K193454 · Philips Medical Systems Nederland B.V. · JAK · Jan 24, 2020 · Radiology
Device Facts
Record ID
K193454
Device Name
IQon Spectral CT
Applicant
Philips Medical Systems Nederland B.V.
Product Code
JAK · Radiology
Decision Date
Jan 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectra may be used to analyse the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology as part of treatment preparation and radiation therapy planning. The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Device Story
Whole-body CT system; utilizes rotating gantry with x-ray tube and detector array to acquire high/low energy transmission data. Spectral reconstruction system processes dual-energy datasets to analyze material attenuation coefficients; generates energy-specific images; quantifies material density, effective atomic number, and electron density; performs calcium suppression. Operated by trained healthcare professionals in clinical settings. Output includes cross-sectional images, material analysis maps, and graphical displays for visualization/analysis of anatomical/pathological structures. Supports diagnostic decision-making in radiology, cardiology, and oncology; facilitates radiation therapy planning and lung cancer screening. Benefits include enhanced material characterization and diagnostic clarity via spectral data.
Clinical Evidence
No clinical study required. Non-clinical performance testing included system-level verification, sub-system requirements specification testing, and usability testing. Clinical workflow and service validation performed via simulated use. Sample clinical images evaluated by certified radiologists confirmed good image quality.
Technological Characteristics
Whole-body CT system with rotating gantry, multi-slice capability, and spectral reconstruction. Components: X-ray HV power supply, X-ray tube, detector array, patient couch, operator console. Connectivity: Workstation hardware/software for acquisition, display, storage, and post-processing. Algorithms: Spectral data processing for electron density and calcium suppression. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-44, IEC 62304, ISO 10993-1, ISO 14971.
Indications for Use
Indicated for diagnostic imaging in radiology, interventional radiology, cardiology, and oncology (treatment/radiation planning) and low-dose lung cancer screening for early detection of nodules in patients meeting established governmental or professional society criteria.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
January 24, 2020
Philips Medical Systems, Nederland B.V. % Nimit Shah Regulatory Affairs Specialist Veenpluis 4-6 Best, 5684 PC THE NETHERLANDS
Re: K193454
Trade/Device Name: IQon Spectral CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: December 11, 2019 Received: December 13, 2019
Dear Nimit Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K193454
Device Name IQon Spectral CT
### Indications for Use (Describe)
The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectra may be used to analyse the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology as part of treatment preparation and radiation therapy planning.
The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl. J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
PHILIPS BG Diagnostic Imaging
## SECTION 5 510(K) SUMMARY
K193454
HA300-006-05 Revision: 01 Status: Approved Page: 1 of 5
CT/AMI
### SECTION 5 510(K) SUMMARY
# 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
- Date Prepared: December 11, 2019
- Manufacturer: Philips Medical Systems Nederland B.V. Veenpluis 4-6. 5684 PC BEST The Netherlands Establishment Registration Number: 3015777306
- Primary Contact Person: Nimit Shah Regulatory Affairs Specialists Phone: 440-869-4409 E-mail: nimit.shah@philips.com
- Secondary Contact Douglas Kentz Person: Director of Regulatory Affairs Phone: (262) 389-7369 E-mail: douglas.kentz@philips.com
- Device: Trade Name: IQon Spectral CT Common name: Computed Tomography x-Ray System Classification Name: Computed Tomography x-Ray System Classification Regulation: 21CFR 892.1750 Classification Panel: Radiology Device Class: ll Primary Product Code: JAK Secondary Product Code: Not Applicable Trade Name: Philips IQon Spectral CT Predicate Device: Manufacturer: Philips Medical Systems (Cleveland), Inc. 510(k) Clearance: K163711 Classification Name: Computed Tomography x-Ray System Classification Regulation: 21CFR §892.1750 Radiology Classification Panel: Class II Device Class: Product Code: JAK
{4}------------------------------------------------
PHILIPS BG Diagnostic lmaqinq
#### CT/AMI
Device Description:
The proposed IQon Spectral CT System is a whole-body computed tomography (CT) X-Ray System featuring a continuously rotating x-ray tube and detectors gantry and multi-slice capability. The acquired x-ray transmission data is reconstructed by computer into cross-sectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment; patient and equipment supports; components; and accessories. The proposed IQon Spectral CT System includes the detector array, which is identical to the currently marketed and predicate device – Philips IQon Spectral CT System (K163711).
