K193507 · Merit Medical Systems, Inc. · MMX · Jan 17, 2020 · Cardiovascular
Device Facts
Record ID
K193507
Device Name
Merit ONE Snare System
Applicant
Merit Medical Systems, Inc.
Product Code
MMX · Cardiovascular
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ONE Snare endovascular snare system is intended for use in the coronary and peripheral vascular system or hollow viscous to retrieve and manipulate foreign objects. Retrieval and manipulation procedures include indwelling venous catheter repositioning, indwelling venous catheter fibrin sheath stripping, and central venous access venipuncture procedure assistance.
Device Story
Endovascular snare system designed for retrieval and manipulation of foreign objects in coronary/peripheral vascular systems or hollow viscous. System comprises nitinol core wire with tapered distal section, formed nitinol cable, gold-plated tungsten coil, and PTFE polymer sleeve; packaged with introducer, torque device, and guiding catheter. Used by clinicians in clinical settings. Device enables mechanical retrieval/manipulation of objects, repositioning of indwelling venous catheters, and fibrin sheath stripping. Output is physical manipulation of objects within the vasculature. Benefits include minimally invasive removal of foreign bodies and assistance in venous access procedures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included snare/catheter size designation, surface inspection, dimensional inspection (OD, ID, tip angle), and simulated use design validation. All results met predetermined acceptance criteria and were comparable to the predicate device.
Technological Characteristics
Nitinol core wire, gold-plated tungsten coil, PTFE polymer sleeve. System includes introducer, torque device, and guiding catheter. Dimensions: 4F, 80cm length, 15mm loop diameter; catheter features 15-degree angle 1cm from distal tip. Sterilization via ethylene oxide. Standards: ISO 11070, ISO 11135, EN 556, ISO 10555-1.
Indications for Use
Indicated for use in coronary and peripheral vascular systems or hollow viscous for retrieval and manipulation of foreign objects, including indwelling venous catheter repositioning, fibrin sheath stripping, and central venous access venipuncture assistance.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Merit One Snare Endovascular Snare System (K122088)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 17, 2020
Merit Medical Systems, Inc. Michael O'Sullivan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, IE Galway
Re: K193507
Trade/Device Name: ONE Snare™ endovascular snare system Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: December 13, 2019 Received: December 18, 2019
# Dear Michael O'Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K193507
#### Device Name
ONE Snare endovascular suare system
## Indications for Use (Describe)
The ONE Snare endovascular snare system is intended for use in the coronary and peripheral vascular system or hollow viscous to retneve and manipulate foreign objects. Retrieval and manipulation procedures include indivelling venous catheter repositioning, indwelling venous catheter fibrin sheath stripping, and central venous access veripouncture procedure assistance.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 70 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
#
| General<br>Provisions | Submitter Name: | Merit Medical Systems, Inc. |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| | Address: | 1600 West Merit Parkway<br>South Jordan, UT 84095 |
| | Telephone Number: | (+353) 91 703700 (3061) |
| | Fax Number: | (+353) 91 680 104 |
| | Contact Person: | Mark Mullaney |
| | Registration Number: | 1721504 |
| | Correspondent Name: | Merit Medical Ireland Ltd. |
| | Address: | Parkmore Business Park West<br>Galway, Ireland |
| | Telephone Number: | (+353) 91 703700 (3224) |
| | Fax Number: | (+353) 91 680 104 |
| | Contact Person: | Michael O'Sullivan |
| | Date of Preparation: | Jan 17th 2020 |
| | Registration Number: | 9616662 |
| | 510(k) Number: | K193507 |
| Subject<br>Device | Trade Name: | ONE Snare® |
| | Common/Usual Name: | One Snare Endovascular Snare System |
| | Classification Name: | 21 CFR 870.5150<br>Device percutaneous retrieval |
| Predicate<br>Device | | Premarket Notification Predicate: (Primary Predicate) |
| | Trade Name: | Merit One Snare Endovascular Snare System |
| | Classification Name: | 21 CFR 870.5150 |
| | Premarket Notification: | K122088 |
| | Manufacturer: | Merit Medical Systems, Inc. |
| Classification | | Class II |
| | | 21 CFR 870.5150 Device percutaneous retrieval |
