K193655 · Self Doctor Care, LLC · NUH · Sep 4, 2020 · Neurology
Device Facts
Record ID
K193655
Device Name
MSLS6QF TENS/PMS Device
Applicant
Self Doctor Care, LLC
Product Code
NUH · Neurology
Decision Date
Sep 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TENS (Transcutaneous Electric Nerve Stimulation): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. Mode 1,3,4,5,6 PMS (Powered Muscle Stimulation): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. Mode 2,6
Device Story
Compact, portable, dual-channel TENS/PMS device; delivers programmed electrical pulses via self-adhesive electrodes to skin. User-operated via buttons/toggle switch; LCD displays mode, intensity, and time. Powered by internal rechargeable Li-battery (DC 3.7V). Includes 6 operation modes for pain relief and muscle stimulation. Used by patients in home/OTC settings. Output intensity and duration are manually controlled. Benefits include temporary pain relief and muscle conditioning.
Clinical Evidence
Bench testing only. Device validated for electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and nerve/muscle stimulator performance (IEC 60601-2-10). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization). Battery safety verified per IEC 62133.
Technological Characteristics
Dual-channel electrical stimulator; ABS housing; self-adhesive electrodes (gel, carbon film, cloth backing). Output: pulsed, rectangular waveform. Power: DC 3.7V Li-battery. Connectivity: USB for charging. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, ISO 10993-5/10, IEC 62133. Sterilization: Not specified.
Indications for Use
Indicated for temporary relief of pain in shoulder, waist, back, upper, and lower extremities due to exercise or household strain (TENS); and stimulation of healthy muscles to improve/facilitate performance (PMS). For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 4, 2020
Self Doctor Care, LLC Wei Wei Manager 8811 Teel Pkwy Ste 100, Unit 6141 Frisco, Texas 75036
Re: K193655
Trade/Device Name: MSLS6QF TENS/PMS Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: April 24, 2020 Received: June 9, 2020
Dear Wei Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Amber Ballard, PhD Acting Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K193655
Device Name MSLS6QF TENS/PMS Device
Indications for Use (Describe)
TENS (Transcutaneous Electric Nerve Stimulation):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. Mode 1,3,4,5,6
PMS (Powered Muscle Stimulation):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. Mode 2,6
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter's Information
Submitter: Self Doctor Care, LLC Address: 8811 Teel Pkwy Ste 100.Unit 6141, Frisco. TX 75036 Contact Person: Wei Wei Tel: 516-289-8425 Email: massagelossage@gmail.com Date of Preparation: 12/26/2019
#### 2. Correspondent's Information
Correspondent: Self Doctor Care, LLC Address: 8811 Teel Pkwy Ste 100, Unit 6141, Frisco, TX 75036 Contact Person: Wei Wei Tel: 516-289-8425 Email: massagelossage@gmail.com
## 3. Subject Device
Device Name: MSLS6QF TENS/PMS Device Common Name: Transcutaneous electrical nerve stimulator (TENS) and Powered MuscleStimulator(PMS) Model: MSLS6QF Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter (OTC) Regulation Description: Transcutaneous electrical nerve stimulator for pain relief Regulation Medical Specialty: Neurology Review Panel: Neurology Product Code: NUH, NGX Regulation Number: 21 CFR 882.5890 Device Class: II Use: Over-The-Counter
## 4. Predicate device
Predicate Device: TENS&PMS, IQ Technologies
{4}------------------------------------------------
510(k) Number: K131290 Use: Over-The-Counter Submitter: IQ Technologies Inc.
## 5. Description of Subject Device
The subject device MSLS60F is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS), intended for the over-the-counter use to temporarily relieve pain and stimulate muscle in different body areas.
This double-channel subject device, which is compact, portable, effectively transfer programmed electrical pulses directly through the self-adhesive electrodes to the suggested area of the body where the electrodes are placed. It delivers a gentle electrical pulse through the connecting wires and electrode pads to the user's skin for pain relief and muscle stimulation.
The subject device has 6 operation modes which can give certain electrical pulse through the 4 pcs of electrodes placed on the skin to help users to enjoy body stimulation. The subject device has the operating elements of ON/OFF Switch, Display screen, Mode Selection key and Intensity Modification keys. The LCD display screen can show selected mode, output intensity of the pulse, and time remaining of an application mode. The subject device could be easily operated through its toggle switch or buttons to manually realize its functions according to the need of users.
