FLUSION-9001, Fluoroscopic C-arm Mobile X-ray System
Applicant
Livermoretech, Inc.
Product Code
OWB · Radiology
Decision Date
Apr 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedical, abdominal, vascular, cardias, critical care and emergency room procedures.
Device Story
Mobile C-arm X-ray system for diagnostic, surgical, and interventional procedures in hospitals (OR, ER, etc.). System comprises high-voltage generator, X-ray tube, collimator, flat panel detector, and mobile cart. Operates by generating X-rays that pass through patient anatomy; detector captures signals, which are processed by PVCARM software for display, storage, and digital manipulation. Operated by clinicians (physicians, technicians). Output provides real-time visualization of anatomical structures (skeletal, respiratory, urinary systems) to guide medical operations. Benefits include improved diagnostic accuracy and procedural guidance during minimally invasive interventions.
Clinical Evidence
Bench testing only. Conformance to EN 60601-1, EN 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, and IEC 60601-2-54. Compliance with 21 CFR 1020 Subchapter J (performance standards for ionizing radiation). No clinical data presented.
Technological Characteristics
Mobile C-arm X-ray system. Rotating anode X-ray tube (300k HU capacity). Imaging via image intensifier or flat panel detector (TFT-based). Motorized collimator. Connectivity via DICOM for PACS integration. Software-based image processing. Sterilization not applicable (non-sterile device).
Indications for Use
Indicated for patients requiring fluoroscopic and radiographic imaging during diagnostic, surgical, and interventional procedures, including digital subtraction angiography, orthopedic, neurological, abdominal, vascular, cardiac, critical care, and emergency room applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
April 3, 2020
LiverMoreTech, Inc. % Mr. Dave Kim Regulatory Affairs MTech Group 7707 Fannin Street. Suite 200 HOUSTON TX 77054
Re: K200022
Trade/Device Name: FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA, IZI Dated: December 31, 2019 Received: January 6, 2020
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K200022
Device Name FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System
### Indications for Use (Describe)
The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedical, abdominal, vascular, cardias, critical care and emergency room procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K200022
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date 510k summary prepared: April 1, 2020
#### I. SUBMITTER
| Submitter's Name | Livermoretech |
|------------------------|---------------------------------------------------|
| Submitter's Address | 801 North Jupiter Rd, Suite 200<br>Plano TX 75074 |
| Submitter's Telephone | Tel: +1-214-257-0113 |
| Contact person | Jay Kim (jay.kim@aspenstate.com) / RA Manager |
| Official Correspondent | Dave Kim (davekim@mtech-inc.net) |
| Address | 7707 Fannin St. Ste 200, V111, Houston, TX 77054 |
| Telephone | +713-467-2607 |
#### II. DEVICE
| Trade/proprietary Name | FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System |
|------------------------|-----------------------------------------------------|
| Regulation Name | Image Intensified Fluoroscopic Mobile X-ray System |
| Regulation Number | 21 CFR 892.1650 |
| Product Code | OWB, OXO, JAA, IZI |
| Regulatory Class | Class II |
#### III. PREDICATE DEVICE
| 510K Number | K032761 |
|-------------------|----------------------------------------------------|
| Manufacturer | United Radiology Systems, Inc |
| Device Name | KMC-950 |
| Regulation Name | Image Intensified Fluoroscopic Mobile X-ray System |
| Regulation Number | 21 CFR 892.1650 |
| Product Code | OWB, OXO, JAA |
| Regulatory Class | Class II |
{4}------------------------------------------------
#### IV. REFRENCE DEVICE
| 510K Number | K180473 |
|-------------------|----------------------------------------------------|
| Manufacturer | ECOTRON Co., Ltd |
| Device Name | ANYVIEW DR SERIES FPD Fluoroscopic Mobile C-Arm |
| Regulation Name | Image Intensified Fluoroscopic Mobile X-ray System |
| Regulation Number | 21 CFR 892.1650 |
| Product Code | OWB, OXO, JAA |
| Regulatory Class | Class II |
#### DEVICE DESCRIPTION: V.
The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System consists of a high voltage (HV) inverter generator, a tube support unit, an X-ray beam limiting device, mobile cart, a detector, operating software, and a tube, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems such as the skull, spinal column, chest, abdomen, extremities, and other body parts. This device is not intended to be used for mammography applications.
FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is a solution to produce radiological images of patient during medical operations. This inverter control X-ray unit visualizes the anatomical structure on screen, which is obtained by X-ray fluoroscopy and a flat panel detector. This system can be applied in emergency room, operation room, cast room or etc. of a hospital.
#### Indications for Use: 21 CFR 807 92 (a) (5) VI.
The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures.
Comparison Table with the Predicate Device for technological characteristics:
| | Predicate device | Proposed device | SE | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Model Name | KMC-950 | FLUSION-9001<br>Fluoroscopic C-arm Mobile X-ray System | | |
| 510(k) Number | K032761 | K200022 | | |
| Indications for Use | The KMC-950 is intended to<br>provide fluoroscopic and<br>radiographic imaging of the<br>patient during diagnostic,<br>surgical and interventional<br>procedures. Clinical<br>applications may include but<br>are not limited to digital<br>subtraction angiography, | The FLUSION-9001<br>fluoroscopic C-arm Mobile X-ray System is intended to<br>provide fluoroscopic and<br>radiographic imaging of the<br>patient during diagnostic,<br>surgical and interventional<br>procedures. Clinical<br>applications may include but | Similar | |
| | | orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures.<br>The system may be used for other RF imaging application at physician's discretion. | are not limited to digital subtraction angiography, orthopedic, neurological, abdominal, vascular, cardiac, critical care and emergency room procedures. | |
| X-ray Tube | Anode Type | Rotating | Rotating | Same |
| | Heat Capacity | 300,000 HU | 300,000 HU | Same |
| | Anode Heat<br>Cooling | 70kHU/min | 60kHU/min | Similar |
| | Focal size | 0.3mm / 0.6mm | 0.3mm / 0.6mm | Same |
| Fluoroscopic<br>Mode | kV range | 40 to 125 kV See below discussion for differences | 40 to 120 kV See below discussion for differences | Similar |
| | mA range | 0.5 to 5 mA. See below discussion for differences | 0.2 to 8mA. See below discussion for differences | Similar |
| | Pulse Fluoro | Yes | Yes | Same |
| | ABS function | Yes | Yes | Same |
| Detector | Manufacturer | Imaging Intensifier:<br>Thales (TH9428HP2) Image Intensifier | Imaging Intensifier:<br>E5830SD-P4A (TOSHIBA)<br>Cleared under K160065 | |
| | Size | 9" | 9" | Same |
| | Magnification | 9" / 6" / 4.5" | 9" / 6" / 4.5" | Same |
| | DQE | 65 | 65 (@ 0 lp/mm) | Same |
| Camera | Type | 1/2" CCD | 1/2" CCD | Same |
| | Active pixels | 512 x 512 | 1024 x 1024 | Better |
| C-arm | Manufacturer | GEMSS medical | Livermoretech Inc. | |
| | SID | 950mm | 1000mm | Similar |
| | Range of C-ram<br>Rail Rotation | 115° | 135° | Wider |
| | Range of the<br>Liner FR-arm<br>Movement | 200 mm | 200 mm | Same |
| | Range of the<br>Linear T-arm<br>Movement | 500 mm | 500 mm | Same |
| | Range of<br>Swing-arm<br>Rotation | ± 12.5° | ± 15° | Wider |
| | Range of Stay-<br>arm Rotation | 360° | 190° | |
| Collimator | Collimator | Motor control / rotation | Motor control / rotation | Same |
{5}------------------------------------------------
# LIVERMORETECH
801 North JupiterRd, Suite 200, Plano, TX 70574 USA
Tel.: +214-257-0113
{6}------------------------------------------------
| | | Subject Device | Subject Device | Subject Device | Reference Device |
|----------------------|-----------|------------------|-----------------------------|----------------|------------------|
| | | | | | |
| Model | | VIVIX-D<br>1212G | Pixium® Surgical<br>3030S-A | 3030DXV | VIVIX-D<br>1212G |
| 510(k) No | | K180473 | K182086 | K132904 | K180473 |
| Detector type | | TFT | TFT | TFT | TFT |
| Active detector size | | 12 x 12 in | 12 x 12 in | 12 x 12 in | 12 x 12 in |
| Total pixel matrix | | 2048 x 2048 | 1956 x 1956 | 1536 x 1536 | 2048 x 2048 |
