K200031 · Avalon Biomedical (Shenzhen) Limited · LBI · Oct 5, 2020 · General Hospital
Device Facts
Record ID
K200031
Device Name
Neonatal Phototherapy System
Applicant
Avalon Biomedical (Shenzhen) Limited
Product Code
LBI · General Hospital
Decision Date
Oct 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Neonatal Phototherapy System, Model: KANGALITE is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates: gestational age ≥38 weeks and weight within 2500-4500 grams. It can be used in the clinical setting or in the home environment. It should be used for neonates whom phototherapy has been prescribed.
Device Story
Wearable phototherapy system for neonatal hyperbilirubinemia; treats neonates in clinical or home settings. System comprises light source, fiber optic panels, swaddle, strap, and disposable covers. Light source emits narrow-band blue LED light (400-500 nm, peak 455±10 nm) transmitted via fiber optic bundle to panels. Bilirubin absorbs blue light, converting to water-soluble waste products for excretion. Device allows kangaroo care; no eye protection required. Healthcare provider prescribes therapy; device operated by clinicians or parents. Output is fixed-intensity blue light; irradiance >30 µW/cm²/nm. Benefits include effective bilirubin reduction and potential for home-based treatment.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and phototherapy performance (IEC 60601-2-50).
Indicated for treatment of unconjugated hyperbilirubinemia in neonates with gestational age ≥38 weeks and weight 2500-4500 grams. Contraindicated in cases of congenital porphyria, family history of porphyria, or concomitant use of photosensitizing drugs/agents.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 5, 2020
Avalon Biomedical (Shenzhen) Limited Eva Zhu Regulatory Affairs Manager 805 Building 4, Yinxing Zhije, 1301-74, GuanGuang Road, XinLan Community Shenzhen, 518110 China
Re: K200031
Trade/Device Name: Neonatal Phototherapy System, Model: KANGALITE Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: Class II Product Code: LBI Dated: August 31, 2020 Received: September 3, 2020
Dear Eva Zhu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200031
Device Name
Neonatal Phototherapy System, Model: KANGALITE
Indications for Use (Describe)
The Neonatal Phototherapy System, Model: KANGALITE is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates: gestational age ≥38 weeks and weight within 2500-4500 grams. It can be used in the clinical setting or in the home environment. It should be used for neonates whom phototherapy has been prescribed.
Contraindication
The Neonatal Phototherapy System, Model: KANGALITE should NOT be used in cases of:
(a) Congenital porphyria or a family history of porphyria
(b) Concomitant use of drugs or agents that are photosensitizers
A use that does not take into account these contraindications may lead to an ineffective treatment or to potential risks to the patient's health.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Avalon Biomedical. The logo consists of a circular pattern of blue dots on the left, followed by the word "AVALON" in a larger, bold, blue font. Below "AVALON" is the word "BIOMEDICAL" in a smaller, lighter blue font. The logo is simple and modern, and the use of blue suggests a sense of trust and reliability.
# 510(k) Summary
# K200031
The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92:
## Date: October 5, 2020
#### 1. Submitter:
Submitter: Avalon Biomedical (Shenzhen) Limited Address: 805, Building 4, Yinxing Zhijie, 1301-74, GuanGuang Road, XinLan Community, GuanLan Street, Longhua District, 518110, Shenzhen, China Contact Person: Eva Zhu, Regulatory Affairs Manager Phone: +86 755 2107 0534 Email: zhul@avalonbiomedical.com
#### 2. De vice Name
Trade or Proprietary Name: Neonatal Phototherapy System, KANGALITE™ Common or Usual Name: Neonatal Phototherapy System Classification Name: Neonatal phototherapy unit
Device Class: II Product Code: LBI Panel: General Hospital Regulation Number: 21 CFR 880.5700
#### 3. Predicate Device:
510(k) Number: K070180 Device Name and Model: Bili-Tx Manufacturer Name: Respironics Inc.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Avalon Biomedical. The logo consists of a circular pattern of blue dots on the left, and the words "AVALON" and "BIOMEDICAL" on the right. The word "AVALON" is in a larger, bolder font than the word "BIOMEDICAL". The color of the text is also blue.
