ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.
Device Story
System of pre-manufactured titanium dental implant abutments; designed for customization via CAD/CAM technology. Inputs: patient-specific digital dental anatomy; device serves as base for prosthetic restorations. Transformation: milling center uses CAD/CAM to shape cylindrical section of blank to patient-specific requirements. Output: customized dental abutment. Used in dental clinics/laboratories; operated by dental professionals. Healthcare provider uses output to support single/multi-unit prostheses; improves patient function/aesthetics. Provided non-sterile; requires steam sterilization before use.
Clinical Evidence
No clinical data included. Substantial equivalence supported by non-clinical performance data: sterilization validation (ISO 17665-1, ISO 17665-2, ISO 14937), biocompatibility (ISO 10993-5, ANSI/AAMI ST72), reverse engineering for compatibility, and static/fatigue testing (ISO 14801).
Technological Characteristics
Material: Ti-6Al-4V ELI (Grade 23) per ASTM F136. Design: Indexed, conical connection. Dimensions: 3.5-5.5 mm implant body diameter; 3.0-5.1 mm restorative interface diameter. Non-coated surface. Provided non-sterile; steam sterilization required. CAD/CAM manufacturing workflow.
Indications for Use
Indicated for patients requiring screw-retained single or cement-retained single or multi-unit dental prosthetic restorations supported by dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 9, 2020
ARUM Dentistry Co. Ltd % Chris Brown Manager Aclivi. LLC 6455 Farley Road Pinckney, Michigan 48169
Re: K193425
Trade/Device Name: Pre-Milled Blank Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 27, 2019 Received: December 10, 2019
Dear Chris Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K193425
Device Name Pre-Milled Blank
### Indications for Use (Describe)
ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.
ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:
| Compatibility Table | | | | | |
|-----------------------|---------|-----------------------------------|---------------------------------------|-------------------------------|-------------------|
| ARUM Pre-Milled Blank | | Implant Platform<br>compatibility | Restorative Platform<br>diameter [mm] | Implant Body<br>diameter [mm] | Abutment<br>Screw |
| Ø10 mm | Ø14 mm | | | | |
| CIHE037 | CIHE038 | NobelActive NP | 3.5 | 3.5 | CSTO001 |
| CIHE039 | CIHE040 | NobelActive RP | 3.9 | 4.3/5.0 | CSTO002 |
| CIHE135 | CIHE136 | NobelActive WP | 5.1 | 5.5 | |
All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-family: Arial, sans-serif;"> <b> <span style="font-size: 10.0pt;">☑</span> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span style="font-family: Arial, sans-serif;"> <b> <span style="font-size: 10.0pt;">☐</span> </b> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K193425 - 510(k) Summary ARUM Dentistry Co., Ltd. Pre-Milled Blank 3/9/2020
#### 5.1 ADMINISTRATIVE INFORMATION
Manufacturer Name ARUM Dentistry Co., Ltd. 1 Building, 44, Techno 8-ro, Yuseong-gu, Daejeon, 34028 Republic of Korea
> Phone: +82-42-721-3644 Fax: +82-42-721-3645
| Official Contact | Heung-Yeon Hwang, Quality Manager |
|------------------|-----------------------------------|
| Email: | hwanghy@arumdentistry.com |
#### 5.2 DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Pre-Milled Blank |
|----------------------------|----------------------------------------------------------------------------------------------|
| Common Name: | Abutment, Implant, Dental, Endosseous |
| Classification Name: | Endosseous dental implant abutment |
| Classification Regulation: | 21 CFR 872.3630 |
| Device Class: | Class II |
| Product Code: | NHA |
| Review Panel: | Dental Products Panel |
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) |
#### 5.3 PREDICATE DEVICE INFORMATION
The devices within this submission are substantially equivalent in indications, intended use and design principles to the following primary predicate and reference devices:
| Primary Predicate | | |
|-------------------|-----------------------------------------|------------------------|
| 510(k) | Device Name | Company Name |
| K190299 | Elos Accurate Customized Abutment | Elos Medtech Pinol A/S |
| Reference Device | | |
| 510(k) | Device Name | Company Name |
| K071370 | NobelActive Internal Connection Implant | Nobel Biocare |
### Primary Predicate
#### 5.4 DEVICE DESCRIPTION
ARUM Dentistry Co., Ltd.'s Pre-Milled Blank abutments are a system of dental implant abutments which have an implant /abutment interface design compatible with the Reference device OEM Nobel Biocare NobelActive implant system. Each Subject device implant abutment has a pre-manufactured implant interface connection interface with a customizable cylindrical section above the implant/abutment interface. The compatible implant body diameters range from 3.5 to 5.5 mm, with restorative interface diameters ranging from 3.0 to 5.1 mm (NP, RP and WP diameters).