The proposed IQon Spectral CT System consists of three main components, that are identical to the currently marketed and predicate device. Philips IQon Spectral CT System (K163711) - a scanner system that includes a rotating gantry, a movable patient couch, and an operator console for control and image reconstruction; a Spectral Reconstruction System; and a Spectral CT Viewer. On the gantry, the main active components are the x-ray high voltage (HV) power supply, the x-ray tube, and the detection system.
In addition to the above components and the software operating them, the proposed IQon Spectral CT System includes workstation hardware and software for data acquisition; image display, manipulation, storage, and filming, as well as post-processing for views other than the original axial images. Patient supports (positioning aids) are used to position the patient.
Indications for Use / The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray Intended Use: transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
> The IQon Spectral CT system acquires one CT dataset – composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyse the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
> This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures and to be used for diagnostic imaging in radiology, interventional radiology, and cardiology and in oncology as part of treatment preparation and radiation therapy planning.
> The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
> The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
> *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl. J Med 2011; 365:395-409) and subsequent literature, for further information.
{5}------------------------------------------------
## SECTION 5 510(K) SUMMARY
| CT/AMI | Page: 3 of 5 |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics | The main components (detection system, the reconstruction algorithm, and the x- ray system) that are used in the proposed IQon Spectral CT System have the same fundamental design characteristics and are based on the same technologies as the currently marketed and predicate device, Philips IQon Spectral CT (K163711)<br>The design, and the fundamental scientific technology provided with the proposed IQon Spectral CT System is identical to the currently marketed and predicate device, Philips IQon Spectral CT (K163711).<br>This 510(k) submission addresses the change of indication for use statement and minor modifications that were implemented in the currently marketed and predicate device Philips IQon Spectral CT (K163711). These minor modifications changes include labelling clarification, hardware obsolescence issues and software defect fixes/improvements, that were implemented since clearance of the predicate device. The additional changes that were introduced are as follows. The following changes do not impact device safety and effectiveness. <ul><li>Updated Indication for Use Statement: This update is only for clarification purposes by adding well known specific indications for use that are covered under the general indication for use statement. The currently marketed predicate device Philips IQon Spectral CT (K163711) is being used for diagnostic imaging in radiology, interventional radiology and cardiology and in oncology as part of treatment preparation and radiation therapy planning. The fundamental scientific technology and design of the proposed IQon Spectral CT system is identical with the currently marketed and predicate device Philips IQon Spectral CT System (K163711). </li><li>Additional Optional Bariatric Couch (Patient Support)</li><li>Hardware Options: 2nd CIRS Rack (Dual CIRS Rack – For Faster reconstruction) and Host Extension Kit (provides the ability to migrate the host from its default location in the control room within the Host Rack to the technical room within the CIRS rack. Both of these hardware options do not affect device safety and effectiveness. </li><li>Electron Density: This is spectral CT image generated from an IQon spectral acquisition. A dedicated algorithm uses spectral data to estimate the electron density (ED) of each voxel. This change does not impact device safety and effectiveness and also does not alter intended use of the device. </li><li>Calcium Suppression Index: This is HU-based spectral CT image generated from an IQon Spectral acquisition. A dedicated algorithm uses spectral data to identify and suppress calcium that normally overlays the underlying tissue. This change does not impact device safety and effectiveness and also does not alter intended use of the device. </li></ul> Based on the information provided above, the proposed IQon Spectral CT System is considered substantially equivalent to the currently marketed and predicate device Philips IQon Spectral CT System (K163711), in terms of fundamental scientific technology. |
| Summary of Non-<br>Clinical Performance<br>Data: | Non-clinical performance testing has been performed on the proposed IQon Spectral CT system and demonstrates compliance with the following International and FDA recognized consensus standards and FDA guidance document(s). <ul><li>IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint) (FDA Recognition Number: 19-4)</li><li>IEC 60601-1-2:2014: (FDA Reorganization Number :19-8)</li><li>IEC 60601-1-3:2008+A1:2013 (FDA Recognition Number: 12-269)</li></ul> |
{6}------------------------------------------------
CT/AMI
- IEC 60601-1-6:2010 +A1: 2013 (FDA Recognition Number: 5-89) ●
- IEC 60601-2-44:2009/AMD2:2016 (FDA Recognition Number: 12-302) ●
- IEC 62304:2006 + A1: 2015 (FDA Recognition Number: 13-79) ●
- ISO 10993-1:2009/Cor.1:2010 (FDA Recognition Number: 2-220) ●
- ISO 14971 2nd Edition. (FDA Recognition Number: 5-40) .
