BOUNDER 300 Power Wheelchair, BOUNDER Plus 300 Power Wheelchair, BOUNDER 450 Power Wheelchair, BIG BOUNDER 600 Power Wheelchair, BIG BOUNDER 1000 Power Wheelchair
K193599 · 21st Century Scientific, Inc. · ITI · Sep 3, 2020 · Physical Medicine
Device Facts
Record ID
K193599
Device Name
BOUNDER 300 Power Wheelchair, BOUNDER Plus 300 Power Wheelchair, BOUNDER 450 Power Wheelchair, BIG BOUNDER 600 Power Wheelchair, BIG BOUNDER 1000 Power Wheelchair
Applicant
21st Century Scientific, Inc.
Product Code
ITI · Physical Medicine
Decision Date
Sep 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the BOUNDER Power Wheelchair and the predicate device is to provide mobility to persons limited to a sitting position.
Device Story
The BOUNDER Power Wheelchair is a rear-wheel drive, non-folding powered mobility device designed for indoor and outdoor use. It consists of a rigid base, two drive motors, two 12V DC batteries, and an electronic controller. Users operate the device via a joystick or equivalent command interface. The device features modular configurations, including options for power tilt, anterior tilt, recline, and seat elevation. It incorporates independent drive wheel suspension and supports weight capacities up to 1000 lbs. The device provides mobility to physically challenged individuals, allowing them to navigate environments independently. It is intended for use by patients who are limited to a sitting position.
Clinical Evidence
No clinical data was required for this submission. Substantial equivalence was demonstrated through bench testing, including design verification and validation performed according to ANSI/RESNA WC-1:2009 and WC-2:2009 standards, and ISO 7176-30:2018.
Technological Characteristics
Materials: steel and aluminum frames, standard foams/covers. Energy: 24V DC (two 12V batteries). Drive: two-motor rear-wheel drive with independent suspension. Controls: joystick/electronic controller (LiNX, R-net). Standards: ANSI/RESNA WC-1:2009, WC-2:2009, ISO 7176-30:2018. Form factor: rigid, non-folding base with anti-tip wheels and front casters.
Indications for Use
Indicated for physically challenged persons limited to a sitting position requiring powered mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
AMYPOWER ALLTRACK Series Power Wheelchair (K092225)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
September 3, 2020
21st Century Scientific, Inc. RD Davidson COO 4931 N Manufacturing Way Coeur d'Alene, Idaho 83815
Re: K193599
Trade/Device Name: BOUNDER 300 Power Wheelchair, BOUNDER Plus 300 Power Wheelchair, BOUNDER 450 Power Wheelchair, BIG BOUNDER 600 Power Wheelchair, BIG BOUNDER 1000 Power Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: August 11, 2020 Received: August 12, 2020
Dear RD Davidson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Heather Dean, Ph.D. Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193599
Device Name
BOUNDER Series Power Wheelchair. Models: BOUNDER 300, BOUNDER 450, BIG BOUNDER 600, BIG BOUNDER 1000
Indications for Use (Describe)
The indication for use of the BOUNDER Power Wheelchair is to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5 510(k) SUMMARY
# BOUNDER Power Wheelchair
## SUBMITTER per 21 CFR 807.92(a)(1):
21st Century Scientific, Inc. 4931 N Manufacturing Way Coeur d'Alene, ID 83815 Phone: (208) 667-8800 Email: rdd(@wheelchairs.com
## Contact Person
RD Davidson, COO
## Manufacturer
21st Century Scientific, Inc. 4931 N Manufacturing Way Coeur d'Alene, ID 83815
## FDA Registration number of Manufacturer
2027797
## DATE PREPARED per 21 CFR 807.92(a)(1):
## November 15, 2019
## PRIOR SUBMISSIONS
The previous 510(k) for the BOUNDER series power wheelchair was K90210.
