K193620 · Opsens, Inc. · DXO · Jun 18, 2020 · Cardiovascular
Device Facts
Record ID
K193620
Device Name
OptoMonitor 3
Applicant
Opsens, Inc.
Product Code
DXO · Cardiovascular
Decision Date
Jun 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2870
Device Class
Class 2
Indications for Use
To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.
Device Story
OptoMonitor 3 is a pressure monitoring system used in cath labs by physicians to measure intravascular blood pressure. It functions as an accessory to Opsens OptoWire pressure guidewires, which utilize fiberoptic sensors to detect pressure. The system processes optical signals from the guidewire and electrical signals from external aortic pressure transducers. It displays real-time pressure curves and numerical values, including mean aortic pressure, mean distal pressure, and hemodynamic indices like fractional flow reserve (FFR) and dPR. The device consists of an Optical Unit, a Display Unit (all-in-one PC), and a Handle Unit. It supports connectivity to hospital DICOM systems. By providing accurate hemodynamic data, it assists clinicians in diagnosing and treating blood vessel disease. The device is non-sterile and non-patient contact.
Clinical Evidence
No clinical data or animal studies were provided. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and software verification and validation.
Technological Characteristics
Fiberoptic pressure sensing; fiber bundle guidewire interface. System components: Optical Unit, Display Unit (all-in-one PC), Handle Unit. Connectivity: DICOM, aortic pressure transducer input (low-level), auxiliary input (high-level). Electrical isolation: Class 2 (double). Operating temperature: 15°C to 30°C. Software: Moderate level of concern.
Indications for Use
Indicated for patients undergoing diagnostic angiography or interventional procedures requiring hemodynamic pressure monitoring in coronary or peripheral blood vessels.
Regulatory Classification
Identification
A catheter tip pressure transducer is a device incorporated into the distal end of a catheter. When placed in the bloodstream, its mechanical or electrical properties change in relation to changes in blood pressure. These changes are transmitted to accessory equipment for processing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 18, 2020
Opsens Inc. Marc Chaunet Director, Regulatory Affairs and Quality System 750 boulevard Du Parc Technologique Quebec, G1P 4S3 Ca
Re: K193620
Trade/Device Name: OptoMonitor 3 Regulation Number: 21 CFR 870.2870 Regulation Name: Catheter Tip Pressure Transducer Regulatory Class: Class II Product Code: DXO Dated: May 18, 2020 Received: May 18, 2020
Dear Marc Chaunet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193620
Device Name OptoMonitor 3
Indications for Use (Describe)
To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/ or any interventional procedures.
Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel desease.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### ഗ 510(k) Summary OptoMonitor 3
#### SUBMITTER 1.
Address: Opsens, Inc.
750, Boulevard du Parc Technologique
Quebec (Quebec) G1P 4S3
Phone: 418.781.0333 ext.: 3408
Fax Number: 418-781-0024
Contact Person: Marc Chaunet, Regulatory Affairs and Quality System Director
Email: marc.chaunet@opsens.com
Date Prepared: June 17, 2020
#### 2. DEVICE
Name of Device: OptoMonitor 3
Common or Usual Name: Pressure Monitor
Classification name: Transducer, pressure, catheter tip (870.2870)
Regulatory Class: II
Product Code: DXO
#### PREDICATE DEVICE 3.
OptoMonitor System cleared via K192340 (cleared on 12/12/2019). This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
#### 4. DEVICE DESCRIPTION
The proposed OptoMonitor 3 is a new version of the OptoMonitor System. This device and its components are considered accessories to Opsens OptoWire™ pressure guidewires and are intended for use with legally marketed pressure guidewires.
The proposed OptoMonitor 3 includes an Optical Unit (OU), a Display Unit (DU), a Handle Unit (HU) and accessories (cables, power supply, etc). These hardware components and device functionalities are equivalent to that of the previous generation OptoMonitor (K192340 (cleared on 12/12/2019).
The device is a non-sterile, non-patient contact device.