## 4. Indications for Use:
The Neonatal Phototherapy System, Model: KANGALITE is intended for the treatment of unconjugated hyperbilirubinemia in the population of neonates: gestational age ≥38 weeks and weight within 2500-4500 grams. It can be used in the clinical setting or in the home environment. It should be used for neonates whom phototherapy has been prescribed.
#### Contraindication
The Neonatal Phototherapy System, Model: KANGALITE should NOT be used in cases of:
(a) Congenital porphyria or a family history of porphyria
(b) Concomitant use of drugs or agents that are photosensitizers
A use that does not take into account these contraindications may lead to an ineffective treatment or to potential risks to the patient's health.
#### 5. Device Description:
The Neonatal Phototherapy System, Model: KANGALITE is a wearable phototherapy system that consists of six components: Light Source with AC Power Cord, Dual Fiber Optic Panels (Standard accessory) /Single Fiber Optic Panel (optional accessory), Swaddle, Strap, Disposable Covers and Table Mounting Clip (optional accessory). The Neonatal Phototherapy System, Model: KANGALITE delivers a narrow band of high-intensity blue light emitting diode (LED), in order to provide phototherapy treatment for neonatal hyperbilirubinemia.
The Neonatal Phototherapy System, Model: KANGALITE Light Source contains a blue LED which emits light in the range 400-500 nm (peak wavelength 455±10 nm). The system gives only one blue light irradiance output level which is not adjustable. This blue light is directed through the optical fiber bundle to the illuminating area of the Fiber Optic Panel where the newborn receives blue light therapy.
The Fiber Optic Panel is inserted into the Disposable Cover and is placed in a soft and comfortable Swaddle, while the baby is wrapped inside to receive phototherapy. Baby is allowed to receive kangaroo care. Besides, compare to conventional phototherapy, no eye protection for the baby is required as blue light will not leak out when properly set up.
The light output (>30 µW/cm²/nm) is sufficient to provide intensive phototherapy which is defined
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Avalon Biomedical. The logo consists of a circular pattern of blue dots on the left, followed by the word "AVALON" in large, bold, blue letters. Below "AVALON" is the word "BIOMEDICAL" in smaller, lighter blue letters.
as in the American Academy of Pediatrics (AAP) standard.
Average irradiance for:
Dual Fiber Optic Panels: 50 µW/cm2/nm (±25%);
Single Fiber Optic Panel: 60 uW/cm2/nm (±25%).
The spectrum range of the blue LED light matches with the spectral absorption of bilirubin. Under blue light irradiation, subcutaneous bilirubin absorbs strongly in the blue region of the light spectrum and converts into less lipophilic waste products which can be excreted through the bile or urine without the need for conjugation.
The Neonatal Phototherapy System, Model: KANGALITE can be directly connected to nominal voltages readily available throughout the world as the Light Source is affixed with a wide-range medical adaptor, rated from 100-240 VAC and 50/60 Hz. It converts the power to 15 VDC to power up the Light Source.
Environment:
Operating temperature/humidity: 59 to 95 °F (15 to 35 °C)/15 to 95 % RH, non-condensing Storage temperature/humidity: -4 to 122 °F (-20 to 50 °C) /15 to 95 %RH, non-condensing Atmospheric pressure: 700-1060 hPa
#### 6. Comparison of technological characteristics with the predicate de vice
#### Comparison Table
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Manufacturer | Avalon Biomedical (Shenzhen)<br>Limited | Respironics Inc. | N/A |
| Regulation No. | 880.5700 | 880.5700 | same |
| Product Code | LBI | LBI | same |
| Classification | II | II | same |
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
