Agilon Strip is indicated for use in voids or gaps of the skeletal system. i.e., the extremities, pelvis, and posteroateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.
Device Story
Agilon Strip is a flexible, resorbable, osteoconductive bone graft substitute. It consists of 1-2mm biphasic calcium salt granules (calcium carbonate with calcium phosphate layer) bound by type I collagen fibers. The collagen binder facilitates shaping and containment of the implant within the surgical site. Used by surgeons in orthopedic procedures to fill osseous defects in extremities, pelvis, and posterolateral spine. The device acts as a scaffold for bone growth; it resorbs over time and is replaced by host bone during the healing process. It provides a mechanical filler that supports bone regeneration without contributing to structural stability.
Clinical Evidence
No human clinical trials. Evidence includes bench testing (granule chemistry, porosity, pore interconnectivity, sterilization, shelf-life) and biocompatibility testing per ISO 10993-1. In vivo performance validated in a critically sized defect model and a single-level posterolateral spinal fusion model. Metrics included histology, histomorphometry, x-ray, and micro-CT. Results showed new bone formation, implant resorption, and bone remodeling comparable to the predicate. Fusion rates in the spinal model were consistent with the predicate and published literature; no adverse events or device-related failures reported.
Technological Characteristics
Flexible, resorbable bone graft substitute. Composition: 1-2mm biphasic calcium salt granules (calcium carbonate/calcium phosphate) bound by type I collagen fibers. Granules are ~65% porous with ~500-micron interconnected pores. Osteoconductive. Sterilized. Standalone implant.
Indications for Use
Indicated for patients with skeletal voids or gaps in extremities, pelvis, or posterolateral spine not intrinsic to bony stability. Defects may be surgical or traumatic. For posterolateral spine use, must be mixed with autograft.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is the Department of Health & Human Services logo. The text is in blue, with "FDA" in a larger font size than the rest of the text.
March 30, 2020
Biogennix, LLC % Elaine Duncan President Paladin Medical, Inc. PO Box 560 Stillwater, Minnesota 55082
Re: K193487
Trade/Device Name: Agilon Strip Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: March 3, 2020 Received: March 4, 2020
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jesse Muir, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193487
Device Name Agilon Strip
Indications for Use (Describe)
Agilon Strip is indicated for use in voids or gaps of the skeletal system. i.e., the extremities, pelvis, and posteroateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities and pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
# I. SUBMITTER
#### Submitted on behalf of:
| Company Name: | BIOGENNIX, LLC |
|-----------------|-------------------------------------------------------------------------------------------------------|
| Address: | 1641 McGaw Ave.<br>Irvine, CA 92614 |
| Telephone: | 949-253-0094 |
| Fax: | 949-266-5800 |
| by: | Elaine Duncan, M.S.M.E., RAC<br>President, Paladin Medical, Inc<br>PO Box 560<br>Stillwater, MN 55082 |
| Telephone: | 715-549-6035 |
| Fax: | 715-549-5380 |
| Contact Person: | Elaine Duncan |
| Date Prepared: | December 13, 2019 |
## II. SUBJECT DEVICE
| Trade Name: | Agilon Strip |
|--------------------|--------------------------------------------------------------|
| Common Name(s): | Bone void filler, Bone graft substitute, Bone graft extender |
| Regulation Number: | 21 CFR 888.3045 |
| Regulation Name: | Resorbable calcium salt bone void-filler device |
| Product Code: | MQV |
| Regulatory Class: | II |
# III. PREDICATE DEVICE
The contents of this submission have demonstrated that Agilon Strip is substantially equivalent to its primary predicate Morpheus (K142828) when used as a bone graft substitute in the extremities and pelvis, and as a bone graft extender in the posterolateral spine.
# IV. DEVICE DESCRIPTION
Agilon Strip is a flexible, resorbable, wicking, osteoconductive bone graft substitute composed of 1-2mm osteoSPAN granules bound by type I collagen fibers to facilitate shaping and containment of the implant.
{4}------------------------------------------------
The osteoSPAN granules in Agilon Strip are approximately 65% porous, biphasic calcium salts with interconnected pores having a nominal cross-section of 500 microns. The primary composition of each granule is calcium carbonate, with a thin laver of calcium phosphate throughout its entire porosity.
## V. INDICATIONS FOR USE
Agilon Strip is indicated for use in voids or gaps of the skeletal system, i.e., the extremities, pelvis, and posterolateral spine, that are not intrinsic to the stability of the bony structure. These osseous defects may be created surgically or from traumatic injury. Agilon Strip may be used alone in the extremities or pelvis but must be mixed with autograft when used in the posterolateral spine. Agilon Strip resorbs and is replaced with bone during the healing process.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Product | 510(k)<br>Number | Granules<br>Composition | Polymer<br>Binder | Collagen | Osteo-<br>conductive |
|--------------|--------------------|------------------------------------------------------|-------------------|----------|----------------------|
| Morpheus | K132377<br>K142828 | Calcium Carbonate/<br>Calcium Phosphate<br>composite | Yes | No | Yes |
| Agilon Strip | K193487 | Identical | No | Yes | Yes |
The function, intended use and technological characteristics of the subject device are substantially equivalent to the predicate device cleared under 510(k) premarket notification K142828.
## VII. PERFORMANCE DATA
Biogennix followed the "Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device: Guidance for Industry and FDA, June 2, 2003, as well as the company's design controls and risk analysis procedures to ensure that Agilon Strip is safe and effective for use.
Bench testing of Agilon Strip was performed to verify that the addition of the type 1 collagen does not affect the granule chemistry, crystallinity, porosity, pore diameter, and pore interconnectivity. Sterilization validation and shelf-life aging studies demonstrate the suitability of the packaging system.
A full evaluation of the biocompatibility of Agilon Strip was conducted in accordance with "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." As a tissue/bone
{5}------------------------------------------------
permanent implant device, Agilon Strip satisfied a battery of tests assessing the following biological effects:
- Cytotoxicity
- Sensitization ●
- Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity ●
- Bacterial Endotoxins
- Subchronic toxicity
- Genotoxicity ●
- . Implantation
In vivo testing of Agilon Strip was conducted using a validated, critically sized defect model. Device performance was assessed at multiple time points against the primary predicate and negative controls using histology, histomorphometry, x-ray, and micro-CT analyses. The critical nature of the model was confirmed by the negative controls. No adverse reactions were noted at the implant site or in distant organs; new bone formation, bone remodeling, and implant resorption for the test materials were confirmed with time.
Agilon Strip was also validated in a clinically relevant, single-level posterolateral spinal fusion model. Device performance was evaluated at multiple time points against the primary predicate and positive control using mechanical, histology, histomorphometry, x-ray, and micro-CT analyses. Fusion rates were the same between all treatment groups at each time point and consistent with published literature. No adverse events or device-related failures were noted at the implantation site or distant organs for any implant group. No abnormal inflammation, redness, swelling, or discoloration was noted at any time point. New bone formation on the host transverse bone processes, as well as evidence of resorption and remodeling of the graft materials, were confirmed with time.
Based on the endpoints and results of these studies, Agilon Strip was concluded to be substantially equivalent to the primary predicate osteoSPAN Morpheus.
## VIII. CONCLUSIONS
The non-clinical data presented in this submission demonstrates that Agilon Strip is substantially equivalent to the predicate device osteoSPAN Morpheus.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.