K193490 · First Light Diagnostics, Inc. · LLH · Jul 8, 2021 · Microbiology
Device Facts
Record ID
K193490
Device Name
SensiTox C. difficile Toxin Test
Applicant
First Light Diagnostics, Inc.
Product Code
LLH · Microbiology
Decision Date
Jul 8, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2660
Device Class
Class 1
Indications for Use
The SensiTox C. difficile Toxin Test is an immunofluorescence assay intended for the qualitative detection of Clostridioides difficile toxins A and/or B in human stool specimens. The test is intended as an aid in the diagnosis of C. difficile infection (CDI) in patients exhibiting symptoms of CDI. Negative results do not preclude toxigenic C. difficile infection. The SensiTox C. difficile Toxin Test should not be used as the sole basis for treatment or other management decisions. The test can only be used with the MultiPath platform.
Device Story
SensiTox C. difficile Toxin Test is an immunofluorescence assay for qualitative detection of C. difficile toxins A and B in human stool. Used with proprietary MultiPath Analyzer in clinical laboratories. Stool samples are processed via spin column; filtrate loaded into single-use fluidic cartridge. Analyzer heats sample to 35°C, splits into 6 aliquots, and mixes with lyophilized fluorescent/magnetic antibody-conjugated beads. Toxins bind to beads; magnetic station uses permanent magnets to draw complexes through a dye-cushion layer to an imaging surface. Non-magnified digital imaging quantifies captured fluorescent particles. Software interprets results as toxin detected or not detected. Results displayed on touchscreen; aid in CDI diagnosis. Benefits include rapid, automated detection of toxigenic C. difficile.
Clinical Evidence
Prospective clinical study at three US sites using 1,046 de-identified, unpreserved stool specimens compared to cellular cytotoxicity neutralization assay (CCNA). Sensitivity: 90.6% (95% CI: 83.1% - 95.0%); Specificity: 95.7% (95% CI: 94.2% - 96.8%). Positive Predictive Value: 68.0%; Negative Predictive Value: 99.0%.
Technological Characteristics
Immunofluorescence assay using proprietary MultiPath detection technology. Single-use fluidic cartridge with lyophilized antibody-conjugated fluorescent and magnetic beads. Heating element (35°C), permanent magnets for particle capture, and non-magnified digital imaging. Standalone MultiPath Analyzer platform. Software-based interpretation of digital images.
Indications for Use
Indicated for qualitative detection of Clostridioides difficile toxins A and/or B in human stool specimens from patients exhibiting symptoms of C. difficile infection (CDI) as an aid in diagnosis.
Regulatory Classification
Identification
A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY AND INSTRUMENT
## I Background Information:
A 510(k) Number
K193490
B Applicant
First Light Diagnostics, Inc.
C Proprietary and Established Names
SensiTox C. difficile Toxin Test
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| LLH | Class I | 21 CFR 866.2660 - Microorganism Differentiation And Identification Device | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the SensiTox C. difficile Toxin Test.
B Measurand:
Clostridioides (Clostridium) difficile toxins A and B
C Type of Test:
Automated qualitative immunoassay
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The SensiTox *C. difficile* Toxin Test in an immunofluorescence assay intended for the qualitative detection of *Clostridioides difficile* toxins A and/or B in human stool specimens. The test is intended as an aid in the diagnosis of *C. difficile* infection (CDI) in patients exhibiting symptoms of CDI. Negative results do not preclude toxigenic *C. difficile* infection. The SensiTox *C. difficile* Toxin Test should not be used as the sole basis for treatment or other management decisions. The test can only be used with the MultiPath platform.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
MultiPath Analyzer
IV Device/System Characteristics:
A Device Description:
The SensiTox *C. difficile* Toxin Test is a cartridge-based immunoassay intended to detect *C. difficile* toxin A and/or toxin B in stool samples from patients suspected of *C. difficile* infection (CDI). Each single-use test cartridge contains the lyophilized reagents required to perform three measurements each for *C. difficile* toxin A and toxin B: Total Signal Measurement, Neutralization Measurement, and Positive Control Measurement. The test kit also includes the following components used to prepare specimens for testing: *C. difficile* Stool Specimen Diluent, *C. difficile* Protease Inhibitor reagent, and Spin Columns for stool specimen manual filtration.
