K200067 · Medacta International S.A. · HRX · Oct 1, 2020 · Orthopedic
Device Facts
Record ID
K200067
Device Name
MectaScope System
Applicant
Medacta International S.A.
Product Code
HRX · Orthopedic
Decision Date
Oct 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Indications for Use
MectaScope System is used to examine human joints such as knee and shoulder from the inside during arthroscopic diagnostic procedures, providing surgeons with magnified images on a monitor.
Device Story
MectaScope System consists of rigid arthroscopes, sheaths, and obturators used for internal joint examination. Device inputs include light delivered via external light-guide cables; optical images captured through the rigid scope. System transforms optical input into magnified visual output displayed on a monitor. Used in surgical settings by surgeons to visualize joint structures. Sheaths and obturators facilitate access and maintain portals. Components are reusable and provided non-sterile. System connects to third-party camera control units and light sources via standardized adapters. Benefits include minimally invasive visualization of joint pathology to assist surgical decision-making.
Clinical Evidence
No clinical data. Evidence consists of non-clinical bench testing including visual field/viewing direction (ISO 8600), concentricity (ISO 18339), fluid passage comparison, cadaver testing, and cleaning/sterilization validation.
Technological Characteristics
Rigid arthroscope system. Materials: stainless steel, EPDM. Light connector per ISO 18339. Available in 30° and 70° directions of view. Reusable. Connects to external camera/light sources via manufacturer-specific adapters. Non-sterile.
Indications for Use
Indicated for examination of human joints (e.g., knee, shoulder) during arthroscopic diagnostic procedures. For use by surgeons.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
October 1, 2020
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Medacta International SA Chris Lussier Senior Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis. Tennessee 38118
# Re: K200067
Trade/Device Name: MectaScope System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: September 1, 2020 Received: September 3, 2020
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Pooja Panigrahi -S
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K200067
Device Name MectaScope System
Indications for Use (Describe)
MectaScope System is used to examine human joints such as knee and shoulder from the inside during arthroscopic diagnostic procedures, providing surgeons with magnified images on a monitor.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory Affairs and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director, Quality and Regulatory, Medacta USA Date Prepared: January 13, 2020 Date Revised: September 29, 2020
#### II. Device
| Device Proprietary Name: | MectaScope System |
|--------------------------|-------------------|
| Common or Usual Name: | Arthroscope |
| Classification Name: | Arthroscope |
| Product Codes: | HRX |
| Regulation Number: | 21 CFR 888.1100 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- Arthroscopic Telescopes, K943052, Karl Storz Endoscopy .
#### IV. Reference Device
- KSEA Large Joint Arthroscopy Set, K963524, Karl Storz Endoscopy ●
#### V. Device Description
The MectaScope System is comprised of rigid scopes use to examine joints, a sheath used to access the surgical site, and obturators to facilitate the initial access of the sheath into the joint cavity in preparation for arthroscopic procedures.
The rigid scopes, manufactured primarily from stainless steel, provide a circular field of view and are available in two (2) different directions of view, 30° and 70°. The sheath, manufactured from stainless steel and EPDM, is introduced into the joint using the obturators and maintains a stable portal during the arthroscopic procedure. The obturators, manufactured with stainless steel and
{4}------------------------------------------------
EPDM, are provided with or without a handle; the obturator with a handle is compatible with a Ø 1.5 mm k-wire. The MectaScope System components are provided non-sterile and are reusable.
The MectaScope scopes are designed to be used with legally marketed camera and control units and are connected to legally marketed available light-guide cables via the light connector (designed in accordance with ISO 18339). Manufacturer specific light guide adapters are provided with the MectaScope System including adapters for Olympus/Storz, Richard Wolf, and ACMI light guide cables.
#### VI. Indications for Use
MectaScope System is used to examine human joints such as knee and shoulder from the inside during arthroscopic diagnostic procedures, providing surgeons with magnified images on a monitor.
#### VII. Comparison of Technological Characteristics
The MectaScope System and the predicate device share the following characteristics:
- arthroscope working length, diameter, field of view, and direction of view; .
- sheath design: ●
- . eyepiece design;
- materials of construction; and
- connection to legally marketed camera systems and light guide cables.
The MectaScope System and the predicate devices are technologically different with respect to the design of connections between system components, camera, and fiber optic cable, as well as sheath and obturator design.
A comparison of key technological features between the subject and predicate device is provided in the table below.
| | MectaScope System | Arthroscopic Telescopes<br>(K943052) |
|-------------------|-------------------|--------------------------------------|
| Arthroscope | | |
| Type of scope | Rigid | Rigid |
| Outer diameter | 4 mm | 4 mm |
| Field of view | circular | circular |
| Direction of view | 30° and 70° | 30° and 70° |
| Reusable | Yes | Yes |
## Discussion
Although slight differences in design with respect to connections and obturator design exist, these differences do not raise different questions of safety and effectiveness as the products utilizing the
{5}------------------------------------------------
same fundamental scientific technology. Design verification testing was undertaken and demonstrates that the subject device performs to specification and as intended.
# VIII. Performance Data
Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed on worst-case device configurations in support of a substantial equivalence determination:
Comparative study
- . Design Validation Report, Scope and Sheath - KARL STORZ Arthroscope Sheath;
- o comparison of fluid passage (volume unit per time) with the MectaScope Rotating Sheath and the KARLSTORZ Arthroscope Sheath, utilizing a rotating spigot plane
Non-Clinical Studies:
- visual field and viewing direction testing per ISO 8600; ●
- concentricity test per ISO 18339;
- cadaver testing; and
- cleaning and sterilization validation. ●
#### IX. Conclusion
The information provided with this submission supports that the MectaScope System is substantially equivalent to the identified predicate device.
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.