The EVA15 Insufflator is intended for use in diagnostic and/or therapeutic endoscopic and laparoscopic procedures to distend the abdomen, colon or thoracic cavity with up to 15 mmHg pressure, by filling it with gas and to evacuate surgical smoke.
Device Story
EVA15 Insufflator is a microprocessor-controlled device for hospital use; provides cavity distension and surgical smoke evacuation during endoscopic/laparoscopic procedures. Inputs: compressed CO2 gas and air. Operation: digital pressure regulation system delivers CO2 to distend cavity; venturi method creates vacuum for smoke evacuation. Three modes: Fixed Flow (0-15 SLPM), Intermittent Pressure (capped at 12 SLPM), and Continuous Pressure (capped at 40 SLPM). Operated by clinicians via membrane panel; connects to patient via standard trocar luer connection. Output: controlled gas flow and smoke removal. Benefits: maintains surgical field visibility and cavity distension; facilitates instrument use. Device is reusable; includes sterile, single-use PVC/polyethylene tubeset.
Clinical Evidence
Bench testing only. Evaluated pressure and flow delivery accuracy under static and dynamic (simulated leak) conditions. Validated smoke evacuation performance and filter efficiency. Confirmed electrical safety (IEC 60601-1) and EMC (IEC 60601-1-2). Sterilization (EtO) and shelf-life (1 year) validated per ISO 11135, ISO 10993-7, and ASTM standards (D4169, F1886, F2096, F88, F1980). No clinical data provided.
Technological Characteristics
Microprocessor-controlled insufflator; dimensions 160x130x330mm; weight 5.5kg. Materials: PVC and polyethylene (tubeset). Power: AC 100-240V. Gas: compressed CO2 and air. Sensing/Actuation: digital pressure regulation and venturi vacuum. Connectivity: standalone. Sterilization: EtO (tubeset). Standards: IEC 60601-1, IEC 60601-1-8, IEC 60601-1-2, ISO 11135, ISO 10993-7.
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic endoscopic and laparoscopic procedures requiring cavity distension (abdomen, colon, or thoracic) and surgical smoke evacuation.
Regulatory Classification
Identification
A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
July 23, 2020
Palliare Ltd. % Paul Dryden Consultant Palliare Ltd c/o ProMedic. LLC 131 Bay Point Dr. NE St. Petersburg, FL 33704
Re: K193520
> Trade/Device Name: EVA15 Insufflator Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: June 23, 2020 Received: June 25, 2020
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jason R. Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
#### 510(k) Summary
| Submitter: | Palliare Ltd.<br>Galway Business Park, Dangan<br>Galway H91 P2DK, Ireland |
|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Official Contact: | John O'Dea, Ph.D., Director<br>Tel: +353-91-516362<br>Email: jodea@palliare.com |
| Contact Person name: | Paul Dryden<br>ProMedic, LLC<br>131 Bay Point Dr NE<br>St. Petersburg, FL 33704 |
| Date Prepared: | July 21, 2020 |
| Proprietary or Trade Name:<br>Regulation number:<br>Regulation Name:<br>Regulatory Class:<br>Product Code: | EVA15 Insufflator<br>21 CFR 884.1730<br>Laparoscopic Insufflator<br>Class II<br>HIF |
| Predicate Device: | K172516 - SurgiQuest AirSeal iFS System |
The predicate devices has no been subject to a design-related recall.
Device Description: The EVA15 insufflator is intended for use in diagnostic and/or therapeutic endoscopic and laparoscopic procedures to distend a cavity by filling it with gas and to evacuate surgical smoke. It is indicated to facilitate the use of various endoscopic and laparoscopic instruments by filling the abdominal or thoracic cavity or rectum with gas to distend it, and by evacuating surgical smoke. The EVA15 Insufflation is intended to use in hospital. It consists of the following major components: (1) a micro-processor-controlled insufflation and smoke evacuation unit and (2) a disposable tube set.
There are 3 operating modes:
- Flow Mode (1) delivers a fixed flow, settable between 0 and 15 standard liters per ● minute (SLPM)
- . Pressure Modes (2) -
- Intermittent Pressure Insufflation a pressure is targeted, but the flow delivered o in targeting that pressure is capped at 12 SLPM (i.e. no more than 12 SLPM will be delivered during intermittent pressure insufflation).
- Continuous Pressure Insufflation a pressure is targeted, but the flow delivered o in targeting that pressure is capped at 40 SLPM (i.e. no more than 40 SLPM will be delivered during continuous pressure insufflation).
The tubeset is a sterile, single-use product. The tubeset is made of PVC and polyethylene. The A VA 15 Insufflator is an active medical device, nonsterile and reusable and is intended to insufflate a body cavity up to 15mmHg and with up to 40 SLPM instantaneous flow. The EVA15 is powered by AC and uses compressed 50 psi CO2 and air gas supplies to supply the pneumatic
{4}------------------------------------------------
circuitry for insufflation and smoke evacuation, respectively.
Principle of Operation: The operating principle employs 2 methods.
A) A digital insufflation pressure regulation system using compressed CO2 gas to deliver CO2 into the patient cavity to be insufflated at the direction and control of the physician; and
B) The use of a venturi method to create a vacuum to evacuate any smoke created during the procedure.
### Indications for Use:
The EVA15 Insufflator is intended for use in diagnostic and/or therapeutic endoscopic and laparoscopic procedures to distend the abdomen, colon or thoracic cavity with up to 15 mmHg pressure, by filling it with gas and to evacuate surgical smoke.
