MRH ALFA Radiographic System, MRH II Radiographic System, MRH IIE Radiographic System
K193637 · Compania Mexicana DE Radiologia Cgr, S.A. DE C.V. · KPR · Jan 28, 2020 · Radiology
Device Facts
Record ID
K193637
Device Name
MRH ALFA Radiographic System, MRH II Radiographic System, MRH IIE Radiographic System
Applicant
Compania Mexicana DE Radiologia Cgr, S.A. DE C.V.
Product Code
KPR · Radiology
Decision Date
Jan 28, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MRH ALFA system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH ALFA radiographic system. The acquisition system comes as a separate system. The MRH ALFA system is not intended for use in mammography, fluoroscopy or angiography. The MRH ALFA system is intended for adults and pediatric patients. The pediatric population is as follow: Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.) Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.) The MRH II system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH II radiographic system. The acquisition system comes as a separate system. The MRH II system is not intended for use in mammography, fluoroscopy or angiography. The MRH II system is intended for adults and pediatric patients. The pediatric population is as follow: Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.) Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.) The MRH IIE system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH IIE radiographic system. The acquisition system comes as a separate system. The MRH IIE system is not intended for use in mammography, fluoroscopy or angiography. The MRH IIE system is intended for adults and pediatric patients. The pediatric population is as follow: Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.) Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
Device Story
General purpose radiographic systems; produce X-ray images of human body; used with conventional film or CR systems; interfaceable with digital detectors (detectors/acquisition systems not included). Components: X-ray tube (dual focal spot), generator, collimator, table, tube stand, wall bucky. Operated by healthcare professionals in clinical settings; patient positioned sitting, standing, or prone/supine. Device produces low radiation dose; output used by clinicians for diagnostic imaging. Modifications from predicate (K182134) include: removal of control panel touch screen; changes to table dimensions, load capacity, and motorized movement; changes to tube stand height and displacement; change to wall bucky casing material.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), radiation safety (IEC 60601-1-3), usability (IEC 60601-1-6), and biocompatibility (ISO 10993-1, -5, -10).
Indicated for producing X-ray images of human body anatomical regions (head, thorax, column, abdomen, extremities, internal organs) in adults and pediatric patients (3-21 yrs, 12-80 kg). Not for mammography, fluoroscopy, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Compania Mexicana De Radiologia CGR, S.A. DE C.V. % Mr. Stuart R. Goldman Senior Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Bldg. 1, Suite 300 AUSTIN TX 78746
January 28, 2020
## Re: K193637
Trade/Device Name: MRH ALFA, MRH II & MRH IIE Radiographic Systems Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR Dated: December 19, 2019 Received: December 27, 2019
## Dear Mr. Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193637
Device Name
MRH ALFA, MRH II & MRH IIE Radiographic Systems
#### Indications for Use (Describe)
1. The MRH ALFA system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH ALFA radiographic system. The acquisition system comes as a separate system.
The MRH ALFA system is not intended for use in mammography, fluoroscopy or angiography. The MRH ALFA system is intended for adults and pediatric patients. The pediation is as follow:
- · Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- · Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
2. The MRH II system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH II radiographic system. The acquisition system comes as a separate system.
The MRH II system is not intended for use in mammography, fluoroscopy or angiography. The MRH II system is intended for adults and pediatric patients. The pediatric population is as follow:
- · Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- · Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
3. The MRH IIE system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH IIE radiographic system. The acquisition system comes as a separate system.
The MRH IIE system is not intended for use in mammography, fluoroscopy or angiography. The MRH IIE system is intended for adults and pediatric patients. The pediatric population is as follow:
- · Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- · Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Special 510(k) Summary MRH ALFA, MRH II & MRH IIE Radiographic Systems
## 1. Submission Sponsor
Compañía Mexicana de Radiología CGR S.A. de C.V. Fraccionamiento Industrial la Noria s/n El Marqués, Querétaro C.P. 76240 México Phone: (55) 54821300 Contact: Tonatiuh Monroy Soberón Title: Design Engineer, Regulatory Specialist
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Stuart R. Goldman Title: Sr. Consultant RA/QA
#### 3. Date Prepared
December 19, 2019
#### 4. Purpose of the Submission
In this Special 510(k) Compañía Mexicana de Radiología (CMR) is making several minor technological modifications to certain components of their previously cleared ARiX RAD Radiographic System (K182134), the predicate device, that will result in the creation of their new MRH ALFA, MRH IIE Radiographic Systems.
