The retrospective study evaluated the standalone performance (sensitivity/specificity) of the AI algorithm and compared the device's notification time against the standard of care notification time derived from clinical radiological reports.
Retrospective study; Clinical imaging database; Sensitivity and specificity; Time to notification
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective standalone performance study; Retrospective analysis of clinical imaging data
Patients undergoing non-contrast CT scans of the brain in an acute setting; Sample Size: 261 scans; Number of Sites: 2
Standard of care (as established from radiological reports)
Sensitivity, specificity, and time to notification
Retrospective study: 261 non-contrast Computed Tomography (NCCT) scans from two clinical sites in the U.S.
>1 (trained neuro-radiologists)
Indications for Use
Viz ICH is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow. Viz ICH uses an artificial intelligence algorithm to analyze images for findings suggestive of a prespecified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes non-contrast CT images of the brain acquired in the acute setting, and sends notifications to a neurovascular or neurosurgical specialist that a suspected intracranial hemorrhage has been identified and recommends review of those images can be previewed through a mobile application. Images that are previewed through the mobile application may be compressed and are for informational purposes only and not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Viz ICH is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. Viz ICH is contraindicated for analyzing non-contrast CT scans that are acquired on scanners from manufacturers other than General Electric (GE) or its subsidiaries (i.e. GE Healthcare). This contraindication applies to NCCT scans that conform to all applicable Patient Inclusion Criteria, are of adequate technical image quality, and would otherwise be expected to be analyzed by the device for a suspected ICH.
Device Story
Software-only parallel workflow tool; analyzes non-contrast CT (NCCT) brain images for intracranial hemorrhage (ICH). Input: DICOM NCCT images from local hospital network via DICOM router. Processing: Backend server uses AI algorithm to detect ICH; sends push notification to neurovascular/neurosurgical specialists. Output: Notification to mobile app; allows preview of images (non-diagnostic). Usage: Hospital networks; clinicians. Benefit: Enables early specialist involvement; reduces time to notification compared to standard of care. Clinicians must view non-compressed images on diagnostic viewer before clinical decisions.
Clinical Evidence
Retrospective study of 261 NCCT scans (47% positive, 53% negative) from two U.S. sites. Primary endpoints: sensitivity and specificity compared to neuro-radiologist ground truth. Results: Sensitivity 93% (87%-97% CI), Specificity 90% (84%-94% CI), AUC 0.96. Average time to notification 0.49 minutes vs 38.2 minutes for standard of care. Stratified performance provided for age, gender, ICH subtype, slice thickness, and volume.
Technological Characteristics
Software-only; DICOM-compliant communication protocol; backend server and mobile application architecture. AI algorithm for image analysis. No image marking or segmentation externalization. Compatible with GE Healthcare NCCT scanners.
Indications for Use
Indicated for hospital networks and trained clinicians to identify and communicate suspected intracranial hemorrhage (ICH) in non-contrast CT (NCCT) brain images acquired in acute settings. Contraindicated for NCCT scans acquired on non-GE scanners.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
March 18, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Viz.ai, Inc. % Mr. Gregory Ramina Director of Regulatory Affairs 555 De Haro St., Suite 400 SAN FRANCISCO CA 94107
Re: K193658
Trade/Device Name: Viz ICH Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: March 2, 2020 Received: March 2, 2020
Dear Mr. Ramina:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K193658
Device Name
#### Viz ICH
Indications for Use (Describe)
Viz ICH is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of care workflow.
Viz ICH uses an artificial intelligence algorithm to analyze images for findings suggestive of a prespecified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes non-contrast CT images of the brain acquired in the acute setting, and sends notifications to a neurovascular or neurosurgical specialist that a suspected intracranial hemorrhage has been identified and recommends review of those images. Images can be previewed through a mobile application.
lmages that are previewed through the mobile application may be compressed and are for informational purposes only and not intended for diagnostic use beyond notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient and relevant discussion with a treating physician before making care-related decisions or requests. Viz ICH is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.