| Indications for<br>use | The Neonatal Phototherapy<br>System, Model: KANGALITE is<br>intended for the treatment of<br>unconjugated hyperbilirubinemia<br>in the population of neonates:<br>gestational age ≥38 weeks and<br>weight within 2500-4500 grams. It<br>can be used in the clinical setting<br>or in the home environment. It<br>should be used for neonates whom<br>phototherapy has been prescribed.<br>Contraindication<br>The Neonatal Phototherapy<br>System, Model: KANGALITE<br>should NOT be used in cases of:<br>(a) Congenital porphyria or a<br>family history of porphyria<br>(b) Concomitant use of drugs or<br>agents that are photosensitizers<br>A use that does not take into<br>account these contraindications<br>may lead to an ineffective<br>treatment or to potential risks to<br>the patient's health. | The Bili-Tx is intended to<br>treat hyperbilirubinemia<br>through phototherapy in a<br>home<br>Or<br>hospital/institutional<br>environment. | Different<br>See comment 1 |
| Contraindications | 1. Congenital porphyria or a family<br>history of porphyria<br>2. Concomitant use of drugs or<br>agents that are<br>photosensitizers | There is no publicly-<br>available information. | Different<br>See comment 1 |
| | Subject device<br>Neonatal Phototherapy System | Predicate device | |
| Device name | Model: KANGALITE | Bili-Tx | Comparison |
| | K200031 | K070180 | |
| | KANGALITE treats neonatal<br>hyperbilirubinemia via blue light<br>phototherapy. The device delivers<br>a narrow band of high-intensity<br>blue light via a blue light<br>emitting diode (LED). Under<br>blue light irradiation,<br>subcutaneous bilirubin absorbs<br>strongly in the blue region of the<br>light spectrum and converts into<br>less lipophilic waste products<br>which can be excreted through the<br>bile or urine without the need for<br>conjugation. | The Bili-Tx phototherapy<br>systemuses blue light<br>emitting diodes (LEDs) to<br>convert bilirubin to waste<br>products that are mostly<br>excreted through urine<br>and stool, thus reducing<br>the bilirubin level in the<br>baby's blood. | |
| Principle of<br>operation | | | Same |
| | 1. Swaddle, the material in contact<br>with patient's body skin is made<br>of 100% cotton<br>2. Disposable Cover, the material<br>in contact with patient's chest and<br>back skin is made of non-woven<br>polypropylene<br>3. Dual Fiber Optic Panels, the<br>material in contact with patient's<br>chest and back skin is made of<br>TPU<br>4. Single Fiber Optic Panel, the<br>material in contact with patient's<br>back skin is made of TPU | There is no publicly-<br>available information. | |
| Patient contact<br>materials<br>(Patient is in supine<br>position) | | | Different<br>See comment 2 |
| Targeted<br>population | Neonates: gestational age ≥38 weeks<br>and weight within 2500-4500 grams. | Neonates | Different<br>See comment 3 |
| Prescription Use | Yes | Yes | Same |
| Over-The-Counter<br>Use | No | No | Same |
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
| Site of use | In the clinical setting or in the<br>home | In the clinical setting or in<br>the home | Same |
| Treatment mode | Use with Fiber Optic Panel | Two different modes<br>1) Use with Fiber Optic<br>Panel<br>2) Overhead treatment | Different<br>See comment 4 |
| Device Components | 1. Light Source (with AC power<br>cord)<br>2. Dual Fiber Optic Panels<br>(Standard accessory) / Single Fiber<br>Optic Panel (optional accessory)<br>3. Swaddle (One size)<br>4. Shoulder Strap<br>5. Table Mounting Clip (optional<br>accessory)<br>6. Disposable Cover × 2 (single-<br>use) | With Fiber Optic Panel:<br>1. Illuminator device with<br>AC power cord<br>2. Fiber Optic Panel<br>3. Disposable Cover<br>4. Illuminator device<br>carrying case (optional)<br>5. System Carrying Case<br>(optional)<br>Other components from<br>overhead phototherapy<br>6. Mounting brace<br>7. Circuit support arm<br>8. Circuit support adaptor | Similar<br>See comment 5 |
