SC+ Haemodialysis Machine, SC+ Dialysate Cartridge, SC+ Blood Tube Set
Applicant
Quanta Dialysis Technologies, Ltd.
Product Code
KDI · Gastroenterology, Urology
Decision Date
Dec 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SC+ Machine and SC+ Dialysate Cartridge: The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, by a trained person who is competent in the use of the device. SC+ Blood Tube Set: The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System
Device Story
System performs hemodialysis using external purified water source; utilizes SC+ Machine, single-use Dialysate Cartridge, and sterile Blood Tubeset. Machine incorporates peristaltic blood pump, pneumatic interface, sensors (blood leak, air-in-blood), and heater/de-aeration module. Dialysate Cartridge uses membrane pumps for mixing/proportioning and fluid removal control. Used in acute/chronic care facilities by trained personnel. Healthcare provider inputs prescription parameters via touchscreen GUI; system monitors therapy via conductivity, pressure, and temperature sensors. Output includes controlled fluid removal and dialysate production. Benefits include effective renal replacement therapy for patients with renal failure.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including dialysate quality, essential performance, electrical/electromagnetic safety, alarm/alert functionality, software/touchscreen validation, biocompatibility, packaging/shelf-life, sterilization/pyrogenicity, and human factors testing. Hemolysis testing compared to reference device showed lower values.
Technological Characteristics
System includes SC+ Machine, Dialysate Cartridge, and Blood Tubeset. Materials: PVC and PP (tubing). Sensing: conductivity, pressure, temperature, blood leak, air-in-blood. Actuation: peristaltic blood pump, membrane diaphragm pumps. Connectivity: touchscreen GUI. Sterilization: sterile, non-pyrogenic (tubeset). Standards: IEC 60601-2-16:2012.
Indications for Use
Indicated for patients with acute and/or chronic renal failure requiring hemodialysis with or without ultrafiltration in acute or chronic care facilities. Requires physician prescription and administration by trained, competent personnel.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
NIPRO BLOOD TUBING SET WITH TRANSDUCER PROTECTOR AND PRIMING SET, MODEL A201-A219, V801-V806, 5M9634, 5M9693 (K072024)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 23, 2020
Quanta Dialysis Technologies Ltd. Chris Rule, Regulatory Manager Tything Road Alcester, Warwickshire B49 6EU UNITED KINGDOM
Re: K193670
Trade/Device Name: SC+ Hemodialysis System, comprised of the SC+ Hemodialysis Machine, SC+ Dialysate Cartridge, and SC+ Blood Tubeset Regulation Number: 21 CFR 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI, FJK Dated: November 20, 2020 Received: November 23, 2020
Dear Chris Rule:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K193670
Device Name
SC+ Hemodialysis System, comprised of the SC+ Hemodialysis Machine, SC+ Dialysate Cartridge, and SC+ Blood Tubeset
Indications for Use (Describe)
SC+ Machine and SC+ Dialysate Cartridge:
The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility. Treatments must be administered under physician's prescription, by a trained person who is competent in the use of the device.
SC+ Blood Tube Set:
The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Quanta. The logo consists of a light blue circle with a smaller circle attached to the bottom right. Below the circle is the word "Quanta" in gray, sans-serif font.
#### б 510(K) SUMMARY
The content of this traditional 510(k) summary is provided in conformance with 21 CFR Part 807.92.
#### 6.1 Submitter's Information
| Submitter's Name: | Quanta Dialysis Technologies Ltd |
|-----------------------------|------------------------------------------------------------------|
| Company:<br>Address: | Tything Road, Alcester, Warwickshire, B49 6EU,<br>United Kingdom |
| Contact person: | Mr. Chris Rule |
| Phone: | +44 (0)1789 400 043 |
| E-mail: | chris.rule@quantadt.com |
| Date of summary preparation | 22nd December 2019 |
#### 6.2 Device Name
SC+ Hemodialysis System, comprised of the SC+ Hemodialysis Machine, SC+ Dialysate Cartridge, and SC+ Blood Tubeset
# 6.2.1 SC+ Machine & Dialysate Cartridge
#### 6.2.1.1 General Device Information
| Trade Name | SC+ Machine & SC+ Dialysate Cartridge |
|---------------------|------------------------------------------------------------------------|
| Common Name | Hemodialysis Delivery System |
| Product Code | KDI |
| Device | Dialyzer, High Permeability With Or Without<br>Sealed Dialysate System |
| Classification Name | High permeability hemodialysis system |
| Regulation Number | 21 CFR §876.5860 |
| Device Class | Class II |
| Review Panel | Gastroenterology/Urology |
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Quanta. The logo consists of a light blue circle with a smaller light blue circle in the lower right quadrant. Below the circle is the word "Quanta" in gray.
