K193579 · Medtronic Navigation · GEI · Mar 24, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193579
Device Name
PlasmaBlade X 3.0S LIGHT
Applicant
Medtronic Navigation
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Surgery System is indicated for cutting and coagulation of soft tissue during General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological procedures.
Device Story
PlasmaBlade X 3.0S LIGHT is a single-use, monopolar RF electrosurgical device; used with a qualified generator as part of a surgery system. Device features a bendable blade, telescoping shaft, integrated suction lumen for smoke/fluid evacuation, and integrated LED illumination. Operated via hand switch or footswitch by surgeons in clinical settings. RF energy enables soft tissue cutting and coagulation. LED light assists in field visualization. Bench and animal testing confirm performance equivalence to predicates regarding thermal damage, blade geometry, and light output. Benefits include precise tissue management and improved visualization during surgical procedures.
Clinical Evidence
No clinical data. Evidence consists of bench testing (design verification, light performance specifications) and animal testing (in-vivo porcine muscle, ex-vivo skin and liver thermal damage assessment) to confirm substantial equivalence.
Technological Characteristics
Monopolar RF electrosurgical device; coated stainless steel blade; telescoping shaft; integrated suction lumen; LED illumination. Sterilization: EtO. Biocompatibility: ISO 10993-1. Electrical safety: IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2. Power output: 0.5W to 50W.
Indications for Use
Indicated for cutting and coagulation of soft tissue in patients undergoing General, Plastic, Reconstructive, ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal, and Neurological surgical procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2020
Medtronic Navigation Gina Cunsolo Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, Colorado 80027
Re: K193579
Trade/Device Name: PlasmaBlade X 3.0S LIGHT Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 20, 2019 Received: December 23, 2019
Dear Gina Cunsolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph. D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193579
Device Name PlasmaBlade X 3.0S LIGHT
Indications for Use (Describe)
The Surgery System is indicated for cutting and coagulation of soft tissue during General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological procedures.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Medtronic Navigation<br>826 Coal Creek Circle<br>Louisville, CO |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Gina Cunsolo<br>Regulatory Affairs Specialist<br>Phone: (603) 842-6210<br>Fax: (603) 742-1488<br>Email: gina.m.cunsolo@medtronic.com |
| Date Summary Prepared: | March 19, 2020 |
| Device Trade Name: | PlasmaBlade X 3.0S LIGHT |
| Common Name: | Electrosurgical Instrument |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR 878.4400) |
| Product Code: | GEI |
| Predicate Devices: | PlasmaBlade X 3.0S (K181257)<br>PlasmaBlade 3.0S (K093695) |
| Reference Device: | PhotonBlade (K162053) |
| Device Description: | The PlasmaBlade™ X 3.0S LIGHT is a single-use,<br>monopolar RF device. It is designed to be used with the<br>qualified Generator as part of the Surgery System. It<br>can be operated with the integrated hand switch or a<br>qualified Footswitch. The PlasmaBlade™ X 3.0S LIGHT<br>consists of a single bendable blade and telescoping<br>shaft that can be configured in both standard and<br>extended length. The finger grip also incorporates a<br>suction lumen for the evacuation of smoke and fluids.<br>The device also has integrated LED-based illumination. |
| Indications for Use: | The Surgery System is indicated for cutting and coagulation of soft tissue during General, Plastic and Reconstructive (including but not limited to skin incisions and development of skin flaps), ENT, Gynecologic, Orthopaedic, Arthroscopic, Spinal and Neurological procedures. |
{4}------------------------------------------------
Substantial Equivalence Comparison Two predicate devices (cleared by Medtronic Advanced Energy) were used for substantial equivalence comparison to the subject device.
