← Product Code [MBI](/productcode/MBI) · K193402

# ALLthread PEEK Suture Anchor (K193402)

_Biomet, Inc. · MBI · Mar 5, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K193402

## Device Facts

- **Applicant:** Biomet, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Mar 5, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The ALLthread PEEK Suture Anchor devices are intended for soft tissue fixation. The ALLthread PEEK Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder. Specific indications for the shoulder include: Bankart repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair.

## Device Story

ALLthread PEEK Suture Anchor is a soft tissue anchor preloaded with nonabsorbable polyethylene surgical sutures. Used by surgeons in shoulder soft tissue reattachment procedures; device secures soft tissue to bone. Subject of 510(k) for line extension including 2 or 3 suture configurations, color-coded suture options (white/black, white/green), and tapered needle options. Device functions as a mechanical fastener; no electronic or software components. Benefits patient by providing stable fixation during healing of soft tissue injuries.

## Clinical Evidence

No clinical data provided. Bench testing only: suture tensile strength, cyclic loading, static pull-out strength, insertion torque, and needle attachment strength testing performed. Bacterial Endotoxins Test (BET) performed per ANSI/AAMI ST 72:2011, meeting limits of ≤20 EU/device.

## Technological Characteristics

PEEK (Polyetheretherketone) suture anchor; nonabsorbable polyethylene surgical sutures. Mechanical threaded fixation. Supplied in double or triple loaded configurations. Sterilization method implied consistent with predicate. No software or energy source.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- ALLthread PEEK Suture Anchor ([K060693](/device/K060693.md))
- ALLthread PEEK Suture Anchor ([K080088](/device/K080088.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 5, 2020

Biomet Inc. Haley Pioch Sr. Regulatory Affairs Specialist 56 East Bell Drive, PO Box 587 Warsaw, Indiana 46581

Re: K193402

Trade/Device Name: ALLthread PEEK Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: MBI, HWC Dated: December 5, 2019 Received: December 6, 2019

Dear Ms. Pioch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laura C. Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193402

Device Name ALLthread PEEK Suture Anchors

#### Indications for Use (Describe)

The ALLthread PEEK Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder. Specific indications for the shoulder include: Bankart repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair.

Type of Use (Select one or both, as applicable)

| <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-----------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the ALLthread PEEK Suture Anchor 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.

| Sponsor:                  | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034                                                                                 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person:   | Haley Pioch<br>Sr. Regulatory Affairs Specialist<br>Telephone: (412-376-2510 extension 214)                                                                                                       |
| Secondary Contact Person: | Jared Cooper<br>Regulatory Affairs Manager<br>Telephone: (574-372-1941)                                                                                                                           |
| Date:                     | December 05, 2019                                                                                                                                                                                 |
| Subject Device:           | Trade Name: ALLthread PEEK Suture Anchor<br>Common Name: Suture Anchor                                                                                                                            |
|                           | Classification Name:                                                                                                                                                                              |
|                           | <ul><li>MBI - Fastener, Fixation, Nondegradable, Soft Tissue, Smooth or Threaded Metallic Bone Fixation Fastener (21 CFR 888.3040)</li><li>HWC- screw, fixation, bone (21 CFR 888.3040)</li></ul> |

Predicate Device(s):

| Primary Predicate3           | 510(k) Number |
|------------------------------|---------------|
| ALLthread PEEK Suture Anchor | K060693       |
| ALLthread PEEK Suture Anchor | K080088       |

| Purpose and Device Description:                               | The ALLthread PEEK Suture Anchors are soft tissue<br>anchors with preloaded nonabsorbable polyethylene surgical<br>sutures used to repair and reattach soft tissue to bone and are<br>supplied in both double and triple loaded suture<br>configurations. The ALLthread PEEK Suture Anchors are<br>designed to reattach soft tissue to bone in procedures in the<br>shoulder.                                                                                                                                                                                |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                               | The purpose of this submission is:<br>To submit a 510(k) for cumulative changes:  Line extension including device configurations with 2 or 3 suture options, a white/black suture option, a white/green suture option and a tapered needle option.   To update labeling in order to narrow the Indications for Use statements, describe the third suture options, and bring the Instructions for Use up to current practices; To ensure that the all of the instrumentation/accessories for use with this system are appropriately associated with a 510(k). |
| Intended Use and<br>Indications for Use:                      | The ALLthread PEEK Suture Anchor devices are intended for soft tissue fixation.<br><br>The ALLthread PEEK Suture Anchors are indicated for use in soft tissue reattachment procedures in the shoulder.<br>Specific indications for the shoulder include: Bankart repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule repair or capsulolabral reconstruction, biceps tenodesis, deltoid repair.                                                                                                                           |
| Summary of Technological<br>Characteristics:                  | The rationale for substantial equivalence is based on consideration of the following characteristics:<br>The rationale for substantial equivalence is based on consideration of the following characteristics:  Intended Use: Identical to predicate Indications for Use: Similar to predicate Materials: Similar to predicate Design Features: Similar to predicate Sterilization: Similar to predicate                                                                                                                                                     |
| Summary of Performance Data:<br>(Nonclinical and/or Clinical) | Non-Clinical Tests:<br>Suture tensile strength testing of the ALLthread PEEK Suture Anchor device was performed to verify the strength of the sutures. The test results indicate that the device modifications do not introduce any new risks to implant performance. Cyclic loading testing of the ALLthread PEEK Suture Anchors was performed to verify cyclic performance of the threaded anchor implant. The test results indicate that the device and accessory modifications do not introduce any new risks to implant performance.                    |

{4}------------------------------------------------

{5}------------------------------------------------

- Static load testing of the ALLthread PEEK Suture ● Anchors was performed to verify the pull out strength of the device. The test results indicate that the device and accessory modifications do not introduce any new risks to device performance.
- Insertion torque testing of the ALLthread PEEK . Suture Anchors and taps was performed to verify the measured insertion torque or the anchors and taps. The test results indicate that the device and accessory modifications do not introduce any new risks to device performance.
- . Needle attachment strength testing of the ALLthread PEEK Suture Anchor devices was performed to verify the minimum force required to detach the suture from needle. The test results indicate that the device and accessory modifications do not introduce any new risks to device performance.
- Bacterial Endotoxins Test (BET) per ANSI/AAMI ● ST 72:2011 as a part of cleaning validation was performed, demonstrating that the device meets the limit of ≤20 Endotoxin units (EU)/Device per USP41-NF36 Chapter <161> Medical Devices -Bacterial Endotoxin and Pyrogen Tests.

## Clinical Tests:

- . None provided
The subject ALLthread PEEK Suture Anchor has the same intended use and similar indications for use as the predicate devices. The subject device has similar technological characteristics to the predicates, and the information provided herein demonstrates that:

- any differences do not raise new questions of safety ● and effectiveness; and
- the subject device is at least as safe and effective as the legally marketed predicate devices.

Substantial Equivalence Conclusion:

---

**Source:** [https://fda-staging.innolitics.com/device/K193402](https://fda-staging.innolitics.com/device/K193402)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com