K193540 · DePuy Orthopaedics, Inc. · JWH · Jun 16, 2020 · Orthopedic
Device Facts
Record ID
K193540
Device Name
TRUMATCH Personalized Solutions
Applicant
DePuy Orthopaedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jun 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Indications for Use
The TruMatch Patient Specific Instruments are intent-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting. The anatomical landmarks necessary for the TruMatch Patient Specific Instruments must be present and identifiable on CT. The TruMatch Patient Specific Instruments are intended for use with Sigma Total Knee Implants and Attune Total Knee Implants and their cleared indications for use. The TruMatch Patient Specific Instruments are intended for single use only.
Device Story
TRUMATCH Personalized Solutions system uses patient CT images to design custom surgical instruments for total knee arthroplasty. Software allows surgeons to plan either mechanical axis or patient-specific alignment techniques. System generates patient-specific guides used intra-operatively to assist in positioning joint replacement components and marking bone for resection. Guides are manufactured based on approved surgical plans. Used in clinical settings by orthopedic surgeons. Benefits include personalized surgical planning and improved accuracy in bone cutting and implant positioning. System is compatible with Sigma and Attune Total Knee Implants.
Clinical Evidence
No clinical tests were conducted. Substantial equivalence demonstrated via non-clinical bench testing, including model creation process verification, cadaver lab studies assessing guide fit, implant alignment, size, and bone resection accuracy, and software verification testing.
Technological Characteristics
Materials: Accura and stainless steel. Sensing/Actuation: Patient-specific guides derived from CT imaging. Energy: None (manual instrumentation). Connectivity: Software-based planning system. Sterilization: Gamma radiation. Packaging: Polyurethane bag in blister tray with Tyvek lid. Shelf life: 185 days.
Indications for Use
Indicated for patients with osteoarthritis requiring primary total knee replacement via measured resection technique. Suitable for patients with mild bone deformities or angular deformities <15 degrees of fixed varus/valgus or flexion. Compatible with patients having prior implants (hip, ankle, soft tissue anchors). Requires identifiable anatomical landmarks on CT.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. To the left of the FDA logo is the symbol of the Department of Health & Human Services, which consists of a stylized human figure.
June 16, 2020
DePuy Orthopaedics, Inc. Melissa Cook Regulatory Affairs Specialist III 700 Orthopaedic Dr. Warsaw, Indiana 46582
Re: K193540
Trade/Device Name: TRUMATCH Personalized Solutions Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OOG Dated: May 15, 2020 Received: May 18, 2020
Dear Melissa Cook:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K193540
Device Name
TRUMATCH Personalized Solutions
#### Indications for Use (Describe)
The TruMatch Patient Specific Instruments are intent-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting. The anatomical landmarks necessary for the TruMatch Patient Specific Instruments must be present and identifiable on CT. The TruMatch Patient Specific Instruments are intended for use with Sigma Total Knee Implants and Attune Total Knee Implants and their cleared indications for use. The TruMatch Patient Specific Instruments are intended for single use only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10pt;">☑</span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <span style="font-size: 10pt;">□</span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
(As required by 21 CFR 807.92)
| Submitter Information | |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DePuy Orthopaedics, Inc. |
| Address | 700 Orthopaedic Dr.<br>Warsaw, IN 46582 |
| Phone number | 574-371-4906 |
| Establishment Registration<br>Number | 1818910 |
| Name of contact person | Melissa Cook |
| Date prepared | June 16, 2020 |
| Name of device | |
| Trade or proprietary name | DePuy TRUMATCH Personalized Solutions |
| Common or usual name | Total knee prosthesis instrumentation |
| Classification name | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented,<br>Polymer/Metal/Polymer |
| Class | II |
| Classification panel | 87 Orthopedics |
| Regulation | 21 CFR 888.3560 |
| Product Code(s) | JWH, OOG |
| Legally marketed device(s) to<br>which equivalence is claimed | TRUMATCH Personalized Solutions, K110397<br>Reference device: TRUMATCH Personalized Solutions, bundled<br>P830055/S217 |