| Attribute | PlasmaBlade X 3.0S<br>LIGHT (Subject -<br>K193579) | PlasmaBlade X<br>3.0S (K181257) | PlasmaBlade<br>3.0S (K093695) |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------|
| Intended<br>Use/Indications<br>for Use | The Surgery System is<br>indicated for cutting<br>and coagulation of<br>soft tissue during<br>General, Plastic and<br>Reconstructive<br>(including but not<br>limited to skin<br>incisions and<br>development of skin<br>flaps), ENT,<br>Gynecologic,<br>Orthopaedic,<br>Arthroscopic, Spinal<br>and Neurological<br>procedures. | Same | Same |
| Mechanism of<br>action/principle<br>od operation | RF Monopolar Energy | Same | Same |
| Power Output | Max power: 50W +/-<br>20% 9Cut 8, Cut 10<br>and Coag 10)<br>Min power: ~0.5W +/-<br>1W at (Cut 1) | Same | Same |
| Sterilization | EtO | Same | Same |
| Biocomp | External<br>communicating device<br><24 hours, tested to<br>ISO 10993-1 | Same | Same |
| Blade Material | Coated Stainless Steel | Same | Same |
| Integrated<br>Suction | Yes | Same | Same |
| LED Light | Yes | No | No |
{5}------------------------------------------------
| Technological<br>Characteristics: | The predicate PlasmaBlade devices and subject PlasmaBlade X 3.0S LIGHT device are intended for RF-based cutting and coagulation. The PlasmaBlade X 3.0S LIGHT device shares the same operational characteristics as the predicate platforms, comprised of a radio-frequency generator which supplies RF power to disposable electrode devices for electrosurgical procedures. The differences between the subject and predicate devices are:<br><br>The PlasmaBlade X 3.0S LIGHT contains an LED light source located at the blade end of the handpiece. Bench testing was completed in order to ensure all light specifications were met. An updated blade geometry to the subject device Bench testing was completed to ensure the new geometry did not affect performance of the blade. Updated blade coating to incorporate material utilized in the previous generation of the PlasmaBlade 3.0S device (K093695). Similar to other devices in the field (i.e. PhotonBlade K162053), the PlasmaBlade X 3.0S LIGHT contains an LED-light source to assist in illuminating the field. |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Non-Clinical Testing: | The design and performance of the new features of the PlasmaBlade X 3.0S LIGHT were verified and validated through bench testing. Medtronic product development processes and the finalized FDA Guidance Document, "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff," issued on August 15, 2016, were utilized to determine applicable bench testing requirements. |
Sterilization, Biocompatibility, IEC and EMC Testing were conducted in compliance with the following FDA recognized consensus standards:
| Standard | Title |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 11135 2nd Ed 2014 | Sterilization Of Health-Care<br>Products - Ethylene Oxide -<br>Requirements For The<br>Development, Validation And<br>Routine Control Of A<br>Sterilization Process For<br>Medical Devices |
| 10993-5 Third Edition<br>2009-06-01 | Biological Evaluation Of Medical Devices - Part<br>5: Tests For In Vitro Cytotoxicity |
{6}------------------------------------------------
| 10993-10 Third Edition<br>2010-08-01 | Biological Evaluation Of Medical Devices - Part<br>10: Tests For Irritation And Skin Sensitization |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 10993-11 Third Edition<br>2017-09 | Biological Evaluation Of Medical Devices - Part<br>11: Tests For Systemic Toxicity |
| ES60601-<br>1:2005/(R)2012 And<br>A1:2012, | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential<br>performance (IEC 60601-1:2005, MOD) |
| 60601-2-2 Edition 6.0<br>2017-03 | Medical electrical equipment - Part 2-2:<br>Particular requirements for the basic safety<br>and essential performance of high frequency<br>surgical equipment and high frequency surgical<br>accessories |
| 60601-1-2 Edition 4.0<br>2014-02 | Medical Electrical Equipment - Part 1-2:<br>General Requirements For Basic Safety And<br>Essential Performance - Collateral Standard:<br>Electromagnetic Disturbances - Requirements<br>And Tests |
### Bench Testing
| Test | Description |
|---------------------|--------------------------------------------------------------------------------------------------------|
| Design Verification | Design Verification was completed to<br>guarantee all new design features functioned<br>appropriately. |
In comparison to the predicate devices, the subject PlasmaBlade X 3.0S LIGHT device was evaluated via bench testing to ensure all features performed in an equivalent manner to the predicates. Bench testing included consideration of the geometry change as well as the addition of the blade coating. All design features were met. In addition, the subject device was tested for multiple light performance specifications, which included allowable activation amounts, lux ranges (minimum and maximum), and light diameter at extension and collapse. All light specifications were met.
### Animal Testing
| Test | Description |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tissue Testing | Thermal damage testing was completed on<br>muscle, skin, and liver in accordance with the<br>guidance entitled, "Premarket Notification<br>(510(k)) Submissions for Electrosurgical Devices<br>for General Surgery," issued August 2016, to<br>ensure substantial equivalence to predicate. |
{7}------------------------------------------------
Similar to the predicate devices, thermal damage testing was completed on a porcine model to ensure the subject device was substantially equivalent to the predicates. The muscle tissue testing was completed in-vivo, while the skin and liver tissue testing was completed ex-vivo. Testing concluded the subject device was substantially equivalent.
| Summary of Clinical<br>Tests: | Clinical Testing was not required for these products. |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The indications for use, technology and performance characteristics of<br>the PlasmaBlade X 3.0S LIGHT devices are equivalent to the predicate<br>devices. Therefore, the PlasmaBlade X 3.0S LIGHT is substantially<br>equivalent to the predicates. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.