| Eye protection for<br>the baby is<br>required during<br>treatment | No | No | Same |
| If Eye Shield<br>includes in the<br>accessories list | No | No | Same |
| TECHNOLOGICAL FEATURES | | | |
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
| Light Source | Single blue LED | Uses blue light emitting<br>diodes (LEDs) | Similar<br>See comment 6 |
| Treatment<br>Wavelength | Blue LED: 400-500 nm<br>Peak: 455 ± 10 nm | Blue LEDs emit light in<br>the range of 400–550 nm<br>(peak wavelength 450–<br>470 nm). | Similar<br>See comment 6 |
| Light Intensity<br>(Irradiance) | Dual Fiber Optic Panel irradiance<br>level:<br>$50 \mu W/cm^2/nm(\pm 25\%)$<br>Single Fiber Optic Panel<br>irradiance level:<br>$60 \mu W/cm^2/nm(\pm 25\%)$ | Standard Panel-Light:<br>$30 \mu W/cm^2/nm$<br>Neonatal Panel-Light:<br>$55 \mu W/cm^2/nm$<br>Overhead Therapy: When<br>the illuminator is<br>positioned 30 cm (12<br>inches) above the<br>neonate: $30 \mu W/cm^2/nm$ | Similar<br>See comment 6 |
| Expected LED life | Nominal lifetime: >45000 hours | Minimum<br>lifetime<br>of<br>30,000 hours | Different<br>See comment 7 |
| Connection of the<br>Fiber optic bundle<br>to Light Source | Direct insert with a "click" sound | Insert and rotate a quarter<br>turn | Different<br>See comment 8 |
| Power Supply to<br>Light Box | AC Power: 100-240 VAC, 50/60<br>Hz, 1.0-0.5 A | AC Power:<br>100-240<br>VAC, 50/60 Hz, 1.0 A | Same |
| Timer | Yes | Yes | Same |
| LIGHT BOX/ LIGHT SOURCE | | | |
| Light box<br>dimensions | 7.5 cm × 7.7 cm × 15.9 cm (Light<br>Source) | 16.10 cm × 7.40 cm<br>(Illuminator) | Similar<br>See Comment 9 |
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
| Weight | < 1.5 kg | < 2.50 lb (1. 3 kg) | |
| FIBER OPTIC PANEL AND SWADDLES | | | |
| Fiber optic panel<br>Dimensions and<br>effective treatment<br>area according to<br>IEC<br>60601-2-50 | Dual Fiber Optic Panels :<br>size of effective illuminated area:<br>16.0 cm × 12.0 cm × 2<br>= 384 cm²<br><br>Single Fiber Optic Panel:<br>size of effective illuminated area:<br>16.0 cm × 12.0 cm<br>= 192 cm² | Model: EG-2000N (Flat<br>Neonatal Panel)<br>Overall<br>Pad<br>Size-<br>Neonatal: 5.00"× 7.00"<br>(12.70 cm × 17.78 cm)<br>Illuminated<br>Area-<br>Neonatal: 4.00" × 6.00"<br>(10.16 cm×15.24 cm) =<br>155 cm²<br><br>Model: EG-2000 (Wrap<br>Around Panel)<br>Overall<br>Pad<br>Size-<br>Standard: 4.00" × 15.00"<br>(10.16 cm × 38.10 cm)<br>Illuminated<br>Area-<br>Standard: 3.00" × 14.00"<br>(7.62 cm × 35.56 cm) =<br>271 cm² | Similar<br>See Comment 10 |
| Swaddle | 100% cotton | N/A | Different<br>See comment 11 |
| PAD TREATMENT AREA | | | |
| Panel material with<br>direct contact to<br>patient | TPU | TPU | Same |
| Use Intent | Multiple patients | Multiple patients | Same |
| NOISE | | | |
| Noise | < 60 dB(A) | < 60 dB (A) | Same |
| Device name | Subject device<br>Neonatal Phototherapy System<br>Model: KANGALITE<br>K200031 | Predicate device<br>Bili-Tx<br>K070180 | Comparison |
| STANDARDS COMPILED | | | |
| Biocompatibility<br>test | Biocompatibility tests performed<br>per ISO 10993 for patient<br>contacting materials (Disposable<br>cover and Swaddle). Both the<br>Disposable cover and Swaddle are<br>considered a skin-contacting<br>device with prolonged exposure.<br>Cytotoxicity per ISO 10993-5<br>Intracutaneous irritation study per<br>ISO 10993-10<br>Sensitization per ISO 10993-10 | There is no publicly-<br>available information. | Different<br>See comment 12 |
| Electrical safety<br>tests | IEC 60601-1 Generalrequirements<br>for basic safety and essential<br>performance | IEC 60601-1 General<br>Requirements for Safety<br>of Medical Electrical<br>Equipment | Same |
| Electromagnetic<br>Compatibility tests<br>(EMI/EMC) | Electromagnetic Compatibility<br>tests (EMI/EMC) performed per<br>IEC 60601-1-2 (4th edition)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.