Viz ICH is contraindicated for analyzing non-contrast CT scans that are acquired on scanners from manufacturers other than General Electric (GE) or its subsidiaries (i.e. GE Healthcare). This contraindication applies to NCCT scans that conform to all applicable Patient Inclusion Criteria, are of adequate technical image quality, and would otherwise be expected to be analyzed by the device for a suspected ICH.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
# Viz.ai, Inc.'s Viz ICH
K193658
- Applicant Name: Viz.ai, Inc. 555 De Haro St Suite 400 San Francisco, CA 94107
- Contact Person: Gregory Ramina Director of Regulatory Affairs 555 De Haro St Suite 400 San Francisco, CA 94107 Tel. (415) 663-6130 Greg@viz.ai
- Date Prepared: March 6, 2020
# Device Name and Classification
| Name of Device: | Viz ICH |
|-----------------------|-----------------------------------------------------------------|
| Common or Usual Name: | Radiological Computer-Assisted Triage and Notification Software |
| Classification Panel: | Radiology |
| Regulation No: | 21 C.F.R. § 892.2080 |
| Regulatory Class: | Class II |
| Product Code: | QAS |
## Predicate Devices
| Manufacturer | Device Name | Application No. |
|------------------------|-------------|-----------------|
| Viz.ai, Inc. (Primary) | ContaCT | DEN170073 |
| Aidoc Medical, Ltd. | BriefCase | K180647 |
### Device Description
Viz ICH is a software-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to an appropriate specialist, such as a neurovascular specialist or neurosurgeon, independent of the standard of care workflow. The system automatically receives and analyzes non-contrast CT (NCCT) studies of patients for image features that indicate the presence of an intracranial hemorrhage (ICH) using an artificial intelligence algorithm, and upon detection of a suspected ICH, sends a notification so as to alert a specialist clinician of the case.
{4}------------------------------------------------
Viz ICH consists of backend and mobile application component software. The Backend software includes a DICOM router and backend server. The DICOM router transmits NCCT images of the head acquired on a local healthcare network to the Backend Server. The Backend Server receives, stores, processes and serves received NCCT scans. The Backend Server also includes an artificial intelligence algorithm that analyzes the received NCCT images for image characteristics that indicate an intracranial haemorrhage (ICH) and, upon detection, sends a notification of the suspected finding to pre-determined specialists.
The Viz ICH Mobile Application software receives notifications generated by the Backend of suspected image findings and allows the notification recipient to view the analyzed NCCT images through a non-diagnostic viewer, as well as patient information that was embedded in the image metadata. Image viewing through the mobile application is for informational purposes only and is not intended for diagnostic use.
## Intended Use / Indications for Use
Viz ICH is a notification-only, parallel workflow tool for use by hospital networks and trained clinicians to identify and communicate images of specific patients to a specialist, independent of standard of care workflow.
Viz ICH uses an artificial intelligence alqorithm to analyze images for findings suggestive of a prespecified clinical condition and to notify an appropriate medical specialist of these findings in parallel to standard of care image interpretation. Identification of suspected findings is not for diagnostic use beyond notification. Specifically, the device analyzes non-contrast CT images of the brain acquired in the acute setting, and sends notifications to a neurovascular or neurosurgical specialist that a suspected intracranial hemorrhage has been identified and recommends review of those images can be previewed through a mobile application.
lmages that are previewed through the mobile application may be compressed and are for informational purposes only and not intended for diagnostic use beyond notification. Notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Viz ICH is limited to analysis of imaging data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis.
Viz ICH is contraindicated for analyzing non-contrast CT scans that are acquired on scanners from manufacturers other than General Electric (GE) or its subsidiaries (i.e. GE Healthcare). This contraindication applies to NCCT scans that conform to all applicable Patient Inclusion Criteria, are of adequate technical image quality, and would otherwise be expected to be analyzed by the device for a suspected ICH.
# Summary of Technological Characteristics
Viz ICH and its primary predicate, ContaCT (DEN170073), use the same process of automatic data identification and transfer to send images from the local hospital network to a
{5}------------------------------------------------
remote server for image processing and analysis, i.e., via a DICOM router that automatically identifies relevant images on a local IT Network and transfers them to a Backend Server using a DICOM compliant communication protocol.
Both Viz ICH and the ContaCT device have Backend Server software that can receive, store, process, and serve images that are forwarded from a DICOM router. The Backend Server software for each device has the same additional software functionality that interacts with the image management architecture, including a notifier module.