#### 6.2.2.1 Predicate Device Information
| Trade Name | Tablo Console |
|-------------------------|------------------------------------------------------------------------------|
| Common Name | Hemodialysis Delivery System |
| 510(k) Clearance Number | K140866 |
| Product Code | KDI |
| Device | Dialyzer, High Permeability<br>With<br>Or<br>Without Sealed Dialysate System |
| Regulation Number | 21 CFR §876.5860 |
| Classification Name | High permeability hemodialysis system |
| Device Class | Class II |
| Review Panel | Gastroenterology/Urology |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the logo for Quanta. The logo consists of a blue circle with a smaller blue circle in the lower right quadrant. Below the circle is the word "Quanta" in gray font.
# 6.2.3 SC+ Blood Tubeset
#### General Device Information 6.2.3.1
| Trade Name | SC+ Blood Tubeset |
|---------------------|-------------------------------------------------------------------|
| Common Name | Blood Tubing Set |
| Product Code | FJK |
| Classification Name | set, tubing, blood, with and without anti-<br>regurgitation valve |
| Regulation Number | 21 CFR §876.5820 |
| Device Class | Class II |
| Review Panel | Gastroenterology/Urology |
#### 6.2.3.2 Predicate Device Information
| Trade Name | Tablo Cartridge |
|-------------------------|-------------------------------------------------------------------|
| Common Name | Blood Tubing Set |
| 510(k) Clearance Number | K140841 |
| Product Code | FJK |
| Device | Set, Tubing, Blood, With And Without Anti-<br>Regurgitation Valve |
| Regulation Number | 21 CFR §876.5820 |
| Classification Name | Hemodialysis system and accessories |
| Device Class | Class II |
| Review Panel | Gastroenterology/Urology |
#### 6.2.3.3 Reference Device Information
| Trade Name | NIPRO<br>BLOOD TUBING SET WITH<br>TRANSDUCER PROTECTOR AND PRIMING<br>SET, MODEL A201-A219, V801-V806,<br>5M9634, 5M9693 |
|-------------|----------------------------------------------------------------------------------------------------------------------------|
| Common Name | Blood Tubing Set |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Quanta. The logo consists of a large, light blue circle with a smaller, light blue circle in the lower right quadrant. Below the circle is the word "Quanta" in gray, sans-serif font.
| 510(k) Clearance Number | K072024 |
|-------------------------|-------------------------------------------------------------------|
| Product Code | FJK |
| Device | Set, Tubing, Blood, With And Without Anti-<br>Regurgitation Valve |
| Regulation Number | 21 CFR §876.5820 |
| Classification Name | Set, Tubing, Blood, with and without Anti-<br>regurgitation valve |
| Device Class | Class II |
| Review Panel | Gasteroenterology/Urology |
#### Device Description 6.3
## 6.3.1 SC+ Hemodialysis System
The SC+ Hemodialysis System is intended for acute and chronic dialysis therapy, with or without ultrafiltration, utilizing Dialysis Water (from standalone Reverse Osmosis (RO) units or a central RO ring main) to produce dialysate. The SC+ Hemodialysis system is for use in patients with arteriovenous (AV) fistula or central venous catheter access.
The system consists of the SC+ Machine, a single use disposable Dialysate Cartridge, and a sterile, single use, disposable Blood Tubeset.
