LUX-Dx Insertable Cardiac Monitor, myLUX Software Mobile Application, LUX-Dx Clinic Assistant Software, Application Server Device Services (DS) - US
K193473 · Boston Scientific Corporation · MXD · Jun 26, 2020 · Cardiovascular
Device Facts
Record ID
K193473
Device Name
LUX-Dx Insertable Cardiac Monitor, myLUX Software Mobile Application, LUX-Dx Clinic Assistant Software, Application Server Device Services (DS) - US
Applicant
Boston Scientific Corporation
Product Code
MXD · Cardiovascular
Decision Date
Jun 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath. The LUX-Dx has not been tested specifically for pediatric use.
Device Story
LUX-Dx is a subcutaneously-implanted cardiac monitor (ICM) inserted in the fourth intercostal space of the left chest wall. It continuously records subcutaneous ECG (S-ECG) signals via electrodes on the device header and battery. An onboard algorithm evaluates S-ECG waveform data for cardiac arrhythmias and marks events when criteria are met. The device transmits data via Bluetooth Low Energy (BLE) to a mobile monitor (myLUX Patient app or LUX-Dx Clinic Assistant app on an OTS mobile device). The mobile app forwards data to the LATITUDE Clarity server via cellular/Wi-Fi. Clinicians access the server to perform remote monitoring, manage patient data, and review marked events. This system enables long-term cardiac rhythm monitoring, assisting physicians in diagnosing arrhythmias and managing patients with cardiac symptoms, potentially improving clinical decision-making and patient outcomes.
Clinical Evidence
Bench testing only. Evidence includes safety risk management, design verification, design validation, and usability testing. Specific testing categories: sterilization, biocompatibility, electromagnetic compatibility, electrical safety, MRI compatibility, mechanical verification, implant tool verification, packaging verification, software verification, algorithm validation, and human factors/usability testing. Pre-clinical evidence includes GLP and non-GLP studies.
Technological Characteristics
Subcutaneously implanted sensor (Model M301). Dimensions: 1.2 cc. Energy source: LiMnO2 battery. Connectivity: BLE to mobile device, cellular/Wi-Fi to cloud server. Sterilization: Supplied sterile. Software: Algorithm-based arrhythmia detection. MR Conditional (1.5T and 3T).
Indications for Use
Indicated for patients with known heart conditions, risk of abnormal heart rhythm, or symptoms suggesting cardiac arrhythmia (dizziness, palpitations, syncope, chest pain, shortness of breath). Not tested for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
June 26, 2020
Boston Scientific Corporation Melissa Klamerus Principal Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112
#### Re: K193473
Trade/Device Name: LUX-DxTM Insertable Cardiac Monitor Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: June 10, 2020 Received: June 12, 2020
Dear Melissa Klamerus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193473
Device Name LUX-Dx™ Insertable Cardiac Monitor
#### Indications for Use (Describe)
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="font-size: 10px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### 1. Submitter
Boston Scientific Corporation 4100 Hamline Avenue North St. Paul, Minnesota 55112-5798
Contact: Melissa Klamerus Principal Regulatory Specialist Phone: (651) 582-6771 Email: melissa.klamerus@bsci.com
Date Prepared: 11 December 2019
### 2. Device
Trade Name: LUX-Dx™ Insertable Cardiac Monitor Common Name: Arrythmia detector and alarm Product Code/Panel: MXD, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
# 3. Predicate Device
Trade Name: Reveal LINQ Insertable Cardiac Monitor Manufacturer: Medtronic Inc Clearance Number: K132649, 14 February 2014 Common Name: Arrythmia detector and alarm Product Code/Panel: DSI, Cardiovascular Device Class and Panel: Class II Classification Regulation: 21 CFR 870.1025
# 4. Indication for Use
The LUX-Dx Insertable Cardiac Monitor (ICM) is intended to monitor and record subcutaneous ECG (S-ECG) for the clinical evaluation and diagnosis of cardiac arrhythmias. The LUX-Dx is indicated for use in patients that have a known heart condition, are at risk of developing an abnormal heart rhythm, or have symptoms that may suggest a heart condition such as dizziness, palpitations, syncope, chest pain, and/or shortness of breath.
