechargeable Lithium-Ion battery; mobile application interface; non-sterile; IEC 60601-1 Class II/Type BF; EN60601-1-2 EMC compliant; biocompatible materials per ISO 10993-1.
Indications for Use
Indicated for adults breathing normally in a sitting position in a healthcare environment for nutritional assessment, optimization of nutritional supplements, quantification of substrate utilization, and aiding in the diagnosis of diseases related to abnormal metabolic parameters. Contraindicated for patients receiving supplementary oxygen.
Regulatory Classification
Identification
An oxygen uptake computer is a device intended to compute the amount of oxygen consumed by a patient and may include components for determining expired gas volume and composition.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2020
TF Health Co. % Roberta (Bobbi) Druyor-Sanchez Regulatory Consultant NDA Partners 1761 W. University Dr., Suite 140 Tempe, Arizona 85281
Re: K200076
Trade/Device Name: Breezing Med Regulation Number: 21 CFR 868.1730 Regulation Name: Oxygen Uptake Computer Regulatory Class: Class II Product Code: BZL Dated: August 28, 2020 Received: September 3, 2020
Dear Roberta (Bobbi) Druyor-Sanchez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
### K200076
Device Name
Breezing MedTM Metabolism Analyzer
#### Indications for Use (Describe)
Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air, it is not intended for patients where supplementary oxygen is being provided. Breezing Med is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# TF Healt
Corporation
Traditional Premarket Notification
Breezing Med™ Metabolism Analyzer
## 510(K) SUMMARY (21 CFR 807.92(C))
1. Applicant
TF Health Co. 1761 W University Dr. Suite 140 Tempe, AZ 85281
Primary Contact Erik Herrmann VP Operations, TF Health 480-629-5360 Email: erik.herrmann@breezing.co
### 2. Device
- Trade Name: Breezing MedTM Metabolism Analyzer ●
- Common Name: Indirect calorimeter ●
- Classification Name: Computer, Oxygen Update (21 CFR 868.1730, product code ● BZL)
- 3. Predicate Device: Express Series (K070858)
## 4. Device Description
Breezing Med™ Metabolism Analyzer is a resting metabolic rate measurement system with an application installed on mobile devices that provides measurements of exhalation rate, oxygen uptake and carbon dioxide production. It is used to determine resting metabolic rates, teach energy balance, and pinpoint accurate caloric intake for weight management.
- 5. Intended Use
Breezing Med™ Metabolism Analyzer is intended for professional use, to determine resting metabolic rates, teach energy balance, and pinpoint accurate caloric intake for weight management.
## 6. Indications for Use
Breezing Med™ Metabolism Analyzer measures metabolic functions including resting energy expenditure and respiratory quotient during resting conditions via direct measurement of oxygen uptake and carbon dioxide production to provide nutritional assessment, to optimize nutritional supplements, and to quantify substrate utilization. The nutritional assessment enables the healthcare professional to further assess energy expenditure and caloric intake for weight
{4}------------------------------------------------
Traditional Premarket Notification
Breezing Med™ Metabolism Analyzer
management, and to aid in the diagnosis of diseases related to abnormal metabolic parameters. Breezing Med 100 is intended for use with adults breathing normally on their own in a sitting position in a healthcare environment. Since the device is designed for the patient to breathe ambient air. it is not intended for patients where supplementary oxygen is being provided. Breezing Med™ is not intended to use as sole means for any diagnosis, and ultimately will provide data for evaluation by the professional.
- 7. Technological Characteristics
Breezing Med™ is a stand-alone and fully integrated mask-like wearable metabolic analyzer based on the principle of indirect calorimetry, which is a gold standard for energy expenditure measurement. Breezing Med 104 is designed for metabolic rate and respiratory quotient measurement and the main outputs include Resting Energy Expenditure (REE), Respiratory Quotient (RQ), Volume of Oxygen consumption (VO2), and Volume of Carbon Dioxide production (VCO2), where REE and RQ are calculated from measured parameters VO2 and VC02.
Breezing Med M measures the expiratory volume similar to the predicate device Medical Graphics Express® Series.