| Pixel pitch | | 145 μm | 154.0 μm | 194.0 μm | 145 μm |
| Frame rate | | 30 fps @ 2x2 | 30 fps | 30 fps | 30 fps @ 2x2 |
| MTF | 1.0 lp/mm | 60% | 55% | 55% | 60% |
| DQE | 0 lp/mm | 69% | 75% | 80% | 69% |
### FPD Image Detector Comparison
#### VII. Discussion of differences
The indications for use, operating principle, technical specifications such as X-ray tube head, image intensifier and FPD of FLUSION-9001 fluoroscopic C-arm Mobile X-ray System are similar to those of the predicate device (K032761) and reference device (K180473). FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System is designed as a set of components (X-ray tube and housing, detector, digital imaging system, collimator, generator etc.) similar to the predicate device KMC-950 (K032761). All digital flat panel detectors available for the subject device have been cleared by FDA as a part of a complete fluoroscopy system (K180473, K182086, K132904). Based on the recognized standard conformity evidences related to electro-, mechanical-, software-, and risk management, Livermoretech certifies that technological characteristics of FLUSION-9001 fluoroscopic C-arm Mobile X-ray System are substantially equivalent to KMC-950, the predicate device. FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System consists of many critical parts same as KMC-950, the predicate device (Collimator, X-ray imager, digital image processing and viewing program, C arm, etc.).
FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System utilizes PVCARM Viewer, a radiological digital image communications system. It provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions related to simple manipulation, enhancement or quantification of images.
- PVCARM receives images from a detector, processes and transfers the images and manages patient's information and the images for doctor. PVCARM enables images such as x-ray images to be stored electronically and viewed on screens.
- PVCARM offers full compliance with DICOM (Digital Imaging and Communications in Medicine) standards to allow the sharing of medical information with other PACS (Picture Archiving and Communication System Server).
{7}------------------------------------------------
Tel.: +214-257-0113
#### VIII. Non clinical testing
Testing was performed successfully according to the following standards:
- EN 60601-1:2006/A1:2013 ●
- EN 60601-1-2:2015 (IEC 60601-1-2: 2014) ●
- IEC 60601-1-3:2008+A1:2013 ●
- IEC 60601-2-43:2010 ●
- IEC 60601-2-54:2009 + A1::2015 ●
Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients:
- "The Content of Premarket Submissions for Software Contained in Medical Devices" ●
- "Guidance for the Submission of 510(k) for Solid State X-ray Imaging devices" ●
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices ● Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
- Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.
All applicable aspects of these guidance documents listed in this 510(k) summary have been addressed.
The device also conforms to the following:
- 21 CFR 1020 Subchapter J: Performance Standards for Ionizing Radiation Emitting Products ●
- 21 CFR 1020.30: Diagnostic x-ray system and their major components ●
- 21 CFR 1020.32: Fluoroscopic Equipment ●
#### IX. Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table, the FLUSION-9001 Fluoroscopic C-arm Mobile Xray System has little difference with its size and user interface as the information in the table. But the system is substantially equivalent to the predicate devices with its design, mechanical and electrical performance as described.
Performance evaluation (test) reports and device inspection report confirmed that the FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System is suitable for its intended use and user instruction of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.