The LUX-Dx has not been tested specifically for pediatric use.
{4}------------------------------------------------
# 5. Device Description
The LUX-Dx ICM sensor device (Model M301) evaluates S-ECG waveform data for indications of cardiac arrhythmias and "marks" the S-EG signal for clinical presentation when the algorithm criteria are met. The device is inserted into the subcutaneous layer of the fourth intercostal space of the left chest wall. The sensor device is powered by an integrated battery. The electrodes used for detecting the S-ECG signal are located on each end of the sensor device, in the header and at the battery. The LUX-Dx system includes the following main components:
- LUX-Dx™ sensor device a subcutaneously-implanted sensor device for cardiac arrhythmia event ● data collection and transmission. In addition, symptom events are collected and transmitted from the sensor device.
- Mobile Monitor (MM) mobile applications (myLUX™ Patient app and LUX-Dx™ Clinic Assistant . app) running on an OTS mobile device that communicates with the LUX-Dx sensor device (using Bluetooth Low Energy (BLE)) and the LATITUDE server (using cellular/Wi-Fi) for collection and transmission of event, patient, and device data.
- . LATITUDE Clarity™ server - a server that communicates with the Mobile Monitor for bidirectional data transmission and provides web access for clinicians to perform remote monitoring activities and manage general patient and system parameters and workflow activities.
- . System Accessories- for insertion of the sensor device, an insertion tool and incision tool are provided. In addition, a magnet is provided to initiate sensor/MM app communication.
| Characteristic | BSC Model M301 LUX-Dx ICM | Medtronic Reveal LINQ LNQII |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The LUX-Dx™ Insertable Cardiac Monitor (ICM)<br>is intended to monitor and record<br>subcutaneous ECG (S-ECG) for the clinical<br>evaluation and diagnosis of cardiac<br>arrhythmias. The LUX-Dx is indicated for use in<br>patients that have a known heart condition,<br>are at risk of developing an abnormal heart<br>rhythm, or have symptoms that may suggest a<br>heart condition such as dizziness, palpitations,<br>syncope, chest pain, and/or shortness of<br>breath. | The Reveal LINQ ICM is an insertable<br>automatically-activated and patient-activated<br>monitoring system that records subcutaneous<br>ECG and is indicated in the following cases:<br>● patients with clinical syndromes or<br>situations at increased risk of cardiac<br>arrhythmias<br>● patients who experience transient<br>symptoms such as dizziness, palpitation,<br>syncope, and chest pain that may suggest a<br>cardiac arrhythmia |
| Usage | Single use | Single use |
| Sterility | Supplied sterile | Supplied sterile |
| Volume Dimension | 1.2 cc | 1.2 cc |
| MR Conditional | 1.5 T and 3T | 1.5 T and 3T |
| Energy Source | Battery powered - LiMnO₂ | Battery powered - LiCFx |
# 6. Substantial Equivalence
{5}------------------------------------------------
| Characteristic | BSC Model M301 LUX-Dx ICM | Medtronic Reveal LINQ LNQII |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Longevity | 3 years | 3 years |
| Remote Monitoring | Yes | Yes |
| Packaging | Sterile kit including ICM device and custom implant tools.<br>Accessories: External, non-sterile devices are packaged separately. | Sterile kit including ICM device and custom implant tools.<br>Accessories: External, non-sterile devices are packaged separately. |
# 7. Summary of Performance Testing
Boston Scientific performed safety risk management activities, design verification, design validation, and usability testing to demonstrate that the LUX-Dx system is substantially equivalent to the predicate device. The system conforms to user needs and intended use.
Performance Testing - Bench:
- . Sterilization
- Biocompatibility
- Electromagnetic compatibility ●
- Electrical safety ●
- MRI compatibility
- Mechanical verification ●
- Implant tool verification
- Packaging Verification
- Software Verification
- Algorithm Validation
- Human Factors/Usability Testing ●
Performance Testing – Pre-clinical:
- . LUX-Dx GLP and non-GLP studies
# 8. Conclusion
Based on the intended use, fundamental technological characteristics, and performance testing, the proposed LUX-Dx system has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.