The subject device is equipped with accessories of the electrodes, lead wires, AC adapter and USB cable. The lead wire is used to connect the electrodes to the main unit; the USB cable is used to connect the AC adapter to the main unit when charging the built-in Li battery; the pads holders are used to storage the electrodes after therapy treatment for the convenience.
The self-adhesive electrodes are important accessories of the subject device, and are contact with the skin surface. It is consists of gel, carbon film, cloth backing ,and electrode connector. The electrode is complying with the biocompatibility standards ISO 10993-5 (Cytotoxicity) and ISO 10993-10 (Irritation and Sensitization).
## 6. Intended Use of Subject Device
TENS (Transcutaneous Electric Nerve Stimulation):
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist,back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. Mode 1,3,4,5,6
PMS(Powered Muscle Stimulation):
It is intended to be used to stimulate healthy muscles in order to improve and facilitate
{5}------------------------------------------------
muscle performance. Mode 2,6
#### 7. Summary of Substantial Equivalence
The following table summarizes the comparison between the subject device and predicate devices, indicating the technical characteristics, specifications, and intended use of the subject device are substantially equivalent to those of the predicate devices.
| | Subject Device | Predicate Device | Judgment | |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|--------------------|
| 510(k) Number | K193655 | K131290 | / | |
| Company Name | Self Doctor Care, LLC | IQ Technologies Inc. | / | |
| Device Name | MSLS6QF | IQ Technologies | / | |
| Operational Principle | Generate small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves and/or muscles are activated | Generate small pulses of electrical current and delivers the pulses to the user's skin through adhesive electrode pads such that the underlying nerves and/or muscles are activated | SE | |
| Intended Use | TENS (Transcutaneous Electric Nerve Stimulation):<br>To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.<br>Mode 1,3,4,5,6<br><br>PMS (Powered Muscle Stimulation):<br>It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.<br><br>It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance. | SE | |
| | | Mode 2,6 | | |
| Prescription/over-the-counter use | | OTC | OTC | SE |
| Power Source | | DC 3.7V Lithium Battery | DC 3.7V Lithium Battery | SE |
| Method of Line Current Isolation | | Type BF Applied Part | Type BF Applied Part | SE |
| Patient Leakage Current | | | | |
| - Normal Condition (μΑ) | | | | |
| - Single Fault Condition (μΑ) | | < 1 μΑ | | |
| | | 6 μΑ | Not publicly available | Note 1 |
| Number of output modes | | 6 | 6 | SE |
| Regulated Current or Regulated Voltage | | Voltage Control | Voltage Control | SE |
| Number of Output Channels | | 2 | 2 | SE |
| Automatic Overload Trip | | No | No | SE |
| Automatic No-Load Trip | | No | No | SE |
| Automatic Shut Off | | Yes | Yes | SE |
| User Override Control | | Yes | Yes | SE |
| | On/Off Status? | Yes | Yes | SE |
| | Low Battery? | Yes | Yes | SE |
| | Voltage/Current | | | |
| Indicator Display | Level? | No | No | SE |
| | | 10 ~ 60 minutes, | 10 ~ 60 minutes, | |
| Timer Range (minutes) | | 10 min/step | 10 min/step | SE |
| Weight(g) | | 40g | 37g | |
| Size(mm) | | 80×41×16 | 80×42×13 | Similar, |
| Housing Materials and Construction | | ABS | ABS and Aluminum alloy | Note 2 |
| Burst Mode | | | | |
| a) Pulses per burst | | | | |
