K200056 · Amo Manufacturng USA, LLC · OOE · May 18, 2020 · Ophthalmic
Device Facts
Record ID
K200056
Device Name
Catalys Precision Laser System
Applicant
Amo Manufacturng USA, LLC
Product Code
OOE · Ophthalmic
Decision Date
May 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CATALYS® Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
Device Story
The Catalys Precision Laser System is an ophthalmic surgical laser used in hospitals and ASCs by surgeons. It utilizes femtosecond laser technology and integrated Optical Coherence Tomography (OCT) to perform cataract surgery procedures. The system takes input from the OCT to map ocular anatomy, including the cornea and crystalline lens. A three-dimensional scanner directs the laser beam to create precise incisions and lens fragmentation via plasma and cavitation. The system includes a single-use Liquid Optics Interface for patient contact. Surgeons use the graphical user interface to monitor the procedure and verify incision placement relative to ocular structures. The device facilitates efficient lens removal and astigmatism management. Software updates in this version include improved workflow, patient exam import, and iris registration to account for cyclorotation.
Clinical Evidence
No clinical data was required. Substantial equivalence was demonstrated through non-clinical performance data, including hardware and software bench testing, subsystem verification, and system-level validation.
Technological Characteristics
Class 4 femtosecond ophthalmic laser (1030 nm wavelength). Features integrated Spectral Domain OCT (15 µm transverse resolution, 1000 A-scans/sec). System uses a microprocessor-controlled 3D scanner and Liquid Optics Interface (single-use). Software-driven workflow with GUI. Connectivity includes encrypted data transfer for patient exams. Sterilization via single-use sterile interfaces.
Indications for Use
Indicated for patients undergoing cataract surgery for removal of the crystalline lens, including anterior capsulotomy, phacofragmentation, and corneal arc cuts/incisions.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 18, 2020
AMO Manufacturing USA, LLC Laarni Ricafort Project Manager, Regulatory Affairs 510 Cottonwood Drive Milpitas, CA 95035
Re: K200056
Trade/Device Name: Catalys Precision Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: April 16, 2020 Received: April 20, 2020
Dear Laarni Ricafort:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tieuvi Nguyen, Ph.D. Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200056
Device Name CATALYS® Precision Laser System
#### Indications for Use (Describe)
The CATALYS® Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intended uses in cataract surgery include anterior capsulotomy, phacofragmentation, and the creation of single plane and multi-plane arc cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|---------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
## [807.92(a)(1)] Submitter Information
| Sponsor/Submitter: | AMO Manufacturing USA, LLC<br>510 Cottonwood Drive<br>Milpitas, CA 95035 USA<br>Phone 408-273-5166<br>Fax:408 273-5966 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Laarni Ricafort<br>Project Manager, Regulatory Affairs<br>Email: lricafor@its.jnj.com<br>Phone: 408-273-4016<br>Fax: :408 273-5966 |
Date Summary Prepared: May 7, 2020
## [807.92(a)(2)] Name of Device
| Device Trade Name: | CATALYS® Precision Laser System |
|------------------------|---------------------------------|
| Common Name: | Ophthalmic laser |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 886.4390 |
| Classification Name: | Ophthalmic Femtosecond Laser |
| Product Code: | OOE |
# [807.92(a)(3)] Legally Marketed Devices
| Primary Predicate Device: | CATALYS® Precision Laser System<br>(K182083, November 9, 2018) |
|-----------------------------|-----------------------------------------------------------------|
| Reference Predicate Device: | LENSAR Laser System (LLS-fs 3D)<br>(K182795, December 21, 2018) |
{4}------------------------------------------------
### [807.92(a)(4)] Device Description
Device Description: Catalys® Precision Laser System ophthalmic surgical laser system used in healthcare facilities such as hospitals, Ambulatory Surgery Centers (ASCs) and surgeon office settings. The System is an electromedical device that contains software. System components include a single-use Liquid Optics™ Interface and optional Mobile Patient Bed.
