← Product Code [KWQ](/productcode/KWQ) · K193494

# Origin™ Anterior Cervical Plate System (K193494)

_Orthopaedic & Spine Development · KWQ · Apr 6, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193494

## Device Facts

- **Applicant:** Orthopaedic & Spine Development
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Apr 6, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Origin Anterior Cervical Plate System is intended for anterior cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

## Device Story

Origin Anterior Cervical Plate System is a spinal fixation implant; consists of titanium plates and bone screws in multiple sizes. Used by surgeons in clinical settings to stabilize cervical spine (C2-T1) via anterior approach. Plate features holes to receive and secure screws to bone. Device provides mechanical stabilization to facilitate fusion; addresses conditions like DDD, trauma, and deformity. Provided non-sterile for single use.

## Clinical Evidence

Bench testing only. Testing included static and dynamic axial compression bending and static torsion testing in accordance with FDA guidance for spinal systems.

## Technological Characteristics

Titanium plate and bone screw system. Non-sterile, single-use. Mechanical fixation via screw-to-plate interface. Dimensions vary to accommodate anatomical requirements.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- TriforeTM Cervical Plating System ([K171112](/device/K171112.md))
- TERRACETM Anterior Cervical Plate System ([K163104](/device/K163104.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the date April 6, 2020. The text is written in a simple, sans-serif font. The date is written in the standard US format, with the month first, followed by the day and year. The text is black and the background is white.

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Orthopaedic & Spine Development % Mr. Roger White President Phiama. Inc. 236 McKinley Park Lane Louisville. Colorado 80027

Re: K193494

Trade/Device Name: Origin™ Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: March 18, 2020 Received: March 20, 2020

Dear Mr. Roger White:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Colin O'Neill, M.B.E. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

#### K193494

Device Name Origin™ Anterior Cervical Plate System

#### Indications for Use (Describe)

The Origin Anterior Cervical Plate System is interior cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvature (i.e., scoliosis, kyphosis), tumor, pseudoarthrosis, and failed previous fusion.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) | <input checked="true" type="checkbox"/> |
|------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)                                                          | <input type="checkbox"/>                |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Origin™ Anterior Cervical Plate System

## 1. Submitter Information

| Submitter:     | Orthopaedic & Spine Development (OSD)                           |
|----------------|-----------------------------------------------------------------|
| Address:       | Technopôle Agroparc – BP 11211<br>84911 Avignon Cedex 9, France |
| Telephone:     | +33 (0)4-32-74-01-10                                            |
| Telefax:       | +33 (0)4-90-80-02-39                                            |
| Contact:       | Roger White<br>rwhite@phiama.com<br>303-550-2451                |
| Date Prepared: | December 16, 2019                                               |

## 2. Device Information

| Trade Name:          | Origin™ Anterior Cervical Plate                 |
|----------------------|-------------------------------------------------|
| Common Name:         | System Anterior Cervical Fixation               |
| Classification:      | System Class II per 21 CFR 888.3060             |
| Classification Name: | Appliance, Fixation, Spinal Intervertebral Body |
| Product Code:        | KWQ                                             |

## 3. Purpose of Submission

The purpose of this submission is to gain clearance for a new Anterior Cervical Fixation System.

#### 4. Predicate Device Information

The Origin Anterior Cervical Plate System described in this submission is substantially equivalent to the following predicates:

| Predicate Device                            | Manufacturer          | 510(k) No. |
|---------------------------------------------|-----------------------|------------|
| TriforeTM Cervical Plating<br>System        | NeuroStructures, Inc. | K171112    |
| TERRACETM Anterior Cervical<br>Plate System | CoreLink              | K163104    |

{4}------------------------------------------------

#### 5. Device Description

Origin Anterior Cervical Plate System is an anterior cervical fixation system comprised of a titanium plate, and bone screws, both in multiple sizes. The Plate features multiple holes designed to receive and secure bone screws. The Plate System is provided non-sterile and is intended for single use.

#### 6. Intended Use

The Origin Anterior Cervical Plate System is intended for anterior cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) defined as neck pain of discogenic origin with degeneration of the disc confirmed by history or radiographic studies, spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvature (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

## 7. Comparison of Technological Characteristics

The substantial equivalence of the Origin Anterior Cervical Plate System to the predicate devices is shown by the similarity in intended use, indications for use, materials, implant dimensions, and performance.

#### 8. Performance Data

Origin Anterior Cervical Plate System was tested in accordance with the recommendations from the FDA Guidance for Spinal Systems 510(K) submission. This testing included static and dynamic axial compression bending and static torsion testing. A review of the performance data demonstrates substantial equivalence to the predicate devices.

#### 9. Conclusion

Based on the indications for use, technological characteristics, biocompatibility, performance results, and comparison to the predicates, the Origin Anterior Cervical Plate System has been shown to be substantially equivalent to the predicate devices identified in this submission and does not present any new issues of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K193494](https://fda-staging.innolitics.com/device/K193494)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).