The Subject device abutments and corresponding abutment screws are all pre-manufactured from Ti-6Al-4V ELI (Grade 23) titanium conforming to ASTM F136 and provided non-sterile to the user.
The Subject device abutments are in two customizable-section diameters for each restorative interface 510k Summary 5- 1
{4}------------------------------------------------
The cylindrical section of the Subject device abutments are intended to be customized by means of CAD/CAM technology to provide basis or support for single or multiple tooth prosthetic restorations. All digitally designed custom abutments from titanium blank abutments are to be sent to an ARUM Dentistry Co., Ltd.-validated milling center for manufacture.
### 5.5 INDICATIONS FOR USE
ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be attached to a dental implant using the included ARUM Dentistry prosthetic screw.
ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:
| ARUM Pre-Milled Blank | | Implant Platform<br>compatibility | Restorative Platform<br>diameter [mm] | Implant Body<br>diameter [mm] | Abutment<br>Screw |
|-----------------------|---------|-----------------------------------|---------------------------------------|-------------------------------|-------------------|
| Ø10 mm | Ø14 mm | | | | |
| CIHE037 | CIHE038 | NobelActive NP | 3.5 | 3.5 | CSTO001 |
| CIHE039 | CIHE040 | NobelActive RP | 3.9 | 4.3/5.0 | CSTO002 |
| CIHE135 | CIHE136 | NobelActive WP | 5.1 | 5.5 | CSTO002 |
# aatihility Tahla
All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling center for manufacture.
### EQUIVALENCE TO MARKETED DEVICE 5.6
The Subject device is substantially equivalent to the predicate device with respect to Indications for Use and technological principles. The Comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Predicate devices.
| Device | Indications for Use Statement | | | | | |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|---------------------------------------|-------------------------------|----------------|---------|
| Subject Device<br>Pre-Milled Blank<br>ARUM Dentistry<br>Co., LTD | ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide<br>support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single<br>restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will be<br>attached to a dental implant using the included ARUM Dentistry prosthetic screw. | | | | | |
| | ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table:<br>Compatibility Table | | | | | |
| | ARUM Pre-Milled Blank | Implant Platform<br>compatibility | Restorative Platform<br>diameter [mm] | Implant Body<br>diameter [mm] | Abutment Screw | |
| | Ø10 mm | Ø14 mm | | | | |
| | CIHE037 | CIHE038 | NobelActive NP | 3.5 | 3.5 | CSTO001 |
| | CIHE039 | CIHE040 | NobelActive RP | 3.9 | 4.3/5.0 | CSTO002 |
| | CIHE135 | CIHE136 | NobelActive WP | 5.1 | 5.5 | |
| | All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling<br>center for manufacture. | | | | | |
| Primary<br>Predicate<br>Elos Accurate<br>Customized<br>Abutment<br>(K190299) | The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for<br>single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental<br>implant using the included Elos Prosthetic screw. | | | | | |
| | The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:<br>Table 1. | | | | | |
# Comparison of Indications for Use Statements
{5}------------------------------------------------
| Device | Indications for Use Statement | | | |
|---------------------------|------------------------------------------------------|-------------------------|------------------------------|----------------------------------|
| Elos Medtech<br>Pinol A/S | Elos Accurate<br>Customized<br>Abutment - Model Type | Platform compatibility | Platform<br>diameter<br>[mm] | Implant Body<br>diameter<br>[mm] |
| | AB-NBR35 | Nobel Replace NP | 3.5 | 3.5 |
| | AB-NBA30 | Nobel CC 3.0 | 3.0 | 3.0 |
| | AB-NBA43 | Nobel CC RP | 3.9 | 4.3 & 4.5 |
| | AB-NBA60 | Nobel CC WP | 5.1 | 5.5 |
| | AB-SBO33 | Straumann Bone Level NC | 3.3 | 3.3 |
| | AB-SBO41 | Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
Except for the device name, reference/model numbers, the list of specific compatible platforms, and company approving the milling facility the Subject and Predicate devices have nearly identical Indications for Use. These differences do not change the intended use of the Subject and Predicate devices to provide support for single or multi-unit prosthetic restorations. The slight differences in the wording "attachment to dental implants in order to provide support for" and "attaching to dental implants in order to provide basis for" are simply different ways to describe the same intended use. The inclusion of the language "screw-retained single restoration and cement-retained single or multi-unit restoration" specifies sub-sets of the types of single and multi-unit restorations included in the primary predicate. Identifying the specific types of single and multi-unit restorations does not change the intended use of the device.
