Digital Thermometer-(Model BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318)
K193621 · Dongguan E-Test Technology Co., Ltd. · FLL · Nov 23, 2020 · General Hospital
Device Facts
Record ID
K193621
Device Name
Digital Thermometer-(Model BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318)
Applicant
Dongguan E-Test Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Nov 23, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home. It can be used for axillary measurement, oral measurement and rectal measurement.
Device Story
Digital thermometer; uses thermistor sensor to detect body temperature via resistance changes; ASIC processes signal; LCD displays temperature; buzzer provides audible feedback; powered by 1.5V button battery. Used by clinicians or patients in hospital or home settings. Provides real-time temperature readings to assist in clinical monitoring and decision-making.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated through bench testing and adherence to recognized standards: IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112-00, ISO 10993-5, and ISO 10993-10.
Technological Characteristics
Thermistor-based sensor; ASIC signal processing; 1.5V button battery power; LCD display; buzzer. Complies with IEC 60601-1, IEC 80601-2-56, IEC 60601-1-2, IEC 60601-1-11, ASTM E1112-00, ISO 10993-5, and ISO 10993-10. Measurement range 32.0°C–42.9°C; accuracy ±0.1°C (35.0–39.0°C) and ±0.2°C otherwise.
Indications for Use
Indicated for measurement and monitoring of human body temperature in patients by clinicians or consumers in hospital or home settings via axillary, oral, or rectal routes.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Ageless Health Medical Digital Thermometer (K172508)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 23, 2020
Dongguan E-Test Technology Co., Ltd % Cassie Lee Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road, Guangzhou Science Park Guangzhou, 510000 China
Re: K193621
Trade/Device Name: Digital Thermometer (Model BT-301, BT-302, BT-303, BT-305, BT-306, BT-308. BT-311. BT-318) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 21, 2020 Received: October 28, 2020
Dear Cassie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193621
Device Name
Digital Thermometer (Model: BT-301, BT-302, BT-305, BT-306, BT-308, BT-311, BT-318)
Indications for Use (Describe)
Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.
It can be used for axillary measurement, oral measurement and rectal measurement.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary -K193621
This summary of 510(K) information is being submitted in accordance with the requirements of
SMDA and 21 CFR 807.92.
Date: 11/20/2020
## 1. Submitter's Information
510(k) Owner's Name: DONGGUAN E-TEST TECHNOLOGY CO., LTD
Establishment Registration Number: Applying
Address: Room 201,301, Building 1, Changping Section No.1, Dongshen Road, Changping Town, Dongguan City, Guangdong, China.
Tel: +86-0769-81158038
Contact Person (including title): Victor Wan (Vice-president)
E-mail: victor@agelh.com
### Application Correspondent:
Contact Person: Ms. Cassie Lee
DONGGUAN E-TEST TECHNOLOGY CO., LTD
Address: Room 201,301, Building 1, Changping Section No.1, Dongshen Road, Changping Town, Dongguan City, Guangdong, China
Tel: +86-0769-81158038 Email: regulatory@glomed-info.com
#### 2. Subject Device Information
| Type of 510(k): | Traditional |
|----------------------|---------------------------------------------------------------------------------------------|
| Common Name: | Thermometer, electronic |
| Trade Name: | Digital Thermometer -(Model BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318) |
| Classification Name: | Thermometer, electronic, clinical |
| 510(K) Number: | K193621 |
| Review Panel: | General Hospital |
{4}------------------------------------------------
| Sponsor: | DONGGUAN E-TEST TECHNOLOGY CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | Digital Thermometer, model: BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318 |
| File No.: | 510(k) Summary |
Product Code: FLL Regulation Number: 880.2910 Requlation Class: 2
#### 3. Predicate Device Information
- 510(K) Number: K172508
Sponsor: Dongguan Ageless Health Industrial Co., Ltd.
Common Name: Thermometer, electronic
Trade Name: Ageless Health Medical Digital Thermometer
Classification Name: Thermometer, electronic, clinical
Review Panel: General Hospital
Product Code: FLL
Regulation Number: 880.2910
Regulation Class: 2
#### Device Description 4.
Digital Thermometers comprises a thermistor for getting temperature signal, a reference resistor for comparing the resistance of the thermistor, a buzzer for sounding effect, an ASIC for processing the target temperature digitally, and a LCD for displaying the temperature result. The design principle of thermometer is based on thermos sensor and ASIC technology. A thermistor is used as thermos sensor. The ASIC gets the sensor's signal from human body, then processes the signal and calculates the result, after that displays the temperature result by a LCD.
