K193539 · Lifecell Corporation · MUU · May 28, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193539
Device Name
REVOLVE ENVI 600 Advanced Adipose System
Applicant
Lifecell Corporation
Product Code
MUU · General, Plastic Surgery
Decision Date
May 28, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
REVOLVE ENVI™ 600 System is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation. REVOLVE ENVI™ 600 System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired: plastic and reconstructive surgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
Device Story
REVOLVE ENVI 600 is a sterile, single-use suction lipoplasty system for harvesting, filtering, and transferring autologous adipose tissue. Used in healthcare facilities by surgeons; requires external vacuum/aspirator. Input: adipose tissue harvested via user-supplied tubing/cannula. Operation: tissue collected in canister with 200 µm mesh filter; manual stirring assembly mixes tissue with Lactated Ringer's solution for washing. Output: processed adipose tissue extracted via bottom port into syringe. Facilitates autologous fat grafting for body contouring; provides filtered tissue without additional manipulation.
Clinical Evidence
Bench testing only. Performance testing included gross/microscopic assessment of adipose samples, tissue viability (Lactate Dehydrogenase assessment), system fluid/vacuum leak testing, tubing tensile strength, hose collapse, and implosion testing per ISO 10079-1:2015. Biocompatibility testing performed per ISO 10993-1:2018 for limited contact (≤ 24 hours) externally communicating tissue devices. Sterilization validated per ISO 11137-2:2013 to SAL 10^-6.
Technological Characteristics
Single-use canister system with 200 µm mesh filter. Materials: unspecified polymers. Principle: mechanical filtration and manual washing. Energy: external vacuum source. Sterilization: gamma irradiation (ISO 11137-2:2013). Connectivity: none. Form factor: 600 mL capacity canister with manual stirring assembly.
Indications for Use
Indicated for patients undergoing aesthetic body contouring or soft tissue aspiration across plastic, reconstructive, gastrointestinal, general, orthopedic, gynecological, thoracic, and laparoscopic surgical specialties. Contraindicated for harvested fat that requires additional manipulation prior to re-implantation.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Predicate Devices
REVOLVE™ Advanced Adipose System (GID 700) (K120902)
Reference Devices
REVOLVE ENVI™ 600 Advanced Adipose System (K163647)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 28, 2020
LifeCell Corporation Bozhana Hegarty Manager, Regulatory Affairs One Millennium Way Branchburg, New Jersey 08876
Re: K193539
Trade/Device Name: REVOLVE ENVI 600 Advanced Adipose System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction Lipoplasty System Regulatory Class: Class II Product Code: MUU Dated: April 29, 2020 Received: April 30, 2020
Dear Bozhana Hegarty:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K193539
Device Name REVOLVE ENVI™ 600 Advanced Adipose System
#### Indications for Use (Describe)
REVOLVE ENVI™ 600 System is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator as a source of suction. If harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.
REVOLVE ENVI™ 600 System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired: plastic and reconstructive surgery, gastrointestinal and affiliated organ surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K193539 p. 1 of 4
#### 510(K) SUMMARY 1
## 1.1 SUBMITTER
#### Name and Address of Submitter:
LifeCell Corporation One Millennium Way Branchburg, NJ 08876 USA
# Contact:
Bozhana Hegarty Manager, Regulatory Affairs Phone: (908) 332-7484 Fax: (908) 325-0020 bozhana.hegarty@allergan.com
## Prepared by and Date:
Bozhana Hegarty April 29, 2020
# 1.2 SUBJECT DEVICE
Trade name: REVOLVE ENVI™ 600 Advanced Adipose System Common Name: Suction Lipoplasty System Classification Name: System, Suction, Lipoplasty (21 C.F.R. §878.5040) Device Class: Class II Product Code: MUU
# 1.3 PREDICATE DEVICE
Predicate Device: REVOLVE™ Advanced Adipose System (cleared as "GID 700" via K120902)
## Reference Device:
REVOLVE ENVI™ 600 Advanced Adipose System (cleared via K163647)
{4}------------------------------------------------
# 1.4 DEVICE DESCRIPTION
REVOLVE ENVI™ 600 Advanced Adipose System (REVOLVE ENVI 600 System) consists of a sterile, single-use canister and components intended to be used for harvesting, filtering and transferring of autologous adipose tissue. Such products have been classified by FDA as Class II devices under 21 CFR \$878.5040. Suction Lipoplasty System. and assigned product code MUU. The subject device should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. The adipose tissue is harvested from the patient using liposuction tubing and cannula that are supplied by the user or the institution. The harvested tissue is collected inside the device which contains an inner basket with 200 um mesh filter that can process up to 600 mL of collected tissue. A manual stirring assembly allows the user to mix the tissue and Lactated Ringer's during the washing step. The processed adipose tissue is removed from the device via an extraction port located on the bottom of the device using a syringe.
