GlassLok is intended for use as a dual cure glass ionomer band cement for use in an orthodontic or dental office.
Device Story
GlassLok is a dual-cure resin-modified glass ionomer cement used by orthodontists or dentists to bond molar bands to enamel or metal substrates. The device is supplied as a two-part system (powder and liquid) mixed at a 2:1 ratio on a mixing pad. The material is hydrophilic, allowing for use in wet or dry fields. Once applied to the band and tooth, it sets chemically within 5 minutes or can be light-cured for 20 seconds. Available in natural or blue shades for visibility during removal. The device provides a mechanical bond to tooth and metal surfaces; it is intended for professional use in clinical settings.
Clinical Evidence
Bench testing only. Performance evaluated via shear bond strength testing against predicate GC Fuji Ortho LC on enamel and metal substrates. Additional testing conducted per ISO 9917-2:2017. Biocompatibility assessed per ISO 10993 and ISO 7405:2018, including cytotoxicity and oral toxicity testing, confirming no adverse local or systemic effects.
Technological Characteristics
Resin-modified glass ionomer cement; powder/liquid formulation. Methacrylate-based liquid. Dual-cure (chemical and light-cure). Hydrophilic properties. Biocompatible. Tested per ISO 9917-2:2017, ISO 10993, and ISO 7405:2018. Shelf life: 2 years at room temperature.
Indications for Use
Indicated for use as a dual cure glass ionomer band cement for bonding molar bands to enamel or metal substrates in orthodontic or dental office settings.
Regulatory Classification
Identification
A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square.
April 27, 2020
Reliance Orthodontic Products Inc. Paula Wendland Regulatory Affairs Manager 1540 West Thorndale Ave. Itasca, Illinois 60143
Re: K193388
Trade/Device Name: GlassLok Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket Adhesive Resin and Tooth Conditioner Regulatory Class: Class II Product Code: DYH Dated: February 25, 2020 Received: February 27, 2020
Dear Paula Wendland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas ''Nandu'' Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
#### K193388
Device Name GlassLok
#### Indications for Use (Describe)
GlassLok is intended for use as a dual cure glass ionomer band cement for use in an orthodontic or dental office.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Page 1 of 1
## PAGE 53
{3}------------------------------------------------
#### K193388
Image /page/3/Picture/1 description: The image shows the logo for Reliance Orthodontic Products, Inc. The logo consists of a circular emblem with the letter "R" inside, followed by the company name written in a cursive font. A horizontal line underlines the company name, adding a touch of elegance to the overall design.
48 · Phone 630-773-4009 · Fax 630
#### Section 5.0 510 (k) Summary
Note: This summary is provided in accordance with 21CFR807.92.
| 510(k) Submitter: | Reliance Orthodontic Products, Inc.<br>Paul Gange, President/Owner |
|-------------------|--------------------------------------------------------------------|
| Address: | 1540 West Thorndale Ave.<br>Itasca, IL 60143 USA |
| Phone Number: | 630-773-4009 |
| Fax Number: | 630-250-7704 |
| Contact Person: | Paula Wendland, Regulatory Affairs Manager |
Date 510(k) Summary was Prepared: 12/4/2019
Medical Device Name:
- Trade name GlassLok .
- Common name Dual Cure Band Cement .
- Classification name -Dental Cement (21CFR872.3275, Product Code . EMA, Class II Device)
## LEGALLY MARKETED DEVICES TO WHICH EQUIVALENCE IS CLAIMED (PREDICATE DEVICE) [807.92(a) (3)]:
GC Fuji Ortho LC 510(k) K981461 . This predicate has not been subject to a design related recall.
{4}------------------------------------------------
## K193388
Image /page/4/Picture/1 description: The image shows the logo for Reliance Orthodontic Products, Inc. The logo features a circular emblem with the letter "R" inside, followed by the company name in an elegant, cursive font. A horizontal line underlines the company name, adding a touch of sophistication to the overall design.
Box 678 1540 West Thorndale Ave · Itaso
### 5.1 DESCRIPTION OF THE APPLICANTS DEVICE:
GlassLok is intended for use as a dual cure glass ionomer band cement for use in an orthodontic or dental office. GlassLok can be used to bond molar bands to an enamel or metal substrate in a wet or dry field due to its hydrophilic properties. GlassLok is available in natural or blue shade (for easy disclosure at removal).
GlassLok is dispensed in a 2 part (powder and liquid) format with bottle dispensing. A 15cc and 25 cc liquid (in blue or natural shade) along with a 15gm or 100gm powder component will be available as separate components. A convenience kit will also be available in Blue or Natural shade in a starter or economy volume.