#### 5. Intended Use
Digital Thermometers are intended for the measurement and monitoring of human body temperature by doctor or consumers in the hospital or home.
It can be used for axillary measurement, oral measurement and rectal measurement.
#### Test Summary 6.
The whole product and manufacturing used for the Digital thermometer are identical to those of the predicate device, which were demonstrated to conform with the following standards:
{5}------------------------------------------------
| Sponsor: | DONGGUAN E-TEST TECHNOLOGY CO., LTD |
|-----------------|--------------------------------------------------------------------------------------------|
| Subject Device: | Digital Thermometer, model: BT-301, BT-302, BT-303, BT-305, BT-306, BT-308, BT-311, BT-318 |
| File No.: | 510(k) Summary |
- . IEC 60601-1 (2005/(R)2012 and A1:2012) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance
- IEC 80601-2-56 (2018) Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
- I IEC 60601-1-2 (2014) Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- I IEC 60601-1-11 (2015) Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . ASTM E1112-00 (2011) Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature
- " ISO 10993-5 (2009) Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro cytotoxicity
- " ISO 10993-10 (2010) Biological Evaluation of Medical Devices - Part 10: Tests for irritation and skin sensitization
#### 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards.
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|-----------------------------------------|------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|--------|
| Company | DONGGUAN E-TEST<br>TECHNOLOGY CO., LTD | Ageless Health Industrial Co.,<br>Ltd. | -- |
| Name and<br>Model | Digital Thermometer, models:<br>BT-301, BT-302, BT-303,<br>BT-305, BT-306, BT-308,<br>BT-311, BT-318 | Ageless Health Medical Digital<br>Thermometer, models: BT-301,<br>BT-302, BT-303, BT-305,<br>BT-306, BT-308, BT-311, BT-318 | -- |
| 510(k)<br>Number | K193621 | K172508 | -- |
| Thermometer<br>Type | Digital Thermometer | Digital Thermometer | SE |
| Intended Use<br>/ Indication for<br>Use | Digital Thermometers are<br>intended for the measurement<br>and monitoring of human body | Ageless Health Medical Digital<br>Thermometers are intended for<br>the measurement and monitoring | SE |
{6}------------------------------------------------
### Sponsor:
DONGGUAN E-TEST TECHNOLOGY CO., LTD Digital Thermometer, model: BT-301, BT-302, BT-305, BT-306, BT-308, BT-311, Subject Device: BT-318 ary
| File No.: | 510(k) Summary |
|-----------|----------------|
|-----------|----------------|
| Elements of<br>Comparison | Subject Device | Predicate Device | Remark |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | temperature by doctor or<br>consumers in the hospital or<br>home.<br>It can be used for axillary<br>measurement, oral measurement<br>and rectal measurement. | of human body temperature by<br>doctor or consumers in the<br>hospital or home.<br>It can be used for axillary<br>measurement, oral measurement<br>and rectal measurement. | |
| Sensor | Thermistor | Thermistor | SE |
| Signal<br>Processing<br>and Display | Using the resistance change of<br>thermal resistor to detect body<br>temperature and displayed<br>through the LCD. | Using the resistance change of<br>thermal resistor to detect body<br>temperature and displayed<br>through the LCD. | SE |
| Power<br>Requirement | 1.5V button battery | 1.5V button battery | SE |
| Measurement<br>Temperature<br>Range | 32.0°C ~ 42.9°C | 32.0°C ~ 42.9°C | SE |
| Measurement<br>Accuracy | 35.0 ~ 39.0°C : +/- 0.1°C<br>The rest: +/- 0.2°C | 35.0 ~ 39.0°C : +/- 0.1°C<br>The rest: +/- 0.2°C | SE |
| Ambient<br>Temperature | 10~35°C | 10~35°C | SE |
| Response<br>Time | 60s | 60s | SE |
| Complied<br>Standard | IEC 60601-1, IEC 80601-2-56,<br>IEC 60601-1-2, IEC 60601-1-11,<br>ASTM E1112, ISO 10993-5, ISO<br>10993-10 | IEC 60601-1, IEC 80601-2-56,<br>IEC 60601-1-2, IEC 60601-1-11,<br>ASTM E1112, ISO 10993-5, ISO<br>10993-10 | SE |
### Conclusion
The subject device Digital Thermometer has all features of the predicate devices. No differences will affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device. All hardware and software of the subject device are based on that of the predicate device (K172508), since no new testing is presented in the submission.
### 8. Summary Prepared Date
20 November 2020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.