The REVOLVE ENVI 600 System is comprised of the following components, intended to be used only as a system:
- Canister (including mesh filter) o
- Syringe Adapter .
- 0 Temperature Strip
- Lactated Ringer's Tubing Set ●
- Vacuum Tubing Set ●
The device is offered in a single size, is packaged in a thermoformed tray and Tyvek® lidding and is sterilized via gamma irradiation. The device is a single-use device to be used in a healthcare facility.
## 1.5 INDICATIONS FOR USE
REVOLVE ENVI™ 600 System is used for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring. The system should be used with a legally marketed vacuum or aspirator apparatus as a source of suction. If harvested fat is to be reimplanted, the harvested fat is only to be used without any additional manipulation.
REVOLVE ENVI™ 600 System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired: plastic and reconstructive surgery, gastrointestinal and affiliated organ surgery, urological surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, and laparoscopic surgery.
{5}------------------------------------------------
# 1.6 TECHNOLOGICAL CHARACTERISTICS
The subject device, REVOLVE ENVI 600 System, and the predicate device, REVOLVE System, are both lipoplasty suction systems used in the aspiration, filtering and transferring of autologous adipose tissue. Both devices have the same intended use, principle of operation and consist of a canister, mesh filter, extraction port and components used to collect, filter and transfer the adipose tissue during lipoplasty procedures. A mesh filter is used to capture the collected tissue in both devices. The vacuum source is user or institution supplied. Various ports on the canister of both devices are provided to access the vacuum source (vacuum port), collect the adipose tissue (liposuction port), and receive the Lactated Ringer's solution during washing (Lactated Ringer's port). The subject device has a dedicated port to transfer the filtered tissue (extraction port), whereas the predicate device utilizes the liposuction port for extraction. Additionally, the subject device has a larger processing volume, new materials of construction and is sterilized via gamma irradiation, compared to electron beam irradiation for the predicate device.
The subject device. REVOLVE ENVI 600 System, is a modified version of the reference device. cleared with the same trade name under K163647. The differences between the subject and reference devices include design modifications of existing parts, addition of new components. and addition of a new supplier.
# 1.7 PERFORMANCE
REVOLVE ENVI 600 System has undergone appropriate performance and biocompatibility testing to ensure safety and effectiveness of the device. Stability testing was completed to support the proposed shelf life. The sterilization of the device is assured using a sterilization method validated in accordance with ISO 11137-2:2013 "Sterilization of health care products -Radiation - Part 2: Establishing the sterilization dose" to provide a Sterility Assurance Level (SAL) of 10-6.
#### Biocompatibility:
In accordance with ISO 10993-1:2018 = "Biological evaluation of medical devices = Part 1: Evaluation and testing within a risk management process," REVOLVE ENVI 600 System is classified as Externally Communicating Device, Tissue, Limited Contact (≤ 24 hours). The subject device has undergone appropriate biocompatibility assessments and testing, and data demonstrates that the device is biocompatible.
{6}------------------------------------------------
#### Bench Testing:
Bench testing was performed to support substantial equivalence to the predicate device. The table below provides a list of performance tests conducted on the subject device to demonstrate substantial equivalence.
| Bench Testing/Assessment | Applicable Standard |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Gross and microscopic assessments of<br>adipose samples | N/A |
| Adipose tissue viability (via Lactate<br>Dehydrogenase assessment) | N/A |
| System fluid leak | N/A |
| System vacuum leak | N/A |
| Tubing connection tensile strength | N/A |
| Hose collapse | ISO 10079-1:2015(E) Medical Suction<br>Equipment Part 1, Section 6.3.1 and<br>Annex A.4 |
| Implosion test | ISO 10079-1:2015(E) Medical Suction<br>Equipment Part 1, Section 6.1.3 and<br>Annex A.3 |
Substantial equivalence data demonstrate that REVOLVE ENVI 600 System is comparable to the predicate REVOLVE System in product performance characteristics relevant to the intended use of the device.
# 1.8 CONCLUSION
In summary, REVOLVE ENVI 600 System, subject of this 510(k), is substantially equivalent in its intended use, principle of operation and performance to its legally marketed predicate device, REVOLVE System (K120902).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.