{4}------------------------------------------------
#### 5. INDICATIONS FOR USE
The OptoMonitor 3, in conjunction with the OptoWire™ pressure guidewire, is indicated for use to measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE 6. DEVICE
The proposed OptoMonitor 3 is substantially equivalent to the OptoMonitor cleared via K192340 on 12/12/2019.
The proposed OptoMonitor 3 is a new version of the predecessor OptoMonitor.
Indications for Use for the OptoMonitor 3 are the same as the predicate device indications K192340 (cleared on 12/12/2019).
The technological characteristics of the proposed OptoMonitor 3 are essentially the same as for the OptoMonitor. The main difference between the subject device and its predecessor resides in the following:
OPTICAL UNIT
- New casing
- New CPU and PCBs. A new software was also validated for the OU. Although, algorithms used for . the FFR calculation remained unchanged from the predicate OptoMonitor.
- Addition of possibility to receive aortic pressure signal (Pa) from a Cathlab Hemodynamic system
- Addition of a low-level signal input (Aortic In) allowing to connect an aortic pressure ● transducer and pass-through to Cathlab Hemodynamic system
- Handle unit redesigned and connector split to separate electrical and optical connections
- Software modifications to support new hardware
- Removal of printer
## DISPLAY UNIT
- Use of commercially available all-in-one PC. .
- . New software to support additional hardware options
- Addition of possibility to connect to hospital DICOM system ●
The changes have been evaluated through the Risk Management Process and no new questions of safety and effectiveness were identified. Existing questions of safety and effectiveness are valid for the proposed device. Any change raises a question concerning whether its performance can be expected to be equivalent with the predicate. Performance testing has confirmed equivalence. No new questions of safety and effectiveness were identified during the execution of Verification and Validation activities.
Therefore, the proposed device, OptoMonitor 3, meets substantial equivalence requirements with regards to the legally marketed predicate OptoMonitor (K192340 cleared on 12/12/2019).
For detailed comparison, refer to the Substantial Equivalence table on the following pages.
{5}------------------------------------------------
| Substantial Equivalence Table | | | | |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | Proposed Device | Predicate Device | Differences | |
| Regulatory Information | Name | OptoMonitor 3 | OptoMonitor | N/A |
| | 510(k)# | K193620 | K192340 | N/A |
| | Predicates | K192340 | K173860<br>K142598<br>K111395<br>K041134 | N/A |
| | Product Code | DXO | DXO | Same |
| | Class | 2 | 2 | Same |
| | Regulation Number | 870.2870 | 870.2870 | Same |
| | Regulation Generic Name | Transducer, pressure, catheter tip | Transducer, pressure, catheter tip | Same |
| Intended use | Regulation Intended Use | Accessory equipment for processing mechanical or electrical property changes in relation to changes in blood pressure. | Accessory equipment for processing mechanical or electrical property changes in relation to changes in blood pressure. | Same |
| | Indications | To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease. | To measure pressure in blood vessels including both coronary and peripheral vessels, during diagnostic angiography and/or other any interventional procedures. Blood pressure measurements provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel disease. | Same |
| | Prescription Use | Rx Only | Rx Only | Same |
| | System Components | Sterile, disposable guidewire<br>Reusable signal processor / monitor<br>Embedded software<br>Connecting cables | Sterile, disposable guidewire<br>Reusable signal processor / monitor<br>Embedded software<br>Connecting cables | Same |
| Technological Characteristics | System Capabilities | Measurement of intravascular blood<br>Pressure including FFR. | Measurement of intravascular blood<br>Pressure including FFR. | Same |