The assay is performed on the MultiPath Analyzer, which is an automated benchtop laboratory instrument intended for use by trained laboratory technicians. The Analyzer includes the following functionality: a barcode scanner for test-specific and specimen-specific information, a pneumatic system that drives fluidics within the test cartridge, a magnetic station that deposits magnetically-tagged fluorescent targeted molecules onto the cartridge imaging surface, an optical system to detect deposited fluorescent particles, and automated software to analyze the test measurement and interpret the results based on predefined signal measurement thresholds for Toxin A and Toxin B total signals and neutralization measurements. The Analyzer uses a touch screen display User Interface to convey test information, status, instructions and user input. Up to five cartridges can be loaded onto the Analyzer at one time.
K193490 - Page 2 of 13
{2}
K193490 - Page 3 of 13
# B Principle of Operation:
The SensiTox C. difficile Toxin Test uses separate monoclonal antibodies that are conjugated to magnetic particles or fluorescent particles to capture and detect antigen, respectively. Labeled antibody-antigen complexes are deposited onto an imaging surface, and the MultiPath Analyzer uses digital imaging to automatically detect and interpret the assay signal.
C Instrument Description Information:
| Modes of Operation | Yes | No |
| --- | --- | --- |
| Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device? | ☑ | ☐ |
| Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission? | ☐ | ☑ |
| Software | | |
| FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types. | ☑ | ☐ |
1. Instrument Name:
MultiPath Analyzer
2. Specimen Identification:
The MultiPath Analyzer scans the two barcodes on the Cartridge: a manufacturer-provided barcode that identifies the test to be performed, along with the reagent lot information and a laboratory-generated barcode affixed by the user to identify the patient specimen.
3. Specimen Sampling and Handling:
Stool specimens are manually diluted into corresponding Specimen Diluent tubes immediately before testing. The diluted specimen is added to a Spin Column tube labeled with the specimen identifier and centrifuged. The eluate is transferred to the Specimen Well of the Cartridge which is then placed in a multi-Cartridge rack that is loaded into the MultiPath Analyzer.
4. Calibration:
The MultiPath Analyzer is calibrated by the manufacturer. Yearly maintenance is performed by qualified service personnel to clean, calibrate, test and verify the functionality of the device.
5. Quality Control:
Internal Controls
The C. difficile Cartridge contains two Internal Positive Control wells, one specific for Toxin A (5x LoD) and one specific for Toxin B (10x LoD), that ensure the sample matrix does not
{3}
negatively impact the performance of the test. An invalid result will be reported if the Internal Positive Control generates a signal below its defined threshold.
## External Controls
The SensiTox C. difficile External Controls Kit containing positive and negative controls is provided separately. The positive control consists of lyophilized purified Toxin A and Toxin B in a buffered solution and the negative control consists of buffered solution. The External Controls are used periodically to verify performance of the reagents and the MultiPath Analyzer. It is recommended that External Controls be tested at least once every 30 days, when a new reagent lot is received in the laboratory, and when the MultiPath Analyzer undergoes installation, repair or maintenance procedures. External Controls should be tested in accordance with all institutional, local, state, federal, and accrediting agency guidelines.
