K193468 · Pioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.) · JDQ · Mar 12, 2020 · Orthopedic
Device Facts
Record ID
K193468
Device Name
Tritium Sternal Cable Plate System
Applicant
Pioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.)
Product Code
JDQ · Orthopedic
Decision Date
Mar 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Tritium® Sternal Cable Plate System is intended for use in the stabilization of fractures of the anterior chest wall including sternal fixation following sternotomy and sternal reconstructive surgical procedures.
Device Story
Sternal fixation system; includes plates, cables, screws, needles, and cable plugs. Designed to enhance stability of sternal closure; utilizes load-sharing concept to distribute lateral force across osteotomy. Implanted via open or minimally invasive approach using manual surgical instruments. Used by surgeons in clinical settings for patients undergoing sternotomy or reconstructive procedures. Provides mechanical stabilization of chest wall fractures; benefits patient by securing sternum post-surgery. Subject of this submission is addition of MR conditional safety labeling.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical MR safety testing per ASTM F2052-15, ASTM F2213-17, ASTM F2182-11a, and ASTM F2119-07.
Technological Characteristics
Materials: Commercially pure titanium Grade IV (ASTM F67), Titanium 6Al-4V alloy (ASTM F136), Custom 470 stainless steel (ASTM F899), Titanium 3Al/2.5V alloy (ASTM F2146). Mechanical fixation via plates, screws, and cable sub-assemblies. Sterilization: Gamma irradiation (sterile implants) and steam sterilization (non-sterile implants/instruments).
Indications for Use
Indicated for stabilization and fixation of anterior chest wall fractures, including sternal fixation post-sternotomy and sternal reconstructive procedures.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
March 12, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Pioneer Surgical Technology, Inc. (dba RTI Surgical, Inc.) % Peter Browne RA Specialist A&E Medical Corporation 5206 Asbury Road, PO BOX 758 Farmingdale, New Jersey 07727
Re: K193468
Trade/Device Name: Tritium Sternal Cable Plate System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ, HRS, HWC Dated: December 10, 2019 Received: December 16, 2019
Dear Peter Browne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Ronald P. Jean, Ph.D. Director (Acting) DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193468
Device Name Tritium® Sternal Cable Plate System
Indications for Use (Describe)
The Tritium® Sternal Cable Plate System is intended for use in the stabilization of fractures of the anterior chest wall including sternal fixation following sternotomy and sternal reconstructive surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K193468 Page 1 of 2
## 510(k) Summary
| Date Prepared | December 10th, 2019 |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner/<br>Manufacturer: | Pioneer Surgical Technology, Inc. dba RTI Surgical, Inc.<br>375 River Park Circle<br>Marquette, MI 49855 USA<br>http://www.rtix.com<br>Establishment Registration #1833824 |
| Contact<br>Person/Consultant: | Peter Browne - RA Specialist<br>A&E Medical Corporation<br>Phone: (732) 378-7337<br>Fax: (732) 938-2399<br>Email: p.browne@aemedical.com |
| Trade name: | Tritium® Sternal Cable Plate System |
| Common name: | Sternal Cable Plate System |
| Classification: | Class II;<br>JDQ; 21 CFR 888.3010; Cerclage, Fixation<br>HRS; 21 CFR 888.3030; Plate, Fixation, Bone<br>HWC; 21 CFR 888.3040; Screw, Fixation, Bone |
| Panel: | Panel Code 87 |
| Predicate: | K161876 Tritium® Sternal Cable Plate System |
| Description: | The Tritium® Sternal Cable Plate System includes implants of various sizes;<br>plates and cable plugs comprised of commercially pure titanium, Grade IV<br>(ASTM F67), and cables and screws comprised of Titanium 6AI 4V Alloy<br>(ASTM F136). The system also includes needles comprised of Custom 470<br>stainless steel, 420 stainless steel (ASTM F899, Custom 470 SST) and leader<br>comprised of Titanium 3AI/ 2.5V Alloy (ASTM F2146).<br><br>The system is designed to enhance the stability and strength of traditional sternal<br>closure techniques. Utilizing a unique load sharing concept, the device can be<br>implanted to distribute lateral force across the osteotomy. The system can be used<br>with traditional monofilament wire or Pioneer Sternal Cable. The device system<br>should be implanted using only the manual surgical instruments designed<br>specifically for this system of implants, which may be implanted via an open or<br>minimally invasive approach. |
| Purpose of<br>submission: | Obtain clearance for the following modification to the predicate K161876<br>Tritium® Sternal Cable Plate System: add magnetic resonance (MR) conditional<br>safety labeling based on completed non-clinical testing. |
| Indications for Use: | The Tritium® Sternal Cable Plate System is intended for use in the stabilization<br>and fixation of fractures of the anterior chest wall including sternal fixation<br>following sternotomy and sternal reconstructive surgical procedures. |
| Summary of<br>Technological<br>Characteristics: | The subject Tritium® Sternal Cable Plate System has the same technological<br>characteristics as the predicate K161876 devices cleared for use in closure of<br>the sternum. Similarities to the predicate device include:<br>- Same indications for use within cardiovascular surgery applications<br>- Same materials, manufacturing processes and biocompatibility:<br>metallic devices, stainless steel and titanium<br>- Same principles of operation and fundamental technology: intended to<br>stabilize and fixate fractures of the anterior chest wall (e.g. sternal<br>fixation) through the use of plates, screws that lock into system plates,<br>and integrated cable sub-assemblies<br>- Same instrumentation: trial plates, screw sizer, cable and plate cutters,<br>tensioner, crimper, screw driver, sterilization case and caddy<br>- Same general surgical technique method:<br>- Same sterility and shelf life: gamma irradiation (sterile implants) and<br>steam sterilization by the user (non-sterile implants and instruments)<br>- Same packaging: double sterile barrier (sterile implants) and<br>polybag/carton (non-sterile implants and instruments)<br>- Same bacterial endotoxin evaluation and limit (20 EU/device)<br>- Same mechanical performance<br>- Same screw features and cable subassembly components<br>The modification to the labeling is supported by non-clinical testing listed<br>below. |
| Discussion of<br>Supporting Clinical<br>Evidence and Non-<br>Clinical Testing: | The following nonclinical tests were submitted and relied on in this premarket<br>notification submission for a determination of substantial equivalence. Testing<br>identified in this summary has all passed acceptance criteria established by the<br>predicate device where applicable.<br><br>MR Safety Evaluation following standards listed below:<br>• ASTM F2052-15, "Standard Test Method for Measurement of Magnetically<br>Induced Displacement Force on Medical Devices in the Magnetic Resonance<br>Environment"<br>• ASTM F2213-17, "Standard Test Method for Measurement of Magnetically<br>Induced Torque on Medical Devices in the Magnetic Resonance Environment"<br>• ASTM F2182-11a, "Standard Test Method for Measurement of Radio<br>Frequency Induced Heating Near Passive Implants During Magnetic Resonance<br>Imaging"<br>• ASTM F2119-07 (Reapproved 2013), “Standard Test Method for Evaluation of<br>MR Image Artifacts from Passive Implants" |
| Conclusion | The subject Tritium® Sternal Cable Plate System was shown to be<br>substantially equivalent to the predicate system. The devices are determined to<br>be MR Conditional based on the results of testing completed according to<br>FDA Guidance document "Establishing Safety and Compatibility of Passive |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.