#### 5. Device Identification
| Trade Name: | MRH ALFA, MRH II & MRH IIE Radiographic Systems |
|-----------------------|-------------------------------------------------|
| Common Name: | Radiographic System |
| Classification Name: | Stationary X-ray System |
| Regulation Number: | 892.1680 |
| Product Code: | KPR, System, X-Ray, Stationary |
| Class: | Class 2 |
| Classification Panel: | Radiology |
{5}------------------------------------------------
## 6. Legally Marketed Predicate Device
| Device name: | ARiX RAD Radiographic System |
|---------------|--------------------------------------------------|
| 510(k) No.: | K182134 |
| Manufacturer: | Compañía Mexicana de Radiología CRG S.A. de C.V. |
## 7. Indications for Use
The indications for use on subject devices are identical to the predicate device, the ARiX RAD Radiographic System, as shown below.
The MRH ALFA system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH ALFA radiographic system. The acquisition system comes as a separate system.
The MRH ALFA system is not intended for use in mammography, fluoroscopy or angiography.
The MRH ALFA system is intended for adults and pediatric patients. The pediation is as follows:
- . Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
The MRH II system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH II radiographic system. The acquisition system comes as a separate system.
The MRH II system is not intended for use in mammography, fluoroscopy or angiography.
The MRH II system is intended for adults and pediatric patients. The pediatric population is as follows:
- . Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
The MRH IIE system is a radiographic system used for producing X-ray images of the human body, including all anatomical regions such as the head, thorax, column, abdomen, extremities and internal organs. The system is designed to be used with conventional film or CR systems and can be interfaced with a digital detector, but neither the X-Ray detector nor the image acquisition system are part of the MRH IIE radiographic system. The acquisition system comes as a separate system.
The MRH IIE system is not intended for use in mammography, fluoroscopy or angiography.
{6}------------------------------------------------
The MRH IIE system is intended for adults and pediatric patients. The pediatric population is as follows:
- Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51 kg (114 lb.)
- Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.)
#### 8. Device Description
The MRH ALFA, MRH II & MRH IIE Radiographic Systems designed and manufactured by CMR are general purpose, computed radiography (CR) X-ray systems used for diagnostic imaging (analog radiology) of adults and pediatric patients for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The devices are designed to meet the requirements of basic radiology procedures and produces a low radiation dose to the patient, while acquiring its images.
## 9. Device Modifications
The following minor technological modifications have been made to the ARiX RAD Radiographic System (K182134), the predicate device. The minor modifications are being made to several main components of the predicate device to create the subject devices. The minor modifications are: 1) removal of the touch screen functionality from the Control Panel, 2) change the dimensions, load capacity and motorized movement feature of the Powered Table Top, 3) change the height, longitudinal displacement and vertical displacement of the X-ray Tube Stand and 4) change the casing material for the Wall Bucky. In addition, the name of the device is also being changed from the ARiX RAD Radiographic System to the MRH ALFA, MRH II & MRH IIE Radiographic Systems to reflect these different changes for marketing purposes.
#### 10. Substantial Equivalence Comparison
Table 5-1 compares the MRH ALFA, MRH IIE Radiographic Systems subject devices to the predicate device ARiX RAD Radiographic System with respect to indications for use, principles of operation, technological characteristics, components/materials, and performance testing. The comparison of the subject devices to the predicate device provides detailed information regarding the basis for the determination of substantial equivalence. The subject devices do not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{7}------------------------------------------------
| Attributes | Subject Devices | Predicate Device | Comparison |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Device Name | MRH ALFA, MRH II & MRH IIE Radiographic Systems | ARIX RAD Radiographic System | - |
| Manufacturer | Compañía Mexicana de Radiología (CMR) | Compañía Mexicana de Radiología (CMR) | Same |
| 510(k) # | Pending | K182134 | - |
| | Regulatory Information | | |
| Product Code | KPR | KPR | Identical |
| Regulation | CFR 892.1680 | CFR 892.1680 | Identical |
| Regulation Name | Stationary x-ray system | Stationary x-ray system | Identical |