| | | FDA Product Code: MMX |
| | | Review Panel: Division of Cardiovascular Devices |
| Intended Use | The ONE Snare endovascular snare system is intended for use in the<br>coronary and peripheral vascular system or hollow viscous to retrieve and<br>manipulate foreign objects. Retrieval and manipulation procedures include<br>indwelling venous catheter repositioning, indwelling venous catheter fibrin<br>sheath stripping, and central venous access venipuncture procedure<br>assistance. | |
| | Special Premarket Notification 510(k) | |
| | sheath stripping, and central venous access venipuncture procedure assistance. | |
| Device<br>Description | The ONE Snare Endovascular Snare System is a snare system that is designed for use in the coronary and peripheral vascular system or hollow viscous to retrieve and manipulate foreign objects. The One Snare wire consists of a Nitinol core with a ground tapered distal section, a formed nitinol cable covered with a gold-plated tungsten coil. The formed cable with coil is attached to the core wire and then covered with a PTFE polymer sleeve. The wires are packaged with an introducer, torque device and guiding catheter – this makes up the “One Snare Endovascular Snare System”. | |
| Comparison<br>to Predicate | The proposed new configuration will consist of the following snare with a 15mm loop diameter which will be 80cm in length. The loop diameter and length fall within the existing approved range for the ONE Snare (Snare loop diameter ranges from 5mm-35mm and Snare length ranges from 65cm-200cm.) Snare dimensions are as follows: 4F, 80cm, 15mm loop. catheter modified with a 15+/-5 degree angle at the catheter tip. The angle will be 1cm from the distal tip. Catheter dimensions: 4F, 65cm, Marker-Banded, with 1cm of the distal tip being angled 15+/- 5 degrees. None of the above changes would be deemed sufficient to affect the technological characteristics of the device and thus the subject One Snare Endovascular Snare System is substantially equivalent to Predicate Device, the Merit ONE Snare Endovascular Snare K122088.<br>The Indications for use of the subject wire are identical to Predicate Device. | |
| | No performance standards have been established under section 514 of the Food, Drug and Cosmetic Act for these devices. However, a battery of tests performed on the One Snare Endovascular Snare System were designed to demonstrate that the device meets critical design specifications as well as clinical performance attributes for its intended use. | |
| | Where appropriate, the tests were based on the requirements of the following documents: | |
| Safety &<br>Performance<br>Tests | ISO 11070 - Sterile, single-use intravascular catheter introducers ISO 11135 - Sterilization of health care products- Ethylene oxide- Requirements for development, validation and routine control of a sterilization process for medical devices processes EN556 - Sterilization of medical devices - Requirements for terminally sterilized medical devices to be labeled "Sterile" ISO 10555-1 - Intravascular Catheters – Sterile and Single-Use Catheters - Part 1: General Requirements | |
| | The ONE Snare Endovascular Snare System was compared to the predicate device for various performance attributes that support substantial equivalence of the device. The difference in assembly between | |
| | • Snare Size Designation<br>• Catheter Size Designation<br>• Catheter Surface Inspection<br>• Catheter OD Inspection<br>• Catheter Tip Length of Angle<br>• Catheter Angle<br>• Catheter Tip ID Inspection<br>• Simulated Use (design validation) | |
| | All test results were comparable to the predicate devices and the subject<br>One Snare Endovascular Snare System met the predeterminded<br>acceptance criteria. This has demonstrated that the subject device is | |
| Summary of<br>Substantial<br>Equivalence | Based on the Indications for Use, design, safety and performance testing,<br>the subject One Snare Endovascular Snare System is substantially<br>equivalent to the cleared Predicate Device, the Merit One Snare<br>Endovascular Snare System [K1220881]. | |
{4}------------------------------------------------
Merit Medical Systems, Inc.
Merit One Snare Endovascular Snare System
Special Premarket Notification 510(k)
{5}------------------------------------------------
the modified device and the cleared device, Merit One Snare Endovascular Snare System [K122088] has raised no new issues with respect to the device.
The following is a list of all testing that was successfully completed:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.