Where Viz ICH and the primary predicate differ, namely the specific condition and image modality analyzed by each device's algorithm. Viz ICH is similar to the secondary predicate. BriefCase (K180647), in that both software algorithms are designed to detect the same condition, intracranial haemorrhage (ICH), in non-contrast Computed Tomography images (NCCTs) of the head. Like the secondary predicate, BriefCase, the Viz ICH algorithm does not externalize any internal segmentation, analysis, or intermediate outputs used in determining if an ICH is present in the NCCT, nor does either algorithm mark the analyzed NCCT image.
Both Viz ICH and the ContaCT predicate support a mobile application that allows a user to receive push notifications, preview related images, and view patient details associated with a series. The Viz ICH mobile application is subject to the same non-diagnostic viewing limitations as the ContaCT predicate and has the same non-diagnostic warning on the image viewing screen as the predicate. Furthermore, the mobile application for Viz ICH and the ContaCT predicate can perform the same image viewing functions (window, pan, level, zoom, scroll through a cine).
| | Viz ICH | ContaCT (DEN170073) | BriefCase (K180647) |
|-------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Technological Characteristics | | | |
| DICOM<br>Compatible | Yes | Yes | Yes |
| Transfer, store,<br>and process<br>DICOM images | Yes | Yes | Yes |
| Data Acquisition | Acquires medical image data<br>from DICOM compliant<br>imaging devices and<br>modalities. | Acquires medical image data<br>from DICOM compliant<br>imaging devices and<br>modalities. | Acquires medical image data<br>from DICOM compliant<br>imaging devices and<br>modalities. |
| Image Analysis | | | |
| Supported<br>Imaging Modality | Computed Tomography, non-<br>contrast (NCCT) | Computed Tomography,<br>contrast-enhanced (CTA) | Computed Tomography, non-<br>contrast (NCCT) |
| Alteration of<br>Original Image | No | No | No |
| Results of Image<br>Analysis | Internal, no image marking | Internal, no image marking | Internal, no image marking |
| Image Viewing Functionality | | | |
| Preview Images | Initial assessment; non-<br>diagnostic purposes | Initial assessment; non-<br>diagnostic purposes | Initial assessment; non-<br>diagnostic purposes |
| View DICOM<br>Data | DICOM Information about the<br>patient, study and current<br>image. | DICOM Information about the<br>patient, study and current<br>image. | DICOM Information about the<br>patient, study and current<br>image. |
| Image viewing | Yes | Yes | Yes |
{6}------------------------------------------------
| and manipulation<br>(window, pan,<br>level, zoom) | | |
|---------------------------------------------------|--|--|
|---------------------------------------------------|--|--|
## Performance Data
Viz.ai conducted a retrospective study to assess the standalone performance of the image analysis algorithm and notification functionality of Viz ICH. The study evaluated the Viz ICH image analysis algorithm in terms of sensitivity and specificity with respect to a ground truth, as established by trained neuro-radiologists, in the detection of intracranial hemorrhage (ICH) in the brain. In addition, the study reported and compared the time to notification for the Viz ICH device with respect to the time to notification for the standard of care as established from available radiological reports.
261 non-contrast Computed Tomography (NCCT) scans (studies) were obtained from two clinical sites in the U.S. There were approximately equal numbers of positive and negative cases (47% of images with ICH and 53% without ICH, respectively) included in the analysis. Sensitivity and specificity were calculated in the image database, comparing the Viz ICH's output to ground truth as established by trained neuro-radiologists. Sensitivity and specificity were 93% (87%-97%) and 90% (84%-94%), respectively. Because the lower bound of each confidence interval exceeded 80%, the study met the pre-specified performance goals of 80% for sensitivity and specificity.
In addition, the area under the receiver operating characteristic curve (AUC) was 0.96 demonstrating the clinical utility and potential benefits of the classifier based on the imaging study results.
{7}------------------------------------------------
Image /page/7/Figure/0 description: The image is a Receiver Operating Characteristic (ROC) curve for Viz ICH. The x-axis represents the false positive rate, and the y-axis represents the true positive rate. The ROC curve has an area of 0.96, and a threshold of 0.1 mL is marked with a dotted vertical line.