{4}------------------------------------------------
## Section 5 ... : 510(k) Summary (continued)
#### DEVICE INFORMATION per 21 CFR 807.92 (a) (2)
Name of Device.................................: BOUNDER Series Power Wheelchair. Models: BOUNDER 300, BOUNDER Plus 300, BOUNDER 450, BIG BOUNDER 600, BIG BOUNDER 1000
| Common or Usual Name............: | Wheelchair, Powered |
|-----------------------------------------|-------------------------------------|
| Classification Name................: | Powered Wheelchair 21 CFR § 890.386 |
| Product Code................: | ITI |
| Classification of New Device......: | Class II |
| Patient Population....................: | Physically Challenged Persons |
#### PREDICATE DEVICE per 21 CFR 807.92 (a) (3)
| Name of Device | : AMYPOWER ALLTRACK Series Power Wheelchair (K092225) |
|-------------------------------------|-------------------------------------------------------|
| Common or Usual Name | : Wheelchair, Powered |
| Classification Name | : Powered Wheelchair 21 CFR § 890.3860 |
| Product Code | : ITI |
| Classifications of Predicate Device | : Class II |
| Patient Population | : Physically Challenged Persons |
## DEVICE DESCRIPTION per 21 CFR 808.92(a)(4)
The subject device is an update to the existing previously cleared BOUNDER Power Wheelchair (K901210). The BOUNDER utilizes components found on most power wheelchairs. These include a rigid or "nonfolding", rear wheel drive power wheelchair base with two drive wheels in the rear with two anti-tip wheels mounted behind and two casters in the front. It has two motors and is powered by two 12V DC Batteries (connected in series), and electronic controller. It also includes a seat, footrests, arms, positioning strap and charger. The design offers a modular approach and can be configured in multiple versions to suit different user needs.
The new models will include: BOUNDER 300, BOUNDER Plus 300, BOUNDER 450, BIG BOUNDER 600 & BIG BOUNDER 1000.
The updated subject version of the BOUNDER Power Wheelchair has the following changes:
- Increase in the weight capacity of the previously cleared device from 500 lb to 1000 lb ●
- . Add Power Tilt, Power Anterior Tilt, Power Recline and Power Seat Elevator options
- Addition of LiNX and R-net electronics
- Add Independent Drive Wheel Suspension .
- Add an Off-Road tire option for BOUNDER 300, BOUNDER Plus 300 and BOUNDER 450 ●
- . Top speed: 8.5 mph
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------|
| Page 2 of 4 | | |
| \\SWORK2014\Drive D - Data\Data\QSM\Design History...\FDA 510K\Submission\Revised 510(k) Summary 2020-09-02.docx | | |
{5}------------------------------------------------
## Section 5 ... : 510(k) Summary (continued)
## INTENDED USE per 21 CFR 807.92(A)(5):
The intended use of the BOUNDER Power Wheelchair and the predicate device is to provide mobility to persons limited to a sitting position.
#### INDICATIONS FOR USE per FORM FDA 3881:
The indication for use of the BOUNDER Power Wheelchair is to provide mobility to persons limited to a sitting position.
## COMPARISON OF DEVICE TECHNOLOGICAL CHARACTERISTICS TO PREDICATE DEVICE per CFR 807.92(A)(6):
The device comparison showed that the subject device is substantially equivalent in design, materials, and operational principles to the previously cleared predicate device in regard to providing mobility to persons limited to a sitting position. There is no difference in intended use and indications for use between the predicate device and the subject device.
This device has similar technological characteristics as the predicate devices utilize steel and aluminum in their frames and components, and standard foams and covers for seat cushions and backs. The user controls the chair by using a joystick or other equivalent command mode through a controller. It includes two motors and operates at 24 Volts using rechargeable batteries as a source of energy. The operating speeds, maneuverability, power modules, hand controls, seat types, and climbing ability are substantially equivalent. Both the BOUNDER Power Wheelchair and the predicate are recommended for indoor and outdoor use. The standard accessories and components are common to all power wheelchair devices.
## BASIS OF SUBSTANTIAL EQUIVALENCE per 21 CFR 807.100(b)(2)(3):
The substantial equivalence of the subject device was determined as per the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" and the technological characteristics which include materials, design, and other device related features, as defined in section 513(i)(1)(B) of the FD&C Act and 21 CFR 807.100(b)(2)(ii)(A). The subject device components are as safe and effective as the predicate device and do not raise different questions of safety and effectiveness. The design verification and validation testing, device comparison, and dimensional analysis demonstrates that the subject device components are substantially equivalent to the predicate device.