## V Substantial Equivalence Information:
A Predicate Device Name(s):
Immunocard Toxins A & B, Model 712050
B Predicate 510(k) Number(s):
K041003
C Comparison with Predicate(s):
| Device & Predicate Device(s): | DEVICE K193490 | PREDICATE K041003 |
| --- | --- | --- |
| Device Trade Name | SensiTox C. difficile Toxin Test | ImmunoCard Toxins A & B |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | The SensiTox C. difficile Toxin Test is an immunofluorescence assay intended for the qualitative detection of Clostridioides difficile toxins A and/or B in human stool specimens. The test is intended as an aid in the diagnosis of C. difficile infection (CDI) in patients exhibiting symptoms of CDI. Negative results do not preclude toxigenic C. difficile infection. The SensiTox C. difficile Toxin Test should not be used as the sole basis for | ImmunoCard Toxins A & B is a rapid, qualitative, horizontal-flow enzyme immunoassay (EIA) for detecting Clostridium difficile toxins A and B in human stool. This assay is used as an aid in the diagnosis of C. difficile-associated disease. |
K193490 - Page 4 of 13
{4}
| Device & Predicate Device(s): | DEVICE K193490 | PREDICATE K041003 |
| --- | --- | --- |
| | treatment or other management decisions. The test can only be used with the MultiPath platform. | |
| Analyte | Same | C. difficile toxins A and B |
| Specimen type | Same | Human Stool |
| Interpretation | Same | Qualitative |
| General Device Characteristic Differences | | |
| Test format | Cartridge-based immunofluorescence immunoassay with automated interpretation | Lateral flow enzyme immunoassay with visual interpretation |
| Antibodies | Capture and Detection: Mouse monoclonal anti-toxin A and B antibodies | Capture: Monoclonal anti-toxin A and goat polyclonal anti-toxin B antibodies Detection: Goat polyclonal anti-toxin A and B antibodies |
| Automated | Yes | No |
| Instrument | Multipath Analyzer | None |
VI Standards/Guidance Documents Referenced:
- IEC 60601-1-2:2014 General Requirements for Basic Safety and Essential performance – Collateral Standard: Electromagnetic Disturbances – Requirements and Tests
- ANSI/AAMI/IEC 62304:2006 & A1:2016 (Consolidated Text): Medical Device Software - Software Life Cycle Processes
- ISO 14971 (3rd Edition): Medical devices — Application of risk management to medical devices
- ISO/IEC 24778:2008: Information technology — Automatic identification and data capture techniques — Aztec Code bar code symbology specification
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
K193490 - Page 5 of 13
{5}
The reproducibility of the SensiTox C. difficile Toxin Test was evaluated at three sites over five days by two operators each day. The sample panels consisted of Stool Specimen Diluent aliquots spiked with both Toxins A and B at the following concentrations: low positive (1-2x LoD), moderate positive (2-4x LoD), high positive (5-8x LoD), and negative (unspiked). The panels were randomized and masked and then distributed to each site. Prior to testing, each operator mixed the designated sample with pooled stool matrix that was prescreened and known to be negative for C. difficile Toxins A and B. The samples were processed according to the labelled test procedure and tested with the SensiTox C. difficile Toxin Test in triplicate for a total of 90 observations per panel member. The qualitative test results for detecting Toxin A (Table 1), Toxin B (Table 2) and each panel member (i.e., detection of Toxin A or B, Table 3. Reproducibility – Table 3) are summarized below.
Overall, 16/360 (4.4%) samples initially produced invalid results in the study. All samples were valid upon retest following labeled instructions. All high positive panel members produced the expected positive results for both toxin A and B (90/90, 100%). For the moderate positive panel members, one sample produced false negative results for both toxin A and B at one site (89/90, 98.9%). Also, one negative sample at the same site produced false positive results for both toxin A and B (89/90, 98.9%). For the low positive panel members, one sample produced a false negative result for toxin A at one site (89/90, 98.9%), and all sites produced the expected positive result for toxin B (90/90, 100%).