> The Catalys® Precision Laser System (also referred to as the Catalys® System or System) is an ophthalmic surgical laser system indicated for use in cataract surgery to create a precise anterior capsulotomy and/or to effect lens fragmentation, thus facilitating efficient lens removal. The System also creates single plane and multi-plane arc-cuts/incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure. The System employs femtosecond ("FS") laser technology with integrated Optical Coherence Tomography ("OCT"), all of which are controlled and monitored by dedicated electronics. The System utilizes a common optical path for the OCT and femtosecond treatment laser (including the three-dimensional scanner and Liquid Optics™ [patient] Interface). As such, the beams are intrinsically co-registered and provide for precise overlap between imaging and treatment beams. Ocular surfaces recognized by the system software include anatomy within the anterior chamber, such as the anterior and posterior corneal surfaces and the anterior and posterior surfaces of the crystalline lens. Detailed axial or sagittal cross-sectional views are available via OCT, to demarcate proposed incisions versus adjacent ocular structures (for example, iris, pupil and limbus).
> The Catalys® Precision Laser System laser classification per 21 CFR 1040.10 and 1040.11 is Class 4.
### [807.92(a)(5)] Intended Use
| Indications for Use: | The Catalys® Precision Laser System is indicated for use in patients<br>undergoing cataract surgery for removal of the crystalline lens.<br>Intended uses in cataract surgery include anterior capsulotomy,<br>phacofragmentation, and the creation of single plane and multi-plane<br>arc cuts/incisions in the cornea, each of which may be performed either<br>individually or consecutively during the same procedure. |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Difference in<br>Indications from | The Catalys® Precision Laser System is indicated for use in patients<br>undergoing cataract surgery for removal of the crystalline lens. |
{5}------------------------------------------------
Intended uses in cataract surgery include anterior capsulotomy. Primary Predicate phacofragmentation, and the creation of single plane and multiplane arc cuts/incisions in the cornea, each of which may be Device performed either individually or consecutively during the same procedure. The subject device and primary predicate device have the same indications for use.
### [807.92(a)(6)] Technical Characteristics
Technological The modified Catalys Precision Laser System is unchanged with regard to its technological characteristics, indications for use, and Characteristics: intended uses. The software revisions in the modified device include updates to the graphical user interface and host to improve the efficiency of the workflow as well as the addition of software modules to support the import of patient exams and iris registration. The revisions in the modified device are constrained to only software changes, of which most are related to advanced astigmatism management and additional improvements to currently implemented features.
> Software verification and validation testing in addition to bench testing was performed to verify the ability of the software to meet its intended use and to ensure that no adverse effects were introduced due to the software changes. This testing included subsystem level verification and regression testing as well as system validation using the latest software, Mobile Patient Bed, and Liquid Optics Interface.
> The following table provides a comparison of the primary predicate device, reference predicate device, and subject device for the purpose of demonstrating substantial equivalence to the predicate devices for its indication for use, intended use, technological characteristics and added new features.
{6}------------------------------------------------
| | Subject Device | Primary Predicate Device | Reference Predicate<br>Device |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | Catalys Precision Laser<br>System | Catalys Precision Laser<br>System | LENSAR Laser<br>System<br>LLS-fs 3D Laser System |
| 510(k) Number | K200056 | K182083 | K182795 |
| Regulation Number | 886.4390 | 886.4390 | 886.4390 |
| Regulation Name | Ophthalmic Laser | Ophthalmic Laser | Ophthalmic Laser |
| Regulatory Class | Class II | Class II | Class II |
| Product Code | OOE | OOE | OOE |