The SC+ Machine consists of a water circuit (heater, de-aeration module, etc) blood leak detector, air in blood detector, a pneumatic interface for the dialysate cartridges, a peristaltic blood pump and various other sensors. The dialysate cartridge contains the following: conductivity monitors, interfaces for pressure and temperature measurement, membrane pumps to perform mixing/proportioning in order to produce dialysis fluid and the controlled removal of fluid from a patient with acute and/or chronic renal failure based on a physician's prescription. The dialysate fluid is manufactured using dialysis water purified externally by reverse osmosis that is heated to approximately 37oC and subsequently de-aerated within the machine before entering the cartridge.
The SC+ Blood Tubeset is a single use, sterile device consisting of an arterial line, a venous line, connections to a standard dialyzer, a saline line, three pressure transducer pods integrated into a single unit, a venous drip chamber, and a line for heparin infusion.
#### 6.4 Indications for Use
## 6.4.1 SC+ Hemodialysis Machine & Dialysate Cartridge
The SC+ Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the text "K193670 Page 5 of 8" in the upper left corner. In the lower right corner is a logo for Quanta. The logo is a blue circle with a smaller blue circle below it, and the word "Quanta" is written in gray below the circles.
facility. Treatments must be administered under physician's prescription, by a trained person who is considered competent in the use of the device.
### 6.4.2 SC+ Blood Tubeset
The SC+ Blood Tubeset is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Blood Tubeset is compatible only with the SC+ Hemodialysis System.
#### Technological Characteristics 6.5
The SC+ Hemodialysis machine and Dialysate Cartridge are predicated against the Outset Tablo Console (K140866).
The SC+ Blood Tubeset is predicated against the Outset Tablo Cartridge (K140841). For mechanical hemolysis testing, the reference devices used for comparison was the Nipro tubing set (K072024).
### 6.5.1 SC+ Machine vs Tablo console (K140866)
The technological characteristics of the SC+ Machine and SC+ Dialysate Cartridge are considered to be equivalent to the predicate device, the Tablo Console (K140866). A summary of the similarities and differences is provided in the table below.
| | Subject<br>Device | Predicate<br>Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------|
| | SC+ | Tablo Console |
| Indicated for use in patients with acute<br>and/or chronic renal failure, with or<br>without ultrafiltration in an acute or<br>chronic care facility | ✓ | ✓ |
| Use of purified water for dialysate<br>production | ✓ | ✓ |
| Use of third-party accessories,<br>including dialyzers, endotoxin retentive<br>filters, acid and bicarbonate | ✓ | ✓ |
| Application of consensus standards | ✓ | ✓ |
Technological Characteristics of the SC+ Machine
{8}------------------------------------------------
K193670 Page 6 of 8
Image /page/8/Picture/2 description: The image shows the logo for Quanta Services. The logo features a stylized letter 'Q' in a bright blue color. The 'Q' is formed by a larger circle with a smaller circle inside it, resembling a lowercase 'q'. Below the symbol is the word "Quanta" in a gray, sans-serif font.
| Device is software controlled and utilize<br>Graphic User Interface (GUI). | ✓ | ✓ |
|---------------------------------------------------------------------------------------------------------------|---|---|
| Design and Construction – Blood pump,<br>alarms, alerts, air detector mechanism,<br>and blood leak detectors. | ✓ | ✓ |
Differences between the subject device and the predicate device include:
- . The subject device utilizes purified water from an external source, for dialysate production, whereas the predicate device internally purifies the water.
- . The subject device incorporates a validated heat disinfection process to clean the water circuit as opposed to a chemical disinfection process utilized by the predicate device.
- The dialysate cartridge pump differs to that of the predicate device. ● The SC+ Machine uses positive displacement pumps contained within the disposable Dialysate Cartridge, which use a combination of the reciprocating action of a membrane diaphragm between two hard surfaces and suitable valves to mix Dialysate Water with concentrates to generate dialysis fluid and to pump dialysis fluid to and from the dialyzer.
- In order to allow for the use of the Dialysate Cartridge with a compliant membrane and to allow for a higher dialysate flow rate, the flow balance control mechanism in the SC+ differs to that of the predicate device. The control and accuracy of the flow balance process has been fully validated to meet the requirements defined in IEC 60601-2-:16:2012.
