Digital Flat Panel X-ray Detector, HAD1717MC is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
Device Story
Wired digital flat panel x-ray detector; uses TFT glass and CsI scintillator to convert incident X-rays into visible light; electron-hole pairs generated in photodiode; charge stored in photodiode capacity; gate pulse processing transfers charges in parallel; custom Read Out IC (ROIC) amplifies and digitizes signals via integrated A/D converter; digitized data transmitted to PC; software acquires, stores, and displays images. Used in clinical settings by radiology staff; enables general projection radiography; provides diagnostic images equivalent to conventional screen-film or CR systems.
Clinical Evidence
Clinical images obtained per FDA guidance for Solid-State X-ray Imaging Devices. Study confirmed subject detector provides images of equivalent diagnostic capability to the predicate device, with results demonstrating similar or slightly better performance.
Technological Characteristics
a-Silicon TFT flat panel detector with CsI scintillator; 17x17 inch active area; 140µm pixel pitch; 3.5 LP/mm resolution; TYP 70% DQE at 0.1 lp; TYP 95% MTF at 0.1 lp. Connectivity via PC workstation. Electrical/mechanical safety per ES 60601-1; EMC per IEC 60601-1-2; performance per IEC 62220-1.
Indications for Use
Indicated for general projection radiographic applications in patients requiring diagnostic imaging, excluding mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
January 31, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym followed by the full name of the agency, "U.S. Food & Drug Administration," on the right. The FDA part of the logo is in blue.
H&abyz % Mrs. April Lee Consultant Withus Group, Inc. 106 Superior IRVINE CA 92620
# Re: K200018
Trade/Device Name: HAD1717MC Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: December 27, 2019 Received: January 3, 2020
Dear Mrs. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200018
Device Name HAD1717MC
Indications for Use (Describe)
Digital Flat Panel X-ray Detector, HAD1717MC is intended for use in general projection radiographic applications wherever conventional screen-film systems may be used. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: overline;">☒</span> Prescription Use (Per 21 CFR 201.66; Select One) | ☐ Over-The-Counter Use (21 CFR 201.66; Select One) |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------|
|----------------------------------------------------------------------------------------------------|----------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for H&abyz. The logo consists of a colorful flower-like graphic on the left and the company name "H&abyz" in a dark gray sans-serif font on the right. The flower graphic has four petals in different colors: green, red, purple, and orange.
## H&abyz
1F, 2-dong, 41-16, Cheoinseong-ro, Namsa-myeon, Cheoin-gu, Yongin-si, Gyeonggi-do, Republic of Korea, 17118 Tel: +82-70-4658-9306
# 510(k) Summary - K200018
# Submitter
H&abyz Nam Kyu Hur 1F, 2-dong, 41-16, Cheoinseong-ro, Namsa-myeon, Cheoin-gu, Yongin-si, Gyeonggi-do, 17118 Republic of Korea Email: hnk@abyzr.com Phone: +82-70-4658-9306
## Device Information
- Trade Name: HAD1717MC
- Classification Name: Stationary X-Ray System ●
- Common Name: Digital Flat Panel X-ray Detector ●
- Product Code: MQB ●
- Panel: Radiology
- Regulation Number: 892.1680 ●
- Device Class: Class II ●
- Date prepared: 01/22/2020
# Primary Predicate Device
- K number: K152094 ●
- Trade Name: GR40CW .
- Classification Name: Stationary X-Ray System
- Common Name: Digital Flat Panel X-ray Detector ●
- Product Code: MQB
- Panel: Radiology
- Regulation Number: 892.1680 ●
# Device Description
This is a wired digital flat panel x-ray detector, a fast and efficient digital radiography system. The detector is used in combination with a TFT glass and scintillator (CSI) and supports automatic trigger signal detection technology that can be used without generator integration.
The incident X-rays are converted into visible light that produces electron hole pairs in a photometer biased by scintillator material. The charge carier is stored in the photodiode capacity. By pulse processing the gates of the TFT lines in the matrix, charges in all the columns are transferred in parallel with the signal output. All signals on the column are amplified by a custom Read Out IC for further processing.
The amplified signal is digitized using an A / D converter integrated into the ROIC. The digitized signal is transmitted to the PC. The device software is new software. When transferred to a PC, the software can acquire and store digitized images and view the images.
### Indications for use
Digital Flat Panel X-ray Detector, HAD1717MC is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for H&abyz. The logo consists of a colorful flower-like graphic on the left and the text "H&abyz" on the right. The flower graphic has five petals in different colors: red, green, purple, and orange. The text is in a simple, sans-serif font and is colored in a dark gray.