Breezing Med "" consists of five components: 1) a device body; 2) a mask; 3) headgear; 4) a sensor cartridge; and 5) an application.
## 8. Performance Data
A verification test of the Breezing Med™ to performance specifications derived from the predicate Medical Graphics Express series was conducted. The result of the test demonstrates substantial equivalence of the proposed to the predicate.
The summary of the Breezing Med™ test results are as follows:
- Both devices measured O2 volumes within the specified accuracy of +/-0.8%. ●
- Both devices measured CO2 volumes within the specified accuracy of +/-0.8%. .
- Both devices measured Flow Rate within +/- 0.3 Lpm < 10 Lpm and +/-3% >= 10 Lpm.
We have also performed testing to ISO10993, IEC 60601-1 General requirements for basic safety and essential performance and IEC 60601-1-2 Electromagnetic disturbances.
## 9. Predicate Device Comparison
The Breezing Med Metabolism Analyzer is substantially equivalent to the Express Series (K070858). Both Breezing Med Me Metabolism Analyzer and its predicate device have a similar intended use. A summary comparison of technological characteristics is provided in Table 5.1.
## Table 5.1 - Breezing Med 111 Metabolism Analyzer and Predicate Comparative Analysis
{5}------------------------------------------------
# TF Health
Corporation
Traditional Premarket Notification
Breezing Med™ Metabolism Analyzer
| Feature | Subject Device | Predicate Device |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Breezing MedTM | Express® Series (indirect<br>calorimeter) |
| Device Manufacturer | TF Health Co. | MGC Diagnostics Corporation |
| 510(k) Reference | This submission | K070858 |
| FDA Product Code | BZL | BZL |
| FDA Classification | Computer, Oxygen Uptake | Computer, Oxygen Uptake |
| FDA Regulation<br>Number | 868.1730 | 868.1730 |
| Device Description | Breezing MedTM is a resting<br>metabolic rate measurement<br>system with an application<br>installed on mobile devices<br>that provides measurements<br>of exhalation rate, oxygen<br>uptake and carbon dioxide<br>production. | The Express® Series is a<br>cardiopulmonary exercise or<br>resting metabolic rate<br>measurement system with an<br>integrated touch screen computer<br>that provides breath by breath<br>measurements of flow, oxygen<br>uptake and carbon dioxide<br>production. |
| Indications for use | Breezing MedTM Metabolism<br>Analyzer measures metabolic<br>functions including resting<br>energy expenditure and<br>respiratory quotient during<br>resting conditions via direct<br>measurement of oxygen<br>uptake and carbon dioxide<br>production to provide<br>nutritional assessment, to<br>optimize nutritional<br>supplements, and to quantify<br>substrate utilization. The<br>nutritional assessment<br>enables the healthcare<br>professional to further assess<br>energy expenditure and<br>caloric intake for weight<br>management, and to aid in<br>the diagnosis of diseases<br>related to abnormal<br>metabolic parameters.<br>Breezing MedTM is intended<br>for use with adults breathing | The Express® Series uses the<br>direct measurement of oxygen<br>uptake to objectively and<br>noninvasively assess cardiac and<br>pulmonary function during<br>exercise. The system can be used<br>to screen for early signs of<br>cardiac and pulmonary<br>dysfunction; differentiate heart<br>and lung disease; assess dyspnea<br>complaints; classify patients<br>according to disease as a guide<br>for patient management; establish<br>an optimal exercise prescription<br>and training program; evaluate<br>the efficiency of prescribed<br>therapy. The Express® Series<br>uses the direct measurement of<br>oxygen uptake and carbon<br>dioxide production to provide<br>nutritional assessment, to<br>optimize nutritional supplements<br>and quantify substrate utilization.<br>The Express® Series can also |
| Feature | Subject Device | Predicate Device |