In the study, the average time to alerting a specialist was 0.49±0.15 minutes, which is substantially lower than the average time to notification seen in the Standard of Care of 38.2±84.3 minutes. This data generally demonstrates that specialists have the opportunity to become involved in the clinical workflow early with notifications from the Viz ICH software.
As part of a secondary analysis, the company stratified the device performance by various confounding variables:
| Performance Stratified by Clinical Site | | | |
|-----------------------------------------|----------------------|----------------------|--|
| Clinical Site | Sensitivity [95% CI] | Specificity [95% CI] | |
| Erlanger | 0.93 [0.85, 0.98] | 0.92 [0.84, 0.97] | |
| Mt. Sinai | 0.92 [0.8, 0.98] | 0.86 [0.75, 0.94] | |
| Performance Stratified by Age | | |
|-------------------------------|----------------------|----------------------|
| Age Range (years) | Sensitivity [95% CI] | Specificity [95% CI] |
| <50 | 0.89 [0.52, 1.0] | 0.95 [0.76, 1.0] |
| 50 - 70 | 0.92 [0.82, 0.97] | 0.9 [0.8, 0.96] |
| >70 | 0.94 [0.84, 0.99] | 0.88 [0.76, 0.95] |
{8}------------------------------------------------
| Performance Stratified by Gender | | | | |
|----------------------------------|----------------------|----------------------|--|--|
| Gender | Sensitivity [95% Cl] | Specificity [95% CI] | | |
| Male | 0.9 [0.8, 0.96] | 0.9 [0.8, 0.96] | | |
| Female | 0.95 [0.86, 0.99] | 0.89 [0.8, 0.95] | | |
| Performance Stratified by ICH Subtype | |
|---------------------------------------|----------------------|
| ICH Subtype | Sensitivity [95% CI] |
| Intraparenchymal Hemorrhage (IPH) | 0.98 [0.91, 1.0] |
| Intraventricular Hemorrhage (IVH) | 1.00 [0.74, 1.0] |
| Subarachnoid Hemorrhage (SAH) | 0.60 [0.26, 0.88] |
| Subdural Hemorrhage (SDH) | 0.85 [0.62, 0.97] |
| Extradural Hemorrhage (EDH) | N/A |
| SDH or EDH | 0.85 [0.62, 0.97] |
| Performance Stratified by Slice Thickness | | | | |
|-------------------------------------------|-------------------|-------------------|--|--|
| Slice Thickness | Sensitivity | Specificity | | |
| 2.5mm <= Slice Thickness < 3.5mm | 0.93 [0.85, 0.98] | 0.92 [0.83, 0.97] | | |
| 3.5mm <= Slice Thickness <= 5.0mm | 0.92 [0.81, 0.98] | 0.88 [0.77, 0.95] | | |
| Performance Stratified by ICH Volume | | |
|--------------------------------------|--------------------------------------|--------------------------------------------|
| Minimal Volume Threshold (mL) | Sensitivity above Threshold [95% CI] | Sensitivity below/equal Threshold [95% CI] |
| 1 | 0.95 [0.89, 0.98] | 0.73 [0.39, 0.94] |
| 5 | 0.97 [0.91, 0.99] | 0.81 [0.64, 0.93] |
| 10 | 0.99 [0.93, 1.0] | 0.84 [0.7, 0.93] |
# Conclusions
Viz ICH is as safe and effective as the primary predicate, ContaCT (DEN170073), in that it has the same intended use and similar indications, technological characteristics and principles of operation. Where minor differences in indications exist between both devices, Viz ICH is similar to the secondary predicate device, BriefCase (K180647), which uses an artificial intelligence algorithm to detect and notify specified clinicians of an intracranial haemorrhage (ICH) in NCCT scans of the head. The minor differences in indications between ContaCT and Viz ICH do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness. In addition, the minor technological differences between the Viz ICH and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that Viz ICH can identify an ICH per the same primary
{9}------------------------------------------------
performance goals, in terms of sensitivity and specificity, as the BriefCase predicate and is as safe and effective as the predicate devices. Thus, Viz ICH is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.