Device Specific Guidance Document:
- Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005)
- Content of Premarket Submissions for Management of Cybersecurity in ● Medical Devices (issued October 2, 2014).
Design Verification planning and testing was conducted at the sub-system and at the system level. The sub-systems are tested against the sub-system requirements specifications (SSRS) and the system level verification is conducted against the system requirement specifications (SRS). System and sub-system verification activities demonstrate the system or sub-systems meet the established system and sub-system level design input requirements. System and sub-system level requirements may be verified by manual test, automated test, inspection/analysis, or any combination of the three. Design verification also includes Image Quality verification and risk analysis risk mitigation testing.
The traceability between the requirements, the hazard mitigations and the test protocols are described in the Traceability Matrix. The Traceability Matrix also shows the overall test results per requirement and per hazard mitigation. The results of the functional and non-functional regression tests as well as the user interface verification are provided in the Traceability Matrix. The detailed results are provided in the Full System Verification Test Report.
Non-Clinical design validation testing covered the intended use and commercial claims as well as usability testing with representative intended users. Validation testing included clinical workflow validation and service validation.
All these tests were used to support substantial equivalence of the proposed IQon Spectral CT System and demonstrate that the proposed IQon Spectral CT System
- complies with the aforementioned international and FDA-recognized . consensus standards and/or FDA device specific guidance document, and;
- Meets the acceptance criteria and is adequate for its intended use. .
Therefore, the Proposed IQon Spectral CT is substantially equivalent to the currently marketed and predicate device Philips IQon Spectral CT (K163711) in terms of safety and effectiveness.
{7}------------------------------------------------
| PHILIPS<br>BG Diagnostic<br>Imaging<br>CT/AMI | SECTION 5 510(K) SUMMARY | HA300-006-05<br>Revision: 01<br>Status: Approved<br>Page: 5 of 5 |
|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Summary of Clinical<br>Performance Data: | The proposed IQon Spectral CT system did not require any external clinical study.<br>The clinical evaluation of workflow was conducted via simulated use testing and is<br>accounted for in the summary of "Non-Clinical Testing" section of the summary. The<br>substantial equivalence to the currently marketed and predicate device Philips IQon<br>Spectral CT (K163711) was demonstrated with the following attributes: | |
| | • | Indication for use; |
| | • | Technological characteristics; |
| | • | Non-clinical performance testing; and |
| | • | Safety and effectiveness. |
| | Sample clinical images were provided with this submission, which were reviewed<br>and evaluated by certified radiologists. All images were evaluated to have good<br>image quality. | |
| Substantial<br>Equivalence<br>Conclusion: | The proposed IQon Spectral CT System is substantially equivalent to the currently<br>marketed predicate device IQon Spectral CT (K163711) in terms of design, features<br>and fundamental scientific technology. The change in indications for use statement<br>is only for moving from general indication for use to specific indication for use for<br>IQon Spectral CT. The updated Indication for Use Statement does not introduce any<br>new risk nor impact safety and effectiveness of the proposed IQon Spectral CT<br>System. | |
| | Additionally, substantial equivalence was demonstrated by non-clinical (verification<br>and validation) performance tests provided in this 510(k) premarket notification.<br>These tests demonstrate that proposed IQon Spectral CT system complies with the<br>requirements specified by Philips Medical Systems Nederland B.V. and the<br>international and FDA-recognized consensus standards and is as safe and effective<br>as its currently marketed and predicate device Philips IQon Spectral CT (K163711)<br>without raising any new safety and/or effectiveness concerns. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.