| b) Bursts per second | | | | |
| c) Burst duration | | | | |
| (seconds) | | | | |
| d) Duty Cycle | | | | |
| [Line(b) x Line (c)] | | N/A (no pulse train or burst) | N/A | |
| ON Time (seconds) | | N/A | N/A | |
| OFF Time (seconds) | | N/A | N/A | |
| | | Self-adhesive Electrodes, | Self-adhesive Electrodes, | |
| | | Lead wires, | Lead wires, | |
| Accessories intended for use with device | | Battery charger, USB cable | Battery charger, USB | SE |
| | | cable | | |
| | Waveform | Pulsed | Pulsed | SE |
| | Shape | Rectangular | Rectangular | SE |
| | Maximum output voltage<br>(Volts +/- 20%) at 500Ω | Mode 1: 45.9<br>Mode 2: 61.6<br>Mode 3: 46.7<br>Mode 4: 39.0<br>Mode 5: 36.1<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 42<br>Mode 2: 63.2<br>Mode 3: 64<br>Mode 4: 34.4<br>Mode 5: 32<br>Mode 6: This mode<br>cycles the above five<br>modes | |
| | Maximum output voltage<br>(Volts +/- 20%) at 2kΩ | Mode 1: 74.2<br>Mode 2: 87.4<br>Mode 3: 60.5<br>Mode 4: 65.1<br>Mode 5: 60.5<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 80.8<br>Mode 2: 94.4<br>Mode 3: 87.2<br>Mode 4: 68<br>Mode 5: 64<br>Mode 6: This mode<br>cycles the above five<br>modes | Similar,<br>Note 3 |
| | Maximum output voltage<br>(Volts +/- 20%) at 10kΩ | Mode 1: 126.0<br>Mode 2: 128.0<br>Mode 3: 88.0<br>Mode 4: 124.7<br>Mode 5: 120.7<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 129<br>Mode 2: 129<br>Mode 3: 96.8<br>Mode 4: 128<br>Mode 5: 119<br>Mode 6: This mode<br>cycles the above five<br>modes | |
| | Maximum output current<br>(mA +/- 20%) at 500Ω | Mode 1: 91.8<br>Mode 2: 123.2<br>Mode 3: 93.4<br>Mode 4: 78.0<br>Mode 5: 72.2<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 84<br>Mode 2: 126.4<br>Mode 3: 128<br>Mode 4: 68.8<br>Mode 5: 64<br>Mode 6: This mode<br>cycles the above five<br>modes | |
| | | | | |
| Maximum output current<br>(mA +/- 20%) at 2KΩ | Mode 1: 37.1<br>Mode 2: 43.7<br>Mode 3: 30.3<br>Mode 4: 32.6<br>Mode 5: 30.3<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 40.4<br>Mode 2: 47.2<br>Mode 3: 43.6<br>Mode 4: 34<br>Mode 5: 32<br>Mode 6: This mode<br>cycles the above five<br>modes | | |
| Maximum output current<br>(mA +/- 20%) at 10KΩ | Mode 1: 12.6<br>Mode 2: 12.8<br>Mode 3: 8.8<br>Mode 4: 12.5<br>Mode 5: 12.1<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 12.9<br>Mode 2: 12.9<br>Mode 3: 9.7<br>Mode 4: 12.8<br>Mode 5: 11.9<br>Mode 6: This mode<br>cycles the above five<br>modes | | |
| Frequency (Hz) | Mode 1: 68.7<br>Mode 2: 12.2~53.4<br>Mode 3: 1.2<br>Mode 4: 96.1<br>Mode 5: 96.1<br>Mode 6: This mode<br>cycles the above five<br>modes | Mode 1: 69.4<br>Mode 2: 12.3~54.3<br>Mode 3: 1.2<br>Mode 4: 100<br>Mode 5: 100<br>Mode 6: This mode<br>cycles the above five<br>modes | | |
| | Mode 1: 14.6<br>Mode 2: 18.7~82.0<br>Mode 3: 840.0<br>Mode 4: 10.4<br>Mode 5: 10.4<br>Mode 6: This mode<br>cycles the above five<br>modes | | | |
| Pulse period (mSec) | | 10~840 | SE | |
| Pulse Width (µSec) | | 100 | SE | |
| Maximum Phase charge<br>(µC) at 500Ω | 17.42 | 16.8 | Similar, | |
| | Mode 1: 3.67 | Mode 1: 3.36 | Note 4 | |
| | Mode 2: 4.93 | Mode 2: 5.06 | | |
| | Mode 3: 3.74 | Mode 3: 5.12 | | |
| | Mode 4: 3.12 | Mode 4: 2.75 | | |
| | Mode 5: 2.89 | Mode 5: 2.56 | | |
| | Mode 6: This mode<br>cycles the above five<br>modes | Mode 6: This mode<br>cycles the above five<br>modes | | |
| Maximum current density<br>(mA/cm²) at 500Ω | | | | |
| | | | | |
| | Mode 1: 0.03556 | | | |
| | Mode 2: 0.0085~0.03726 | Not publicly available | | |
| | Mode 3: 0.00063 |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.