The Resolution 360 ULTRA Clip is indicated for clip placement within the Gastrointestinal (GI) tract for the purpose of: 1. Endoscopic marking 2. Hemostasis for: -Mucosal/sub-mucosal defects <3cm -Bleeding ulcers -Arteries <2mm -Polyps <1.5cm in diameter -Diverticula in the colon -Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection 3. Anchoring to affix jejunal feeding tubes to the wall bowel; and Anchoring to affix fully covered sophageal self-expanding metal stents to the wall of the esophagus 4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
Device Story
Resolution 360 ULTRA Clip is a sterile, single-use endoscopic device for GI tract procedures. It consists of a delivery system (handle and catheter) and a deployable clip. Operated by a physician via a forward-viewing endoscope (≥2.8mm working channel), the device allows for clip rotation and repositioning (jaws open/close up to five times) before final deployment at the lesion site. The clip is used for marking, hemostasis, anchoring feeding tubes/stents, and closing small perforations. By providing mechanical closure of tissue or vessels, the device aids in managing bleeding and structural defects, potentially reducing the need for more invasive surgical interventions.
Clinical Evidence
Bench testing only. Evaluations included clip assembly, repeated open/close cycles, opening gap, retention force, opening/closing forces, scope compatibility/usability, endoscope passibility, and MR conditionality testing per FDA guidance.
Technological Characteristics
Sterile, single-use, radiopaque endoscopic clip. Delivery system: stainless steel and polyester. Clip: stainless steel capsule/arms, Cobalt Chrome yoke, styrene tension breaker. 235cm working length. 17mm clip opening. MR Conditional. Sterilized via Ethylene Oxide (ISO 11135-1, ISO 10993-7).
Indications for Use
Indicated for patients requiring endoscopic marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, arteries <2mm, polyps <1.5cm, colonic diverticula, prophylactic post-resection bleeding reduction), anchoring of jejunal feeding tubes or esophageal stents, and supplemental closure of luminal perforations <20mm in the GI tract.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with an eagle and text around it. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of each other, and the word "ADMINISTRATION" below them. The logo is clean and professional, and it is likely used on official FDA documents and websites.
September 1, 2020
Boston Scientific Corporation Elena Hennessey Fellow, Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K193424
Trade/Device Name: Resolution 360™ ULTRA Clip Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: II Product Code: PKL Dated: July 31, 2020 Received: August 3, 2020
## Dear Elena Hennessey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193424
Device Name Resolution 360 ULTRA Clip
Indications for Use (Describe)
The Resolution 360 ULTRA Clip is indicated for clip placement within the Gastrointestinal (GI) tract for the purpose of:
1. Endoscopic marking
2. Hemostasis for:
-Mucosal/sub-mucosal defects <3cm
-Bleeding ulcers
-Arteries <2mm
-Polyps <1.5cm in diameter
-Diverticula in the colon
-Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection
3. Anchoring to affix jejunal feeding tubes to the wall bowel; and Anchoring to affix fully covered sophageal self-expanding metal stents to the wall of the esophagus
4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K193424 Page 1 of 3
## SECTION 5 510(K) SUMMARY
#### 510(k) SUMMARY
#### 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Elena Hennessey Fellow Regulatory Affairs Specialist Tel: 508-683-4347 Date Prepared: December 9, 2019
#### 2. Proposed Device:
Trade Name: Resolution 360™ ULTRA Clip Common Name: Hemostatic Metal Clip for the GI Tract Regulation Name: Hemorrhoidal Ligator Regulation Number: 876.4400 Product Code: PKL Classification: Class II
### 3. Predicate Device:
### Primary Predicate Device:
Trade Name: Resolution 360™ Clip Common Name: Hemostatic Metal Clip for the GI Tract Regulation Name: Hemorrhoidal Ligator Regulation Number: 876.4400 Product Code: PKL Classification: Class II 510(k) Clearance Number: K151802
And
### Secondary Predicate Device:
Trade Name: SureClip MAX Repositionable Hemostasis Clip Common Name: Hemostatic Metal Clip for the GI Tract Regulation Name: Hemorrhoidal Ligator Regulation Number: 876.4400 Product Code: PKL Classification: Class II 510(k) Clearance Number: K182556
{4}------------------------------------------------
And
# Reference Device:
Trade Name: Resolution™ Hemostasis Clipping Device Common Name: Hemostatic Metal Clip for the GI Tract Regulation Name: Hemorrhoidal Ligator Regulation Number: 876.4400 Product Code: PKL Classification: Class II 510(k) Clearance Number: K122660 and K142973
# 4. Device Description:
The Resolution 360TM ULTRA Clip is a sterile device consisting of a pre-loaded, radiopaque, single-use, endoscopic clipping device consisting of two main components: the delivery system and the clip.