| Reason for 510(k) submission | Device modification - This submission seeks clearance for changes to the<br>software that enable the TRUMATCH surgical plan and Guides to be<br>designed for a patient specific alignment surgical technique. |
| Device description | The TRUMATCH Personalized Solutions System is comprised of patient-<br>specific instruments and the software utilized to design the patient-specific<br>surgical plans and instruments. The predicate TRUMATCH System creates<br>a surgical plan based on a mechanical axis surgical technique. Changes to the<br>Planning software component allow the surgeon to select a mechanical axis<br>surgical technique or a patient-specific alignment surgical technique. Upon<br>approval of the surgical plan, the TRUMATCH Guides are manufactured<br>according to the plan. |
| Intended use of the device | Intended for patients with osteoarthritis who:<br>Meet the criteria for primary total knee replacement performed with a measured resection technique. Have mild bone deformities and/or angular deformities less than 15 degrees of fixed varus/valgus, or flexion. Have some previous implants, such as hip implants, ankle implants, and soft tissue anchors. Contralateral knee replacement is acceptable as long as the contralateral knee is flexed away (not within the same medial/lateral axis) from the knee of interest during the CT scan. |
| Indications for use | The TRUMATCH Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting. The anatomical landmarks necessary for the creation of the TRUMATCH Patient Specific Instruments must be present and identifiable on CT. The TRUMATCH Patient Specific Instruments are intended for use with Sigma Total Knee Implants and Attune Total Knee Implants and their cleared indications for use. The TRUMATCH Patient Specific Instruments are intended for single use only. |
{4}------------------------------------------------
{5}------------------------------------------------
| SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE | | |
|------------------------------------------------------------|--|--|
| COMPARED TO THE PREDICATE DEVICE | | |
| Characteristics | Subject Device:<br>TRUMATCH Personalized Solutions | Predicate Device: TRUMATCH<br>Personalized Solutions<br>(K110397) |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Intended Use | Intended for patients with osteoarthritis who:<br>Meet the criteria for primary total knee replacement performed with a measured resection technique. Have mild bone deformities and/or angular deformities less than 15 degrees of fixed varus/valgus, or flexion. Have some previous implants, such as hip implants, ankle implants, and soft tissue anchors. Contralateral knee replacement is acceptable as long as the contralateral knee is flexed away (not within the same medial/lateral axis) from the knee of interest during the CT scan. | Same |
| Material | Accura and Stainless Steel | Same |
| Image Inputs | CT Images | Same |
| Software<br>Packages | TRUMATCH 3.0 System software | TRUMATCH 2.5 System software |
| Size | Patient-specific | Same |
| Implant<br>Compatibility | Used with the implantation of the<br>ATTUNE and SIGMA Total Knee<br>Implants | Same |
| Packaging, Sterilization and Shelf Life | | |
| Sterile Method | Gamma | Same |
| Packaging | Polyurethane bag within a single blister<br>tray, sealed with Tyvek lid, placed in unit<br>carton | Same |
| Shelf Life | 185 Days | 120 days |
{6}------------------------------------------------
The subject TRUMATCH Personalized Solutions is the same as the predicate TRUMATCH Personalized Solutions (K110397) in intended use, design, and materials. The subject and predicate devices are intended to be used as patient-specific surgical instrumentation in total knee arthroplasty. The subject device utilizes the same software system components as the reference device TRUMATCH Personalized Solutions (bundled P830055/S217). The subject device is modified to enable either a mechanical axis alignment or patient specific alignment surgical technique.
# PERFORMANCE DATA
## SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE
The following tests were performed to demonstrate substantial equivalence of safety and efficacy with the predicate device:
- Model creation process verification for Patient Specific Alignment Guides ●
- Cadaver lab with simulated surgery for assessment of guide fit, implant alignment and size, and bone resection accuracy analysis
- Software verification testing ●
# SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
No clinical tests were conducted to demonstrate substantial equivalence.
# CONCLUSION DRAWN FROM NON-CLINICAL AND CLINICAL DATA
The subject device TRUMATCH Personalized Solutions System is substantially equivalent to the predicate TRUMATCH Personalized Solutions System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.