The ExcelsiusGPS is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody spacers, and intracranial devices including biopsy needles, electrodes, and tubes.
Device Story
ExcelsiusGPS Cranial 1.0 Module is a computer-assisted surgical navigation system. Inputs include preoperative/intraoperative CT, MRI, or fluoroscopy images. System performs registration between virtual patient images and physical patient anatomy using a dynamic reference base and optical tracking camera. Surgeon plans trajectories on images; system tracks compatible instruments in real-time, displaying position on images. Used in OR by surgeons for stereotactic procedures. Output is visual guidance for instrument placement. Benefits include precise navigation to cranial targets, aiding in biopsy, electrode, and tube placement. System manages data, user, and safety functions.
Clinical Evidence
No clinical trials on human patients. Evidence includes non-clinical system, software, and instrument verification/validation; surgical simulations on phantom cranial sawbone models; and human cadaveric quantitative validation. Cadaveric study demonstrated system accuracy in navigating cranial devices to targets. Navigational accuracy results: mean position <1.5mm (99% CI ≤2mm) and mean angle <2.0° (99% CI ≤2°). Biocompatibility per ISO 10993-1 and electrical safety per IEC 60601-1 confirmed.
Technological Characteristics
Optical tracking system; uses preoperative/intraoperative CT, MRI, and fluoroscopy. Components include dynamic reference base, positioning camera, and navigated instruments with unique array patterns/reflective markers. Software-based navigation; connectivity for data/network management. Biocompatible materials per ISO 10993-1. Electrical safety per IEC 60601-1. System supports adult and pediatric stereotactic procedures.
Indications for Use
Indicated for adult and pediatric patients requiring stereotactic neurosurgery for the placement of intracranial devices including biopsy needles, electrodes, and tubes, as well as spinal and orthopedic bone screws and interbody spacers.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
July 10, 2020
Globus Medical Inc. Jennifer Antonacci Group Manager, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403
Re: K200047
Trade/Device Name: ExcelsiusGPS Cranial 1.0 Module Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, HAW Dated: June 11, 2020 Received: June 12, 2020
Dear Jennifer Antonacci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200047
Device Name ExcelsiusGPS Cranial 1.0 Module
#### Indications for Use (Describe)
The ExcelsiusGPS is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or guide to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody spacers, and intracranial devices including biopsy needles, electrodes, and tubes.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: ExcelsiusGPS Cranial 1.0 Module
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403<br>610-930-1800 |
|------------------|-----------------------------------------------------------------------------------------|
| Primary Contact: | Jennifer Antonacci, Ph.D.<br>Group Manager, Regulatory Affairs |
| Date Prepared: | July 10, 2020 |
| Device Name: | ExcelsiusGPS® Cranial 1.0 Module |
| Common Name: | Computer-assisted surgical device |
- Classification: Per 21 CFR as follows: §882.4560 Stereotaxic instrument Product Code(s): OLO, HAW Regulatory Class: II
Primary Predicate: ExcelsiusGPS® (K171651)
- Other Predicates: ExcelsiusGPS® Spine 1.1 Interbody Module (K191100) Medtronic StealthStation S8 (K162309) ROSA ONE Brain (K182417)
# Purpose:
The purpose of this submission is to request clearance of the ExcelsiusGPS® Cranial 1.0 Module software and navigated instruments for access, preparation, and placement of cranial devices.
# Device Description:
The ExcelsiusGPS® Cranial Module includes hardware and software that enables real time surgical navigation using radiological patient images (MRI, CT, and fluoroscopy), using a dynamic reference base and positioning camera. The navigation system determines the registration or mapping between the virtual patient (points on the patient images) and the physical patient (corresponding points on the patient's anatomy). Once this registration is created, the software displays the relative position of a tracked instrument on the patient images. As an aid to visualization, the surgeon can plan trajectories for instrument placement on the patient images prior to surgery. Registration provides the necessary information to provide visual assistance to the surgeon during freehand navigation. During surgery, the system tracks the position of GPS compatible instruments in or on the patient anatomy and continuously updates the instrument position on patient images utilizing optical tracking. System software is responsible for all
{4}------------------------------------------------
navigation functions, data storage, network connectivity, user management, case management, and safety functions. ExcelsiusGPS® surgical instruments include non-sterile, re-usable instruments and sterile instruments that are operated manually or with the use of the positioning system.
The ExcelsiusGPS® Cranial Module is designed to assist with stereotactic procedures that include quidance to cranial targets for instrument navigation and device placement. Instruments consist of end effector instruments, registration instruments, navigated instruments, patient positioning instruments, and surgical instruments. End effector instruments include instruments to the Interchangeable Guide End Effector. Registration and navigated instruments incorporate unique array patterns with reflective markers, and are used to track patient anatomy and surgical instruments. Patient positioning instruments aid in patient fixation. Surgical instruments are used to access and prepare the local site and place devices, such as needles, electrodes, and tubes. The ExcelsiusGPS® Cranial Module is suitable for use in adult and pediatric patient populations in which the use of stereotactic neurosurgery may be appropriate.
### Indications for Use:
The ExcelsiusGPS® is intended for use as an aid for precisely locating anatomical structures and for spatial positioning and orientation of an instrument holder or quide tube to be used by surgeons for navigating and/or quiding compatible surqical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on CT scans or fluoroscopy. The system is indicated for the placement of spinal and orthopedic bone screws and interbody spacers, and intracranial devices including biopsv needles, electrodes, and tubes.