1F, 2-dong, 41-16, Cheoinseong-ro, Namsa-myeon, Cheoin-gu, Yongin-si, Gyeonggi-do, Republic of Korea, 17118 Tel: +82-70-4658-9306
# Summary of Technological Characteristics
| | Subject Device | Primary Predicate |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | H&abyz | Samsung Electronics Co., Ltd. |
| 510(k) Number | NA | K152094 |
| Model | HAD1717MC | GR40CW |
| Indications for Use | Digital Flat Panel X-ray Detector,<br>HAD1717MC is intended for use in<br>general projection radiographic<br>applications wherever conventional<br>screen-film systems or CR systems may<br>be used. This device is not intended for<br>mammographic applications. | The GR40CW Digital X-ray Imaging<br>System is intended for use in general<br>projection radiographic applications<br>wherever conventional screen-film<br>systems or CR systems may be used.<br>This device is not intended for<br>mammographic applications. |
| Appearance | Image: Subject Device Appearance | Image: Primary Predicate Appearance |
| Detector | | |
| Panel Type | a-Silicon, TFT | a-Silicon, TFT |
| Scintillator | CsI | CsI |
| Detector Area | 17 X 17 inches (427 x 427mm) | 17 X 17 inches (425mm X 425mm) |
| Pixel Pitch (µm) | 140 | 140 |
| High Contrast Limiting<br>Resolution(LP/mm) | 3.5 | 3.5 |
| DQE | TYP. 70% at 0.1 lp | TYP. 70% at 0.1 lp |
| MTF | TYP. 95% at 0.1 lp | TYP. 95% at 0.1 lp |
| Imaging Workstation | | |
| CPU | Intel Core 2 | Intel® Xeon® E5-1620 |
| Memory (RAM) | 4GB | 8GB |
| Storage (HDD) | 500GB | 1TB |
| Monitor | 1,920 x 1,080 pixels | 21 inch (1,920 X1,080) |
| CIB(Control Interface Box) | | |
| Max. Signal Input<br>Voltage | N/A | 400V DC/AC |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for H&abyz. The logo consists of a colorful flower-like graphic on the left and the text "H&abyz" on the right. The flower graphic has four petals in red, green, purple, and orange. The text is in a dark gray color.
## H&abyz
#### 1F, 2-dong, 41-16, Cheoinseong-ro, Namsa-myeon, Cheoin-gu, Yongin-si, Gyeonggi-do, Republic of Korea, 17118 Tel: +82-70-4658-9306
| Similarities | The subject and predicate devices are same and similar in indications, design,<br>technology, functions, and principle of operation. | | | | |
|--------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| Differences | Detector<br>Area | As the imaging area and number of pixels is calculated by effective<br>pixels, there may have some differences. However, this difference<br>does not raise different questions of safety and effectiveness than<br>the predicate. | | | |
| | CPU | Subject medical device CPU has faster processing speed than the<br>predicate device's CPU, and the faster processing speed does not<br>contribute any adverse impacts to the device's safety and<br>performance. | | | |
| | Memory<br>(RAM) | Subject medical device memory size is smaller than the predicate<br>device's memory size, and the memory size does not contribute<br>any adverse impacts to the device's safety and performance. | | | |
| | Storage<br>(HDD) | Subject medical device Storage has smaller capacity than the<br>predicate device's Storage, and the small capacity does not<br>contribute any adverse impacts to the device's safety and<br>performance. | | | |
| | Max. Signal<br>Input<br>Voltage | Subject medical device uses the TRIGGER line instead of the CIB.<br>The CIB is managed by the x-ray system. The CIB only carries the<br>signal, and the HAD1717MC's Trigger cable also carries the signal.<br>Therefore, this difference does not raise different questions of<br>safety and effectiveness than the predicate. | | | |
# Non-clinical Testing
1) Electrical Safety and Electromagnetic Compatibility
Electrical, mechanical, environmental safety and performance testing were conducted according to standard ES 60601-1(2006/A1:2013), and EMC testing was conducted according to IEC 60601-1-2(2015). All test results were satisfying with the standards.
# 2) Performance data
Non-clinical testing data was provided in conformance to the FDA "Guidance for the Submission of 510(k)'s for Solid-State X-ray Imaging Devices", which includes MTF and DQE measurements as tested by IEC 6220-1. All test results were satisfying with the IEC6220-1 standards.
# Clinical Data
In clinical data, clinical images were obtained in accordance with FDA guidance for the submission of 510(k)'s for Solid State X-ray Imaging Devices. The study confirmed that the subject x-ray detector provides images of equivalent diagnostic capability to the predicate device and its results demonstrate similar equivalence or slightly better.
### Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification H&abyz concludes that Digital Flat Panel X-ray Detector is similar equivalent in comparison with the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.