| Class | II | II | Identical |
| Review Panel | Radiology | Radiology | Identical |
| Intended Use | Radiographic system used for producing X-ray images of | Radiographic system used for producing X-ray images of | Identical |
| | the human body. | the human body. | |
| Indications for Use | The MRH ALFA system is a radiographic system used for<br>producing X-ray images of the human body, including all<br>anatomical regions such as the head, thorax, column,<br>abdomen, extremities and internal organs. The system is<br>designed to be used with conventional film or CR systems<br>and can be interfaced with a digital detector, but neither<br>the X-Ray detector nor the image acquisition system are<br>part of the MRH ALFA radiographic system. The<br>acquisition system comes as a separate system.<br>The MRH ALFA system is not intended for use in<br>mammography, fluoroscopy or angiography.<br>The MRH ALFA system is intended for adults and<br>pediatric patients. The pediatric population is as follows:<br>• Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51<br>kg (114 lb.)<br>• Adolescent (12-21 yrs.), approximately 51 kg (114<br>lb.) to 80 kg (176 lb.)<br>The MRH II system is a radiographic system used for | The ARIX RAD system is a radiographic system used for<br>producing X-ray images of the human body, including all<br>anatomical regions such as the head, thorax, column,<br>abdomen, extremities and internal organs. The system<br>is designed to be used with conventional film or CR<br>systems and can be interfaced with a digital detector,<br>but neither the X-Ray detector nor the image acquisition<br>system are part of the ARIX RAD radiographic system.<br>The acquisition system comes as a separate system.<br>The ARIX RAD system is not intended for use in<br>mammography, fluoroscopy or angiography.<br>The ARIX RAD system is intended for adults and<br>pediatric patients. The pediatric population is as follows:<br>• Child (3-12 yrs.), approximately 12kg (26 lb.) to<br>51 kg (114 lb.)<br>• Adolescent (12-21 yrs.), approximately 51kg (114<br>lb.) to 80kg (176 lb.) | Identical |
| Attributes | Subject Devices | Predicate Device | Comparison |
| | anatomical regions such as the head, thorax, column,<br>abdomen, extremities and internal organs. The system is<br>designed to be used with conventional film or CR systems<br>and can be interfaced with a digital detector, but neither<br>the X-Ray detector nor the image acquisition system are<br>part of the MRH II radiographic system. The acquisition<br>system comes as a separate system. | | |
| | The MRH II system is not intended for use in<br>mammography, fluoroscopy or angiography. | | |
| | The MRH II system is intended for adults and pediatric<br>patients. The pediatric population is as follows: | | |
| | Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51<br>kg (114 lb.) | | |
| | Adolescent (12-21 yrs.), approximately 51 kg (114<br>lb.) to 80 kg (176 lb.) | | |
| | The MRH IIE system is a radiographic system used for<br>producing X-ray images of the human body, including all<br>anatomical regions such as the head, thorax, column,<br>abdomen, extremities and internal organs. The system is<br>designed to be used with conventional film or CR systems<br>and can be interfaced with a digital detector, but neither<br>the X-Ray detector nor the image acquisition system are<br>part of the MRH IIE radiographic system. The acquisition<br>system comes as a separate system. | | |
| | The MRH IIE system is not intended for use in<br>mammography, fluoroscopy or angiography. | | |
| | The MRH IIE system is intended for adults and pediatric<br>patients. The pediatric population is as follows: | | |
| | Child (3-12 yrs.), approximately 12 kg (26 lb.) to 51<br>kg (114 lb.) | | |
| Attributes | Subject Devices | Predicate Device | Comparison |
| | Adolescent (12-21 yrs.), approximately 51 kg (114 lb.) to 80 kg (176 lb.) | | |
| Technology Overview | General diagnostic X-ray system consisting of a tube-head and collimator assembly, with a generator, generator control, and an X-ray table. | General diagnostic X-ray system consisting of a tube-head and collimator assembly, with a generator, generator control, and an X-ray table. | Identical |
| Mechanism of Action | X-ray tube<br>(Dual Focal Spot) | X-ray tube<br>(Dual Focal Spot) | Identical |
| Configuration | Column mount | Column mount | Identical |
| Positioning Controls | Enhanced | Enhanced | Identical |
| Collimator | Manual<br>(made by Ralco) | Manual<br>(made by Ralco) | Identical |
| Anatomical Location | Full body | Full body | Identical |
| Power Supply | 220 VCA | 220 VCA | Identical |
| Generator | High frequency: 32, 50, 60 and 80 kW<br>(made by CMR) | High frequency: 32, 50, 60 and 80 kW<br>(made by CMR) | Identical |
| Exposure Voltage Ranges for Available Generators | 40 - 125 kVp or 40 - 150 kVp…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.