### Technological Characteristics:
The ExcelsiusGPS® Cranial Module and associated instruments have similar technological characteristics to the predicate devices including the main system components, workflow, user interface, software features, and design. The ExcelsiusGPS® Cranial Module is comparable to the predicates in terms of intended use, fundamental scientific technology, technological characteristics and principle of operation.
| Device | Subject<br>ExcelsiusGPS®<br>Cranial Module | Predicate<br>ExcelsiusGPS®<br>(K171651, K191100) | Predicate<br>StealthStation S8<br>(K162309) | Predicate<br>ROSA ONE Spine<br>(K182417) |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of<br>operation | - Intraoperative/<br>preoperative<br>images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking<br>of navigated<br>instruments | - Intraoperative/<br>preoperative<br>images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking<br>of navigated<br>instruments | - Intraoperative/<br>preoperative<br>images<br>- Patient<br>registration<br>- Surgical planning | - Intraoperative/<br>preoperative<br>images<br>- Patient registration<br>- Surgical planning<br>- Guidance of<br>instruments |
| Device | Subject<br>ExcelsiusGPS®<br>Cranial Module | Predicate<br>ExcelsiusGPS®<br>(K171651, K191100) | Predicate<br>StealthStation S8<br>(K162309) | Predicate<br>ROSA ONE Spine<br>(K182417) |
| | - Guidance of<br>instruments | - Guidance of<br>instruments | - Real-time tracking<br>of navigated<br>instruments | |
| Input images | 3D preoperative<br>exam<br>3D intraoperative<br>exam<br>2D intraoperative<br>exam | 3D preoperative<br>exam<br>3D intraoperative<br>exam<br>2D intraoperative<br>exam | 3D preoperative<br>exam<br>3D intraoperative<br>exam<br>2D intraoperative<br>exam | 3D preoperative<br>exam<br>3D intraoperative<br>exam<br>2D intraoperative<br>exam |
| Integrated<br>planning<br>software | ExcelsiusGPS®<br>Planning and<br>Navigation<br>Application Software | ExcelsiusGPS®<br>Planning and<br>Navigation<br>Application Software | Stealthstation<br>Cranial | ROSANNA BRAIN |
| Save/load<br>planning | Yes | Yes | Yes | Yes |
| Merge images<br>functionality | Yes | Yes | Yes | Yes |
| Image-guided | Yes | Yes | Yes | Yes |
| Patient<br>registration<br>method | Intra-Op CT:<br>Registration<br>Fixture<br>Fluoroscopy:<br>Registration<br>Fixture<br>Stereotactic<br>Localizer<br>Registration | Intra-Op CT:<br>Registration<br>Fixture<br>Pre-Op CT:<br>Fluoroscopic to<br>Pre-Op CT Merge<br>Fluoroscopy:<br>Registration<br>Fixture | Point-to-point<br>registration with<br>anatomical<br>markers or<br>skin/bone<br>fiducials<br>Optical or<br>electromagnetic<br>trace merge<br>Intra-Op CT:<br>Calibrated CT<br>gantry to camera<br>merge<br>Stereotactic<br>Localizer<br>Registration | Point-to-point<br>registration with<br>anatomical markers<br>or skin/bone<br>fiducials<br>Stereotactic Localizer<br>Registration<br>Optical surface<br>registration |
| Accuracy<br>verification on<br>anatomical<br>landmarks | Yes | Yes | Yes | Yes |
| Real time<br>display of<br>instrument<br>position | Yes | Yes | Yes | Yes |
| Patient<br>fixation | Reference is<br>attached to patient<br>head holder<br>- Head is fixed to<br>patient stabilization<br>stand | Reference is fixed to<br>patient's bony<br>structure such as a<br>long bone, iliac<br>crest, spinous<br>process, vertebra,<br>etc. for tracking<br>system | Reference is<br>attached to patient<br>head holder | Head is fixed onto<br>robot arm's flange |
#### Comparison of Principles of Operation and Technological Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Testing:
Verification and validation testing was conducted on ExcelsiusGPS® Cranial Module to confirm that the device meets performance requirements under the indications for use and to ensure safety and efficacy of the system:
- Non-clinical system, software, and instrument verification and validation -● demonstrated compliance with user needs and corresponding design inputs
- Surgical simulations conducted on phantom models cranial sawbone . models and registration matrix were used to quantify accuracy in a controlled setting
- Human cadaveric quantitative validation under clinically relevant scenarios - demonstrated system accuracy in navigating cranial devices to the desired location on patient images
- . Compliance conformity assessments per:
- o IEC 60601-1 Medical electrical equipment. General requirements for basic safety and essential performance, 2005, Amendment 1, 2012
- o IEC 60601-1-6 Medical Electrical Equipment Part 1-6, General Requirements for Basic Safety and Essential Performance - Usability
- Navigational accuracy of the subject device system with mean, standard . deviation, and 99% confidence interval (CI). Results demonstrate the subject device meets the accuracy acceptance criteria for position and angle as follows:
| Position | | Angle | |
|----------|--------|--------|--------|
| Mean | 99% CI | Mean | 99% CI |
| < 1.5mm | ≤ 2mm | < 2.0° | ≤ 2° |
# Biocompatibility:
The biocompatibility evaluation for ExcelsiusGPS® has been conducted in accordance with FDA Guidance for Industry and FDA Staff, "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process'," June 16, 2016. The evaluation confirms that ExcelsiusGPS® meets biocompatibility requirements.
### Electrical Safety and Electromaqnetic Compatibility:
Testing was performed to assure compliance with recognized safety standard: IEC 60601-1:2012 standard for electrical safety. Electromagnetic compatibility was not affected.
### Software Verification and Validation Testing:
Software validation and verification testing was performed in accordance with the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). The software for this device is considered a "MAJOR" level of concern.
{7}------------------------------------------------
# Basis of Substantial Equivalence:
ExcelsiusGPS® Cranial Module has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.