| Indications for Use | The Catalys® Precision<br>Laser System is indicated<br>for use in patients<br>undergoing cataract<br>surgery for removal of the<br>crystalline lens. Intended<br>uses in cataract surgery<br>include anterior<br>capsulotomy,<br>phacofragmentation, and<br>the creation of single plane<br>and multi-plane arc<br>cuts/incisions in the<br>cornea, each of which may<br>be performed either<br>individually or<br>consecutively during the<br>same procedure. | The Catalys® Precision<br>Laser System is indicated<br>for use in patients<br>undergoing cataract<br>surgery for removal of the<br>crystalline lens. Intended<br>uses in cataract surgery<br>include anterior<br>capsulotomy,<br>phacofragmentation, and<br>the creation of single plane<br>and multi-plane arc<br>cuts/incisions in the<br>cornea, each of which may<br>be performed either<br>individually or<br>consecutively during the<br>same procedure. | The LENSAR Laser<br>System - fs 3D (LLS-fs 3D)<br>with Streamline™ is<br>an ophthalmic surgical<br>laser indicated for use:<br>• in the creation of an<br>anterior capsulotomy;<br>• in patients undergoing<br>surgery requiring<br>laser-assisted<br>fragmentation of the<br>cataractous lens;<br>• in the creation of full<br>and partial thickness<br>single-plane and multiplane<br>arc cuts/<br>incisions in the<br>cornea;<br>• in patients undergoing<br>ophthalmic surgery or<br>other treatments<br>requiring pocket<br>cuts/incisions in the<br>cornea;<br>• in the creation of a<br>corneal flap in patients<br>undergoing treatment<br>requiring initial<br>lamellar resection of<br>the cornea; |
| System Type | Opthalmic Femtosecond<br>Laser with Spectral<br>Domain OCT | Opthalmic Femtosecond<br>Laser with Spectral<br>Domain OCT | Not applicable |
| Laser Mechanism of<br>Action | Plasma, Cavitation | Plasma, Cavitation | Not applicable |
| Treatment Laser<br>Wavelength (nm) | 1030 ±5 | 1030 ±5 | Not applicable |
| Output Power, Max | Per ISO 15004-2:2007<br>limits | Per ISO 15004-2:2007<br>limits | Not applicable |
| Maximum Pulse<br>Energy (µJ) | 10 | 10 | Not applicable |
| Repetition Rate<br>(kHz) | 9-120 | 9-120 | Not applicable |
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
| | | | |
| Attribute | | | |
| | Catalys Precision Laser<br>System | Catalys Precision Laser<br>System | LENSAR Laser<br>System<br>LLS-fs 3D Laser System |
| Pulse Duration (fs) | < 600 | < 600 | Not applicable |
| Spot Size; diameter<br>(μm) | 5 | 5 | Not applicable |
| System controls | Microprocessor with<br>Graphical User Interface | Microprocessor with<br>Graphical User Interface | Not applicable |
| Patient Contact<br>Interface | Suction-ring type interface<br>devices (marketed as<br>Liquid Optics™ Interface)<br>Sterile and Single-use<br>Cleared on K141079 &<br>K170322 | Suction-ring type interface<br>devices (marketed as<br>Liquid Optics™ Interface)<br>Sterile and Single-use<br>Cleared on K141079 &<br>K170322 | Not applicable |
| | LOI<br>External (mm): 21.6<br>Internal (mm): 14.1 | LOI<br>External (mm): 21.6<br>Internal (mm): 14.1 | |
| LOI Suction Ring<br>Seal Diameters<br>(mm) | LOI-12<br>External (mm):19<br>Internal (mm):12 | LOI-12<br>External (mm):19<br>Internal (mm):12 | Not applicable |
| | 0180-1401<br>External (mm): 21.6<br>Internal (mm): 14.1 | 0180-1401<br>External (mm): 21.6<br>Internal (mm): 14.1 | |
| | 0180-1201<br>External (mm): 19<br>Internal (mm): 12 | 0180-1201<br>External (mm): 19<br>Internal (mm): 12 | |
| | OCT Axial<br>Resolution (µm) | 30 | |
| OCT transverse<br>Resolution (µm) | 15 | 15 | Not applicable |
| Scan speed (A-<br>scans/sec) | 1000 | 1000 | Not applicable |
| A-scan depth (nm) | 2 | 2 | Not applicable |
| Optical Source | 820-930 | 820-930 | Not applicable |
| Optical Power | ANSI Class 1 < 3.48mW at<br>cornea | ANSI Class 1 < 3.48mW at<br>cornea | Not applicable |
| Iris Imaging | Live iris view | Live iris view | Not applicable |
| Trajectory Timing<br>Synchronization | FPGA coordinates from<br>one<br>non--reentrant VI | FPGA coordinates from<br>one<br>non--reentrant VI | Not applicable |
| Communication<br>Method for<br>Watchdog for Host<br>PC with the Mobile<br>Patient Bed Pairing | Direct FPGA Interface | Direct FPGA Interface | Not applicable |
| Software Features | Built-in Nomogram:<br>Inclusion of nomogram<br>formula calculator for<br>arcuate incisions | The items listed under the<br>"subject device" column<br>are new features. All<br>other software features not | Not applicable |
| Attribute | Subject Device | Primary Predicate Device | Reference Predicate<br>Device |
| | Catalys Precision Laser<br>System…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.