The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.
Device Story
Electric breast pump (single model KP033_01; double model KP053_01) for home use by lactating women. Device consists of pump unit, tubing, and milk collection kit (funnel, cushion, valve, bottle). Internal chip controls vacuum pump and solenoid valve to generate negative pressure. User operates device to express milk; vacuum levels and cycle speeds are adjustable. Output is expressed breast milk collected in bottle. Benefits include lactation maintenance and relief of engorgement.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance assessments (vacuum pressure, cycle rates, backflow protection, battery life, and use-life testing). All parameters met predefined acceptance criteria.
Technological Characteristics
Li-ion battery-powered diaphragm pump. Components: funnel, silicone cushion, valve, tubing, milk bottle. Adjustable vacuum (9 levels) and cycle speeds. Backflow protection mechanism included. Software-controlled vacuum/solenoid operation. Tested per IEC 60601-1 and IEC 60601-1-2. Biocompatibility per ISO 10993.
Indications for Use
Indicated for lactating women to express and collect breast milk, alleviate engorgement, maintain lactation, and provide milk for future feedings during mother-baby separation. Intended for single-user home use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
September 11, 2020
Guangdong Horigen Mother & Baby Products Co., Ltd Jun Deng Vice President No. 18, Pingnan Industrial Zone, Mianbei Street, Chaoyang District Shantou, 515100 China
Re: K193449
Trade/Device Name: Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053_01) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: August 8, 2020 Received: August 10, 2020
Dear Jun Deng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193449
Device Name Electric Breast Pump (model KP033 01)
Electric Double Breast Pump (model KP053 01)
Indications for Use (Describe)
The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary - K193449
#### I. SUBMITTER:
510(k) Owner:
Guangdong Horigen Mother & Baby Products Co., Ltd. No. 18, Pingnan Industrial Zone, Mianbei Street, Chaoyang District, 515100 Shantou, Guangdong, PEOPLE'S REPUBLIC OF CHINA
Primary Contact Person: Jun Deng Title: Vice president Phone: +86 (754) 83613668 EXT.836 Email: dengjun@horigen.cn
Date summary prepared: September 8, 2020
#### II. DEVICE
Device Name: Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053 01) Common Name: Powered Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Product Code: HGX (Pump, Breast, Powered) Regulatory Class: II
#### PREDICATE DEVICE III.
The predicate device is the Electric Breast Pump (K182413).
This predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are electric breast pumps that are intended for lactating women to express and collect breast milk. The Electric Breast Pump and Electric Double Breast Pump are intended for a single user and are to be used in the home environment.
The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053_01) are Li-ion-battery powered breast pumps that include following main components: the expressed milk collection kit, the pump unit, and tubing. The expressed milk collection kit includes a funnel cover, silicone cushion, funnel
{4}------------------------------------------------
adapter, membrane, valve body, valve membrane cap, and milk bottle. The pump unit contains a chip that controls the work of the vacuum pump and solenoid valve to draw air inside of the funnel adapter to form negative pressure. The tubing has two connectors on the two ends. The user attaches the connector to the air vent of the membrane cap and the connector to the pump unit separately to ensure a tight fit.
# DEVICE MODELS:
- KP033_01: Single pump ●
- KP053_01: Double pump
#### INDICATIONS FOR USE V.
The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053 01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033 01), Electric Double Breast Pump (model KP053 01) are intended for a single user.
| DEVICE | Subject Device | Primary Predicate Device |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) | Electric Breast Pump |
| Indications for Use | The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. The Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) are intended for a single user. | The Electric Breast Pump (K182413) is intended to express and collect milk from the mother's breast, to alleviate engorgement of the breast, maintain the ability of lactation, and provide mother's milk for future feedings when separation of mother and baby occurs. |
### VI. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT AND PREDICATE DEVICE
{5}------------------------------------------------
| K1934- | | |
|----------------------------------------------------------------------------------|-------------------------------------------------------|---------------------|
| Classification Product<br>Code | HGX | HGX |
| Single User | Yes | Yes |
| Single/Double Pump | Single or Double Pump | Single |
| Environment of Use | Home | Home |
| Pump Type | Diaphragm pump | Diaphragm pump |
| Adjustable Vacuum<br>Levels | 9 Levels | 9 Levels |
| Backflow Protection<br>(prevention of<br>backflow of liquid into<br>pump/tubing) | Yes | Yes |
| Maximum Expression<br>Pressure | 247.5mmHg | 247.5mmHg |
| Stimulation Pressure<br>range | KP033_01 (30-120, mmHg)<br>KP053_01 (37.5-112.5mmHg ) | 30-120 mmHg |
| Stimulation Velocity | 100 Cycles/min | 100 cycles/min |
| Expression Velocity | 26-60 Cycles/min | 20 – 90 cycles/ min |
| Expression Pressure<br>Range | 120 -247.5 mmhg | 75-247.5mmhg |
{6}------------------------------------------------
The intended use of the subject and predicate devices is the same (express and collect milk from breasts of lactating women). The subject device has different adjustable stimulation levels and a different power source as compared to the predicate device. There are minor differences in stimulation and expression pressure ranges and cycle speeds of stimulation and expression. However, these differences do not raise different questions of safety and effectiveness.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility
A biocompatibility evaluation for the subject device was conducted in accordance with the International Standard ISO 10993-1:2009, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity - (ISO 10993-5: 2009)
- Sensitization - (ISO 10993-10:2010)
- . Skin Irritation - (ISO 10993-10:2010)
# Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device in accordance with IEC 60601-1:2012 (electrical safety) and IEC 60601-1-2:2014 (EMC).
## Software
Software verification and validation for the subject device was conducted as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (2005).
## Performance testing
Performance testing was conducted on the subject device, including the following assessments:
- . Minimum and maximum vacuum pressure testing in single and double pumping mode
- Assessment of technical specifications for stimulation mode and expression mode at ● all cycle and vacuum settings
- . Assessment of working current, pressure, cycle rate and noise
- Backflow protection mechanism testing
- Battery specification validation and performance .
- . Use life testing of vacuum level and battery performance
{7}------------------------------------------------
All of the tested parameters met the predefined acceptance criteria.
# VIII. CONCLUSION
The nonclinical performance data described above demonstrate that the Electric Breast Pump (model KP033_01), Electric Double Breast Pump (model KP053_01) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.