Table 1. Reproducibility - Toxin A
| Toxin A Panel | Site 1 | | Site 2 | | Site 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | # | % | # | % | # | % | # | % |
| Negative | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| Low | 30/30 | 100% | 29/30 | 96.7% | 30/30 | 100% | 89/90 | 98.9% |
| Moderate | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| High | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
Table 2. Reproducibility - Toxin B
| Toxin B Panel | Site 1 | | Site 2 | | Site 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | # | % | # | % | # | % | # | % |
| Negative | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| Low | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
| Moderate | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| High | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
Table 3. Reproducibility - Per Panel Member
| Toxin B Panel | Site 1 | | Site 2 | | Site 3 | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | # | % | # | % | # | % | # | % |
| Negative | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| Low | 30/30 | 100% | 29/30 | 96.7% | 30/30 | 100% | 89/90 | 98.9% |
| Moderate | 30/30 | 100% | 30/30 | 100% | 29/30 | 96.7% | 89/90 | 98.9% |
| High | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% |
2. Linearity:
Not applicable
K193490 - Page 6 of 13
{6}
3. Analytical Specificity/Interference:
# Interfering substances
Endogenous and exogenous substances commonly found in stool were evaluated for their potential effect on the performance of the SensiTox C. difficile Toxin Test. Negative pooled stool and contrived positive pooled stool containing $15\mathrm{ng / mL}$ of Toxin A (4x LoD) and 300 $\mathrm{pg / mL}$ of Toxin B (6x LoD) were tested unspiked (positive and negative control) as well as spiked with the potential interferents shown in Table 4 and tested in triplicate.
All negative samples produced the expected negative results in the presence of each substance. One initial invalid result was observed in the presence of Metronidazole at 40 mg/mL (1/115, 0.9% initial invalid rate); however, the sample produced the valid expected result upon repeat testing (0% final invalid rate). Out of six sample replicates tested with Vancomycin at 50 mg/mL, one false negative Toxin A result and one false negative Toxin B result were observed. No assay interference was observed in the presence of Vancomycin at 40 mg/mL or in the presence of the other substances at the concentrations tested.
Table 4. Potential Interfering Substances
| Potential Interfering Substance | Concentration Tested |
| --- | --- |
| Nystatin | 500 U/mL (5% w/v) |
| Barium Sulphate | 50 mg/mL (5% w/v) |
| Hydrocortisone | 0.5 mg/mL (5% w/v) |
| Phenylephrine (Preparation H) | 0.1 mg/mL (5% w/v) |
| Calcium Carbonate (Tums) | 10.4 mg/mL (5% w/v) |
| Aluminum Hydroxide / Magnesium Hydroxide (Sunmark antacid) | 1 mg/mL (5% v/v) |
| Loperamide Hydrochloride (Imodium) | 3.3 μg/mL (5% v/v) |
| Bismuth Subsalicylate (Pepto Bismol) | 0.2 mg/mL (5% v/v) |
| Sennosides (Senokot) | 0.6 mg/mL (5% w/v) |
| Metronidazole in DMSO | 50 mg/mL (5% w/v) |
| Vancomycin | 40 mg/mL (4% w/v) |
| Mucin | 50 mg/mL (5% w/v) |
| DMSO | 10% v/v |
| Whole Blood | 40% v/v |
# Analytical Specificity
The analytical specificity of the SensiTox C. difficile Toxin Test was evaluated by testing a panel of microorganisms commonly found in stool samples (Table 5). Cultured organisms were spiked into negative pooled stool or spiked into contrived positive stool containing $15\mathrm{ng / mL}$ Toxin A (4x LoD) and $300~\mathrm{pg / mL}$ Toxin B (6x LoD). Bacteria and yeast were tested at $1\times 10^{6}$ CFU/mL and viruses were tested at $1\times 10^{5}$ PFU/mL, except for Norovirus $(7\times 10^{4}$ PFU/mL) and Echovirus $(4.1\times 10^{4}$ PFU/mL). All microorganisms were tested in triplicate with the SensiTox C. difficile Toxin Test. No false positive results were observed. Initial invalid results were obtained for four samples (4/135, $3.0\%$ initial invalid rate), and all samples produced the expected valid result upon repeat (0% final invalid rate).