- The SC+ Machine calculates the net fluid removal rate from the . required mass for removal and the desired treatment time up to a maximum as stated in IEC 60601-2-16:2012. The accuracy of net fluid removal rate is independent of the dialysate flow rate. Although this is different from the predicate device, the SC+ System has undergone extensive Net Fluid Removal testing and conforms to the same standards as the predicate device.
In summary, the SC+ machine has the same intended use and is considered equivalent with regards to the technological characteristics as the predicate. The SC+ system has been fully validated and, therefore, where there are minor differences, the differences do not raise any new questions with regards to safety or effectiveness.
### 6.5.2 SC+ Blood Tubeset vs Tablo Cartridge (K140841)
The SC+ Blood Tubeset is predicated against the Outset Tablo Cartridge (K140841).
Technological Characteristics of the SC+ Blood Tubeset
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the text "K193670 Page 7 of 8" in the upper left corner. In the lower right corner is a logo for Quanta. The logo is a blue circle with a smaller circle below it, and the word "Quanta" is written in gray below the circles.
| | Subject<br>Device | Predicate<br>Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------|
| | SC+ | Tablo<br>Cartridge |
| Intended Use: To provide extracorporeal<br>access during hemodialysis | ✓ | ✓ |
| Materials: Primary fluid path materials are<br>Polyvinyl Chloride (PVC) and Polypropylene<br>(PP). | ✓ | ✓ |
| Design & Construction: Polyvinyl Chloride<br>(PVC) tubing of various lengths and<br>diameters, with color coded pinch clamps,<br>color coded injection ports, heparin line,<br>saline line, and pressure monitoring<br>components. | ✓ | ✓ |
| Sterility: Sterile, single use, non-pyrogenic. | ✓ | ✓ |
| Priming Volume: | ≤165ml | ≤200 mL |
| Needle configuration: Double needle | ✓ | ✓ |
The SC+ Blood Tubeset is considered to have equivalent technological characteristics to the predicate device. The minor differences between the devices include:
- . Subject Device is mounted directly onto the blood platen of the SC+ Machine. The predicate has an organizer to facilitate interface with the machine's front panel.
- The pressure measuring component of the SC+ Blood Tubeset is ● different from that of the Predicate as it has been designed specifically to fit the SC+ Machine. The SC+ Blood Tubeset has been validated as part of the system and there is no impact on safety or performance.
- The maximum prime volume in the SC+ Blood Tubeset is less than . that of the Predicate. This has no impact on performance and reduces the total amount of blood loss in the event the circuit needs to be discarded without returning blood back to the patient.
Comparison to the reference device demonstrated that the SC+ Blood tube set had lower hemolysis values.
None of these differences raise questions with regards to safety or effectiveness.
#### 6.6 Performance Data
Extensive design verification and validation activities have been performed on the SC+ System. Performance testing, developed in accordance with appropriate FDA quidance documents and relevant standards, has been
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the text 'K193670' and 'Page 8 of 8' on the left side. On the right side, there is a blue circular logo with a smaller circle inside. Below the logo, the word 'Quanta' is written in gray.
performed on the SC+ System to support the determination of substantial equivalence, testing has included:
- Performance testing for dialysate quality, essential performance testing, ● and all the key functions/design features/components.
- . Testing to confirm compliance with electrical and electromagnetic safety standards and performance of alarms and alerts.
- Performance testing for software and the touchscreen. .
- Biocompatibility testing. .
- Packaging and shelf life testing. ●
- Sterilization validation including pyrogenicity testing for the blood fluid path.
- . Human factors testing.
#### Conclusion 6.7
The Performance Testing demonstrates that the SC+ Hemodialysis System (including the SC+ Blood Tubeset) meets all performance specifications and complies with applicable standards and FDA Guidance Documents. The SC+ Hemodialysis System is considered to be substantially equivalent to the predicate devices and the minor differences in the technological characteristics of the SC+ System and the predicate devices do not raise any new or different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.