| | sitting position in a<br>healthcare environment.<br>Since the device is designed<br>for the patient to breathe<br>ambient air, it is not intended<br>for patients where<br>supplementary oxygen is<br>being provided. Breezing<br>MedTM is not intended to use<br>as sole means for any<br>diagnosis, and ultimately will<br>provide data for evaluation<br>by the professional. | of cardiac output using real time<br>oxygen uptake measurements. |
| Target population | Adults | Unspecified |
| Use Environment | Professional healthcare<br>facilities | Professional healthcare facilities |
| Device<br>Measurements | Metabolic functions during<br>resting conditions | Non-invasive assessment of the<br>cardiopulmonary response to<br>exercise or measurement of<br>energy expenditure using indirect<br>calorimetry. |
| Measurement<br>Mode(s) | Mixing chamber | Breath by breath |
| Output Parameters | Main outputs include resting<br>energy expenditure (REE),<br>respiratory quotient (RQ),<br>oxygen consumption (VO2),<br>carbon dioxide production<br>(VCO2), and respiratory<br>parameters. | Main outputs include resting<br>energy expenditure (REE),<br>respiratory quotient (RQ), oxygen<br>consumption (VO2), carbon<br>dioxide production (VCO2), and<br>respiratory parameters. |
| Major Separate<br>System Components | Device body, disposable<br>mask, headgear, sensor<br>cartridge, app in a mobile<br>device | Express® Series indirect<br>calorimeter, face tent/face mask,<br>printer, mobile cart |
| Flowmeter<br>Technology | Fixed-orifice differential<br>pressure pneumotach | Bi-directional Pitot Tube Flow<br>Sensor |
| O2 Sensor<br>Technology | Colorimetry | Galvanic |
| CO2 Sensor<br>Technology | Colorimetry | Non-dispersive infrared (NDIR) |
| User Interface | App on Mobile device | Touchscreen display |
| Software | Application | Intuitive icon-based software |
| Sterility | Non-sterile | Non-sterile |
| Feature | Subject Device | Predicate Device |
| Biocompatibility | All user contact components<br>are biocompatible and used<br>in already legally marketed<br>devices with the same<br>intended use | All patient contact components<br>are biocompatible and used in<br>already legally marketed devices<br>with the same intended use |
| Mechanical Safety | No mechanical safety risk<br>identified. | No mechanical safety risk<br>identified. |
| Chemical Safety | No chemical safety risks<br>identified. | No chemical safety risks<br>identified. |
| Anatomical Sites | Patient's mouth and nose,<br>through face mask | Patient's mouth and nose, through<br>face tent/face mask |
| Human factors | Requires patient to keep<br>mask sealed. | Requires patient to keep face<br>tent/face mask sealed. |
| Energy used | External or internal power<br>supply | External power supply |
| Energy delivered | No energy delivered to the<br>patient. | No energy delivered to the<br>patient. |
| Battery | Lithium-Ion, rechargeable | NA |
| Compatibility with<br>environment | No known issues. | No known issues. |
| Compatibility with<br>other devices | No known devices for<br>compatibility issues. | No known devices for<br>compatibility issues. |
| Standards Electrical<br>safety | IEC 60601-1: class II /<br>Internal Electric Power<br>Source - type BF | Unspecified |
| Standards: EMC<br>Testing | EN60601-1-2 | Unspecified |
| Thermal Safety | No known thermal safety<br>issues | No known thermal safety issues |
| Radiation Safety | No known radiation safety<br>issues. | No known radiation safety issues. |
| Sensor cartridge | One-time use sensor<br>cartridge for simultaneously<br>O2 and CO2 sensing | O2 and CO2 sensors are for long<br>time use (e.g. 1 year), and change<br>during maintenance periods |
| Dimensions | 80 x 110 x 160 mm | 240 x 190 x 267 mm* |
| Weight | 170 gm | 9.2 lbs. * |
| Measurement Range | Flow: 0-150 l/min<br>O2: 10-21%<br>CO2: 0-10% | Flow: 0-40 l/min*<br>O2: 5-85%*<br>CO2: 0-10% * |
| Feature | Subject Device | Predicate Device |
| Measurement<br>Accuracy<br>Error/Resolution | Flow Accuracy: <3% or 10<br>ml/min…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.