Table 5. Potential Interfering Microbes
| Microorganism | Microorganism |
| --- | --- |
| Adenovirus | Enterovirus |
K193490 - Page 7 of 13
{7}
| Microorganism | Microorganism |
| --- | --- |
| Aeromonas hydrophila | Escherichia coli |
| Bacillus cereus | Escherichia coli sero:0157 |
| Bacillus subtilis | Escherichia coli type 026:H4 |
| Bacteroides fragilis | Helicobacter pylori |
| Campylobacter jejuni | Klebsiella oxytoca |
| Campylobacter coli | Norovirus |
| Candida albicans | Peptostreptococcus anaerobius |
| Clostridium difficile (non-toxigenic) | Proteus vulgaris |
| Clostridium haemolyticum | Pseudomonas aeruginosa |
| Clostridium novyi | Rotavirus |
| Clostridium perfringens | Salmonella enterica (typhimurium) |
| Clostridium septicum | Serratia liquefaciens |
| Clostridium sordellii¹ | Shigella dysenteriae |
| Clostridium sporogenes | Shigella flexneri |
| Coxsackie virus | Shigella sonnei |
| Cytomegalovirus | Staphylococcus aureus |
| Echovirus | Staphylococcus epidermidis |
| Enterobacter aerogenes | Vibrio cholera |
| Enterobacter cloacae | Vibrio parahaemolyticus |
| Enterococcus faecalis | |
¹ The potential for purified Clostridium sordellii toxin to cross-react was not evaluated. It is unknown if C. sordellii toxin concentration in the 10⁶ CFU/mL preparation that was tested falls below the limit of detection for the SensiTox C. difficile Toxin Test.
## Microbial Interference
Microbial interference was evaluated for the SensiTox C. difficile Toxin Test by using the same panel of microorganisms from in the analytical specificity study (Table 5) spiked into contrived positive stool containing 15 ng/mL Toxin A (4x LoD) and 300 pg/mL Toxin B (6x LoD). Toxins A and B were detected in all samples indicating lack of interference. However, one initial invalid result was observed (1/135, 0.7% initial invalid rate), and the sample produced the expected valid result upon retest per the study protocol and assay instructions (0% final invalid rate).
## Hook Effect
A study was conducted with the SensiTox C. difficile Toxin Test to determine if high concentrations of target C. difficile toxins interfere with the test result (hook effect). Diluted toxins (Toxin A dilutions from 50 μg/mL – 2.5 ng/mL, Toxin B dilutions from 1.3 μg/mL – 66 pg/mL) were spiked into pooled negative stool and three replicates per dilution were tested with the SensiTox C. difficile Toxin Test. No false negative results were observed for either toxin at any concentration tested.
## 4. Assay Reportable Range:
Not applicable
K193490 - Page 8 of 13
{8}
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
## Quality Control
The SensiTox C. difficile Toxin Test uses internal positive controls (individual wells specific to each toxin and contained within the test cartridge) as well as external positive and negative controls. Quality control procedures monitor the accuracy and precision of the analytical process. It is recommended that External Controls be tested at least once every 30 days, when a new reagent lot is received in the laboratory, and when the MultiPath Analyzer undergoes installation, repair or maintenance procedures. External controls are required but not provided with the SensiTox C. difficile Toxin Test.
A minimum of one positive and one negative external control were used each day prior to testing clinical specimens or performing analytical studies. Controls were processed in the same manner as clinical specimens and results interpreted in accordance with the SensiTox C. difficile Toxin Test software assay parameters. To pass, a positive control must detect both Toxins A and B while the negative control must generate a negative result for both, according to the instructions for use of the external control kit. During the clinical study, a total of 343 external controls were run (169 positive and 174 negative controls). Of the 343 controls tested, 10 generated invalid results (2.9% invalid rate). Of the 333 valid QC results, two were QC failures (i.e., incorrect result). Per the study protocol, when this occurred another control was tested and passed which made the clinical specimen data acceptable for analysis on those days.
## MultiPath Analyzer
Additional assay and instrument parameters are monitored during a test run on the MultiPath Analyzer. These include sample input errors, instrument cartridge processing errors, and instrument system messages.