| | Pressure Sensing & Signal Transmission Technology | Fiberoptic sensor & fiber bundle embedded in guidewire. Monitor Senses pressure from Fiberoptic sensor. | Fiberoptic sensor & fiber bundle embedded in guidewire. Monitor Senses pressure from Fiberoptic sensor. | Same |
| | Operating Temperature (Monitor) | 15°C to 30°C | 15°C to 30°C | Same |
| | Transport Temperature (Monitor) | -25°C to 60°C | -25°C to 60°C | Same |
| | Operating Relative Humidity (Monitor) | 10% to 85% non-condensing | 10% to 85% non-condensing | Same |
| | Storage Temperature (Monitor) | Room Temperature | Room Temperature | Same |
| | Proposed Device | Predicate Device | Differences | |
| Operating Pressure | 70 to 106 kPa | 70 to 106 kPa | Same | |
| Pressure Range | -30 to 300 mmHg | -30 to 300 mmHg | Same | |
| Pressure Accuracy | +/- 1 mmHg plus +/- 1% of<br>reading<br>(pressure range -30 to 50<br>mmHg) or +/- 3% of reading<br>(pressure range 50 to 300<br>mmHg) | +/- 1 mmHg plus +/- 1% of<br>reading<br>(pressure range -30 to 50<br>mmHg) or +/- 3% of reading<br>(pressure range 50 to 300<br>mmHg) | Same | |
| Thermal Zero Shift | <0.3 mmHg/deg C | <0.3 mmHg/deg C | Same | |
| Zero Drift | <1 mmHg/h | <1 mmHg/h | Same | |
| Electrical Isolation | Class 2 (double isolation) | Class 2 (double isolation) | Same | |
| User Interface | Touchscreen | Touchscreen | Same | |
| Auto-zeroing | Yes | Yes | Same | |
| Real Time Curves | Aortic instantaneous pressure,<br>aortic mean pressure, distal<br>instantaneous pressure, distal<br>mean pressure | Aortic instantaneous pressure,<br>aortic mean pressure, distal<br>instantaneous pressure, distal<br>mean pressure | Same | |
| Real Time Numerical<br>Values | Mean aortic pressure, mean<br>distal pressure, mean Pd/mean<br>Pa; FFR, dPR | Mean aortic pressure, mean<br>distal pressure, mean Pd/mean<br>Pa; FFR, dPR | Same | |
| Recording Values | Instantaneous Pa, Pd and<br>Pd/Pa; mean Pa; mean Pd;<br>mean Pd/mean Pa; FFR, dPR | Instantaneous Pa, Pd and<br>Pd/Pa; mean Pa; mean Pd;<br>mean Pd/mean Pa; FFR, dPR | Same | |
| Display Monitor | LCD | LCD | Same | |
| Aortic Input | Low Level (5μV/V/mmHg) | High Level (100 mmHg/V) | Low level attributed<br>to Aortic input for<br>OpM3 | |
| Distal Input | OptoWire (optical) | OptoWire (optical) | Same | |
| AUX Input | High Level (100 mmHg/V) | N/A | High level input<br>added to AUX In for<br>OpM3 | |
| Distal pressure<br>output | Low Level (5μV/V/mmHg) | Low Level (5μV/V/mmHg) | Same | |
| Hardware<br>components | Signal Conditioner Unit (SCU),<br>the Display Unit (DU), The<br>Handle Unit (HU) and<br>accessories (cables, power<br>supply, etc) | Signal Conditioner Unit (SCU),<br>the Display Unit (DU), The<br>Handle Unit (HU) and<br>accessories (cables, power<br>supply, etc) | Same | |
| Connected devices | OptoWire™ | OptoWire™ | Same | |
K 1 9 3 6 2 0
{6}------------------------------------------------
{7}------------------------------------------------
#### PERFORMANCE DATA 7.
The following performance data were provided in support of the substantial equivalence determination.
Electrical safety and electromagnetic compatibility (EMC) was assessed with respect to the software change and the system was found to comply with IEC 60601-1, and IEC 60601-1-2 standard for EMC.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
No new questions of safety and effectiveness were identified during review of Risk Management documentation or execution of Verification and Validation activities.
No animal studies or clinical investigations are included with this submission.
#### 8. CONCLUSIONS
The results from these tests mentioned above demonstrate that the technological and performance characteristics of the proposed OptoMonitor 3 is comparable to the predicate device, support substantial equivalence of the device that is the subject of this 510(k).
The results of the verification/validation tests and the risk analysis have demonstrated that the additional features incorporated in the OptoMonitor 3 device do not add any new questions of safety and efficacy and is therefore substantially equivalent to the predicate OptoMonitor System (K192340 cleared on 12/12/2019).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.