| Design Parameter | Subject Device | Primary Predicate | Equivalence Discussion |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Pre-Milled Blanks<br>ARUM Dentistry Co., Ltd | Elos Accurate Customized Abutment<br>(K190299) | |
| FDA Classification &<br>Product Code | Endosseous dental implant abutment,<br>21 CFR 872.360, NHA | Endosseous dental implant abutment,<br>21 CFR 872.360, NHA | Substantially Equivalent |
| Materials | Titanium Ti-6Al-4V ELI | Titanium Ti-6Al-4V ELI | Substantially Equivalent |
| Surface Finish | Non-coated | Non-coated<br>Optional DLC coating on screws | Substantially Equivalent<br>The optional coating does not change<br>the intended use of the device. |
| Implant Interface | Indexed | Indexed | Substantially Equivalent |
| Connection type | Conical | Flat top and conical | Substantially Equivalent<br>Different connection interfaces do<br>not change the intended use of the<br>device. |
| Abutment Interface<br>diameter | 3.5-5.1 mm | 3.0-6.0 mm | Substantially Equivalent<br>Slight differences in interface<br>diameters do not change the<br>intended use of the device. |
| Prosthesis Attachment | Cement-retained | Cement-retained | Substantially Equivalent |
| Restoration | Single-unit & Multi-Unit | Single-unit & Multi-Unit | Substantially Equivalent |
| Mode of operation | Provide support for single and multi-<br>unit prostheses. | Provide support for single and multi-<br>unit prostheses. | Substantially Equivalent |
| Sterility | Provided non-sterile | Provided non-sterile | Substantially Equivalent |
| Sterilization method | Steam Sterilization | Steam sterilization | Substantially Equivalent |
| Abutment Design<br>Parameters | Minimum wall thickness: 0.5 mm<br>Maximum post height: 13 mm<br>Maximum angulation: 30°<br>Maximum diameter: 14 mm<br>Minimum post height: 4 mm<br>Minimum gingiva height: 0.5 mm<br>Maximum gingiva height: 4 mm | Minimum wall thickness: 0.4 mm<br>Maximum post height: 13 mm<br>Maximum angulation: 30°<br>Maximum diameter: 12 mm<br>Minimum post height: 4 mm<br>Minimum gingiva height: 0.5 mm<br>Maximum gingiva height: 5mm | Substantially Equivalent<br>A slightly larger minimum wall<br>thickness, and slightly smaller<br>maximum gingival height of the<br>Subject device does not change the<br>intended use of the device. |
# Comparison of Technological Characteristics
Minor differences in abutment interface geometry due to compatible implant systems or restorative interface diameters do not introduce new risk nor change the intended use of the device to provide support for single and multi-unit prosthetic restorations.
{6}------------------------------------------------
Overall, the Technological Characteristics, mode of operation and materials of the Subject device are substantially equivalent to that of the Predicate device.
### 5.7 NON-CLINICAL PERFORMANCE DATA
Non-clinical performance data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1, ISO 17665-2, and ISO 14937; biocompatibility according to ISO 10993-5 and ANSI/AAMI ST72; reverse engineering of the OEM implant bodies, OEM abutments and OEM abutment screws to confirm compatibility; static and fatigue testing according to ISO 14801. No clinical data is included in this submission.
### 5.8 CONCLUSION
Slight variations in the wording of the Indications for the Subject and Predicate devices do not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations.
Slight differences in the compatible implant system abutments available in the Subject and Predicate device systems does not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations.
The slight differences in the Technological Characteristics of the Subject and Predicate devices do not change the intended use of the devices to provide support for single or multi-unit prosthetic restorations or introduce new risk or concerns. Differences in Subject device abutment design parameters were validated with respect to intended use through Performance testing.
Overall, the Pre-Milled Blank Subject device is substantially equivalent to the Predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.