## Specimen Stability
A study was conducted to evaluate the stability of stool samples that have been stored refrigerated at $2 - 8^{\circ}\mathrm{C}$ for analysis with the SensiTox C. difficile Toxin Test. Seven individual stool samples and one pool of two stool samples were collected and confirmed negative for C. difficile A and B toxins with the SensiTox C. difficile Toxin Test. Sample aliquots were spiked with each target toxin individually at $15\mathrm{ng / mL}$ (4x LoD) for Toxin A and $250~\mathrm{pg / mL}$ (5x LoD) for Toxin B. The samples were stored refrigerated and a minimum of three replicates each were tested with the SensiTox C. difficile Toxin Test at different time points. Spiked samples were tested after 0, 24, 48, 72, 96, and 100 hours of storage, and unspiked negative samples were tested after 0, 96, and 100 hours of storage.
Although there were trends of decreasing assay toxin signal average values and increasing neutralization ratio average values in spiked samples over time, the values were well within the cutoff threshold values to support a 72 hours stability claim when stored refrigerated at $2 - 8^{\circ}\mathrm{C}$.
6. Detection Limit and Analytical Reactivity:
## Limit of Detection
The limit of detection (LoD) for Toxins A and B was determined by spiking known concentrations of purified Toxin A and B into negative pooled stool and testing with the
K193490 - Page 9 of 13
{9}
SensiTox C. difficile Toxin Test. Concentrations ranging from 1.5 to 3.5 ng/mL for Toxin A and 30 to 100 pg/mL for Toxin B were tested using three reagent lots across two analyzers. Twenty replicates per analyte dilution and per reagent lot were tested for a total of 60 replicates per analyte concentration. The LoD for each toxin was defined as the lowest concentration that generated a Detected result ≥ 95% of the time. The LoD for Toxin A was 3.5 ng/mL and the LoD for Toxin B was 50 pg/mL.
## Analytical reactivity (Inclusivity)
A study was conducted to demonstrate reactivity of the SensiTox C. difficile Toxin Test with C. difficile A and B toxins sourced from a diverse collection of clinically relevant C. difficile strain ribotypes and toxinotypes (Table 6). Toxin A was purified from six ribotypes and tested at 15 ng/mL (4x LoD). Toxin B was purified from eight ribotypes and tested at 300 pg/mL (6x LoD). As a positive control, Toxins A and B were purified from the wildtype strain (ribotype 087), which was the standard toxin used throughout product development. All samples produced the expected positive result with the SensiTox C. difficile Toxin Test. One initial invalid result was observed in the study (1/49, 2.0% initial invalid rate), and this sample produced the valid expected result upon repeat testing (0% final invalid rate).
Table 6. C. difficile strain ribotypes source of purified toxin used in Inclusivity Study
| Target Toxin | C. difficile Ribotype | C. difficile Toxinotype | Clinical Isolate |
| --- | --- | --- | --- |
| Toxin A | 001 | 0 | L1228 |
| | 002 | 0 | D2177 |
| | 014 | 0 | D1321 |
| | 027 | III | 1412 |
| | 078 | V | 312 |
| | 106 | 0 | B5343 |
| Toxin B | 001 | 0 | L1228 |
| | 002 | 0 | D2177 |
| | 014 | 0 | D1321 |
| | 017 | VIII | NCTC 13569 |
| | 027 | III | 1412 |
| | 036 | X | CCUG 20309 |
| | 078 | V | 312 |
| | 106 | 0 | B5343 |
## 7. Assay Cut-Off:
Assay parameter thresholds were established based on the evaluation of negative stool specimens, contrived stool specimens spiked with C. difficile A and B toxins, and clinical specimens characterized by cellular cytotoxicity neutralization assay (CCNA) for C. difficile toxins. This included approximately 200 spiked specimens, 500 initial CCNA specimens, and an additional 300 CCNA specimens for confirmation.