## DESIGN VERIFICATION AND VALIDATION TESTING DATA
Design verification and validation testing was performed on the subject device and was found to meet performance requirements. Further, it is substantially equivalent to the predicate device identified throughout this submission and does not raise any new questions of safety and effectiveness. The acceptance criteria for the full verification of the design and acceptance criteria for each section of the Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing standard have been met.
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------|
| Page 3 of 4 | | |
| \\SWORK2014\\Drive D - Data\\Data\\QSM\\Design History\\...\\FDA 510K\\Submission\\Revised 510(k) Summary 2020-09-02.docx | | |
{6}------------------------------------------------
## Section 5 ... : 510(k) Summary (continued)
Non-Clinical Test per 21 CFR 807.92(b)(1) Rehabilitation Engineering and Assistive Technology Society of North America (RESNA) testing was performed to demonstrate that the BOUNDER Power Wheelchair meets the performance requirements and is substantially equivalent to the predicate device identified throughout this submission and do not raise any new questions of safety and effectiveness.
The following testing was performed according to FDA recognized consensus standards:
- · ANSI/RESNA WC-1:2009 Section 1: Determination of Static Stability
- · ANSIRESNA WC-2:2009 Section 2: Determination of Dynamic Stability of electrically Powered Wheelchairs
- · ANSI/RESNA WC-2:2009 Section 3: Determination of Effectiveness of Brakes
- · ANSI/RESNA WC-2:2009 Section 4: Energy Consumption of Electrically Powered Wheelchairs and Scooters for Determination of Theoretical Distance Range
- · ANSI/RESNA WC-1:2009 Section 5: Determination of Dimensions, Mass and Maneuvering Space
- · ANSI/RESNA WC-2:2009 Section 6: Determination of Maximum Speed, Acceleration and Deceleration of Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 7: Method of Measurement of Seating and Wheel Dimensions
- · ANSI/RESNA WC-1:2009 Section 8: Requirements and Test Methods for Static, Impact and Fatigue Strengths
- · ANSI/RESNA WC-2:2009 Section 9: Climatic Tests for Electrically Powered Wheelchairs
- · ANSI/RESNA WC-2:2009 Section 10: Determination of Obstacle-Climbing Ability of Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 11: Test Dummies
- · ANSI/RESNA WC-1:2009 Section 13: Determination of Coefficient of Friction of Test Surfaces
- · ANSI/RESNA WC-2:2009 Section 14: Power and Control Systems for Electrically Powered Wheelchairs
- · ANSI/RESNA WC-1:2009 Section 15: Requirements for Information Disclosure, Documentation and Labeling
- · ANSI/RESNA WC-1:2009 Section 16: Resistance to Ignition of Upholstered Parts
- · ANSI/RESNA WC-2: 2009 Section 21: Requirements and Test Methods for Electromagnetic Compatibility of Electrically Powered Wheelchairs and Motorized Scooters
- · ANSI/RESNA WC-1:2009 Section 22: Set-up Procedures
- ISO 7176-30:2018[E] Part 30: Wheelchairs for Changing Occupant Posture
Verification testing demonstrated that the BOUNDER Power Wheelchair is substantially equivalent to the marketed predicate device.
#### Animal Study
Animal testing was not required for this submission.
#### Clinical Testing
Clinical testing was not required for this submission.
#### CONCLUSIONS per 21 CFR 807.92(b)(3)
The subject device has the same intended use and technological characteristics as the predicate device. The design verification and validation data support the safety and performance of the subject device and demonstrate that the subject device will perform as intended in the specified use conditions. Therefore, the subject BOUNDER Power Wheelchair is comparable with the previously cleared AMYPOWER ALLTRACKS Series Power Wheelchair and demonstrates a basis of substantial equivalence.
| 510(k) SUMMARY | BOUNDER POWER WHEELCHAIR | SUBMITTED: 2020-09-02 |
|------------------------------------------------------------------------------------------------------------------|--------------------------|-----------------------|
| Page 4 of 4 | | |
| \\SWORK2014\Drive D - Data\Data\QSM\Design History...\FDA 510K\Submission\Revised 510(k) Summary 2020-09-02.docx | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.