CorticaLINK Spinal Fusion Platform

K193465 · Spine Smith Holdings, LLC · NKB · Jan 10, 2020 · Orthopedic

Device Facts

Record IDK193465
Device NameCorticaLINK Spinal Fusion Platform
ApplicantSpine Smith Holdings, LLC
Product CodeNKB · Orthopedic
Decision DateJan 10, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CorticaLINK Spinal Fusion Platform is intended for non-cervical pedicle fixation as an adjunct to fusion for the following acute and chronic instabilities or deformities for the T1-S1 vertebrae in skeletally mature patients: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) spondylolisthesis trauma (i.e., fracture or dislocation) spinal stenosis curvatures (i.e., scoliosis, kyphosis, and/or lordosis) tumor pseudoarthrosis and failed previous fusion

Device Story

CorticaLINK Spinal Fusion Platform; suite of spinal fixation implants including screws, rods, and locking caps. Used to stabilize spine as adjunct to fusion following surgical decompression. Implants accommodate varying patient anatomies and surgeon preferences. Operated by surgeons in clinical/OR settings. Provides mechanical stabilization of T1-S1 vertebrae. Benefits patient by facilitating spinal fusion in presence of instability or deformity.

Clinical Evidence

Bench testing only. Mechanical performance evaluated per ASTM F1717 (static/dynamic compression bending and static torsion). Results demonstrate substantial equivalence to predicate device.

Technological Characteristics

System components: screws, rods, locking caps. Materials: Ti6Al4V (ASTM F136) for tulips, screw shafts, and rods; CP Grade 4 Titanium (ASTM F67) for push plates; CoCr (ASTM F75) for rods. Non-sterile; steam autoclave sterilization by healthcare professional. Mechanical fixation system.

Indications for Use

Indicated for skeletally mature patients requiring non-cervical pedicle fixation (T1-S1) as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma (fracture/dislocation), spinal stenosis, spinal curvatures (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Spine Smith Holdings, LLC Nickolas G. Kriska Director of Engineering 4719 South Congress Ave. Austin, Texas 78745 Re: K193465 Trade/Device Name: CorticaLINK Spinal Fusion Platform Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB Dated: December 10, 2019 Received: December 16, 2019 Dear Mr. Kriska: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for RPJ (Ronald Jean) Vacant Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K193465 Device Name CorticaLINK Spinal Fusion Platform #### Indications for Use (Describe) The CorticaLINK Spinal Fusion Platform is intended for non-cervical pedicle fixation as an adjunct to fusion for the following acute and chronic instabilities or deformities for the T1-S1 vertebrae in skeletally mature patients: · degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies) - · spondylolisthesis - · trauma (i.e., fracture or dislocation) - spinal stenosis - · curvatures (i.e., scoliosis, kyphosis, and/or lordosis) - tumor - · pseudoarthrosis - · and failed previous fusion | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary ### ADMINISTRATIVE INFORMATION | Manufacturer Name: | Spine Smith Holdings, LLC<br>4719 South Congress Ave.<br>Austin, TX 78745<br>Telephone (512) 637-6745 | |-------------------------|-------------------------------------------------------------------------------------------------------| | Official Contact: | Nickolas G. Kriska<br>Director of Engineering | | Date Prepared: | December 10, 2019 | | Trade/Proprietary Name: | CorticaLINK Spinal Fixation Platform | | Classification Panel: | Orthopedics | | Common Name: | Thoracolumbosacral Pedicle Screw System | | Classification: | Class II per 21 CFR §888.3070 | | Product Code(s): | NKB | ## ESTABLISHMENT REGISTRATION NUMBER The Spine Smith Holdings, LLC is listed under the establishment registration number 3006404071. ### INTENDED USE The CorticaLINK Spinal Fusion Platform is intended for non-cervical pedicle fixation as an adjunct to fusion for the following acute and chronic instabilities or deformities for the T1-S1 vertebrae in skeletally mature patients: - degenerative disc disease (defined as back pain of discogenic origin with o degeneration of the disc confirmed by history and radiographic studies) - spondylolisthesis ● - trauma (i.e., fracture or dislocation) ● - spinal stenosis - curvatures (i.e., scoliosis, kyphosis, and/or lordosis) - o tumor - pseudoarthrosis - and failed previous fusion ● ### DEVICE DESCRIPTION The CorticaLINK Spinal Fusion Platform is a comprehensive suite of fixation implants which can be used to stabilize the spine as an adjunct to fusion following surgical decompression. The platform is comprised of screws, rods and locking caps, with the screws being offered in several different lengths, diameters, and thread pitches to accommodate varying anatomies, pathologies, and surgeon preferences. {4}------------------------------------------------ The systems components are manufactured using standard manufacturing processes. The tulips, screw shafts, and one family of rods are of medical grade Titanium alloy (Ti6Al4V) that complies with ASTM F136 - Standard Specifications for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications. The push plate is of Commercially Pure Grade4 Titanium (CP Ti) per ASTM F67 - Standard Specifications for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, R50400, R50500 & R50700). The other family of rods are of Cobalt Chromium (CoCr) per ASTM F75 - Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075). The implant components are provided clean and non-sterile. These devices are supplied in a rigid sterilization tray and are to be sterilized by a healthcare professional using a Steam Autoclave in accordance with the instructions for use provided by Spine Smith Holdings. LLC, as well as the instructions provided by the Autoclave manufacturer. # EQUIVALENCE TO MARKETED PRODUCT Spine Smith Holdings, LLC has submitted information to demonstrate that, for the purposes of FDA's requlation of medical devices, the CorticaLINK Spinal Fusion Platform is substantially equivalent to the predicate device based on a comparison including the following characteristics: - FDA Product Code(s) - Intended Uses ● - Anatomical Region - Implant Materials - Product Features # PREDICATE DEVICE(S) - LinkSPINE. Inc. CorticaLINK Spinal Fusion Platform (K160722) (Primary Predicate) . # PERFORMANCE TESTING The worst-case subject device was tested according to ASTM F1717 in static and dynamic compression bending as well as static torsion. Results of the testing demonstrate substantially equivalent mechanical performance as compared to the predicate(s). # CONCLUSION Spine Smith Holdings, LLC has submitted information to demonstrate that, for the purposes of FDA's requlation of medical devices, the CorticaLINK Spinal Fusion Platform modifications substantially meets the performance criteria established for the cleared parent device. The modified device is used to treat the same indications for use, utilizes the same scientific and operational principles, and is manufactured using the same manufacturing practices from identical materials as the parent device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...