GlassLok powder is dispensed to a mixing pad using a measuring scoop and mixed with GlassLok liquid in a ratio of 2:1. The mixed GlassLok composite is then applied to the inside of an orthodontic band and applied to a properly prepared tooth structure. Working Time is 2 minutes and setting time is 5 minutes once the working time has expired. GlassLok can also be cured with a curing light for 20 seconds, if required for polymerization.
Associated accessories include a measuring scoop. Spatula and mixing pad.
## 5.2 INDICATIONS FOR USE AND POPULATION:
GlassLok is intended for use as a dual cure glass ionomer band cement for use in an orthodontic or dental office.
#### 5.3 PREDICATE DEVICE:
GC Fuji Ortho™ LC 510(k) submission (K981461) dated 06/25/1998 is similar in intended use, handling and technology compared to the device described in this submission.
{5}------------------------------------------------
#### K193388
Image /page/5/Picture/1 description: The image shows the logo for Reliance Orthodontic Products, Inc. The logo features a circular emblem with the letter "R" inside, followed by the company name in a stylized font. A horizontal line underlines the company name, adding a touch of elegance to the overall design.
Toll Free 1-800 323-4348 · Phone 630-773-4009 · Fax 630-2 P.O. Box 678, 1540 West Thorndale Ave. · Itasca, IL
5.4 COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
| Property | GlassLok | Fuji OrthoTM LC |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/<br>Indications for<br>Use | Dual Cure Glass<br>Ionomer Band Cement | Light Cure Orthodontic<br>Bonding Adhesive /<br>Glass Ionomer dental<br>cement. |
| Chemical<br>Composition | Resin Modified Glass<br>Ionomer Powder /<br>Methacrylate based<br>liquid | Resin Modified Glass<br>Ionomer Powder /<br>Methacrylate based<br>liquid |
| Product<br>Features | Bonds to metal and<br>enamel surfaces.<br>Band cement<br><br>Bonds in a wet field<br>Contains Fluoride<br>Biocompatible | Bonding metal brackets<br>and attachments.<br>Band cementation only<br>when extra band<br>retention is desired<br>Bonds in a wet field<br>Fluoride releasing<br>Biocompatible |
| Shelf Life | 2 years at room<br>temperature | 2 years at room<br>temperature |
| Dispensing<br>Method | Bottle – Powder / Liquid | Bottle – Powder / Liquid |
| Method of Cure | Dual Cure (Light Cure<br>and Chemical Cure) | Triple-Cure (Chemical<br>Cure / Light Cure / Acid<br>Based) |
#### 5.5 PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination:
GlassLok was tested against GC Fuji Ortho™ LC using a performance test method for shear bond strength. Testing resulted in similar or better
{6}------------------------------------------------
# K193388 SECTION 5.0
Image /page/6/Picture/1 description: The image shows the logo for Reliance Orthodontic Products, Inc. The logo features a circular emblem with the letter "R" inside, followed by the company name in a stylized font. The text is underlined, giving it a classic and professional appearance. The overall design is clean and elegant, conveying a sense of quality and reliability.
Free 1-800-323-4348 · Phone 630-773-4009 · Fax 630-2 P.O. Box 678, 1540 West Thorndale Ave. · Itasca, IL · 60143
performance of GlassLok when compared to GC Fuji Ortho LC for shear bond strength to enamel and metal.
In addition, testing per ISO 9917-2:2017 was conducted on both GlassLok and GC Fuji Ortho LC resulting in similar results between the two products.
A biocompatibility evaluation was also conducted in accordance with ISO 10993 and ISO 7405:2018 for the composition of the GlassLok device and determined that the device did not directly or through the release of material constituents produce adverse local or systemic effects; it is not carcinogenic or mutagenic, and will not produce adverse reproductive or developmental effects. As a supplement to the biocompatibility evaluation, both cytotoxicity and oral toxicity testing was conducted on GlassLok.
## 5.6 CONCLUSION:
Based on nonclinical testing conducted for shear bond strength and testing conducted per ISO 9917-2:2017 resulting in similar performance, substantial equivalence was demonstrated between the GlassLok and the GC Fuji Ortho LC devices. In conjunction with biocompatibility demonstrated via ISO 7405:2018 and ISO 10993 evaluations, evidence has been submitted to demonstrate GlassLok is safe and equivalent to or better than the GC Fuji Ortho LC device in terms of performance as a band cement.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.