## 8. Accuracy (Instrument):
Not applicable
K193490 - Page 10 of 13
{10}
K193490 - Page 11 of 13
9. Carry-Over:
A study was conducted to determine if carry-over contamination occurs within runs or between runs with the SensiTox C. difficile Toxin Test on the MultiPath Analyzer. The study used negative pooled stool samples and high positive samples contrived by spiking both Toxin A and B into negative stool at a concentration of 1.3 µg/mL (more than 370X LoD for Toxin A and 26,000X LoD for Toxin B). In total, 13 high positive and 13 negative samples were prepared in individual test cartridges and run on one MultiPath Analyzer using an alternating pattern within-run and between-run for a total of five runs. No evidence of carry-over effect was observed as all negative samples in the study produced the expected Not Detected result for both toxins.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable
2. Matrix Comparison:
Not applicable
C Clinical Studies:
1. Clinical Sensitivity:
The performance of the SensiTox C. difficile Toxin Test was evaluated in a prospective clinical study performed at three geographically diverse sites in the US. The study tested left-over, de-identified, unpreserved liquid or soft stool specimens sequentially collected from patients suspected of C. difficile infection. Stool specimens were stored up to 72 hours refrigerated at 2 – 8 °C prior to testing with the investigational device. Test performance was evaluated in comparison to the results of the cellular cytotoxicity neutralization assay which detects the presence of C. difficile toxin in fecal sample filtrates. Samples were stored and tested in accordance with the assay instructions and specifications.
Specimen inclusion and exclusion criteria were defined in the protocol. Specimens that met the study inclusion criteria and that did not meet any of the exclusion criteria were included in the analysis. Out of 1107 enrolled samples, 27 samples were excluded from the analysis due to sample processing or shipping errors and 20 samples were excluded due to storage outside of the sample stability timeframe. Results from a total of 1060 samples were available for analysis; 621 (59%) collected from females, 438 (41%) collected from males, and one collected from an individual whose gender was unknown. The age distribution is shown in Table 7.
Table 7. Age Distribution
| Age | n | % |
| --- | --- | --- |
| 2-21 | 40 | 3.8% |
{11}
| Age | n | % |
| --- | --- | --- |
| 22-59 | 440 | 41.5% |
| >60 | 580 | 54.7% |
| Total | 1060 | 100% |
Of the 1060 samples, 14 samples (1.3%) required repeat testing with the SensiTox C. difficile Toxin Test due to initial invalid results, per the protocol. However, samples were not retested because they were either unavailable or exceeded the 72-hour sample stability timeframe. The remaining 1046 samples with valid results were included in the performance analysis.
The performance of the SensiTox C. difficile Toxin Test compared to the cellular cytotoxicity neutralization assay is shown in Table 8.
Table 8. Performance of the SensiTox C. difficile Toxin Test
| | Cellular Cytotoxicity Neutralization Assay | | | |
| --- | --- | --- | --- | --- |
| | | Detected | Not Detected | Total |
| SensiTox C. difficile Toxin Test | Detected | 87 | 41 | 128 |
| | Not Detected | 9 | 909 | 918 |
| | Total | 96 | 950 | 1046 |
| Sensitivity (95% CI) | | 90.6% (83.1 – 95.0%) | | |
| Specificity (95% CI) | | 95.7% (94.2 – 96.8%) | | |
| Positive Predictive Value (95% CI) | | 68.0% (59.5 – 75.4%) | | |
| Negative Predictive Value (95% CI) | | 99.0% (98.1 – 99.5%) | | |
2. Clinical Specificity:
See Section VII.C.1.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
D Clinical Cut-Off:
Not applicable
E Expected Values/Reference Range:
In the clinical study, a total of 1107 left-over, de-identified, unpreserved liquid or soft stool specimens sequentially collected from patients suspected of C. difficile infection were enrolled from three geographically diverse sites in the US. Of these specimens, 1046 provided valid test results for analysis of which 96 were positive by the cellular cytotoxicity neutralization assay comparator for an overall observed prevalence of $9.2\%$ (96/1046). The overall percentage of positive results observed with the SensiTox C. difficile Toxin Test was $12.2\%$ (128/1046).
K193490 - Page 12 of 13
{12}
F Other Supportive Instrument Performance Characteristics Data:
Not applicable
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K193490 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.