The delivery system consists of a handle and delivery catheter. The delivery system is constructed using stainless steel, and polyester materials. The delivery system will allow for the device to rotate at the distal end. The Resolution 360™ ULTRA Clip delivery system is offered in a 235cm working length.
The clip consists of a stainless steel capsule, and clip arms, a Cobalt Chrome Yoke, and a styrene tension breaker. The clip is deployed from the delivery system during use. The Resolution™ 360 ULTRA Clip jaws are engineered such that they can be opened and closed up to five times prior to deployment, aiding in repositioning of the clip at the lesion site. Re-opening, closing, and rotation capability may be limited by clinical circumstances and patient anatomy. There are no associated accessories included with this device. The clip jaws will be available with a 17mm clip opening.
## 5. Indications for Use:
The Resolution 360 ULTRA Clip is indicated for clip placement within the gastrointestinal (GI) tract for the purpose of:
- 1. Endoscopic marking
- 2. Hemostasis for:
- Mucosal/sub-mucosal defects < 3 cm ●
- Bleeding ulcers ●
- Arteries < 2 mm ●
- Polyps < 1.5 cm in diameter ●
- Diverticula in the colon ●
- Prophylactic clipping to reduce the risk of delayed bleeding post lesion ● resection
- 3. Anchoring to affix jejunal feeding tubes to the wall of the small bowel; and Anchoring to affix fully covered esophageal self-expanding metal stents to the wall of the esophagus
{5}------------------------------------------------
- 4. As a supplemental closure method of luminal perforations < 20 mm that can be treated conservatively
## 6. Technological Characteristics:
The proposed Resolution™ 360 ULTRA Clip has different technological characteristic compared to the primary predict Resolution 360™ Clip (K151802). However, both the proposed and the proposed and primary predicate devices can pass through forward viewing endoscopes with a working channel equal to or greater than 2.8 mm to the target position. In addition, the materials of the components used to manufacture the deployed clip components of the proposed and primary predicate are identical.
The proposed Resolution 360 ULTRA Clip has identical indications for use as the primary predicate Resolution 360 Clip. The proposed device has the identical intended use and is placed using the identical methodology as the predicate devices. The proposed device will be available with a 17mm clip opening which is similar to the secondary predicate SureClip MAX Repositionable Hemostasis Clip (K182556) and functions in the same manner as the primary predicate Resolution 360 Clip (K151802). However, the proposed devices deployed clip length is slightly longer than that the predicate devices.
The materials of the proposed Resolution 360 ULTRA Clip are identical to the primary predicate device Resolution 360 Clip.
## 7. Performance Data:
The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing", ISO 11135-1 "Sterilization of Health Care products -Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals",
The following bench tests were performed on the Resolution™ 360 ULTRA Clip: Clip Assembly Repeated Open/Close: Clip Opening Gap: Retention Force: Clip Opening Force: Clip Close Force; Scope Compatibility/Usability; and Endoscope Damage - Clip Passibility.
In addition, the proposed device was evaluated for Magnetic Resonance to support MR Conditionality for the proposed device. Magnetic Resonance (MR) testing along with scientifically based rationale for clinically relevant acceptance criteria consistent with the recommendations contained in the FDA Guidance for Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment (issued December11, 2014) were completed. The results from the test data and scientific rationale have determined the Resolution 360™ ULTRA Clip to be an MR Conditional device.
## 8. Conclusion:
Boston Scientific Corporation has demonstrated that the proposed Resolution 360 ULTRA Clip is substantially equivalent to the currently cleared Resolution 360 Clip (K151802) and SureClip MAX Repositionable Hemostasis Clip (K182556) as the performance of the proposed device meets the requirements of its pre-defined acceptance criteria and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.