Comparison of Technological Characteristics of the subject and predicate devices are presented in Table 1 below.
{5}------------------------------------------------
| Table 1 - Comparison - Subject vs. Predicate | | | |
|----------------------------------------------|--|--|--|
|----------------------------------------------|--|--|--|
| | Subject Device: EVA15 Insufflator<br>(K193520) | Predicate Device: SurgiQuest AirSeal iFS System<br>(K172516) | Comparison |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Manufacturer | Palliare Ltd. | SurgiQuest Inc. (Acquired by ConMed Co. in 2015) | |
| Classification | 21 C.F.R. § 884.1730 (Laparoscopic<br>Insufflator), Product Code HIF | 21 C.F.R. § 884.1730 (Laparoscopic Insufflator),<br>Product Code HIF, GCJ | Same |
| Fundamental<br>scientific<br>technology | Digital insufflation pressure<br>regulation system using compressed<br>CO2 gas. Venturi smoke evacuation. | Digital insufflation pressure regulation system using<br>compressed CO2 gas<br>Smoke evacuation using custom trocar generating<br>negative pressure | Similar |
| Patient connection | Standard Trocar luer connection | Customized trocar | Similar - each connects<br>to cavity using a trocar |
| Indications for Use | The EVA15 Insufflator is intended for<br>use in diagnostic and/or therapeutic<br>endoscopic and laparoscopic<br>procedures to distend the abdomen,<br>colon or thoracic cavity with up to 15<br>mmHg pressure, by filling it with gas<br>and to evacuate surgical smoke. | The SurgiQuest AirSeal® iFS System is intended for use<br>in diagnostic and/or therapeutic endoscopic procedures<br>to distend a cavity by filling it with gas, to establish and<br>maintain a path of entry for endoscopic instruments and<br>to evacuate surgical smoke.<br>It is indicated to facilitate the use of various<br>thoracoscopic and laparoscopic instruments by filling<br>the abdominal or thoracic cavity with gas to distend it,<br>by creating and maintaining a gas sealed obstruction-<br>free instrument path and by evacuating surgical smoke.<br>This instrument can also be used to insufflate the rectum<br>and colon to facilitate endoscopic observation, diagnosis<br>and treatment. The trocar of the AirSeal® iFS System is<br>indicated for use with or without visualization. | Similar except the<br>subject device does not<br>include a trocar. |
| | Subject Device: EVA15 Insufflator | Predicate Device: SurgiQuest AirSeal iFS System<br>(K172516) | Comparison |
| Gas Delivery<br>Modes | Fixed Flow<br>Intermittent Pressure (Standard)<br>Insufflation<br>Continuous Pressure Insufflation | Intermittent Pressure (Standard and Smoke Evac Modes)<br>Insufflation<br>Continuous Pressure Insufflation (Airseal Mode) | Similar<br>Subject device also has<br>a fixed flow mode |
| Smoke Evacuation | Available in all modes. Operates<br>continuously or may be activated<br>on/off using foot pedal. | Available in Airseal (continuous pressure) and Smoke<br>Evacuation modes. Operates continuously. | Similar |
| Flow Range | 0-40 SLPM | 0-40 SLPM | Same |
| Pressure Range | 7-15 mmHg | 5-20 mmHg | Similar<br>Subject device has a<br>smaller range |
| Accessories | Tubeset | Tubeset, Custom Trocar | Similar |
| Dimensions | 160x130x330mm | 420x220x470mm | Similar<br>Subject is smaller |
| Weight | 5.5kg | 26.0kg | Similar<br>Subject is smaller |
| Power Source | AC 100-240V | AC 100/115/230V | Similar |
| Tubeset<br>Sterilization | EtO | EtO | Same |
| User Interface | Membrane Panel | Touchscreen | Similar |
{6}------------------------------------------------
# Discussion of Differences
The differences in technological characteristics do not raise different questions of safety and effectiveness when compared to the predicate.
{7}------------------------------------------------
## Summary of Non-clinical Performance Testing
The subject devices comply with voluntary standards for electrical safety, and electromagnetic compatibility. The following data were provided in support of the substantial equivalence determination:
### Bench
- . Static condition
- Pressure and flow delivery accuracy; the device was demonstrated to meet o predefined acceptance criteria regarding pressure and flow
- Dynamic condition ●
- o Simulated leak; the device was demonstrated to meet predefined acceptance criteria under simulated leak conditions
- Smoke evacuation; the device was demonstrated to meet predefined acceptance o criteria regarding smoke evacuation
- . Efficiency of Smoke Evacuation Filter; the device was demonstrated to meet predefined acceptance criteria regarding the smoke evacuation filter
Software was validated per Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.
Electrical / EMC
- . Electrical Safety
- o IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) IEC 60601-1-8: 2012 o
The device was demonstrated to meet the requirements of the standard.
- EMC
- IEC 60601-1-2:2014 o
The device was demonstrated to meet the requirements of the standard.
### Sterility and Aging Effect
Sterilization of the tubeset is via EtO and was validated per ISO 11135:2014/Amd.1:2018 Annex E, Half cycle approach. In addition, testing to ISO 10993-7:2008/Amd 1:2019, ASTM D4169-16, ASTM F1886 (visual inspection), ASTM F2096 (leak detection), ASTM F88 (seal strength) and ASTM F-1980-16 (aging) supported a shelf-life of 1 year.
### Substantial Equivalence Conclusion
The performance testing described above demonstrate that the subject device is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.